(238 days)
Not Found
No
The description focuses on the mechanical and structural features of the device for stone management and instrument placement, with no mention of AI or ML capabilities. The performance studies are based on physical testing, not algorithmic performance.
No.
Explanation: The device is primarily designed for mechanical intervention (bypassing and entrapping calculi, preventing migration, facilitating instrument placement) during urological procedures. It does not appear to treat or cure a disease itself, but rather assists in procedures to address a condition.
No
The device is intended for interventional purposes such as bypassing and entrapping calculi, preventing retrograde migration, and facilitating instrument placement during procedures, not for identifying the nature or cause of a medical condition.
No
The device description clearly describes a physical device with a braided structure, guidewire-like features, and mechanical actions (deployment, retrieval, stenting compatibility), indicating it is a hardware device, not software-only.
Based on the provided information, the XenX device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly describes a device used within the body (endoscopically) to manipulate and manage calculi in the urinary tract. This is an in vivo procedure.
- Device Description: The description details a physical device that is inserted into the body and performs mechanical functions (blocking, facilitating placement).
- Lack of IVD Characteristics: There is no mention of the device being used to examine specimens outside the body (in vitro) to provide information about a physiological state, health, or disease. It doesn't analyze biological samples like blood, urine, or tissue.
IVD devices are used to perform tests on samples taken from the body, not to directly interact with structures within the body in the way the XenX device does.
N/A
Intended Use / Indications for Use
The XenX device is intended to be used endoscopically to bypass and entrap calculi from the urinary tract, to prevent retrograde migration of calculi during laser lithotripsy,and to facilitate the placement of endourological instruments during diagnostic or interventional procedures.
Product codes (comma separated list FDA assigned to the subject device)
OCY, FFL
Device Description
The XenX device combines several features in one device which facilitate an endoscopic procedure. The XenX easily tracks past ureteral kidney stone and is navigated under fluoroscopy like a urologic guidewire. Once in place, a self expandable braided structure is deployed to block the ureteral lumen and prevent stone particle migration towards the kidney. The braided structure enables a constant irrigation flow for clear uretroscopic vision during the procedure. Following stone fragmentation and if required the physician can choose to propagate a urinary stent over the device for ureteral placement.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Urinary tract
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The following nonclinical testing were performed to demonstrate the performance and safety of XenX as compared to its predicate devices
- Tip flexibility test
- Deployment/retrieval force
- Stenting compatibility
- Tensile strength
- Pushability test
- Radial expansion
- Particle sieving
- Radiopacity test
Testing showed that the XenX is as safe, effective and performs as well as, or better than the predicates.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
K082536, K970121, K052048, K863081
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 876.1500 Endoscope and accessories.
(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
0
page 1 of 5
510(K) SUMMARY (06/11/12) [As required by 21 CFR §§ 807.87 and 807.92] XenXTM
510(k) Number K113692
AUG 9 2012
Applicant's Name:
Xenolith Medical Ltd. 1 Leshem Street, PO Box 720 Kiryat-Gat, 82000 Israel Phone: +972-8-6811761 Fax: +972-8-6811763
Contact Person:
Mr. Ofer Zigman Head of Research and Development Xenolith Medical 1 Leshem Street, PO Box 720 Kiryat Gat, 82000 Israel Phone: +972-8-6811761 Cell: +972-52-6447263 Email: ozigman@xenolithmedical.com
US Agent
Shoshana (Shosh) Friedman, RAC Push-Med LLC 1914 JN Pease Place Charlotte, NC 28269 Phone: 704-899-0092 704-430-8695 Cell: Email: shosh@pushmed.com
Trade Name:
ХепХтм
Classification Name:
Endoscopic guidewire, gastroenterology-urology AND dislodger, stone, basket, ureteral, metal
Classification:
FDA has classified these types of devices as class II devices (product codes OCY and FFL) and they are reviewed by the Gastroenterology/Urology panel.
1
Basis for Submission:
New device
Predicate Devices:
- Sensor Nitinol Guidewire, Boston Scientific Corp, 510(k) Exempt .
- Roadrunner PC Wire Guide, Cook Urological Inc, K082536 .
- . Stone Cone Nitinol Urological retrieval coil, Boston Scientific Corp. K970121
- Accordion urological occluding guidewire, Percsys, K052048 .
- . NTrap stone entrapment and extraction device, Cook Urological Inc, K863081
Device Description:
The XenX device combines several features in one device which facilitate an endoscopic procedure. The XenX easily tracks past ureteral kidney stone and is navigated under fluoroscopy like a urologic guidewire. Once in place, a self expandable braided structure is deployed to block the ureteral lumen and prevent stone particle migration towards the kidney. The braided structure enables a constant irrigation flow for clear uretroscopic vision during the procedure. Following stone fragmentation and if required the physician can choose to propagate a urinary stent over the device for ureteral placement.
Intended Use:
The XenX device is intended to be used endoscopically to bypass and entrap calculi from the urinary tract, to prevent retrograde migration of calculi during laser lithotripsy,and to facilitate the placement of endourological instruments during diagnostic or interventional procedures.
Technological Characteristics:
The XenX consist of a braided tubular mesh compacted into a guidewireshaped device.
Performance Data
The following nonclinical testing were performed to demonstrate the performance and safety of XenX as compared to its predicate devices
- Tip flexibility test .
- Deployment/retrieval force ●
- Stenting compatibility .
- Tensile strength .
- Pushability test .
- Radial expansion .
- Particle sieving .
- Radiopacity test ●
2
K 113692
page 3 of 5
Testing showed that the XenX is as safe, effective and performs as well as, or better than the predicates.
Substantial Equivalence
Characteristic | Sensor | Roadrunner | NTrap | Stone cone | Accordion | XenX |
---|---|---|---|---|---|---|
Intended use | Intended to | |||||
facilitate the | ||||||
placement of | ||||||
endourological | ||||||
instruments | ||||||
during | ||||||
diagnostic or | ||||||
interventional | ||||||
procedures. | Used for | |||||
ureteral access | ||||||
to establish a | ||||||
tract, and | ||||||
assist in the | ||||||
placement, | ||||||
replacement | ||||||
and exchange | ||||||
of devices | ||||||
during | ||||||
urological | ||||||
procedures | ||||||
includes use in | ||||||
a torturous or | ||||||
kinked ureter | ||||||
traversing a | ||||||
large stone in | ||||||
route to the | ||||||
kidney or in | ||||||
cases | ||||||
demanding | ||||||
enhanced | ||||||
control and | ||||||
high | ||||||
radiopacity. | Used as an | |||||
endoscopic | ||||||
entrapment and | ||||||
extraction | ||||||
device for | ||||||
calculi and | ||||||
other foreign | ||||||
bodies in the | ||||||
urinary tract, | ||||||
and to minimize | ||||||
stone migration | ||||||
during laser, | ||||||
electrohydraulic | ||||||
or pneumatic | ||||||
lithotripsy | Intended to be | |||||
used | ||||||
endoscopically | ||||||
to entrap | ||||||
and remove | ||||||
calculi and | ||||||
other foreign | ||||||
objects from | ||||||
the urinary | ||||||
tract. | Intended to be | |||||
used | ||||||
endoscopically | ||||||
to entrap and | ||||||
remove calculi | ||||||
and other | ||||||
foreign objects | ||||||
from the | ||||||
Urinary tract | ||||||
and to guide | ||||||
instrumentation | ||||||
within the | ||||||
ureteral tract. | The XenX | |||||
device is | ||||||
intended to be | ||||||
used | ||||||
endoscopically | ||||||
to bypass and | ||||||
entrap calculi | ||||||
from the | ||||||
urinary tract, | ||||||
to prevent | ||||||
retrograde | ||||||
migration of | ||||||
calculi during | ||||||
laser | ||||||
lithotripsy,and | ||||||
to facilitate the | ||||||
placement of | ||||||
endourological | ||||||
instruments | ||||||
during | ||||||
diagnostic or | ||||||
interventional | ||||||
procedures | ||||||
Location of | ||||||
use | Urinary tract | Urinary tract | Urinary tract | Urinary tract | Urinary tract | Urinary tract |
Technological | ||||||
characteristics | Boston | |||||
Scientific | ||||||
offers the | ||||||
Sensor | ||||||
guidewire | ||||||
with PTFE | ||||||
coatings; | ||||||
hydrophilic | ||||||
coatings for | ||||||
reduced | ||||||
friction; | ||||||
Nitinol core | ||||||
for kink | ||||||
reduction and | ||||||
a flexible tip to | ||||||
facilitate | ||||||
navigation. | Cook Wire | |||||
Guides consist | ||||||
of a Nitinol | ||||||
core for kink | ||||||
resistance | ||||||
coated with a | ||||||
polymer | ||||||
sleeve, and | ||||||
hydrophilically | ||||||
coated to | ||||||
reduce | ||||||
friction. | NiTi extractor | |||||
retrieval basket, | ||||||
interlaced and | ||||||
interweaved | ||||||
together wires, | ||||||
creating a | ||||||
special shape | ||||||
that will not | ||||||
allow urinary | ||||||
tract stones or | ||||||
other objects to | ||||||
fall out at the | ||||||
extraction stage | ||||||
as well as | ||||||
ensuring their | ||||||
easy capture, it | ||||||
is manually | ||||||
deployed by | ||||||
retracting the | The Stone | |||||
Cone Nitinol | ||||||
urological | ||||||
retrieval coil | ||||||
consists of a | ||||||
nitinol core | ||||||
wire with a | ||||||
PTFE coating. | ||||||
The wire | ||||||
assembly is | ||||||
housed in a | ||||||
sheath. The | ||||||
sheath is | ||||||
provided to | ||||||
straighten the | ||||||
device coil | ||||||
during | ||||||
placement and | ||||||
withdrawal; | The | |||||
ACCORDION | ||||||
Urological | ||||||
Occluding | ||||||
Guidewire | ||||||
consists of a | ||||||
film membrane | ||||||
pre-loaded | ||||||
within a two- | ||||||
part guidewire | ||||||
it is activated by | ||||||
a removable | ||||||
handle. The | ||||||
folded | ||||||
membrane acts | ||||||
to entrap stone | ||||||
fragments | ||||||
during | ||||||
lithotripsy. The | The XenX | |||||
device | ||||||
incorporates a | ||||||
nitinol core | ||||||
wire, a floppy, | ||||||
polyurethane- | ||||||
coated | ||||||
hydrophilic tip | ||||||
for easy stone | ||||||
passage and | ||||||
kink resistance | ||||||
with a very | ||||||
thin and | ||||||
flexible nitinol- | ||||||
based self- | ||||||
expandable | ||||||
braid/sieve | ||||||
structure as an | ||||||
integral part of | ||||||
the guide wire. | ||||||
Characteristic | Sensor | Roadrunner | NTrap | Stone cone | Accordion | XenX |
sheath. | manually | coated Nitinol | The braid is | |||
deployed by | tip is offered to | housed in a | ||||
retracting the | enable easy | sheath and is | ||||
cone out of | stone passage | manually | ||||
sheath. | deployed out | |||||
of the sheath. | ||||||
Mode of | ||||||
operation | Inserted into | Inserted into | Inserted into | Inserted in the | Inserted into | Inserted in the |
the ureter past | the ureter past | the ureter in its | ureter in its | the ureter in its | ureter in its | |
the stone to | the stone to | closed | closed | straightened | un-deployed | |
establish a | establish a | configuration; | configuration, | un-deployed | guidewire | |
tract and | tract and assist | once past the | once past the | film | mode, once | |
assist in the | in the | stone the | stone the | configuration | past the stone | |
placement, | placement | retention basket | retention cone | similar to a | the self | |
replacement | replacement | is deployed to | is deployed to | guidewire, once | expandable | |
and exchange | and exchange | prevent stone | prevent stone | past the stone | braid is | |
of devices | of devices | migration | migration | the multi fold | deployed, | |
during | during | during | during | film is formed. | braid conforms | |
urological | urological | lithotripsy; | lithotripsy; | The device's | to and fills the | |
procedures | procedures. | larger stone | ||||
fragments can | larger stone | film conforms | ureter to | |||
be swept into | fragments can | |||||
be swept into | to and fills the | prevent | ||||
the bladder | the bladder | ureter to | ||||
prevent | retrograde | |||||
migration of | ||||||
following | following | retrograde | stone | |||
treatment. | treatment. | migration of | fragments. | |||
stone | ||||||
fragments. | ||||||
Following | ||||||
fragmentation; | ||||||
The device can | ||||||
sweep larger | ||||||
stone fragments | ||||||
into the bladder | ||||||
Material made | Nitinol core | Nitinol core, | Nitinol core | Nitinol core | Nitinol core | Nitinol core |
wire PTFE | Platinum coil, | wire, | wire PTFE | wire, PTFE | wire | |
coat, | Hydrophilic | Nitinol woven | coated, | coated, | Nitinol braid | |
Polyurethane | coat, | mesh basket, | Polymeric | Hydrophilic tip, | mesh, | |
coat, | Polyurethane | Polyimide outer | sheath | Polymeric | Polyurethane | |
Hydrophilic | jacket | sheath | occlusion film | coat, Tungsten | ||
coating, | filled | |||||
Tungsten filled | radiopaque tip, | |||||
radiopaque tip | Hydrophilic | |||||
Nitinol braided | ||||||
mesh coating, | ||||||
Polyimide | ||||||
outer sheath | ||||||
Advancement | Advancement | Inserted past | Inserted past | Inserted past | Inserted while | |
Principal of | ||||||
operation | of the flexible | of the flexible | occluding stone | occluding | stone while | braid is |
tip past the | tip past the | while basket is | stone while | occluding film is | retracted | |
stone until | stone until | retracted within | cone is | in its straight | within outer | |
located at the | located at the | outer sheath, | retracted | configuration, | sheath, the | |
desired | desired | followed by | within outer | followed by | flexible tip is | |
basket | sheath, | folding the | advanced past | |||
anatomical | anatomical | |||||
Characteristic | Sensor | Roadrunner | NTrap | Stone cone | Accordion | XenX |
location | location | deployment | followed by | |||
cone | ||||||
deployment. | occluding film. | the stone until | ||||
the desired | ||||||
anatomical | ||||||
location is | ||||||
reached, | ||||||
followed by | ||||||
deployment of | ||||||
occluding | ||||||
braid. |
XenX™ - Page 5-4
3
Image /page/3/Picture/0 description: The image shows the text "K113692" on the top line. The second line of text says "page 4 of 5". The text is written in a handwritten style.
4
K113692
page 5 of 5
The XenX combines guide-wire and stone retention functionalities in one device. In its guidewire configuration it is substantially equivalent in its introduction, function, intended use and technology to the Sensor and Roadrunner guidewires and to the Accordion device. Once the self-expanding braid is deployed the XenX acts as a retention device and is substantially equivalent to the Stone-cone, Ntrap and Accordion devices in all characteristics excluding stone particle retrieval (sweeping functionality) which is not claimed as one of XenX features. This makes the XenX safer and less invasive in comparison with its predicate retention devices. Using the XenX, stone particles can be disintegrated by lithotripsy down to 1mm size, allowing them to easily pass through the urine system, hence the sweeping function becomes irrelevant.
Conclusion:
Xenolith Medical Ltd. believes that, based on the information provided in this submission, XenX™ is substantially equivalent to its predicate devices without raising any new safety and/or effectiveness issue.
5
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Ofer Zigman Head of Research and Development Xenolith Medical 1 Leshem, P.O Box 720 KIRYAT GAT 82000 ISRAEL
AUG 9 2012
Re: K113692
Trade/Device Name: XenXTM Regulation Number: 21 CFR8 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: OCY, FFL Dated: July 24, 2012 Received: July 30, 2012
Dear Mr. Zigman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
6
Page 2 -
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Benjamin K-tinka
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
7
INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K113692
Device Name: ХепХтм
Indications for Use:
The XenX device is intended to be used endoscopically to bypass and entrap calculi from the urinary tract, to prevent retrograde migration of calculi during laser lithotripsy,and to facilitate the placement of endourological instruments during diagnostic or interventional procedures
Prescription Use _X (Per 21 CFR 801 Subpart D) AND/OR
Over the Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRHAffice of Device Evaluation (ODE)
Helent Leeman
ivision Sign-10(k) Num
XenX™ - Page 4:2