(121 days)
Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)that meets all of the requirements of ASTM standard D 5250-06.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Acceptance Criteria and Device Performance
| Acceptance Criteria Category | Specific Criteria | Reported Device Performance |
|---|---|---|
| Intended Use | A disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. | Same |
| Material Formulation | Poly vinyl chloride | Same |
| Manufacturing | PU Treatment (lack of donning powder) | Same |
| Dimensions | ASTM Standard D 5250-06 | Meets ASTM D5250-06 |
| Physical Properties | ASTM Standard D 5250-06 | Meets ASTM D5250-06 |
| Freedom from Pinhole | 21 CFR 800.20 | Meets 21 CFR 800.20 and FDA requirements for waterleak test on pinhole AQL. |
| Powder Residual | ASTM Standard D 5250-06 and D6124-06 | Meets ASTM D5250-06 and D6124-06 |
| Biocompatibility | No Primary Skin Irritation No Dermal sensitization | Same |
| Overall Comparison | Substantially equivalent to predicate device (K032908) in safety, effectiveness, and performance. | "It can be concluded that the Powder Free Vinyl Patient Examination Gloves meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims. It is as safe, as effective, and performed as well the legally marketed device identified in (a)(3)." |
Study Information for Device Performance
Based on the provided 510(k) summary, the device's performance was evaluated through non-clinical testing. There was no mention of a clinical study being conducted or relied upon for this specific submission, as stated: "Clinical data is not needed for gloves or for most devices cleared by the 510(k) process."
Here's the detailed breakdown of the non-clinical testing:
-
Sample size used for the test set and the data provenance:
- The document does not specify the exact sample size used for each non-clinical test (e.g., how many gloves were tested for pinholes, or for physical properties).
- Data Provenance: Not explicitly stated beyond "non-clinical testing data that was submitted, referenced or relied on". The manufacturer is in China, so the testing was likely conducted by or for the manufacturer. Whether the data is retrospective or prospective for the current submission is implied as prospective testing to demonstrate compliance with current standards.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not provided in the document. For non-clinical, objective tests (like physical properties or pinhole detection), "experts" in the sense of clinical specialists establishing ground truth are typically not required. Instead, adherence to standardized test methods (e.g., ASTM standards) by qualified laboratory personnel is expected.
-
Adjudication method for the test set:
- Adjudication methods (like 2+1, 3+1) are relevant for studies involving human interpretation or subjective assessments. For the non-clinical, objective tests described (e.g., ASTM standards, 21 CFR 800.20 for freedom from pinholes), adjudication in this context is not applicable. Pass/fail criteria are defined by the standards themselves.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices involving human readers, which is not the case for patient examination gloves.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not applicable. This device is not an algorithm or AI system. Its performance is based on intrinsic material and manufacturing characteristics.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The "ground truth" for the non-clinical tests is derived from established industry standards and regulatory requirements.
- ASTM D5250-06: Specifies requirements for vinyl patient examination gloves.
- ASTM D6124-06: Specifies test methods for residual powder on medical gloves.
- 21 CFR 800.20: Pinhole requirements for patient examination gloves (water leak test).
- ISO10993-10: Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization (for biocompatibility).
- The "ground truth" for the non-clinical tests is derived from established industry standards and regulatory requirements.
-
The sample size for the training set:
- Not applicable. This is a physical medical device (patient examination gloves), not an AI algorithm that requires a training set.
-
How the ground truth for the training set was established:
- Not applicable, as there is no training set for this device.
{0}------------------------------------------------
li
510(K) Summary
"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."
"The assigned 510(k) number is: K 1383 ." (applicant leave blank)
Premarket Notification [510(k)] Summary
| [(a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the |
|---|
| submitter's name, address, phone and fax numbers, name of contact person, and date the |
| summary was prepared : |
| Submitter's name : | Hebei Changsheng Plastic Products Co.,Ltd. |
|---|---|
| Submitter's address : | No 99, Huangjin Avenue, Julu County, Xingtai,Hebei Province, China, 055250 |
| Phone number : | 86-319-4361966 |
| Fax number : | 86-319-4362371 |
| Name of contact person: | Mrs. Renxia Han |
| Date the summary was prepared: | April. 5th, 2012 |
[(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known
| Device Name: | Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) | |
|---|---|---|
| Proprietary/Trade name: | Powder Free Vinyl Patient Examination Gloves | |
| Common Name: | Patient examination glove | |
| Classification Name: | Patient examination glove | |
| Device Classification: | I | |
| Regulation Number: | 21 CFR 880.6250 | |
| Panel: | General Hospital (80) | |
| Product Code: | LYZ |
[(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .
Class I* Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) that meets all of the requirements of ASTM standard D 5250-06.
Predicate device : FUGUAN (Brand) Powder-Free Vinyl Patient Examination Gloves, Shijiazhuang Fuguan Plastic Products Co., Ltd.. K032908 .
[(a)(4)] A description of the device
Device Description : Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)that meets all of the requirements of ASTM standard D 5250-06.
{1}------------------------------------------------
[(a)(5)] The summary describes the intended use of the device
Device Intended Use: Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
[(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.
| Characteristics | K 032908 | Device in submission |
|---|---|---|
| Intended use | A disposable device intendedfor medical purposes that isworn on the examiner's handor finger to preventcontamination between patientand examiner. | Same |
| Material Formulation | Poly vinyl chloride | Same |
| Manufacturing | PU Treatment ( Lack ofdonning powder) | Same |
| Dimension | ASTM standard D 5250-00 | ASTM standard D 5250-06 |
| Physical Properties | ASTM standard D 5250-00 | ASTM standard D 5250-06 |
| Freedom from pinholes | 21 CFR 800.20 | 21 CFR 800.20 |
| Powder Residual | ASTM standard D 5250-00 | ASTM standard D 5250-06 |
| and D6124-00 | and D6124-06 | |
| Biocompatability | No Primary Skin IrritationNo Dermal sensitization | SameSame |
It can be concluded that the Powder Free Vinyl Patient Examination Glovest is as safe, as effective, and performed as well the legally marketed device identified K032908.
[(b)(1)] A brief discussion of the non-clinical testing data that was submitted, referenced or relied on to demonstrate that the Subject Device is safe and effective, and whose performance meets the requirements of its pre-defined acceptance criteria and intended uses.
Powder Free Vinyl Patient Examination Gloves. Clear (non-colored)meet requirements per ASTM D5250-06, per ASTM D6124-06, per 21 CFR 800.20 and ISO10993-10. , it is safe and effective, and it's performance meets the requirements of its pre-defined acceptance criteria and intended uses.
[{b)(2)| A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .
Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.
((b)(3)] The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performed as well or better than the legally marketed device identified in (a)(3).
It can be concluded that the Powder Free Vinyl Patient Examination Gloves meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims . It is as safe,as effective, and performed as well the legally marketed device identified in (a)(3).
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Richard Lu Hebei Changsheng Plastic Products Company, Limited No 99 Huangjin Avenue, Julu County Xingtai, Hebei Province China 055250
APR 27 2012
Re: K113837
Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) Regulation Number: 21 CFR 880.6250 . Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: April 16, 2012 Received: April 16, 2012
Dear Mr. Lu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2- Mr. Lu
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem /default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
INDICATIONS FOR USE
Applicant: Hebei Hongtai Plastic Products Co.,Ltd.
510(k) Number (if known):_*
K43837
Device Name: Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)
Indications For Use:
Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth P. Clavie-Wille
(Division Sign-Off) Division of Anesthesiology, General Hospital infection Control, Dental Devices
510(k) Number: _______________________________________________________________________________________________________________________________________________________________
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.