K Number
K113375
Manufacturer
Date Cleared
2013-01-25

(436 days)

Product Code
Regulation Number
882.1320
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For muscle reeducation by application of external stimulation to the muscles necessary for pharyngeal contraction.

Device Description

The devices are single patient, multiple-use [up to three times] cutaneous electrodes for the application of electrical stimulation. The lead wire is attached to the electrode and is designed to be connected to a transcutaneous external muscle stimulator. The cutaneous electrodes are available in two sizes; one for adult patients and one for youth patients. Both sizes are composed of identical materials. The cutaneous electrodes and lead wire do not contain active electronics, software or firmware. The lead wire connects to the cutaneous electrotherapy device. The cutaneous electrodes are composed of materials commonly used in this application: polyethylene, medical grade adhesives, 'spunlace' non-woven, carbon film, silver, polyester, and hydrogel.

AI/ML Overview

This document is a 510(k) summary for the eSWALLOW USA Electrodes and Leadwire, seeking substantial equivalence to a predicate device. The information provided heavily focuses on bench testing and comparison to the predicate, and explicitly states that no clinical studies were required. Therefore, many of the requested criteria related to clinical performance, ground truth, and expert evaluation cannot be fully addressed from the provided text.

Here's an attempt to answer based on the available information:

1. Table of Acceptance Criteria and Reported Device Performance

Since this is a 510(k) submission primarily relying on substantial equivalence to a predicate device and bench testing, "acceptance criteria" here refers to the standards met and characteristics matching the predicate.

Acceptance Criteria (based on predicate equivalence and standards)Reported Device Performance
BiocompatibilityDemonstrated per ISO-10993 for cytotoxicity, primary skin irritation, and delayed contact hypersensitivity, per FDA G-95 requirements for skin-contact devices of limited duration use. Material components (polyethylene, medical grade adhesives, spunlace, carbon film, silver, polyester, hydrogel) are commonly used and deemed biocompatible.
Electrical Performance (Impedance & Dispersion)Tested per ANSI/AAMI EC12:2000/[R] 2010, Disposable ECG Electrodes, for impedance and dispersion testing. (Specific numerical results are not provided, only that testing was done to the standard).
Intended UseSame as predicate: For muscle reeducation by application of external stimulation to the muscles necessary for pharyngeal contraction. Device description states intent for application of electrical current to patient skin.
Indications for UseSame as predicate: For muscle reeducation by application of external stimulation to the muscles necessary for pharyngeal contraction.
Technological CharacteristicsSimilar to predicate: Single-patient, multiple-use, self-adhering, disposable electrodes and lead wires. To be used with NMES Device. External, anterior pharyngeal region body site. Electrotherapy mode of action. Non-sterile. Similar electrode sizes (Adult: 0.85 inch round electrode; pediatric: 0.688 inch round electrode vs. Predicate Adult: 0.875 inch round; pediatric: 0.6875 inch round). Adhesive butterfly shape, round lead wires with snap connectors. Packaging: sealed pouch. No active electronics, software, or firmware.
Safety (implied by material and design similarity)"The electrodes and lead wires are made of the same biocompatible materials and have similar technological characteristics as previously cleared devices and therefore raise no new issues of safety."

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: No clinical test set. The study primarily relies on bench testing for material biocompatibility and electrical performance, and comparison to a predicate device. The document does not specify a "sample size" for the bench tests in terms of number of electrodes or leadwires tested, beyond stating that testing was "done by the contract manufacturer."
  • Data Provenance: N/A for a clinical test set. Bench testing was performed by a contract manufacturer.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • N/A. No clinical test set or human-in-the-loop performance evaluation was conducted that would require expert ground truth.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • N/A. No clinical test set requiring adjudication was performed.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. This device (electrodes and leadwires) does not involve "AI assistance" or "human readers" in the context of diagnostic interpretation. It is a physical medical device for electrical stimulation delivery.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • N/A. This device is not an algorithm. Bench testing of the physical components (impedance, dispersion) could be considered "standalone" evaluation of the device's electrical characteristics, but not in the conventional sense of an algorithm's performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Bench Testing: The "ground truth" for the bench tests would be the specifications and requirements defined by the referenced standards (ISO-10993 for biocompatibility, ANSI/AAMI EC12:2000/[R] 2010 for electrical performance). The results of these tests (e.g., demonstrating no cytotoxicity, meeting impedance limits) are compared against these standards.
  • Substantial Equivalence: The primary "ground truth" for the 510(k) submission's approval is the established safety and effectiveness of the identified predicate device (K083756, SpectraMed, Inc. Guardian 150 electrode). The new device's characteristics are compared directly to this cleared device.

8. The sample size for the training set:

  • N/A. This device is not an AI/ML algorithm that requires a "training set."

9. How the ground truth for the training set was established:

  • N/A. This device is not an AI/ML algorithm that requires a "training set."

{0}------------------------------------------------

K113375

Image /page/0/Picture/1 description: The image shows the word "eSWALLOW" in a stylized font. The "e" is in lowercase, while the rest of the word is in uppercase. There is a registered trademark symbol after the word "SWALLOW", and the word "USA" is written after the trademark symbol.

JAN 2 5 2013

ﺮ. ۱۰ ﺑﻪ ۱۰

510(k) Summary

flying

This summary of information is being submitted in accordance with the requirements of SMDA 1990 and CFR 807.92.

510|k] Owner's Name eSWALLOW USA, LLC

Address:

106 Hidden Drive Scottsboro, Alabama 35769

William R. Ingram, Managing Principal

Phone: Toll Free: Fax: Contact:

Summary Date:

January 18, 2013

256-571-0443

866-964-3102

256-571-7539

eSWALLOW USA, LLC

{1}------------------------------------------------

. さら『『Trade Name:

eSWALLOW USA Electrodes and Leadwire Common Name: Cutaneous electrode; leadwire for connection of cutaneous electrode Classification Name: 21 CFR 882.1320, Cutaneous Electrodes

Classification: Class II .

Legally marketed device to which equivalence is claimed [Predicate Device]: K083756, SpectraMed, Inc. Guardian 150 electrode

Description of Electrodes / Leadwire

The devices are single patient, multiple-use [up to three times] cutaneous electrodes for the application of electrical stimulation. The lead wire is attached to the electrode and is designed to be connected to a transcutaneous external muscle stimulator. The cutaneous electrodes are available in two sizes; one for adult patients and one for youth patients. Both sizes are composed of identical materials.

The cutaneous electrodes and lead wire do not contain active electronics, software or firmware. The lead wire connects to the cutaneous electrotherapy device.

The cutaneous electrodes are composed of materials commonly used in this application: polyethylene, medical grade adhesives, 'spunlace' non-woven, carbon film, silver, polyester, and hydrogel.

Indications for Use

The electrodes and lead wires are intended for muscle reeducation by application of external stimulation to the muscles necessary for pharvngeal contraction.

Comparison to Predicate Device

Many devices for external, transcutaneous electrotherapy have been cleared by FDA for multiple intended uses. In this submission only the electrodes and lead wires to be used with a previously cleared NMES device are described.

ParameterDevicePredicate Device
Device NameeSWALLOW USA Electrodes andLeadwireGuardian 150 Electrode
ManufacturerPepin Manufacturing, Inc.SpectraMed [originally cleared bySelectiveMed Components, Inc.]
510[k] #K113375K083756
ClassIISame
21 CFR number882.1320Same
CodeGXYSame
Product TypeCutaneous electrodes / lead wiresSame

Summary Comparison Table of New Device to Predicate Device

{2}------------------------------------------------

ParameterDevicePredicate Device
DescriptionSingle-patient, multiple-use, self-adhering, disposable electrodes and lead wiresSame
To be used withNMES DeviceSame
Intended UseFor application of electrical current to patient skinSame
Indications for UseFor muscle reeducation by application of external stimulation to the muscles necessary for pharyngeal contractionSame
End User; settingHealthcare personnel; office setting and home useSame
Body SiteExternal, anterior pharyngeal regionSame
Mode of ActionElectrotherapySame
Standards metANSI/AAMI EC12:2000/[R] 2010Same
Pt-contacting MaterialsBiocompatible per ISO 10993Same
Component materialsPolyethylene, medical grade adhesives, spunlace, carbon film, silver, polyester, and hydrogelSilver / silver chloride carbon conductor; proprietary blue hydrogel, medical grade adhesives
SterilityNon-sterileSame
Electrode size, shapeAdult: 0.85 inch round electrode; pediatric: 0.688 inch round electrode; adhesive butterfly shape; round lead wires with snap connectorsAdult: 0.875 inch round electrode; pediatric: 0.6875 inch round electrode; adhesive butterfly shape; round lead wires with snap connectors
PackagingSealed pouchSame

Bench Testing

Biocompatibility of the patient-contacting materials was demonstrated by the source manufacturer per ISO-10993 for cytotoxicity, primary skin irritation, and delayed contact hypersensitivity, per FDA G-95 requirements for skin-contact devices of limited duration use.

Testing of the electrodes and lead wires was done by the contract manufacturer per ANSI/AAMI EC12:2000/[R] 2010, Disposable ECG Electrodes, for impedance and dispersion testing.

Clinical Testing

{3}------------------------------------------------

No clinical studies were required for electrodes and lead wires. The electrotherapy device with which these electrodes and lead wires are used was previously cleared under K092202.

Device Comparison Statement

The eSWALLOW electrodes and lead wire have the same intended use and indications for use as the predicate device. The electrodes and lead wires are made of the same biocompatible materials and have similar technological characteristics as previously cleared devices and therefore raise no new issues of safety. Based on the assessment of performance data and the characteristic comparison, the eSWALLOW electrodes and lead wire are substantially equivalent to the legally marketed predicate device.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the caduceus.

Public Health Service

January 25.2013

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

eSwallow™ USA, LLC c/o Mr. William R. Ingram Managing Principal 106 Hidden Drive Scottsboro, Alabama 35769

Re: K113375

Trade/Device Name: eSwallow USA Electrodes and Lead Wires Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: November 1, 2012 Received: December 3, 2012

Dear Mr. Ingram:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{5}------------------------------------------------

Page 2 - Mr. William R, Ingram

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Joyce M. Whang

Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices

Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): K113375

Device Name: _ eSWALLOW USA Electrode and Lead Wires

Indications For Use:

For muscle reeducation by application of external stimulation to the muscles necessary for pharyngeal contraction.

Prescription Use __ × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Jovce M. W

Division Sign Off) Division of Neurological and Physical Medicine Devices (DNPMD)

510(k) Number K113375

Page 1 of 1

§ 882.1320 Cutaneous electrode.

(a)
Identification. A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.(b)
Classification. Class II (performance standards).