(14 days)
ORTHOLOC™ Bone Screws are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.
The ORTHOLOC™ Bone Screws subject to this premarket notification include the nonlocking 3.5mm low profile cortical bone screws and non-locking 3.5 and 4.0 cortical bone screws.
These screws are manufactured from titanium alloy and have a solid core. The implants are single use only devices.
This document is a 510(k) Pre-market Notification for the ORTHOLOC™ Bone Screws. It describes mechanical testing performed to demonstrate substantial equivalence to predicate devices, rather than a study on algorithm performance or clinical utility in the context of medical imaging or AI.
Therefore, many of the requested elements for describing acceptance criteria and a study to prove a device meets acceptance criteria (especially those related to AI, ground truth, experts, and reader studies) are not applicable to this submission.
However, I can extract the relevant information regarding the non-clinical evidence provided:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Perform at least as well as legally marketed predicate devices in terms of mechanical strength (specifically pull-out force) | "The results show that the subject ORTHQLOC™ Bone Screws can be expected to perform at least as well as the legally marketed predicate ORTHOLOCTM Bone Screws and Wright Compression Screws." (Specifically from worst-case pull out testing) |
| Be manufactured from biocompatible material (Titanium alloy) | Device is manufactured from titanium alloy. |
| Maintain the same indications for use, diameter, and size range as predicates | "The subject screws are identical in indication for use, diameter, size range and material to the predicates..." |
| No new types of safety and effectiveness questions | "...differ only in screw neck radii. No new types of safety and effectiveness questions can be expected." |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: Not explicitly stated as a number of screws. The term "worst-case pull out testing" implies a representative sample was tested, but the exact count is not provided.
- Data Provenance: Not applicable. This is not clinical data but mechanical testing performed by the manufacturer. The location of the testing laboratory is not specified but would typically be internal to the manufacturer or a third-party testing facility. It is a prospective test in the sense that the new device was manufactured and then tested.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. This submission does not involve clinical data or medical images requiring expert interpretation or ground truth establishment in that context. The "ground truth" here is the physical mechanical strength measured in a laboratory setting.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. No human adjudication method is relevant for mechanical pull-out testing. The "adjudication" is based on comparing measured force values against predicate device performance or established standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This is not a study involving human readers or AI.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This is not an algorithm or AI device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The "ground truth" for the non-clinical testing is the measured mechanical strength (specifically pull-out force) of the screws, as determined through standardized laboratory testing procedures. The comparison is made against the measured mechanical performance of legally marketed predicate devices.
8. The sample size for the training set
- Not Applicable. This is not an AI/machine learning device requiring a training set.
9. How the ground truth for the training set was established
- Not Applicable. This is not an AI/machine learning device requiring a training set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "WRIGHT." in a bold, sans-serif font. Below the word is a stylized graphic consisting of three angled shapes. At the top of the image, there is a handwritten number "K112772(1/2)".
510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
OCT - 7 2011
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the URTHOLOC™ Bone Screws.
| A.1. Submitted By: | Wright Medical Technology, Inc.5677 Airline RdArlington, TN 38002 |
|---|---|
| Date: | August 29, 2011 |
| Contact Person: | Kellen HillsRegulatory Affairs Specialist(901) 290-5816 |
| A.2. Proprietary Name: | ORTHOLOC™ Bone Screws |
| Common Name: | Bone Screw |
| Device Classification Regulation: | 21 CFR 888.3040-Class II |
| Device Product Code & Panel: | HWC: Screw, Fixation Bone87 Orthopedics |
| A.3. Predicate Device: | K102429-ORTHOLOC™ 3Di AnkleSystem and ORTHOLOC™ Bone Screws |
K082320-WrightTM Compression Screws
A.4. Device Description
The ORTHOLOC™ Bone Screws subject to this premarket notification include the nonlocking 3.5mm low profile cortical bone screws and non-locking 3.5 and 4.0 cortical bone screws.
These screws are manufactured from titanium alloy and have a solid core. The implants are single use only devices.
A.5. Intended Use
ORTHOLOC™ Bone Screws are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.
Wright Medical Technology, Inc. 5677 Airline Road Arlington, TN 38002 901.867.9971 phone
. ・ ...
Plating
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "WRIGHT." in a bold, sans-serif font. Below the word is a graphic design of three horizontal lines that are stacked on top of each other. Above the word "WRIGHT." is the text "K 112772 (2/2)".
A.6. Technological Characteristics Comparison
The subject ORTHOLOC™ Bone Screws are identical to the predicate ORTHOLOC™ Bone Screws with the exception of the radii at the transition from the screw head to the screw shaft. This change resulted in approximately one thread being removed from the distal portion of the screw.
B.1. Substantial Equivalence - Non-Clinical Evidence
Substantial equivalence is shown through worst-case pull out testing, materials information, and comparison of design characteristics. The results show that the subject ORTHQLOC™ Bone Screws can be expected to perform at least as well as the legally marketed predicate ORTHOLOCTM Bone Screws and Wright Compression Screws.
B.2. Substantial Equivalence - Clinical Evidence
N/A
B.3. Substantial Equivalence - Conclusions
Substantial equivalence is shown through worst-case pull out testing, materials information, and comparison of design characteristics. The subject screws are identical in indication for use, diameter, size range and material to the predicates, and differ only in screw neck radii. No new types of safety and effectiveness questions can be expected. From the evidence given in the Premarket Notification, the subject devices can be expected to perform at least as well as the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three parallel lines forming the wings of the serpent. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Wright Medical Technology, Inc % Kellen Hill Regulatory Affairs Specialist 5677 Airline Road Arlington, Tennessee 38002
OCT - 7 2011
Re: K112772
Trade/Device Name: ORTHOLOC"™ Bone Screws Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class H Product Code: HWC Dated: September 12, 2011 Received: September 23, 2011
Dear Kellen Hill:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and lisuing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 – Kellen Hill
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (QS) regulation (SF OF to Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Parl 801), please If you desire specific advice for your cevree on var acconsideratives/ucm l 1589.html for go to http://www.fda.gov/Aboutr1787Ccmcrs0fffcs) Office of Connyliance. Also, please
the Center for Devices and Radiological Health's (CDRH's) workst natification" (21CFF) P the Center for Devices and Radionogical Health Steans to premarket notification" (21CFR Part note the regulation entitled, "Misolanding of reference to promised in MDR regulation (21 CFR Part 803), please go to
CFR Part 803), please go to Illip://www.ida.2017/01/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11/11
You may obtain other general information on your responsibilities under the Act from the You may obtain oner general mornation on your responsion.
Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address (800) 038-2041 of (2017) 720 Devices/Resourcesfor You/Industry/default.h1m.
Sincerely yours,
Per
E. Melkerson
Mark N. Melkers Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K || 2772 (11)
Indications for Use
510(k) Number (if known):
Device Name: ORTHOLOC™ Bone Screws
Indications For Use:
ORTHOLOC™ Bone Screws are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mutheal Threis for mxm
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K112772
1 of 1
.
N/A