A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Classified by FDA's General Hospital and Personal Use Device panel as Class I, 21 CFR 880.6250, Power Free Vinyl Patient Examination Glove, 80 LYZ, and meets all requirements of ASTM standard D-5250-06e1.
This 510(k) submission describes the non-clinical testing performed for the Clean Medical Manufacturing Co., Ltd. Powder Free Vinyl Patient Examination Gloves to demonstrate substantial equivalence to a predicate device.
Here's an analysis of the provided information:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Standard/Test) | Reported Device Performance |
|---|---|
| Physical and Dimensions Testing | Meets requirements based on ASTM D-5250-06e1. |
| Pinholes (Water Fill Test) | Meets requirements with AQL 2.5, inspection level I (FDA 1000 ml. Water Fill Test). |
| Primary Skin Irritation | No primary skin irritant reactions. |
| Skin Sensitization | No sensitization reactions (allergic contact dermatitis). |
| Residual Powder | Meets "powder-free" claims (contains no more than 2 mg powder per glove) based on ASTM D-6124-06. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state a single distinct "test set" sample size with specific data provenance for the overall device evaluation. Instead, it refers to various non-clinical tests with their own sampling plans:
- Physical and Dimensions Testing: "Inspection level S-2, AQL 4.0." This is a sampling plan, implying a sample size was drawn based on AQL (Acceptable Quality Limit) principles for manufacturing quality control. Specific sample numbers are not provided.
- Water Fill Test (Pinholes): "Samplings of AQL 2.5, inspection level I." Again, a sampling plan based on AQL, not a fixed sample size.
- Primary Skin Irritation, Skin Sensitization, Residual Powder Tests: No specific sample sizes mentioned for these tests, only that they were "conducted with results showing..." or "conducted to insure...".
Data Provenance: The data appears to be prospective as it involves the manufacturer (Clean Medical Manufacturing Co., Ltd.) performing tests on their own gloves during production and for the 510(k) submission. The country of origin for the manufacturing and presumably the testing is China (Shijiazhuang, Hebei).
3. Number of Experts Used to Establish Ground Truth and Qualifications
Not Applicable. This document describes the evaluation of a medical device (patient examination gloves) through non-clinical laboratory testing and adherence to established standards (ASTM, FDA regulations). There is no mention of "experts" being used to establish a "ground truth" in the way one would for diagnostic imaging or clinical assessment. The ground truth is defined by the technical specifications and performance metrics outlined in the standards themselves.
4. Adjudication Method for the Test Set
Not Applicable. As no expert ground truth or subjective assessment is described, there is no adjudication method involved. The tests are objective and defined by measurement against established standards.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. This type of study (MRMC) is typically performed for diagnostic devices where human readers interpret results (e.g., radiologists reading images) and the AI's impact on their performance is evaluated. This submission is for patient examination gloves, which do not involve human interpretation in this manner.
6. Standalone Performance Study (Algorithm Only)
Yes, in spirit, but not an "algorithm." The testing described is standalone performance of the physical device itself. The "device" (the glove) is tested in isolation against established performance criteria. There is no AI algorithm involved in the function or evaluation of these gloves. The performance data presented in the table (physical properties, pinhole rate, biocompatibility, powder content) reflects the standalone performance of the glove.
7. Type of Ground Truth Used
The ground truth used is primarily based on established industry standards and regulatory requirements. Specifically:
- ASTM D-5250-06e1: For physical and dimensions testing.
- FDA 1000 ml. Water Fill Test: For pinhole detection.
- ASTM D-6124-06: For residual powder testing.
- Biocompatibility testing guidelines: For primary skin irritation and skin sensitization.
These standards define the acceptable criteria for the glove's performance and safety characteristics.
8. Sample Size for the Training Set
Not Applicable. This submission is for physical medical devices (gloves) and does not involve an AI algorithm or machine learning model that would require a "training set." The manufacturing process for gloves is typically validated through quality control measures, not machine learning training.
9. How the Ground Truth for the Training Set Was Established
Not Applicable. As there is no training set, there is no ground truth established for one. The "ground truth" relevant to this submission refers to the performance specifications defined by regulatory bodies and consensus standards against which the manufactured gloves are tested to ensure quality and safety.
{0}------------------------------------------------
EXHIBIT #1 Page 1 of 2
510(K) SUMMARY
This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR §807.92.
K112743 The assigned 510(K) number is:
1. Owner's Identification:
Mr. Wei Zhang Clean Medical Manufacturing Co., Ltd. No. 35 Fengshou Xi Road Shijiazhuang, Hebei China 050000 Date Summary Prepared: June 4, 2010 Tel: 86-311-83980319
2. Name of the Device:
Clean Medical Manufacturing Co., Ltd. Powder Free Vinyl Patient Examination Gloves
3. Predicate Device Information:
Better Care Plastic Products Co., Ltd Powder Free Vinyl Patient Examination Glove
4. Device Description:
Classified by FDA's General Hospital and Personal Use Device panel as Class I, 21 CFR 880.6250, Power Free Vinyl Patient Examination Glove, 80 LYZ, and meets all requirements of ASTM standard D-5250-06e1.
5. Intended Use:
A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
6. Comparison to Predicate Devices:
Clean Medical Manufacturing Co., Ltd.'s Powder Free Vinyl Patient Examination Gloves is substantially equivalent in safety and effectiveness to the Better Care Plastic Products Co., Ltd's Powder Free Vinyl Patient Examination Glove.
{1}------------------------------------------------
7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as Follows:
The standards used for Clean Medical Manufacturing Co., Ltd.'s glove production are based on ASTM D-5250-06e1. All testing meets requirements for physical and dimensions testing conducted on gloves. Inspection level S-2, AQL 4.0.
The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, inspection level I, meeting these requirements.
Primary Skin Irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.
A Residual Powder Test that based on ASTM D-6124-06 for powder at finished inspection is conducted to insure that our gloves meet our "powder-free" claims (contain no more than 2 mg powder per glove).
8. Labeling:
There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels.
9. Discussion of Clinical Tests Performed:
Not Applicable - There is no hypoallergenic Claim.
10. Conclusions:
Clean Medical Manufacturing Co., Ltd.'s Powder Free Vinyl Patient Examination - Cloud Mcdical Manattaring Cong E.a.661 standard as well as applicable 21 CFR Offerences, and meets pinhole FDA requirements, biocompatibility requirements and Icreleneces, and moots pinkers ata in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited. The device herein mentioned is as safe, as effective, and performs as well as or better than the predicate device.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized emblem. The emblem consists of three overlapping, stylized bird-like shapes, possibly representing the department's mission related to health and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002
Clean Medical Manufacturing Company, Limited C/O Mr. Aaron Richard Clean Medical Supplies, Incorporated 18633 E. Gale Avenue, #105 City of Industry, California 91710
DEC 1 6 2011
Re: K112743
Trade/Device Name: Powder Free Vinyl Patient Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: November 16, 2011 Received: November 25, 2011
Dear Mr. Richard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Richard
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies, You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.hum.
Sincerely yours.
Anthony D. Nwa
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Attachment A
INDICATION FOR USE
k 112743 510 (k) NUMBER (IF KNOW):
Clean Medical Manufacturing Co., Ltd.
DEVICE NAME:
APPLICANT:
Powder Free Vinyl Patient Examination Gloves_
INDICATIONS FOR USE:
A patient examination glove is disposable device intended for medical purpose that is A patient examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use (Part 21 CFR 801 Subpart D) AND/ OR
Over-The-Counter-Use _ (21CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrent of CDRH, Office of Devi
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: 《112743
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.