K Number
K112644
Manufacturer
Date Cleared
2012-12-19

(464 days)

Product Code
Regulation Number
884.6160
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Seaforia™ Sperm Separation Assistance System Disposable Kit is intended for preparing motile sperm using the swim-up (SU) separation method and holding sperm for use in assisted reproductive procedures such as intrauterine insemination (IUI), intracervical insemination (ICI) and in-vitro fertilization (IVF). The Seaforia™ Sperm Separation Assistance System Disposable Kit is indicated for incubation periods not exceeding one (1) hour.

The Seaforia ™ Sperm Separation Assistance System Heating Device is indicated for use only with the Seaforia ™ Sperm Separation Assistance System Disposable Kit to maintain the temperature (37°C ±0.5) of sperm during the swim-up (SU) separation procedure.

Device Description

The SeaforiaTM Sperm Separation Assistance System separates motile/progressively motile sperm cells from debris and non-motile sperm cells from the original semen sample for further use in assisted reproductive procedures, including intrauterine insemination (IUI), intra-cervical insemination (ICI) and in-vitro fertilization (IVF). The system provides conditions such as temperature 37°C ±0.5 and effective surface area for the separation of motile human sperm cells. Note: The separation media is not included with the Seaforia" Sperm Separation Assistance System. LOTUS BIO® requires that the media used with the Seaforia'™ Sperm Separation Assistance System be a media cleared for use in the United States that is indicated for sperm handling, washing, or swim-up procedures with a formulation compatible with use outside of a CO2 environment.

The system is composed of the following units:

  • . Seaforia™ Sperm Separation Assistance System Disposable Kit - includes designated containers (separators) for the separation media available in various volumes and labware accessories such as syringes, polypropylene tips, pipette and vials. Five versions of the kit will be available to accommodate sperm samples ranging from 0.5-6 ml (Versions 0.5-2 ml. 2.5-3.0 ml. 3.5-4.0 ml, 4.5-5.0 ml, and 5.5-6.0 ml).
  • Seaforia™ Sperm Separation Assistance System Heating Device capable . of maintaining the sperm sample at the required temperature (37℃ ±0.5).

The sperm separation procedure takes place within the separator and is based on the original swim-up technique. Media is layered over the liquefied semen sample and during a subsequent incubation period of 30 minutes in the Seaforia™ Sperm Separation Assistance System Heating Device, the motile sperm cells migrate from the semen layer into the media.

AI/ML Overview

Acceptance Criteria and Study for the Seaforia™ Sperm Separation Assistance System

This response outlines the acceptance criteria and the study conducted to demonstrate the performance of the Seaforia™ Sperm Separation Assistance System, based on the provided 510(k) summary.


1. Table of Acceptance Criteria and Reported Device Performance

The provided document describes performance testing for both the disposable kit and the heating device components of the Seaforia™ Sperm Separation Assistance System.

Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
Disposable Kit
HSSA (Human Sperm Survival Assay)≥ 70% motility following 1 hour of exposureHSSA ≥ 70% motility following 1 hour
Endotoxin (LAL)< 20.0 EU/Device< 20.0 EU/ Device
CytotoxicityNon-cytotoxic (per ISO 10993-5:2009)Non-cytotoxic
SterilitySAL 10⁻⁶ via gamma irradiationSAL 10⁻⁶ via gamma irradiation
Heating Device
Temperature Maintenance37°C ± 0.5°CMet the acceptance criteria (temperature maintenance at 37 ± 0.5°C)
Overall System EfficacyYield an end product with improved characteristics (motile sperm with progressive motility and improved normal morphology) relative to the original semen sample.Media overlaying semen in the separators was enriched with motile sperm cells characterized by progressive motility and improved percentage of normal morphology.

2. Sample Size and Data Provenance

  • Sample Size for Test Set: The document does not explicitly state the sample size used for the efficacy testing of sperm preparation. For HSSA and LAL testing, these are described as "lot specific testing parameters per release activities," implying ongoing quality control rather than a single fixed test set for this 510(k).
  • Data Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective.

3. Number and Qualifications of Experts for Ground Truth

The document does not mention the use of experts to establish ground truth for the test set. The performance data seems to be based on laboratory measurements and analytical methods rather than expert assessment of subjective outcomes.


4. Adjudication Method

No adjudication method is mentioned or implied, as the performance testing focuses on objective laboratory measurements rather than subjective assessments requiring dispute resolution.


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done. The device is for separating sperm, not for diagnostic interpretation by human readers.


6. Standalone Performance Study (Algorithm Only)

The Seaforia™ system is a physical device for sperm separation, not an algorithm. Therefore, a standalone (algorithm only) performance study is not applicable. The efficacy testing described is for the device's physical function.


7. Type of Ground Truth Used

The ground truth for the performance testing appears to be based on:

  • Laboratory measurements/analytical methods: For HSSA (motility), LAL (endotoxin), cytotoxicity assays, and thermal verification.
  • Direct observation/quantification: For the "Efficacy testing of sperm preparation," which assessed characteristics like progressive motility and normal morphology in the separated sperm. This is an objective characteristic of the sperm sample.

8. Sample Size for Training Set

The document does not refer to a "training set" as the device is not an AI/ML-based system requiring such data. The performance testing described validates the physical and biological characteristics of the separated sperm and the heating device's function.


9. How Ground Truth for Training Set was Established

Given that there is no AI/ML component, there is no "training set" and therefore no ground truth established for a training set.

{0}------------------------------------------------

DEC 1 9 2012

510(k) Summary

510(k) Number:

510(K) Owner:

Submitter (US Agent):

Contact:

Manufacturer:

Date Prepared:

Trade name:

Common name:

Classification name:

Product Code(s):

K112644

Lotus Bio® (NYMPHAEA) Ltd. Jabotinsky 135a Ramat Gan 52563, Israel

Lori Kahler The RC Insight Group 461 Main Street, Suite 200 Pawtucket, RI 02860 Telephone: 401-722-1811 / 202-715-6242 Fax: 202-686-2877 Email: Ikahler@rcinsightgroup.com

Dr. Vered Shany, DMD, MBA CEO Telephone: +972 - 777-00-1669 Fax: +972 - 777-00-1670 Email: Vered@LotusBio.com

Lotus Bio® (NYMPHAEA) Ltd. Jabotinsky 135a Ramat-Gan 52563, Israel

17 December 2012

Seaforia™ Sperm Separation Assistance System

Assisted Reproduction Labware. Assisted Reproduction Accessories

21 C.F.R. §884.6160 , 21 C.F.R. § 884.6120

MQK (labware, assisted reproduction), MQG (accessories, assisted reproduction)

Classification:

Class II

{1}------------------------------------------------

Legally Marketed Device - Claiming Equivalence:

RI Migration Sedimentation Chamber (MSC) Research Instruments Ltd K112413

MTG- Minitüb Heated Stage System Zander Medical Supplies, Inc. K002971

Summary Description of the Device:

The SeaforiaTM Sperm Separation Assistance System separates motile/progressively motile sperm cells from debris and non-motile sperm cells from the original semen sample for further use in assisted reproductive procedures, including intrauterine insemination (IUI), intra-cervical insemination (ICI) and in-vitro fertilization (IVF). The system provides conditions such as temperature 37°C ±0.5 and effective surface area for the separation of motile human sperm cells. Note: The separation media is not included with the Seaforia" Sperm Separation Assistance System. LOTUS BIO® requires that the media used with the Seaforia'™ Sperm Separation Assistance System be a media cleared for use in the United States that is indicated for sperm handling, washing, or swim-up procedures with a formulation compatible with use outside of a CO2 environment.

The system is composed of the following units:

  • . Seaforia™ Sperm Separation Assistance System Disposable Kit - includes designated containers (separators) for the separation media available in various volumes and labware accessories such as syringes, polypropylene tips, pipette and vials. Five versions of the kit will be available to accommodate sperm samples ranging from 0.5-6 ml (Versions 0.5-2 ml. 2.5-3.0 ml. 3.5-4.0 ml, 4.5-5.0 ml, and 5.5-6.0 ml).
  • Seaforia™ Sperm Separation Assistance System Heating Device capable . of maintaining the sperm sample at the required temperature (37℃ ±0.5).

The sperm separation procedure takes place within the separator and is based on the original swim-up technique. Media is layered over the liquefied semen sample and during a subsequent incubation period of 30 minutes in the Seaforia™ Sperm Separation Assistance System Heating Device, the motile sperm cells migrate from the semen layer into the media.

Indications for Use/Intended Use:

The Seaforia™ Sperm Separation Assistance System consists of the following components:

{2}------------------------------------------------

The Seaforia™ Sperm Separation Assistance System Disposable Kit (Versions 0.5-2 ml, 2.5-3.0 ml. 3.5-4.0 ml. 4.5-5.0 ml. and 5.5-6.0 ml)

The Seaforia TM Sperm Separation Assistance System Disposable Kit is intended for preparing motile sperm using the swim-up (SU) separation method and holding sperm for use in assisted reproductive procedures such as intrauterine insemination (IUI), intracervical insemination (ICI) and in-vitro fertilization (IVF). The Seaforia™ Sperm Separation Assistance System Disposable Kit is indicated for incubation periods not exceeding one (1) hour.

The Seaforia TM Sperm Separation Assistance System Heating Device

The Seaforia TM Sperm Separation Assistance System Heating Device is indicated for use only with the Seaforia TM Sperm Separation Assistance System Disposable Kit to maintain the temperature (37°C ±0.5) of sperm during the swim-up (SU) separation procedure.

Substantial Equivalence Summary:

The SeaforiaTM Sperm Separation Assistance System is substantially equivalent to the Research Instruments Ltd Migration Sedimentation Chamber (K112413) and to the MTG-Minitüb Heated Stage System (K002971).

The SeaforiaTM Sperm Separation Assistance System Disposable Kit and the Research Instruments Ltd Migration Sedimentation Chamber (K112413) have equivalent intended use and indications for use as assisted reproduction labware to aid in the separation of motile sperm from liquefied semen for use in assisted reproductive medicine applications. The SeaforiaTM Sperm Separation Assistance System Disposable Kit and the RI Migration Sedimentation Chamber (MSC) are both disposable devices and are made of materials commonly used in assisted reproduction labware. All components of the Seaforia™ Sperm Separation Assistance System Disposable Kit are made of polypropylene except for the pipettes which are made from a low density polyethylene, while the RI MSC is made from polystyrene. The main difference between the SeaforiaTM Sperm Separation Assistance System Disposable Kit and the RI MSC is that the disposable kit contains designated containers for different semen volume and accessories (such as syringes, polypropylene tips, pipette, and vials) whereas the RI MSC device is a single device rather than a kit. Both the Seaforia™ Sperm Separation Assistance System Disposable Kit and the RI MSC are indicated for use in preparing sperm in assisted reproduction procedures. Both the Seaforia™ Sperm Separation Assistance System Disposable Kit and the RI MSC use a comparable swim-up procedure and collection principal. The only minor difference in the design between the predicate device and the proposed device is the Seaforia™ System method of loading and collecting that requires the use of designated containers and accessories to aid in sperm separation procedure (such as syringes, polypropylene tips, pipette, and vials). The differences in the design do not impact the swim-up procedure and collection method. Regarding the Seaforia™ Sperm Separation Assistance System Disposable Kit HSSA specification: as the Seaforia™ Sperm Separation Assistance System Disposable Kit is

{3}------------------------------------------------

indicated for incubation periods of 30 minutes (should not exceed one (1) hour); the HSSA assessment was completed at 1 hour. The predicate's assessment was completed at 24 hours; however, the predicate device is also indicated for sample storage and the sperm is pre-loaded with culture media two hours prior to the moment of insemination. The proposed device has a short incubation period and is not indicated for sample storage. The HSSA assessment completed at 1 hour is appropriate for the use of the proposed device.

In addition, the Seaforia™ Sperm Separation Assistance System is also for use in assisted reproductive procedures such as intra-cervical insemination (ICI) whereas the RI MSC is not indicated for use with ICI. This difference would not represent a new intended use as intra-cervical insemination (ICI) is a common method of assisted reproduction.

The SeaforiaTM Sperm Separation Assistance System Heating Device and the MTG-Minitüb Heated Stage System (K002971) are equivalent with respect to intended use, indications for use and technological characteristics as a heating device used to warm biological material. The MTG-Minitüb Heated Stage System (K002971) has a wider variation of temperature range (ambient temperature through 55°C with a control accuracy +/- 0.2°C) while the Seaforia™ Heating Device maintains a temperature of 37°C ±0.5. The differences do not affect the safety and effectiveness of the device and does not impact the use of the identified predicate to support an equivalence decision.

The following detailed substantial equivalence table compares the Seaforia"M Sperm Separation Assistance System with the predicate devices.

New DevicePredicate DevicePredicate Device
510(k) #K002971K112413
CompanyLotus Bio®(NYMPHAEA) Ltd.Zander Medical Supplies, Inc.Research Instruments Ltd.
NameSeaforia™MTG-Minitüb Heated Stage SystemRI MSC
Classification/RegulationClass II / 884.6160,884.6120Class II / 884.6120Class II / 884.6160
Product CodeIndications forUse - DisposableKitDisposable Kit: MQK(labware, assistedreproduction HeatingDevice: MQG (accessories,assisted reproduction);)The Seaforia™ SpermSeparation AssistanceSystem Disposable KitThe Seaforia ™ SpermSeparation AssistanceSystem Disposable Kit isintended for preparingmotile sperm using theswim-up (SU) separationmethod and holding spermfor use in assistedreproductive proceduressuch as intrauterineinsemination (IUI), intra-cervical insemination (ICI)and in-vitro fertilizationMQG (accessories. assistedreproduction)n/aMQK (labware, assistedreproduction)The RI MSC is intendedto prepare sperm bymigration-sedimentationmethod for the assistedreproduction techniquesof intracytoplasmic sperminjection (ICSI), in vitrofertilization (IVF), andintrauterine insemination(IUI).
New DevicePredicate DevicePredicate Device
(IVF). The Seaforia™Sperm SeparationAssistance SystemDisposable Kit is indicatedfor incubation periods notexceeding one (1) hour.
Indications forUse - HeatingDeviceThe Seaforia™ SpermSeparation AssistanceSystem Heating DeviceThe Seaforia ™ SpermSeparation AssistanceSystem Heating Device isindicated for use only withthe Seaforia ™ SpermSeparation AssistanceSystem Disposable Kit tomaintain the temperature(37°C ±0.5) of spermduring the swim-up (SU)separation procedure.Maintaining the temperatureof biological material likegametes and embryos at acertain temperature isessential to cover multipleapplications in reproductivemedicine, biology and otherareas.n/a
DisposableMaterialsPolypropylene; lowdensity polyethylene(pipette only)n/aMedical gradepolystyrene - non treated
Design Features -Disposable KitThe Disposable Kitcontains designatedcontainers (for differentsemen volume) andaccessories (such assyringes, polypropylenetips, pipette, and vials).The kits are suitable forvariable semen volume 0.5-6ml. The containers aremade of polypropylene.n/aCylindrical container withinternal gallery and well.Optically clear, with flatbase.
Design Features -Heating DeviceSwitch on/off toggle:PC+ABS GE C2950 plasticprotective case.Temperature range 37°C±0.5. Output 5.0 VDC ±5%, 1000mAControl Units: metal casingprotects electronic controlunit - Digital displays andtouch pads control andmonitor the temperature.Temperature range: ambienttemperature through 55℃.Control accuracy +/- 0.2℃.All control units can beswitched between 200-240V/50Hz or 100-120V/60Hz.n/a
SterileDisposable Kit: Yes, Sterile(SAL 10-6), via gammairradiationHeating Device: n/an/aYes, Sterile (SAL 10-6),via gamma irradiation
HSSAHSSA ≥ 70% motilityfollowing 1 hourn/aHSSA ≥ 70% motility at24hr
CytotoxicityCytotoxicity tested per ISO10993-5:2009.n/aNot known
Endotoxin (LAL)Disposable Kit: Testednon-pyrogenic by LALAcceptance Criteria: <20.0EU/ DeviceHeating Device: n/an/aTested non-pyrogenic byLAL: <0.5 EU/Device
New DevicePredicate DevicePredicate Device
ContraindicationsNot recommended for original semen samples containing less than 30% progressive motility.n/aNot recommended for sperm samples with total motility rate < 40%.

Substantial Equivalence Table

{4}------------------------------------------------

:

.

.

:

.

  1. September 19.

:

.

{5}------------------------------------------------

Technical Characteristics:

The Seaforia ™ Sperm Separation Assistance System Disposable Kit

The Seaforia " Sperm Separation Assistance System Disposable Kit contains 4 designated separator containers (for different semen volume) and accessories (three syringes, 3 polypropylene tips, 1 pipette, and 1 vial). The kits are suitable for variable semen volume 0.5-6ml. The separator containers are made of polypropylene.

The Seaforia" Sperm Separation Assistance System Disposable Kit (separators and accessories) are sterilized by gamma irradiation to achieve SAL of 106. The kit components are non-pvrogenic as tested.by the limulus amebocvte lysate (LAL) assay, non-cytotoxic as tested by the cytotoxicity assay, and non-sperm toxic as tested by the human sperm survival assay (HSSA). Endotoxin testing/LAL (<20 EU/device) and HSSA testing (≥ 70% motility following 1 hour of exposure) are conducted as lot specific testing parameters per release activities.

The Seaforia "" Sperm Separation Assistance System Heating Device

The SeaforiaTM Sperm Separation Assistance System Heating Device is a warming unit with a pre-set temperature of 37°C±0.5. The Heating Device structure is designed to fit 3 separators at the same time. The Heating Unit is powered by SV 1000mA and is intended for multiple use.

Electromagnetic Compatibility/Electrical Safety:

Medical electrical equipment testing has been completed in accordance to IEC 60601-1. Medical Electrical Equipment - Part 1 General Requirements for Safety, 1988: Amendment 1, 1991. Amendment 2, 1995. Electromagnetic Compatibility (EMC) testing was conducted in accordance with applicable parts of AAMI/ANSI/IEC 60601-1-2, Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests (Editions 2:2001 with Amendment 1: 2004).

Performance Testing - Bench:

Performance testing on the Seaforia™ Sperm Separation Assistance System was completed to assure that the device performed as intended in accordance with design and functional specifications. The following performance testing was conducted:

{6}------------------------------------------------

    1. Efficacy testing of sperm preparation using the Seaforia™ Sperm Separation Assistance System
      The test was designed to confirm that by using the Seaforia™ Sperm Separation Assistance System, the separation process yields an end product with improved characteristics relative to the original semen sample. Results showed that media overlaving semen in the separators was enriched with motile sperm cells characterized by progressive motility and improved percentage of normal morphology, leaving most round non-sperm cells and immotile sperm cells behind. Therefore, testing confirmed that separation using the Seaforia "14 Sperm Separation Assistance System resulted in an end product with improved characteristics as compared to the original semen sample.
    1. Thermal verification test for the Seaforia™ Sperm Separation Assistance System Heating Device
      Thermal verification testing was conducted to validate the thermal performance of the Seaforia Sperm Separation Assistance System Heating Device. Testing showed that devices met the acceptance criteria (temperature maintenance at 37 ± 0.5°C).

Summary of Sterilization and Shelf Life:

The SeaforiaTM Sperm Separation Assistance System Disposable Kit is sterilized by gamma-irradiation overkill VDmax method at 25 KGv. The heating device has also undergone validation testing to support the cleaning and disinfection procedures included in device labeling.

The Seaforia™ Sperm Separation Assistance System Disposable Kit has a shelf-life of 1 vear from the date of manufacture and should be stored at room temperature.

Conclusion:

The SeaforiaTM Sperm Separation Assistance System Disposable Kit and the Research Instruments Ltd Migration Sedimentation Chamber (K112413) have equivalent intended use, indications for use and technological characteristics for use in preparing sperm for use in assisted reproductive procedures. The Seaforia™ Sperm Separation Assistance System Heating Device and the MTG-Minitüb Heated Stage System (K002971) are equivalent with respect to intended use, indications for use and technological characteristics as a heating device for use in assisted reproductive procedures.

The Seaforial " Sperm Separation Assistance System is substantially equivalent to the predicate devices which have equivalent intended use, indications for use and technological characteristics. There are no differences between the Seaforia™ Sperm Separation Assistance System and the predicate devices that would raise new types of

{7}------------------------------------------------

safety or effectiveness issues. The Seaforia™ Sperm Separation Assistance System is as safe and effective as the legally marketed predicate devices.

.

:

.

:

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hamnshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 19, 2012

Lotus Bio® (NYMPHAEA) Ltd. % Ms. Lori Kahler President The RC Insight Group 461 Main Street, Suite 217 PAWTUCKET RI 02860

Re: K112644

Trade/Device Name: SeaforiaTM Sperm Seperation Assistance System Regulation Number: 21 CFR§ 884.6160 Regulation Name: Assisted reproduction labware Regulatory Class: II Product Code: MQK, MQG Dated: November 10, 2012 Received: November 27, 2012

Dear Ms. Kahler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{9}------------------------------------------------

Page 2 - Ms. Lori Kahler

You must comply with all the Act's requirements, including, but not limited to: registration and i ou must compy with an the 1100 of 2017 medical device reporting (reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements medied. abviet related ad vestems (QS) regulation (21 CFR Part 820); and if applicable, the as oct to the many tradiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

(2) CPN P att 800) produces/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

1.0 Indications for Use

510(k) Number (if known): K112644

Device Name:

SeaforiaTM Sperm Separation Assistance System

Indications for Use:

The Seaforia™ Sperm Separation Assistance System consists of the following components:

The Seaforia™ Sperm Separation Assistance System Disposable Kit (Versions 0.5-2 ml, 2.5-3.0 ml, 3.5-4.0 ml, 4.5-5.0 ml, and 5.5-6.0 ml)

The Seaforia ™ Sperm Separation Assistance System Disposable Kit is intended for preparing motile sperm using the swim-up (SU) separation method and holding sperm for use in assisted reproductive procedures such as intrauterine insemination (IUI), intracervical insemination (ICI) and in-vitro fertilization (IVF). The Seaforia™ Sperm Separation Assistance System Disposable Kit is indicated for incubation periods not exceeding one (1) hour.

The Seaforia TM Sperm Separation Assistance System Heating Device

The Seaforia ™ Sperm Separation Assistance System Heating Device is indicated for use only with the Seaforia ™ Sperm Separation Assistance System Disposable Kit to maintain the temperature (37°C ±0.5) of sperm during the swim-up (SU) separation procedure.

Prescription Use ア (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Beniamin R. Fisher - S 2012.12.19 15:10:46 -05'00'

(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices

510(k) NumberK112644
------------------------

§ 884.6160 Assisted reproduction labware.

(a)
Identification. Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.(b)
Classification. Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.