K Number
K120286
Date Cleared
2012-06-19

(140 days)

Product Code
Regulation Number
884.5300
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

TROJAN® HER PLEASURE™ (Ribbed) ECSTASY® Latex Condom with Lubricant is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).

Device Description

The TROJAN® HER PLEASURE™ (Ribbed) ECSTASY® Latex Condom with Lubricant is a male condom consisting of a sheath of natural rubber latex with a silicone-based lubricant, designed with ribs and a larger bulbous end than the predicate device. The condom has a nominal length of 200 mm and a maximum nominal flat-width of 52-54 mm. measured 30 mm from the open end. The bulbous portion at the closed end of the condom has a flat-width of 73 mm. The condom is consistent with the specifications in ASTM D3492-08 for condoms having a shaped profile about the closed end.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information based on the provided text, focusing on the performance characteristics of the TROJAN® HER PLEASURE™ (Ribbed) ECSTASY® Latex Condom:

1. A table of acceptance criteria and the reported device performance

The document primarily references ASTM D3492-08 Standard Specifications for Rubber Contraceptives (Male Condoms) for physical testing and presents consumer-use performance data (slip/breakage rates) for comparison to other marketed condoms. The implicit acceptance criterion for the consumer-use data is comparability to these existing products.

Acceptance Criteria / Performance MetricReported Device Performance (TROJAN® HER PLEASURE™ (Ribbed) ECSTASY®)Comparator Performance (Control Condoms)Notes
Physical TestingMet specifications of ASTM D3492-08N/A (Standard specification)Three lots tested. The specific numerical thresholds for burst pressure, volume, etc., as per ASTM D3492-08, are not detailed in the provided text, only that the device "met specifications."
Consumer-Use PerformanceThe acceptance criterion for these metrics is implied to be "comparable to currently marketed latex condoms."
Slippage Rate1.99%1.09%, 0.99%, 0.00%
Breakage Rate0.46%0.11%, 0.54%, 0.00%
Total Failure Rate2.45%1.20%, 1.53%, 0.00%Total Failure Rate = Slippage Rate + Breakage Rate. The document concludes these results "support that the slip and break rates...is comparable to currently marketed latex condoms."
BiocompatibilityConsidered safe for consumer useN/AStudies included in vitro cytotoxicity extract, vaginal irritation, penile irritation, acute systemic toxicity, sensitization, and bacterial reverse mutation assay.
Shelf-lifeInitial expiration date set at 36 monthsN/AWill be verified through real-time stability for five years. Guided by 21 CFR §801.435.

2. Sample sized used for the test set and the data provenance

  • Physical Testing: Three (3) lots of the device were tested.
  • Consumer-Use Performance:
    • Test Set Sample Size: Over 1250 couples participated in the study.
      • 500+ couples used over 1300 TROJAN® ECSTASY® condoms (including the 510(k)-subject).
      • 750+ couples used more than 2200 currently marketed TROJAN® latex condoms (control).
    • Data Provenance: The study is described as a "branded consumer preference study." It is a prospective study conducted to gather real-world use data. The country of origin is not explicitly stated, but given the manufacturer's location (Princeton, NJ) and submission to the FDA, it is highly likely to be U.S.-based.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This type of information is not applicable to this device and its studies. The "ground truth" for condom performance studies relies on direct observation of user experience (slippage and breakage) and adherence to objective physical performance standards. There are no "experts" in the traditional sense establishing a ground truth as there would be for medical imaging or diagnostic devices.

4. Adjudication method for the test set

This is not applicable. The consumer-use study involved direct reporting of events (slippage/breakage) by the users themselves. Physical testing is based on objective laboratory measurements against a standard. No adjudication by experts is a part of this process.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. An MRMC study is relevant for diagnostic devices where human interpretation of medical images or data is involved, often with AI assistance. This submission concerns a physical medical device (condom), not an AI-powered diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. The device is a physical product, not an algorithm. Therefore, "standalone" algorithm performance is not a relevant concept here.

7. The type of ground truth used

  • Consumer-Use Performance: The ground truth for slip/breakage rates is based on direct observational data reported by couples during actual use.
  • Physical Testing: The ground truth is established by adherence to the objective, quantitative specifications outlined in ASTM D3492-08.

8. The sample size for the training set

This is not applicable. Condom performance studies, as described, do not typically involve "training sets" in the context of machine learning or AI algorithm development. The "training" for such a device effectively comes from its design and manufacturing process, which are regulated by quality systems. The consumer-use study is a validation, not a training activity.

9. How the ground truth for the training set was established

This is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

TROJAN® HER PLEASURE™ (Ribbed) ECSTASY® Latex Condom with Lubricant 510(k) Notification

June 13, 2012

K120286

age

1 of 2

(Revised)

Submitted by:

Church & Dwight Co., Inc. 469 North Harrison Street Princeton, NJ 08543

Contact Person:

Joseph Ciccone Manager, Regulatory Affairs Phone: (609) 497-7251 Fax: (609) 497-7179 eMail: JOSEPH.CICCONE@CHURCHDWIGHT.COM

Date Prepared:

June 13, 2012

Proprietary Name:

TROJAN® HER PLEASURE™ (Ribbed) ECSTASY® Latex Condom with Lubricant

Common Name:

Natural Rubber Latex Condom with Lubricant

Classification Name:

Condom [21CFR §884.5300] HIS

Predicate Device:

TROJAN® (TM-TDB) Latex Condom with Lubricant (K912901) [Secondary Brand Name HER PLEASURE™ added when introduced to market]

Description of Device: The TROJAN® HER PLEASURE™ (Ribbed) ECSTASY® Latex Condom with Lubricant is a male condom consisting of a sheath of natural rubber latex with a silicone-based lubricant, designed with ribs and a larger bulbous end than the predicate device. The condom has a nominal length of 200 mm and a maximum nominal flat-width of 52-54 mm. measured 30 mm from the open end. The bulbous portion at the closed end of the condom has a flat-width of 73 mm. The condom is consistent with the specifications in ASTM D3492-08 for condoms having a shaped profile about the closed end.

Intended Use of the Device: The 510(k)-subject TROJAN® HER PLEASURE™ (Ribbed) ECSTASY® Latex Condom with Lubricant has the same intended use as the predicate TROJAN® HER PLEASURE™ Latex Condom with Lubricant (K912901). TROJAN® HER PLEASURE™ (Ribbed) ECSTASY® Latex Condom with Lubricant is used for contraception and for prophylactic purposes (to help prevent pregnancy and other sexually transmitted infections).

(continued)

{1}------------------------------------------------

TROJAN® HER PLEASURE™ (Ribbed) ECSTASY® Latex Condom with Lubricant 510(k) Notification

June 13, 2012

Technological Characteristics: The 51Q(k)-subject TROJAN® HER PLEASURE™ (Ribbed) ECSTASY® Latex Condom with Lubricant has a similar contour-shaped condom design (shape, lubricated condoms with an integral formed ring at the open-end) and material (natural rubber latex) as the predicate HER PLEASURE™ device reviewed and cleared under 510(k) K912901. The 510(k)-subject condom has been modified to add ribs on the wall of the condom, add more room in the bulbous area by increasing the width, and remove the nipple-end at the condom. The nominal flat-width remains 52-54 mm, measured at 30 mm from the open-end of the condom as specified by ASTM D3492-08. Labeling for the 510(k)-subject device is consistent with the Special Controls provisions of 21 CFR §884.5300 and that of the predicate device. The primary purpose of the added ribbing and added room in the bulbous area of the condom is based on preferred aesthetics by the consumer.

Summary of Studies

Biocompatibility Studies - Biocompatibility studies applicable to the TROJAN® HER PLEASURE™ (Ribbed) ECSTASY® Latex Condom with Lubricant were performed on the final 510(k)-subject device. These studies include in vitro cytotoxicity extract test; vaginal irritation test; penile irritation test; acute systemic toxicity; sensitization test; bacterial reverse mutation assay. TROJAN® HER PLEASURE™ (Ribbed) ECSTASY® Latex Condom with Lubricant is considered safe for consumer use under normal and reasonably forseeable misuse conditions.

Consumer-use performance data - FDA's recognition of D3492-08 notes that, "FDA will also carefully evaluate any condom whose nominal width exceeds 58 mm" (recognition #22, 09/08/2009). Slip/Break data from a branded consumer preference study which included the 510(k)-subject condom and three other currently marketed TROJAN® latex condoms was made part of this 510(k) for FDA's review. More than 1250 couples participated in this study (500+ using over 1300 TROJAN® ECSTASY® condoms and 750+ using more than 2200 currently marketed TROJAN® latex condoms). The slippage rate for the TROJAN® HER PLEASURE" (Ribbed) ECSTASY® Condom was 1.99%, and for the control condoms was 1.09%, 0.99% and 0.00%. The breakage rate for the TROJAN® HER PLEASURE™ (Ribbed) ECSTASY® Condom was 0.46%, and for the control condoms was 0.11%, 0.54%, and 0.00%. The total failure rate for TROJAN® HER PLEASURE™ (Ribbed) ECSTASY® Latex Condom was 2.45%, and for the control condoms was 1.20%, 1.53%, and 0.00%. These results support that the slip and break rates of the TROJAN® HER PLEASURE™ (Ribbed) ECSTASY® Latex Condom is comparable to currently marketed latex condoms.

Physical testing data - Three (3) lots of the TROJAN® HER PLEASURE™ (Ribbed) ECSTASY® Latex Condom with Lubricant were tested and met specifications of ASTM D 3492-08 Standard Specifications for Rubber Contraceptives (Male Condoms).

Shelf-life - Stability of the 510(k)-subject device was established from results of physical testing data using a protocol that followed 21 CFR $801.435 as a guide. Based on the evaluation of the results of the physical testing data, the expiration date has been initially set at 36 months and will be then verified through real-time stability through five (5) years in compliance with FDA expiration date labeling requirements in 21 CFR §801.435.

Accordingly, when compared to the predicate HER PLEASURE™ male latex condom, the data from the performance and biocompatibility studies demonstrate that the TROJAN® HER PLEASURE™ (Ribbed) ECSTASY® Latex Condom with Lubricant is substantially equivalent to the predicate.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of a bird, possibly an eagle, with its wings spread.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JUN 1 9 2012

Mr. Joseph Ciccone Manager, Regulatory Affairs Church & Dwight Co., Inc. 469 North Harrison Street PRINCETON NJ 08543

Re: K120286

Trade/Device Name: TROJAN® HER PLEASURE™ (Ribbed) ECSTASY® Latex Condom with Lubricant

Regulation Number: 21 CFR& 884.5300

Regulation Name: Condom Regulatory Class: II Product Code: HIS Dated: April 13, 2012 Received: April 16, 2012

Dear Mr. Ciccone:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Bayard R. Ewks.

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number:

Device Name:

Indications for Use:

K120286

TROJAN® HER PLEASURE™ (Ribbed) ECSTASY® Latex Condom with Lubricant

TROJAN® HER PLEASURE™ (Ribbed) ECSTASY® Latex Condom with Lubricant is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).

Prescription Use

OR . Over-the-Counter Use _ X

(Per 21 CFR §8001.109)

Concurrence of CDRH, Office of Device Evaluation (ODE)

uctive, Gastro-Renal, and

Image /page/4/Picture/14 description: The image shows the logo for Trojan brand condoms. The logo features the word "TROJAN" in bold, sans-serif font, with the tagline "BRAND CONDOMS" in smaller letters underneath. To the right of the wordmark is a stylized depiction of a Trojan warrior's head, complete with a helmet. The logo is black and white.

Church & Dwight Co., Inc.

§ 884.5300 Condom.

(a)
Identification. A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.(b)
Classification. (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.