K Number
K113157
Manufacturer
Date Cleared
2012-03-13

(139 days)

Product Code
Regulation Number
888.3070
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Internal fixation implants are load-sharing devices intended to stabilize and maintain alignment until normal healing occurs. Implants are not intended to replace normal body structures or bear the weight of the body in the presence of incomplete bone healing.

The Aileron Interspinous Fixation System is a posterior, non-pedicle supplemental fixation device, intended for use in the non-cervical spine (T1-S1). It is intended for plate fixation/attachment to the spinous processes for the purpose of achieving, in conjunction with autogenous bone graft, single level supplemental fusion in the following conditions: (1) degenerative disc disease (is defined as back pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies), (2) trauma (i.e., fracture or dislocation), (3) spinal tumor. (4) spondylolisthesis. The Aileron Interspinous Fixation System is not intended for stand-alone use.

The ARX® Spinal System, PILOT® Spinal System, PILOT Posterior Lumbar Plating System, and CONOUEST® Spinal System, when properly used, are intended for posterior pedicle screw fixation of the non-cervical posterior spine in skeletally mature patients. They provide stabilization and immobilization of spinal segments as an adjunct to fusion.

When used as posterior spine thoracic/lumbar systems, the ARX Spinal System, PILOT Spinal System, PILOT Posterior Lumbar Plating System, and CONQUEST Spinal System are indicated for one or more of the following: (1) degenerative disc disease (is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). (2) trauma (i.e. fracture or dislocation), (3) curvatures (scoliosis, kyphosis, and/or lordosis). (4) spinal tumor. (5) failed previous fusion. (6) pseudoarthrosis, (7) spinal stenosis, (8) spondvlolisthesis.

Device Description

The Aileron Interspinous Fixation System is a posterior, non-pedicle supplemental fixation device, intended for use in the non-cervical spine (T1-S1). Implants are manufactured from titanium alloy per ASTM F136 and are available in a range of sizes to suit the individual pathology and anatomical conditions of the patient.

AI/ML Overview

The provided text describes a medical device, the Aileron Interspinous Fixation System, and states that its substantial equivalency to predicate devices was demonstrated through performance data. However, it does not contain information about acceptance criteria or a study that specifically proves the device meets those criteria, as typically understood in the context of clinical or diagnostic performance. Instead, the "Performance Data" section details mechanical testing to demonstrate substantial equivalency.

Therefore, many of the requested fields cannot be filled from the provided text.

Here's a breakdown of what can be extracted and what cannot:

1. Table of acceptance criteria and reported device performance:

Acceptance Criteria (Mechanical Testing)Reported Device Performance (Mechanical Testing)
Static Compression (per ASTM F1717)Presented to demonstrate substantial equivalency
Dynamic Compression (per ASTM F1717)Presented to demonstrate substantial equivalency
Static Axial Grip (per ASTM F1798)Presented to demonstrate substantial equivalency

Note: The document states "subsequent engineering analysis was presented," implying the device met the requirements/comparisons set by these ASTM standards to demonstrate substantial equivalency with predicate devices. Specific quantitative acceptance criteria or raw performance values are not provided.

2. Sample size used for the test set and the data provenance:

  • Sample size for test set: Not applicable or specified for mechanical testing. This typically refers to patient data, which is not part of this "Performance Data" section.
  • Data provenance: Not applicable. Mechanical testing is performed in a lab setting, not on patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. Ground truth, in this context, usually refers to clinical diagnostic accuracy, which is not assessed here. The "ground truth" for mechanical testing is adherence to ASTM standards and comparison to predicate devices, which would be evaluated by engineers.

4. Adjudication method for the test set:

  • Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, an MRMC study was not done. This study solely focuses on the mechanical performance of a spinal implant, not on AI-assisted diagnostic capabilities.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is not an AI algorithm.

7. The type of ground truth used:

  • For the mechanical testing, the "ground truth" implicitly refers to the established standards of ASTM F1717 and ASTM F1798, and the performance characteristics of the predicate devices for comparison.

8. The sample size for the training set:

  • Not applicable. This is not an AI model requiring a training set.

9. How the ground truth for the training set was established:

  • Not applicable.

{0}------------------------------------------------

K113157
page 1 of 2

510(k) Summary Aileron Interspinous Fixation System

MAR 1 3 2012

Submitted By:Life Spine, Inc.2401 W. Hassell Road, Suite 1535Hoffman Estates, IL 60169Telephone: 847-884-6117Fax: 847-884-6118
510(k) Contact:Randy LewisLife Spine2401 W. Hassell Road, Suite 1535Hoffman Estates, IL 60169Telephone: 847-884-6117Fax: 847-884-6118
Date Prepared:December 5th, 2011
Trade Name:Aileron Interspinous Fixation System
Common Name:Interspinous Process Fixation System
Classification:MNH, 888.3070, Class II, Spondylolisthesis SpinalFixation Device SystemMNI, 888.3070, Class II, Pedicle Screw Spinal SystemKWP, 888.3050, Class II, Spinal Interlaminal Fixation Orthosis
Predicate Device:Life Spine Interspinous Fixation System (K100407), Lanx Spinal FixationSystem (K090252) and the CD HORIZON Spinous Process Plate(K032037)

Device Description:

The Aileron Interspinous Fixation System is a posterior, non-pedicle supplemental fixation device, intended for use in the non-cervical spine (T1-S1). Implants are manufactured from titanium alloy per ASTM F136 and are available in a range of sizes to suit the individual pathology and anatomical conditions of the patient.

{1}------------------------------------------------

Intended Use of the Device:

Internal fixation implants are load-sharing devices intended to stabilize and maintain alignment until normal healing occurs. Implants are not intended to replace normal body structures or bear the weight of the body in the presence of incomplete bone healing.

K113157

ige 2 of 2

The Aileron Interspinous Fixation System is a posterior, non-pedicle supplemental fixation device, intended for use in the non-cervical spine (TI-S1). It is intended for plate fixation/attachment to the spinous processes for the purpose of achieving, in conjunction with autogenous bone graft, single level supplemental fusion in the following conditions: (1) degenerative disc disease (is defined as back pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies), (2) trauma (i.e., fracture or dislocation), (3) spinal tumor. (4) spondylolisthesis. The Aileron Interspinous Fixation System is not intended for stand-alone use.

The ARX® Spinal System, PILOT® Spinal System, PILOT Posterior Lumbar Plating System, and CONOUEST® Spinal System, when properly used, are intended for posterior pedicle screw fixation of the non-cervical posterior spine in skeletally mature patients. They provide stabilization and immobilization of spinal segments as an adjunct to fusion.

When used as posterior spine thoracic/lumbar systems, the ARX Spinal System, PILOT Spinal System, PILOT Posterior Lumbar Plating System, and CONQUEST Spinal System are indicated for one or more of the following: (1) degenerative disc disease (is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). (2) trauma (i.e. fracture or dislocation), (3) curvatures (scoliosis, kyphosis, and/or lordosis). (4) spinal tumor. (5) failed previous fusion. (6) pseudoarthrosis, (7) spinal stenosis, (8) spondvlolisthesis.

Technological Characteristics:

The Aileron Interspinous Fixation System is substantially equivalent to the predicate systems in terms of design, materials, and indications for use.

Material:

The Aileron Interspinous Fixation System is 6AL-4V-ELI titanium manufactured according to ASTM F136. The device is comprised of a variety of non-sterile titanium, single use components.

Performance Data:

Static and Dynamic compression testing per ASTM F1717, Static Axial Grip testing per ASTM F1798, and subsequent engineering analysis was presented to demonstrate the substantial equivalency of the Aileron Interspinous Fixation System.

Conclusion:

The information presented demonstrates the substantial equivalency of the Aileron Interspinous Fixation System to the predicate devices.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of stylized lines.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

MAR 1 3 2012

Life Spine, Inc. % Mr. Randy Lewis RA/QA Manager 2401 W. Hassell Road, Suite 1535 Hoffman Estates, Illinois 60169

Re: K113157

Trade/Device Name: Aileron Interspinous Fixation System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: II Product Code: KWP, MNI, MNH Dated: February 20, 2012 Received: February 22, 2012

Dear Mr. Lewis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

Page 2 - Mr. Randy Lewis

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours

For Roth Asler-Dir.

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

KII3157 510(k) number (if known):

Device Name: Aileron Interspinous Fixation System

Internal fixation implants are load-sharing devices intended to stabilize and maintain alignment until normal healing occurs. Implants are not intended to replace normal body structures or bear the weight of the body in the presence of incomplete bone healing.

The Aileron Interspinous Fixation System is a posterior, non-pedicle supplemental fixation device, intended for use in the non-cervical spine (T1-S1). It is intended for plate fixation/attachment to the spinous processes for the purpose of achieving, in conjunction with autogenous bone graft, single level supplemental fusion in the following conditions: (1) degenerative disc disease (is defined as back pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies), (2) trauma (i.e., fracture or dislocation), (3) spinal tumor, (4) spondylolisthesis. The Aileron Interspinous Fixation System is not intended for stand-alone use.

The ARX® Spinal System. PILOT® Spinal System, PILOT Posterior Lumbar Plating System, and CONQUEST® Spinal System, when properly used, are intended for posterior pedicle screw fixation of the non-cervical posterior spine in skeletally mature patients. They provide stabilization and immobilization of spinal segments as an adjunct to fusion.

When used as posterior spine thoracic/lumbar systems, the ARX Spinal System, PILOT Spinal System, PILOT Posterior Lumbar Plating System, and CONQUEST Spinal System are indicated for one or more of the following: (1) degenerative disc disease (is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), (2) trauma (i.e. fracture or dislocation), (3) curvatures (scollosis, kyphosis, and/or lordosis), (4) spinal tumor, (5) failed previous fusion, (6) pseudoarthrosis, (7) spinal stenosis, (8) spondylolisthesis.

Prescription Use _ x __ (Part 21 CFR 801 Subpart D) . And/Or ,

Over-the-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

(Division Sign-Off)

Livision of Surgical, Orthopedie, and Restorative Devices

K113157 510(k) Number_

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.