K Number
K120116

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-10-01

(262 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SERF Ablation System is intended for the coagulation of soft tissue during percutaneous, laparoscopic and intraoperative surgical procedures.
The SERF Ablation System is intended for the coagulation and ablation of soft tissue during percutaneous, laparoscopic and intraoperative surgical procedures.

Device Description

The SERF Ablation System and SERF Ablation Needle provide heat to unwanted tissue targeted for ablation and thermal necrosis. The SERF System has these main components:

  1. SERF Ablation System console which houses the RF generators, fluid pump, and system user interface.
  2. System user interface, an alphanumeric display for menus and messages and a control panel which displays the set-point of each parameter of the therapy.
  3. SERF Ablation Needle, a single-use, sterile device intended to penetrate tissue to locate the point of therapy initiation based on visual guidance. The SERF Needle is electrically connected to the system and delivers RF energy and warm saline to the tissue through perforations in the shaft near the needle tip.
  4. A sterile single-use syringe for use with the SERF system is commercially available.
  5. Return electrode pads and cables for use with the SERF System are commercially available, The SERF system is a monopolar electrosurgical device.
AI/ML Overview

The provided text is a 510(k) summary for the Thermedical SERF Ablation System, which is a medical device for tissue ablation. It outlines the device's intended use, technological characteristics, and performance testing to demonstrate substantial equivalence to predicate devices.

However, the 510(k) summary does not contain the level of detail requested for acceptance criteria and study design elements typical for an AI/ML medical device submission. Specifically, it is lacking:

  • A table of specific acceptance criteria with corresponding performance metrics. The document states general categories of testing (e.g., "Needle performance characterization and function tests of electrical impedance and fluid leak") but not quantitative targets or results that would be used as acceptance criteria.
  • Information regarding a test set sample size, provenance, or details of a ground truth establishment process (experts, adjudication, exact type of ground truth). The studies mentioned are primarily engineering validation (electrical safety, EMC, software validation) and preclinical (animal studies) and a human study for return electrode pad thermal rise, not diagnostic performance studies that would involve a test set with ground truth.
  • Any mention of a multi-reader multi-case (MRMC) study or standalone algorithm performance. This is because the device is a physical RF ablation system, not an AI/ML algorithm that interprets medical images or data.
  • Details about training sets or how ground truth for a training set would be established. Again, this is not an AI/ML device.

Given the nature of the device (a physical radiofrequency ablation system) and the regulatory pathway chosen (510(k) for substantial equivalence to existing predicate devices), the "acceptance criteria" discussed in the document are primarily related to product performance specifications, safety standards, and biocompatibility, rather than diagnostic accuracy or algorithmic performance.

Therefore, I cannot directly answer your prompt using the provided text as it pertains to an AI/ML device. The document describes a traditional medical device submission, focusing on engineering validation and animal/limited human studies for safety and performance of the physical system, not an AI/ML model's diagnostic or predictive capabilities.

If you are looking for an example of a 510(k) summary for a medical device that includes AI/ML components, the information provided here for the Thermedical SERF Ablation System would be insufficient.

{0}------------------------------------------------

510(k) Summary

K120116
PAGE 1 OF 3

Company: Thermedical 150 Bear Hill Road Waltham, MA 02451-1036 United States of America

OCT 1 2012

FDA Establishment #: not yet assigned

Contact: Katharine M Stohlman, Vice President of Regulatory Affairs Email: k.stohlman@thermedical.com Office: +1 (781) 373-3688 x118 Cell: +1 (978) 764-5154 FAX: +1 (781) 373-2520

Proprietary: Thermedical SERF Ablation System and Accessories

Common: System, Ablation, Radio-frequency and Accessories

Classification: General and Plastic Surgery, GEI, 21 CFR 878.4400

Date Prepared: September 05, 2012

Indications for Use of the SERF™ System

The SERF Ablation System is intended for the coagulation of soft tissue during percutaneous, laparoscopic and intraoperative surgical procedures.

Predicate Devices

The Thermedical SERF Ablation System is substantially equivalent to the following currently marketed devices:

  • AngioDynamics / RITA Model 1500 Electrosurgical RF Generator - Class II -21CFR878.4400, which has been the subject of a cleared 510(k) with FDA log number K993944, K021329 and others.
  • 트 Covidien / ValleyLab Cool-Tip RF Generator, Cool-Tip RF System and Accessories - Class II - 21CFR878.4400, which has been the subject of a cleared 510(k) with FDA log number K984552, K052796 and K053290.

Technological Characteristics

The SERF Ablation System and SERF Ablation Needle provide heat to unwanted tissue targeted for ablation and thermal necrosis. The SERF System has these main components:

    1. SERF Ablation System console which houses the RF generators, fluid pump, and system user interface. The system is designed to comply with Electrical Safety (IEC60601-1), Electromagnetic Compatibility (IEC60601-1-2), and Particular Requirements for the Safety of High Frequency Surgical Equipment (IEC60601-2-2). Product testing was conducted to evaluate conformance to product specifications. Software is validated.

{1}------------------------------------------------

    1. System user interface, an alphanumeric display for menus and messages and a control panel which displays the set-point of each parameter of the therapy.
    1. SERF Ablation Needle, a single-use, sterile device intended to penetrate tissue to locate the point of therapy initiation based on visual guidance. The SERF Needle is electrically connected to the system and delivers RF energy and warm saline to the tissue through perforations in the shaft near the needle tip. All patient contacting materials are commonly used and have been tested for biocompatibility in this application.
    1. A sterile single-use syringe for use with the SERF system is commercially available. The syringe mounts on the pump and delivers sterile saline through a fluid delivery tube to the Ablation Needle.
    1. Return electrode pads and cables for use with the SERF System are commercially available, The SERF system is a monopolar electrosurgical device. The RF circuit is closed through a return path including the return electrode pads which are plugged into the SERF system.

The System is programmed for the desired power level, and procedure time by the operator. Saline infusion rate and target saline temperature are also programmed. The system controls power by monitoring the voltage and current.

Performance Testing

The SERF Ablation System and accessories have been extensively tested to demonstrate suitability for clinical use as follows:

  • . Bioburden and sterility testing of the manufactured needle
  • l Sterilization dose substantiation and process validation of the manufactured needle
  • 트 Environmental control and package shelf life validation of the sterilized, single-use needle
  • . Biocompatibility of the sterilized, needle against requirements of tissue contacting / communicating device
  • I Needle performance characterization and function tests of electrical impedance and fluid leak
  • 1 Software validation for software of Maior concern level
  • B Electrical safety testing according to IEC 60601-1
  • Electromagnetic compatibility testing according to IEC 60601-1-2
  • = High frequency surgical equipment safety testing according to IEC 60601-2-2
  • 시 System console performance characterization and function tests of fluid flow rate, thermometry accuracy, pump response, power display accuracy, tipping force, and system alarm thresholds.
  • . Animal studies of ablation lesions performed at academic labs:
      1. in-vivo study of ablation lesions under different system control settings,
      1. chronic study of lesion pathology,
      1. explanted tissue study of SERF system ablation compared with two predicate systems, and

{2}------------------------------------------------

K120116
PAGE 3 OF 3

  1. thermal rise study of return electrode pads.
  • Non-significant risk human study of thermal rise of return electrode pads

Substantial Equivalence

The SERF Ablation System has the same intended use and indications for use as other previously 510(k) cleared RF ablation systems. The technological characteristics of the SERF Ablation System are also substantially similar to other previously cleared ablation systems. Although there are minor differences in technological characteristics between the SERF Ablation System and the predicates, these differences do not raise any new types of safety or effectiveness issues.

The SERF Ablation System is compared with the Covidien™ Cool-Tip™ and the AngioDynamics™ RITA™ predicate devices. Comparison of intended use, technological features, the therapy applicator, the control methodology, and the resulting ablative lession demonstrates equivalence as confirmed by performance testing.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is an abstract image of an eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

OCT 1 202

Thermedical % Janice M. Hogan Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, PA 19103

K120116 Re: Trade/Device Name: SERF Ablation System and SERF Ablation Needle Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: September 14, 2012 Received: September 14, 2012

Dear Ms. Hogan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{4}------------------------------------------------

Page 2 - Ms. Janice M. Hogan

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Sincerely yours,

Mark N. Melkerson

Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Statement

510(k) Number: K120116

Device Name: Saline-enhanced Radiofrequency (SERF) Ablation System and Accessories

Indications for Use:

The SERF Ablation System is intended for the coagulation and ablation of soft tissue during percutaneous, laparoscopic and intraoperative surgical procedures.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Signature

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K120116

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.