K Number
K120351

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-05-03

(87 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CareSens N POP Blood Glucose Monitoring System consists of a measuring meter, test strips and control solutions. The CareSens N POP Blood Glucose Meter is intended for use with the CareSens N Single Blood Glucose Test Strips for the quantitative measurement of glucose in capillary whole blood from the fingertip and alternative sites such as the forearm, palm, thigh and calf. Alternative site testing should be used only during steady-state blood glucose conditions. The CareSens Control Solutions are for use with the CareSens N POP Meter and CareSens N Single Test Strips to check that the meter and the test strips are working together properly and that the test is performing correctly.

The CareSens N POP Blood Glucose Monitoring System is intended for self testing, in vitro (outside the body) testing at home by people with diabetes as an aid in monitoring the effectiveness of a diabetes control program. The system is intended to be used by a single person and should not be shared. It is not intended for use on neonates and is not for the diagnosis or screening of diabetes.

Device Description

The CareSens N POP Blood Glucose Monitoring System (BGMS) measures the glucose level in whole blood sample by the meter using a small electrical current generated in the test strips. The system consists of the following: the CareSens N POP Meter, CareSens N Single Test Strips, CareSens Control Solutions with two different glucose concentrations ("Control A" and "Control B" ranges, sold separately), Lancing Device, Lancets, User's manual, Quick reference guide and Logbook.

AI/ML Overview

Here's an analysis of the provided text regarding the CareSens N POP Blood Glucose Monitoring System's acceptance criteria and studies:

Understanding the Context:

It's important to note that this document is a 510(k) summary, specifically a "Special 510(k)". This type of submission is for modifications to a legally marketed device that do not significantly alter its fundamental scientific technology or intended use. Therefore, the performance data presented here largely relies on demonstrating that the modified device performs equivalently to the predicate device (CareSens N BGMS), rather than undergoing extensive de novo clinical trials.


Acceptance Criteria and Device Performance

The document explicitly states that the CareSens N POP BGMS has the same fundamental scientific technology, intended use, operating principle, and operating ranges as its predicate device. This implies that the acceptance criteria for performance are primarily based on demonstrating equivalence, with additional validation for newly added features.

1. Table of Acceptance Criteria and Reported Device Performance

Feature/MetricAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance
Intended UseSame as predicate: Quantitative measurement of glucose in capillary whole blood for self-testing at home by people with diabetes as an aid in monitoring diabetes control. Not for neonates or diagnosis/screening of diabetes."Same" (as predicate device K083468)
EnzymeGlucose Oxidase"Same"
Measurement PrincipleAmperometric method"Same"
Test PrincipleElectro-chemical reaction"Same"
Sample TypeCapillary whole blood"Same"
Test Time5 seconds (as per predicate)"Same"
Sample VolumeMinimum 0.5 µL (as per predicate)"Same"
Test Range20 - 600 mg/dL (as per predicate)"Same"
Temperature Range50 - 104°F (10 - 40°C) (as per predicate)"Same"
Operating Humidity10 - 90% (as per predicate)"Same"
Hematocrit Range20 - 60% (as per predicate)"Same"
PowerTwo (2) 3.0 V lithium battery (CR2032) (as per predicate)"Same"
Battery Life1,000 tests (as per predicate)"Same"
Memory CapacityIncreased to 500 blood glucose values (from 250 in predicate)500 blood glucose values
Averaging OptionsExpanded to 1, 7, 14, 30 and 90 days (from 14 days in predicate)1, 7, 14, 30 and 90 days
Disinfection EffectivenessComplete inactivation of live virus (HBV) on materials; no impact on performance or external materials after 5 years simulated use.CLOROX GERMICIDAL Wipes validated for complete inactivation. 260 pre-cleaning and disinfection cycles (simulating 5 years) had no effect on performance or external materials.

Key takeaway on acceptance criteria: For most core performance attributes, the acceptance criterion is "demonstrated equivalence" to the predicate device. For new features (memory, averaging options, disinfection), specific validation tests were conducted to confirm proper operation and robustness.


Study Details

The document primarily describes validation testing for the modifications from the predicate device, relying on the predicate's extensive testing for the core functionalities.

2. Sample size used for the test set and the data provenance

  • Disinfection Study: 260 samples (meters and lancing devices) for pre-cleaning and disinfection cycles, simulating 5 years of use.
  • Other performance characteristics (e.g., accuracy, precision): The document does not explicitly state sample sizes for test sets related to the core glucose measurement performance. It states that the "performance, safety and effectiveness show results similar to the predicate device" because the device uses the same test strips and control solutions, and the fundamental scientific technology is identical. This strongly implies reliance on the predicate device's existing performance data rather than new, large-scale clinical test sets for glucose accuracy.
  • Data Provenance: Not specified for any performance testing. Given the manufacturer's location (Seoul, Korea), it's highly probable that any performance testing was conducted in Korea. The disinfection study was performed by an "outside commercial testing service," location not specified.
  • Retrospective or Prospective: Not specified. For core performance, it would be largely retrospective (relying on predicate data). For disinfection testing and new feature validation, it would be prospective testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable/Not mentioned. For glucose monitoring systems, the "ground truth" for glucose levels is typically established using a reference laboratory method (e.g., YSI analyzer), not expert consensus. The document does not detail how ground truth was established for any glucose measurement performance.

4. Adjudication method for the test set

  • Not applicable/Not mentioned. Adjudication methods are typically used in studies where human readers interpret images or cases. This is a blood glucose monitoring system.

5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done. If so, what was the effect size of how much human readers improve with AI vs without AI assistance.

  • Not applicable. This device is a blood glucose monitoring system, not an AI-assisted diagnostic imaging or interpretation device that would involve human readers.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Yes, implicitly. Blood glucose meters operate as standalone devices, providing a quantitative measurement. The performance claims for the "CareSens N POP BGMS" are for the device itself, providing an output "concentration of glucose." There isn't a human-in-the-loop aspect for the core measurement function.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • For glucose measurement, the ground truth would typically be established by a laboratory reference method (e.g., YSI glucose analyzer), which is the standard for blood glucose monitoring systems. However, the document does not explicitly state this, mostly relying on the equivalence to the predicate where such methods would have been used.
  • For the disinfection study, the ground truth was the presence/absence and viability of HBV (Hepatitis B Virus) inoculated on the materials.

8. The sample size for the training set

  • Not applicable. This device is a traditional electrochemical blood glucose meter, not an AI/machine learning system that requires a "training set" in the conventional sense. Its algorithm (converting electrical current to glucose concentration) is based on electrochemical principles and calibration, not learned from a training dataset.

9. How the ground truth for the training set was established

  • Not applicable (as it's not an AI/ML device with a training set).

{0}------------------------------------------------

K120350

Image /page/0/Picture/1 description: The image shows a logo with the text "I-sens" in a stylized font. The "I" is bold and slightly larger than the rest of the text. The logo is partially encircled by a circular graphic element, which appears to be a stylized representation of a globe or a similar round object. The overall design is simple and modern.

i-SENS, Inc. 465-6 Wolgye-dong, Nowon-gu, Seoul 139-845, Korea

Tel. 82-2-916-6191

3 2012

510(k) Summary

(As required by 21 CFR 807.92)

Type of 510(k):

Special 510(k)

i-SENS, Inc.

Submitted By:

465-6, Wolgye-dong, Nowon-gu, Seoul 139-845, Korea Tel.) +82-2-916-6191 Fax) +82-2-942-2514 www.i-sens.com .

Contact Person:

Dr. Hyun Joon Oh Tel.) +82-33-903-0760 Fax) +82-33-748-6191 e-mail: hjoh@j-sens.com

Prepared Date: .

February 1, 2012

Device Name:

Common Name:

Trade name: CareSens N POP Blood Glucose Monitoring System Glucose Test System

Regulatory Information: l ) Regulation section: 21 CFR 862.1345 Glucose Test System, 21 CFR 862.1660, Quality control material

  1. Classification:

Class II, Class I

  1. Product Code:

CGA - glucose oxidase, glucose NBW - system, test, blood glucose, over the counter JJX - Quality control material

  1. Panel:

Clinical Chemistry (75)

5 - 1

{1}------------------------------------------------

i-SENS, Inc. 465-6 Wolgye-dong, Nowon-gu, Seoul 139-845, Korea Tel. 82-2-916-6191

Intended Use:

The CareSens N POP Blood Glucose Monitoring System consists of a measuring meter, test strips and control solutions. The CareSens N POP Blood Glucose Meter is intended for use with the CareSens N Single Blood Glucose Test Strips for the quantitative measurement of glucose in capillary whole blood from the fingertip and alternative sites such as the forearm, palm, thigh and calf. Alternative site testing should be used only during steady-state blood glucose conditions. The CareSens Control Solutions are for use with the CareSens N POP Meter and CareSens N Single Test Strips to check that the meter and the test strips are working together properly and that the test is performing correctly.

The CareSens N POP Blood Glucose Monitoring System is intended for self testing, in vitro (outside the body) testing at home by people with diabetes as an aid in monitoring the effectiveness of a diabetes control program. The system is intended to be used by a single person and should not be shared. It is not intended for use on neonates and is not for the diagnosis or screening of diabetes.

Device Description:

The CareSens N POP Blood Glucose Monitoring System (BGMS) measures the glucose level in whole blood sample by the meter using a small electrical current generated in the test strips. The system consists of the following: the CareSens N POP Meter, CareSens N Single Test Strips, CareSens Control Solutions with two different glucose concentrations ("Control A" and "Control B" ranges, sold separately), Lancing Device, Lancets, User's manual, Quick reference guide and Logbook.

Substantial Equivalence Information:

  1. Predicate Device name: CareSens N Blood Glucose Monitoring System 2) Predicate 510(k) Number : K083468

  2. Comparison with Predicate Device:

The modified CareSens N POP BGMS has the following features that are same to the predicate device:

  • intended use
  • I operating principle

5 - 2

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a logo with the text "I-sens" in a stylized font. The "I" is lowercase and has a circular design element to the left of it. The word "sens" is in a bolder, sans-serif font. The logo appears to be for a company or product named "I-sens."

i-SENS, Inc.

465-6 Wolgye-dong, Nowon-gu, Seoul 139-845, Korea Tel. 82-2-916-6191

  • fundamental scientific technology i
  • I operating ranges

The test strips and the control solutions are the same as those of the predicate system, CareSens N Test Strips and CareSens Control Solutions.

The modifications from the predicate device are as follows:

  • physical design
  • 동 user interface
  • software of meter 을
  • software of displaying averages stored in the meter's memory

* Similarities and Differences between candidate and predicate device

FeaturesCareSens N BGMS (K083468, Predicate Device)CareSens, N.POP BGMS (Candidate Device)
Intended UseIt is intended to be used for quantitative measurement of glucose in fresh capillary whole blood as an aid to monitor the effectiveness of diabetes control in people with diabetes.Same
EnzymeGlucose OxidaseSame
Measurement principleAmperometric methodSame
Test principleElectro-chemical reactionSame
SampleCapillary whole bloodSame
Test Time5 secondsSame
Sample volumeMinimum 0.5 µLSame
Test Range20 - 600 mg/dLSame
Temperature50 - 104°F (10 - 40℃)Same
Operating Humidity10 - 90%Same
Hematocrit range20 - 60%Same
PowerTwo(2) 3.0 V lithium battery (CR2032)Same

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a logo with the text "i-sen" next to a graphic. The graphic is a circular design with a pattern of curved lines, resembling a stylized globe or network. The text "i-sen" is in a simple, sans-serif font, with the "i" in lowercase and the "sen" in lowercase as well.

Type of Test:

Tel. 82-2-916-6191 '

Battery life1,000 testsSame
Memory Capacity250 blood glucose values500 blood glucose values
Averaging14 days1, 7, 14, 30 and 90 days

Quantitative, Amperometric method, Glucose oxidase (Aspergillus sp.)

The reagent on the test strip produces a small electrical current using glucose Test Principle: as a substrate in the blood sample. The Meter converts the electrical current to a concentration of glucose.

The CareSens N POP BGMS has the same fundamental scientific technology Technological Characteristics: as the predicate CareSens N BGMS.

Assessment of Performance Characteristics: When compared with CareSens N BGMS, the basic features of the CareSens N POP BGMS (including the intended use, operating principle, and fundamental scientific technology) are all the same except for the meter's appearance and output value. The CareSens N POP BGMS uses the same test strips (just a different name) and control solutions being used in the CareSens N POP BGMS. So, its performance, safety and effectiveness show results similar to the predicate device. Therefore, to check the features newly added to the CareSens N POP, validation testing was conducted and each feature was checked through the test results on whether or not they operated properly.

Summary of Pre-cleaning and Disinfection:

Disinfection studies were performed on the CareSens N POP meter and lancing device by an outside commercial testing service to determine the robustness of the meter and lancing device to the recommended pre-cleaning and disinfection protocol, and its effectiveness in preventing the spread of blood-borne pathogens, using hepatitis B virus (HBV). CLOROX GERMICIDAL Wipes (EPA Reg. No: 67619-12) was validated for complete inactivation of live virus inoculated on the materials of the meter and lancing device. We have also demonstrated that 260 each of pre-cleaning and disinfection cycles designed to simulate 5 years of use has affected neither

ર - 4

{4}------------------------------------------------

i-SENS, Inc.

Image /page/4/Picture/1 description: The image shows a logo with the text "I-sens" in bold, sans-serif font. To the left of the text is a circular graphic with three concentric rings, resembling an eye or a target. The logo has a slightly distressed or textured appearance, giving it a somewhat vintage or worn look.

465-6 Wolgye-dong, Nowon-gu, Seoul 139-845, Korea Tel. 82-2-916-6191

the performance nor on the external materials for the meter and lancing . deviçe.

Conclusion: ·

Based on the information provided in this submission, the CareSens N POP BGMS is substantially equivalent to the predicate CareSens N BGMS. The CareSens N POP BGMS has met the performance, safety, and effectiveness of the device for its intended use.

ર - ર

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, represented by three curved lines that suggest the shape of a bird in flight.

10903 New Hampshire Avenue Silver Spring, MD 20993

I-Sens. Inc. c/o Hvun Joon Oh Division Manager, Quality Assurance 465-6 Wolgye-dong Nowon-gu Seoul, Republic of Korea 139-845

MAY - 3 2012

K120351 Re:

Trade/Device Name: CareSens N POP Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW, CGA Dated: April 2, 2012 Received: April 4, 2012

Dear Hyun Joon Oh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{6}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical

Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance...

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm

Sincerely yours,

C.

Counney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use Form

510(k) Number (if known):

Device Name: _ CareSens N POP Blood Glucose Monitoring System

Indications for Use:

The CareSens N POP Blood Glucose Monitoring System consists of a measuring meter, test strips and control solutions. The CareSens N POP Blood Glucose Meter is intended for use with the CareSens N Single Blood Glucose Test Strips for the quantitative measurement of glucose in capillary whole blood from the fingertip and alternative sites such as the forearm, palm, thigh and calf. Alternative site testing should be used only during steady-state blood glucose conditions. The CareSens Control Solutions are for use with the CareSens N POP Meter and CareSens N Single Test Strips to check that the meter and the test strips are working together properly and that the test is performing correctly.

The CareSens N POP Blood Glucose Monitoring System is intended for self testing, in vitro (outside the body) testing at home by people with diabetes as an aid in monitoring the effectiveness of a diabetes control program. The system is intended to be used by a single person and should not be shared. It is not intended for use on neonates and is not for the diagnosis or screening of diabetes.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

R. Chm

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K120351

Page 1 of

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.