(128 days)
For in vitro diagnostic use with the IMMULITE® 2000 Analyzer -- for the quantitative measurement of allergen-specific IgE in human serum, as an aid in the clinical diagnosis of IgE-mediated allergic disorders. The test results are to be used in conjunction with clinical findings and other laboratory tests.
IMMULITE® 2000 3gAllergy™ Specific IgE is a solid-phase, two-step, chemiluminescent immunoassay that exploits liquid phase kinetics in a bead format. The allergens are covalently bound to a soluble polymer matrix, which in turn is labeled with a ligand. The use of an amino acid co-polymer amplifies the amount of allergen that the matrix can support.
Here's a breakdown of the acceptance criteria and study information for the IMMULITE® 2000 3gAllergy™ Specific IgE Assay, based on the provided text:
Acceptance Criteria and Device Performance
The study evaluates the analytical performance and clinical performance of the device.
Analytical Performance Acceptance Criteria:
The key acceptance criteria for analytical performance are related to precision, linearity, and specificity. While explicit quantitative criteria were provided for precision, linearity had implied criteria based on excellent regression statistics, and specificity was demonstrated by successful inhibition.
| Acceptance Criteria Category | Specific Criteria | Reported Device Performance | Comments |
|---|---|---|---|
| Precision | Intra-assay Variability: Less than 15% CV for all allergen lots. | All allergens met this criterion. For example, nOle e 1 (Lot 201) had within-run CVs ranging from 3.95% to 6.23%. | This indicates consistent results within a single test run. |
| Inter-assay Precision: Less than 15% CV for all allergen lots. | All allergens met this criterion. For example, nOle e 1 (Lot 201) had total CVs ranging from 5.85% to 7.03%. | This indicates consistent results across multiple test runs over different days. | |
| Lot-to-Lot Precision: Less than 20% CV for all 5 allergens. | All 5 allergens met this criterion. For example, combined nOle e 1 lots showed total CVs up to 6.94%. | This demonstrates consistency of results when using different batches (lots) of the allergen. | |
| Negative Control: Average negative control remained negative (< 0.10 kU/L) during precision testing. | The average negative control remained negative (< 0.10 kU/L) throughout precision testing on all allergens. | This confirms proper function and absence of non-specific binding for negative samples. | |
| Linearity | Excellent regression statistics (slope close to 1, intercept close to 0, R close to 1) comparing observed to expected data across a range of concentrations. | The reported R values were all 0.995 or higher, slopes were close to 1 (e.g., 0.997 for nOle e 1, 0.975 for Cat Serum Albumin), and intercepts were close to 0. | This indicates that the assay accurately measures IgE concentrations proportionally across its analytical range. |
| Specificity (Inhibition) | Demonstrate concentration-dependent inhibition by relevant inhibitor extract, with a target of 50% inhibition at the highest inhibitor concentration. No significant inhibition by unrelated allergen extracts. | The inhibition plots demonstrated concentration-dependent inhibition and the target 50% inhibition was met for all tested allergens. Inhibition with unrelated extracts did not generate any significant inhibition. | This confirms that the assay specifically detects the intended allergen-specific IgE and does not show significant cross-reactivity with unrelated allergens. |
Clinical Performance Acceptance Criteria:
The acceptance criteria for clinical performance are based on the agreement of the device's results with clinical diagnosis, measured by sensitivity, specificity, and overall agreement.
| Acceptance Criteria Category | Specific Criteria | Reported Device Performance | Comments |
|---|---|---|---|
| Clinical Agreement | Sensitivity, Specificity, and Overall Agreement with clinical diagnosis within acceptable confidence intervals (though no explicit numerical target for the confidence interval width or minimum performance metric was given, the reported values demonstrate acceptable performance given the context of the device). | Sensitivity: 76.7% (Lower CI: 71%, Upper CI: 83%) | This indicates the proportion of individuals with allergic disorders that the test correctly identifies as positive. |
| Specificity: 95.8% (Lower CI: 94%, Upper CI: 98%) | This indicates the proportion of individuals without allergic disorders that the test correctly identifies as negative. | ||
| Agreement: 90.5% (Lower CI: 88%, Upper CI: 93%) | This indicates the overall accuracy of the test compared to clinical diagnosis. |
Study Details
2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Analytical Test Set Sample Size:
- Precision: For each allergen lot and each positive sample level, N=80 measurements were performed (2 aliquots per run, 2 runs per day, 20 days). For combined lot precision, N=240 (80 measurements x 3 lots) where applicable. Negative controls were also tested.
- Linearity: 11 or 12 data points (depending on the allergen) from 2-fold serial dilutions, including undiluted (neat) samples.
- Specificity (Inhibition): Not explicitly stated how many unique serum samples were used "a single serum sample or pool of sera". For cross-reactivity, one positive sample was tested with three unrelated allergen extracts.
- Clinical Test Set Sample Size:
- Total Patients: 693 individuals were tested (193 clinically positive, 500 clinically normal).
- Data Provenance: Not explicitly stated for either analytical or clinical studies. However, the manufacturer (Siemens Healthcare Diagnostics Inc. and Siemens Healthcare Diagnostics Products, LTD) has locations in Tarrytown, NY, and Llanberis, Gwynedd, UK. It is likely that the data originates from studies conducted or coordinated by these facilities, potentially involving samples from different geographic locations. The document does not specify whether the data was retrospective or prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
- Analytical Studies: No external experts were used; ground truth was established by the experimental design (e.g., known concentrations for linearity, known positive/negative samples for precision and specificity).
- Clinical Studies: The ground truth for clinical performance was "clinical documentation of presence or absence of signs, symptoms and other diagnostic evidence of allergen sensitivity." The number and qualifications of experts (e.g., allergists, physicians) involved in making these clinical diagnoses are not specified in the provided text.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Clinical Studies: The document does not describe any specific adjudication method for establishing the "clinical diagnosis." It refers to "clinical findings and other laboratory tests," implying a standard clinical diagnostic process rather than a formalized adjudication among multiple experts for the purpose of the study.
- Analytical Studies: Adjudication is not applicable as ground truth is based on known analytical properties.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an in-vitro diagnostic assay for quantitative measurement of allergen-specific IgE in human serum. It is an automated laboratory test, not an imaging device or AI-assisted diagnostic tool that would involve human readers or AI assistance in interpretation. The study evaluates the performance of the analytical device itself.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- This is a standalone diagnostic assay. The IMMULITE® 2000 3gAllergy™ Specific IgE Assay, when run on the IMMULITE® 2000 Analyzer, operates as an automated system. The performance data presented (precision, linearity, specificity, and clinical agreement) reflects the "algorithm only" or "device only" performance, without human interpretive input for the final result beyond operating the analyzer and reviewing the quantitative output. The results are quantitative IgE levels, not qualitative interpretations requiring human judgment (e.g., image reading).
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Analytical Studies: The ground truth was based on known properties of the samples and experimental design:
- Precision: Replicates of the same sample, the true value of which is the mean of the replicates.
- Linearity: Gravimetrically prepared dilutions with known expected concentrations.
- Specificity: Known inhibitor extracts and known unrelated allergens.
- Clinical Studies: The ground truth for clinical performance was "clinical documentation of presence or absence of signs, symptoms and other diagnostic evidence of allergen sensitivity." This can be classified as a form of clinical diagnosis and outcomes data, likely based on patient history, physical examination, symptom evaluation, and other relevant diagnostic tests, rather than a single definitive pathology result.
8. The sample size for the training set
- Not applicable / Not explicitly stated. This document describes a 510(k) submission for new allergens to an existing IMMULITE® 2000 3gAllergy™ Specific IgE assay. It details the validation and performance studies for these additional allergens. The IMMULITE® 2000 system and the 3gAllergy™ assay are commercially available, validated technologies. The concept of a "training set" as understood in machine learning (where algorithms are "trained" on data) is not directly applicable here. The assay itself uses established immunochemical principles; it's not a machine learning algorithm that undergoes a training phase with a distinct dataset.
9. How the ground truth for the training set was established
- Not applicable. As explained above, this device does not involve an "AI algorithm" or "machine learning" in the traditional sense that would require a separate training set and ground truth establishment for that training phase. The analytical and clinical performance studies are for validation of the assay for the specific new allergens.
{0}------------------------------------------------
SUMMARY OF SAFETY AND EFFECTIVENESS
JAN - 6 2012
Assigned 510(k) Number
The assigned 510(k) number is: K112523.
Sponsor Name and Address
Siemens Healthcare Diagnostics Inc. 511 Benedict Avenue Tarrytown, NY 10591 (Office) 914-524-3270 (Fax) 914-524-2101
Contact
Garo Mimaryan Technical Regulatory Affairs Specialist III Phone: (914) 524-3270 Fax: (914) 524-2101 garo.mimaryan@siemens.com
Device Name
| Trade name: | IMMULITE® 2000 3gAllergy™ Specific IgE Assay |
|---|---|
| Classification: | Class II |
| Classification Names: | Radioallergosorbent (RAST) Immunological Test System |
| Regulation Number: | 866.5750 |
| Product Code: | DHB |
| Catalog Numbers: | L2KUN6 (600 tests) |
Description of Device
IMMULITE® 2000 3gAllergy™ Specific IgE is a solid-phase, two-step, chemiluminescent immunoassay that exploits liquid phase kinetics in a bead format. 12
(U.S. Patent No. 4,778,751) It represents a significant advance over conventional methods relying on allergens attached to a solid-phase support, such as a paper disk.
The allergens are covalently bound to a soluble polymer matrix, which in turn is labeled with a ligand. The use of an amino acid co-polymer amplifies the amount of allergen that the matrix can support.
Incubation Cycles: 2 × 30 minutes.
1 El Shami AS, Alaba O. Liquid-phase in vitro allergen-specific IgE assay with in situ immobilization. Adv Biosci 1989;74.191-201.
2 Alaba O, El Shami AS. Evaluation of non-specific IgE binding: comparison of two in vitro allergen assays. Adv Biosci 1989:74:203-14.
{1}------------------------------------------------
Indications for Use
For in vitro diagnostic use with the IMMULITE® 2000 Analyzer -- for the quantitative measurement of allergen-specific IgE in human serum, as an aid in the clinical diagnosis of IgEmediated allergic disorders. The test results are to be used in conjunction with clinical findings and other laboratory tests.
Establishment Information
IMMULITE® 2000 3gAllergy Specific IgE assay is manufactured by Siemens Healthcare Diagnostics Inc. at the following locations:
Siemens Healthcare Diagnostics Products, LTD Glyn Rhonwy, Llanberis, Gwynedd, LL55 4EL, United Kingdom Telephone Number: 44 1286 878187 188408 Fax Number: 44 1286 87818718 FDA Establishment Registration: 9610633
Siemens Healthcare Diagnostics Inc. 5210 Pacific Concourse Drive Los Angeles, CA 90045-8200 Telephone Number: (310)-645-8200 Fax Number: (310)-645-9999 FDA Establishment Registration: 3005250747
Predicate
The purpose of this 510(k) submission is for clearance of five additional specific allergens, named in the table below, to be used with the IMMULITE® 2000 3gAllergy™ Specific IgE assay on the IMMULITE® 2000 analyzer.
| 1 | A482 (Native Ole e 1) | 4 | E220 (Cat Serum Albumin, Fel d 2) |
|---|---|---|---|
| 2 | A753 (Native Art v 1) | 5 | E221 (Dog Serum Albumin, Can f 3 ) |
| 3 | A89 (Native Bet v 1) |
FDA clearance was previously obtained for the assay kit and 106 specific allergens (K093987, K100910 and K101572).
Please note that the FDA clearances indicated above were in the name of Diagnostic Products Corporation which was acquired by Siemens Medical Solutions Diagnostics in July 2006. Siemens Medical Solutions Diagnostics was renamed Siemens Healthcare Diagnostics Inc. on January 1, 2008.
{2}------------------------------------------------
| Similarities | ||
|---|---|---|
| Description Item | IMMULITE Device | Predicate Device |
| Indication for Use | For in vitro diagnostic use with the IMMULITE®2000 Analyzer — for the quantitative measurementof allergen-specific IgE in human serum, as an aidin the clinical diagnosis of IgE-mediated allergicdisorders. The test results are to be used inconjunction with clinical findings and otherlaboratory tests. | Same |
| Number ofCalibrators | Eight | Same |
| Sample Matrix | Serum | Same |
| Antibody | Monoclonal murine anti-human IgE conjugated toalkaline phosphatase | Same |
| Basic Principle | Chemiluminescent Immunoassay | Same |
| Sample Volume | 50 μl | Same |
| Process Time | 65 minutes | Same |
| IncubationTemperature | 37°C | Same |
| Differences | ||
| Modifications | Allergenic proteins purified from native sources | Native whole allergens |
Precision
Precision studies were performed in accordance with Clinical Laboratory Standard Institute (CLSI) guidance: Evaluation of Precision Performance of Quantitative Methods; Approved Guideline-Second Edition. CLSI document EPS-A2 (ISBN 1-56238-542-9). CLSI, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898 USA, 2004, assaying two aliquots of each test sample in two runs per day on 20 different days (Positives 1-3). Additional studies for Positives 4 and 5 were run with two runs per day over 20 days. Analysis of variance was used to estimate the within-run and total precision.
Three allergen lots were tested using three positive samples and one negative sample. Intra-assay and inter-assay precision for the positive samples were evaluated by calculating the kU/L dose percent coefficients of variation (%CV) for each positive sample. Non-specific binding (NSB) was monitored by testing the negative control sample. Representative precision claims for each allergen tested are presented below:
{3}------------------------------------------------
Precision statistics using instrument 13474 and 13476 for the 5 individual allergens, noting allergen lot used are presented in the following tables:
| Within-Run | Total | |||||||
|---|---|---|---|---|---|---|---|---|
| Sample | N | Mean | SD | CV | SD | CV | ||
| (kU/L) | (kU/L) | % | (kU/L) | % | ||||
| Allergen = nOle e 1, Lot 201 | ||||||||
| Positive #1 | 80 | 0.50 | 0.027 | 5.46 | 0.030 | 5.98 | ||
| Positive #2 | 80 | 1.91 | 0.119 | 6.23 | 0.134 | 7.03 | ||
| Positive #3 | 80 | 17.34 | 0.686 | 3.95 | 1.013 | 5.85 | ||
| Allergen = nOle e 1, Lot 202 | ||||||||
| Positive #1 | 80 | 0.52 | 0.028 | 5.48 | 0.033 | 6.38 | ||
| Positive #2 | 80 | 1.97 | 0.095 | 4.84 | 0.117 | 5.96 | ||
| Positive #3 | 80 | 17.80 | 0.702 | 3.94 | 1.068 | 6.00 | ||
| Allergen = nOle e 1, Lot 203 | ||||||||
| Positive #1 | 80 | 0.51 | 0.027 | 5.18 | 0.035 | 6.90 | ||
| Positive #2 | 80 | 2.00 | 0.111 | 5.55 | 0.139 | 6.93 | ||
| Positive #3 | 80 | 17.82 | 0.907 | 5.09 | 1.156 | 6.48 | ||
| Within-Run | Total | |||||||
| Sample | N | Mean(kU/L) | SD(kU/L) | CV% | SD(kU/L) | CV% | ||
| Allergen = nArt v 1, Lot 201 | ||||||||
| Positive #1 | 80 | 1.28 | 0.040 | 3.15 | 0.066 | 5.15 | ||
| Positive #2 | 80 | 5.13 | 0.196 | 3.83 | 0.288 | 5.61 | ||
| Positive #3 | 80 | 52.54 | 1.676 | 3.19 | 2.671 | 5.08 | ||
| Positive #4 | 80 | 0.44 | 0.015 | 3.41 | 0.021 | 4.67 | ||
| Positive #5 | 80 | 0.46 | 0.018 | 3.94 | 0.033 | 7.10 | ||
| Allergen = nArt v 1, Lot 202 | ||||||||
| Positive #1 | 80 | 1.28 | 0.137 | 10.65 | 0.144 | 11.25 | ||
| Positive #2 | 80 | 5.16 | 0.227 | 4.39 | 0.293 | 5.68 | ||
| Positive #3 | 80 | 52.43 | 1.862 | 3.55 | 2.873 | 5.48 | ||
| Positive #4 | 80 | 0.43 | 0.017 | 4.04 | 0.028 | 6.49 | ||
| Positive #5 | 80 | 0.46 | 0.012 | 2.72 | 0.032 | 7.02 | ||
| Allergen = nArt v 1, Lot 203 | ||||||||
| Positive #1* | 80 | 1.53 | 1.697 | 111.20 | 1.710 | 112.08 | ||
| Positive #1** | 78 | 1.32 | 0.156 | 11.67 | 0.144 | 10.87 | ||
| Positive #2 | 80 | 5.10 | 0.225 | 4.40 | 0.307 | 6.02 | ||
| Positive #3 | 80 | 52.13 | 2.253 | 4.32 | 2.839 | 5.45 | ||
| Positive #4 | 80 | 0.43 | 0.015 | 3.51 | 0.026 | 5.94 | ||
| Positive #5 | 80 | 0.47 | 0.016 | 3.35 | 0.033 | 7.03 | ||
| *Two Outliers detected (Day 12, Run 2, Rep Number 2) (Day: 15, Run: 1,Rep Number 1) | ||||||||
| **No outliers used in calculation | ||||||||
| Within-Run | Total | |||||||
| Sample | N | Mean | SD | CV | SD | CV | ||
| (kU/L) | (kU/L) | % | (kU/L) | % | ||||
| Allergen = Dog Serum Albumin (Can f 3), Lot 201 | ||||||||
| Positive #1 | 80 | 47.84 | 1.056 | 2.21 | 1.979 | 4.14 | ||
| Positive #2 | 80 | 23.10 | 0.950 | 4.11 | 1.170 | 5.07 | ||
| Positive #3 | 80 | 0.87 | 0.024 | 2.81 | 0.064 | 7.35 | ||
| Positive #4 | 80 | 0.44 | 0.015 | 3.50 | 0.040 | 9.17 | ||
| Positive #5 | 80 | 0.46 | 0.025 | 5.46 | 0.034 | 7.24 | ||
| Allergen = Dog Serum Albumin (Can f 3), Lot 202 | ||||||||
| Positive #1 | 80 | 48.19 | 6.043 | 12.54 | 6.252 | 12.97 | ||
| Positive #2 | 79 | 23.05 | 0.751 | 3.26 | 1.041 | 4.52 | ||
| Positive #3 | 80 | 0.87 | 0.028 | 3.21 | 0.057 | 6.54 | ||
| Positive #4 | 80 | 0.44 | 0.015 | 3.35 | 0.045 | 10.19 | ||
| Positive #5 | 80 | 0.48 | 0.020 | 4.09 | 0.029 | 6.01 | ||
| Allergen = Dog Serum Albumin (Can f 3), Lot 203 | ||||||||
| Positive #1 | 80 | 48.83 | 1.248 | 2.55 | 2.119 | 4.34 | ||
| Positive #2 | 80 | 22.97 | 0.809 | 3.52 | 1.291 | 5.62 | ||
| Positive #3 | 80 | 0.84 | 0.027 | 3.28 | 0.065 | 7.78 | ||
| Positive #4 | 80 | 0.44 | 0.016 | 3.58 | 0.044 | 9.88 | ||
| Positive #5 | 80 | 0.48 | 0.020 | 4.23 | 0.034 | 7.11 |
:
Allergen Precision Claims
{4}------------------------------------------------
:
:
・
{5}------------------------------------------------
:
.
.
.
:
・
،
{6}------------------------------------------------
| Within-Run | Total | |||||
|---|---|---|---|---|---|---|
| Sample | N | Mean | SD | CV | SD | CV |
| (kU/L) | (kU/L) | % | (kU/L) | % | ||
| Allergen = nBet v 1, Lot 201 | ||||||
| Positive #1 | 80 | 2.42 | 0.102 | 4.22 | 0.103 | 4.23 |
| Positive #2 | 80 | 9.24 | 0.304 | 3.29 | 0.352 | 3.81 |
| Positive #3 | 80 | 23.78 | 0.898 | 3.77 | 1.038 | 4.36 |
| Positive #4 | 80 | 0.39 | 0.020 | 5.16 | 0.024 | 6.28 |
| Positive #5 | 80 | 0.41 | 0.018 | 4.34 | 0.025 | 6.17 |
| Allergen = nBet v 1, Lot 202 | ||||||
| Positive #1 | 80 | 2.42 | 0.104 | 4.30 | 0.107 | 4.43 |
| Positive #2 | 80 | 9.25 | 0.309 | 3.34 | 0.381 | 4.11 |
| Positive #3 | 80 | 23.78 | 0.717 | 3.02 | 0.987 | 4.15 |
| Positive #4 | 80 | 0.38 | 0.015 | 4.00 | 0.024 | 6.28 |
| Positive #5 | 80 | 0.41 | 0.017 | 4.18 | 0.028 | 6.83 |
| Allergen = nBet v 1, Lot 203 | ||||||
| Positive #1 | 80 | 2.36 | 0.095 | 4.04 | 0.109 | 4.63 |
| Positive #2 | 80 | 9.09 | 0.284 | 3.12 | 0.344 | 3.78 |
| Positive #3 | 80 | 22.94 | 0.724 | 3.16 | 0.828 | 3.61 |
| Positive #4 | 80 | 0.38 | 0.015 | 3.86 | 0.020 | 5.38 |
| Positive #5 | 80 | 0.41 | 0.015 | 3.77 | 0.030 | 7.32 |
. . . . .
:
:
.
:
:
1
:
{7}------------------------------------------------
| Within-Run | Total | ||||||
|---|---|---|---|---|---|---|---|
| Sample | N | Mean | SD | CV | SD | CV | |
| (kU/L) | (kU/L) | % | (kU/L) | % | |||
| Allergen = Cat Serum Albumin (Fel d 2), Lot 201 | |||||||
| Positive #1 | 80 | 0.91 | 0.031 | 3.41 | 0.055 | 6.03 | |
| Positive #2 | 80 | 36.38 | 1.173 | 3.22 | 1.365 | 3.75 | |
| Positive #3 | 80 | 45.81 | 2.032 | 4.44 | 2.032 | 4.44 | |
| Positive #4 | 80 | 0.36 | 0.013 | 3.58 | 0.021 | 5.80 | |
| Positive #5 | 80 | 0.38 | 0.014 | 3.74 | 0.023 | 6.12 | |
| Allergen = Cat Serum Albumin (Fel d 2), Lot 203 | |||||||
| Positive #1 | 80 | 0.92 | 0.027 | 2.94 | 0.063 | 6.91 | |
| Positive #2 | 80 | 35.86 | 0.853 | 2.38 | 1.046 | 2.92 | |
| Positive #3 | 80 | 44.62 | 1.260 | 2.82 | 1.556 | 3.49 | |
| Positive #4 | 80 | 0.36 | 0.012 | 3.47 | 0.026 | 7.22 | |
| Positive #5 | 80 | 0.38 | 0.018 | 4.70 | 0.028 | 7.25 | |
| Allergen = Cat Serum Albumin (Fel d 2), Lot 204 | |||||||
| Positive #1 | 80 | 0.97 | 0.027 | 2.77 | 0.067 | 6.93 | |
| Positive #2 | 80 | 36.41 | 0.942 | 2.59 | 1.125 | 3.09 | |
| Positive #3 | 80 | 45.38 | 1.362 | 3.00 | 1.685 | 3.71 | |
| Positive #4 | 80 | 0.35 | 0.018 | 5.18 | 0.023 | 6.36 | |
| Positive #5 | 80 | 0.38 | 0.019 | 4.98 | 0.025 | 6.68 | |
| Within-Run | Total | ||||||
| Sample | N | Mean | SD | CV | SD | CV | |
| (kU/L) | (kU/L) | % | (kU/L) | % | |||
| Allergen = nOle e 1 (Lots 201, 202, 203) | |||||||
| Positive #1 | 240 | 0.51 | 0.030 | 5.88 | 0.034 | 6.67 | |
| Positive #2 | 240 | 1.96 | 0.115 | 5.87 | 0.136 | 6.94 | |
| Positive #3 | 240 | 17.65 | 0.863 | 4.89 | 1.075 | 6.09 | |
| Allergen = nArt v 1 (Lots 201, 202, 203) | |||||||
| Positive #1* | 240 | 1.36 | 0.992 | 72.94 | 0.999 | 73.46 | |
| Positive #1** | 239 | 1.30 | 0.134 | 10.31 | 0.151 | 11.62 | |
| Positive #2 | 240 | 5.13 | 0.214 | 4.17 | 0.297 | 5.79 | |
| Positive #3 | 240 | 52.37 | 1.905 | 3.64 | 2.737 | 5.23 | |
| Positive #4 | 240 | 0.44 | 0.016 | 3.65 | 0.025 | 5.75 | |
| Positive #5 | 240 | 0.46 | 0.016 | 3.46 | 0.033 | 7.06 | |
| *Outlier Detected (Day: 12, Run: 2, Rep Number: 2) | |||||||
| **No outliers used in calculation | |||||||
| Allergen = Dog Serum Albumin (Can f 3) (Lots 201, 202, 203) | |||||||
| Positive #1 | 240 | 48.29 | 3.571 | 7.39 | 3.825 | 7.92 | |
| Positive #2 | 240 | 23.04 | 0.829 | 3.60 | 1.146 | 4.97 | |
| Positive #3 | 240 | 0.86 | 0.058 | 6.74 | 0.063 | 7.33 | |
| Positive #4 | 240 | 0.44 | 0.015 | 3.44 | 0.042 | 9.61 | |
| Positive #5 | 240 | 0.48 | 0.022 | 4.67 | 0.032 | 6.79 | |
| Sample | N | Mean | Within-Run | Total | |||
| (kU/L) | SD (kU/L) | CV % | SD (kU/L) | CV % | |||
| Allergen = nBet v 1 (Lots 201, 202, 203) | |||||||
| Positive #1 | 240 | 2.40 | 0.101 | 4.19 | 0.108 | 4.49 | |
| Positive #2 | 240 | 9.19 | 0.299 | 3.25 | 0.353 | 3.84 | |
| Positive #3 | 240 | 23.50 | 0.784 | 3.34 | 0.998 | 4.25 | |
| Positive #4 | 240 | 0.38 | 0.017 | 4.41 | 0.022 | 5.85 | |
| Positive #5 | 240 | 0.41 | 0.017 | 4.08 | 0.028 | 6.76 | |
| Allergen = Cat Serum Albumin, Fel d 2 (Lots 201, 203, 204) | |||||||
| Positive #1 | 240 | 0.93 | 0.028 | 3.04 | 0.066 | 7.12 | |
| Positive #2 | 240 | 36.22 | 0.988 | 2.76 | 1.181 | 3.26 | |
| Positive #3 | 240 | 45.27 | 1.588 | 3.51 | 1.778 | 3.93 | |
| Positive #4 | 240 | 0.36 | 0.015 | 4.11 | 0.023 | 6.43 | |
| Positive #5 | 240 | 0.38 | 0.017 | 4.44 | 0.025 | 6.68 |
.
.
.
:
.
. .
،
·
·
{8}------------------------------------------------
:
Lot-to-Lot variability is summarized in the following tables for each of the 5 allergens.
{9}------------------------------------------------
Precision of nOle e 1, nArt v 1, Dog Serum Albumin (Can f 3), nBet v 1, and Cat Serum Albumin (Fel d 2) meet the criteria for acceptable intra- assay precision, inter- assay precision, and lot-to-lot precision. Intra-assay variability is less than 15% for all allergen lots, inter-assay precision is less than 15% for all allergen lots, and lot-to-lot precision is less than 20% for all 5 allergens. The average negative control remained negative (< 0.10 kU/L) throughout precision testing on all allergens.
{10}------------------------------------------------
Linearity
For each allergen, two samples were diluted in 2-fold serial dilution to five (5) levels. The undiluted (neat) and diluted samples were tested with the specific allergen to demonstrate linearity at concentrations within the assay limits. Regression statistics for each allergen comparing observed to expected data are presented below.
| Allergen | Regression Equation | N | Slope 95% CI | Intercept 95% CI | R | Range Tested(kU/L) |
|---|---|---|---|---|---|---|
| nOle e 1 | y = 0.997x + 0.0993 | 11 | 0.974 - 1.020 | 0.0234 - 0.1753 | 0.999 | 0.14 - 9.43 |
| nArt v 1 | y = 1.004x + 0.3830 | 12 | 0.972 - 1.035 | -0.1692 - 0.9352 | 0.998 | 1.11 - 40.00 |
| Dog SerumAlbumin | y = 0.999x - 0.0832 | 12 | 0.986 - 1.012 | -0.1906 - 0.0242 | 1.000 | 0.31 - 23.15 |
| nBet v 1 | y = 0.994x + 0.1429 | 12 | 0.972 - 1.016 | -0.0569 - 0.3428 | 0.999 | 0.60 - 22.24 |
| Cat SerumAlbumin | y = 0.975x + 0.0230 | 12 | 0.928 - 1.023 | -0.6815 - 0.7275 | 0.995 | 0.52 - 42.40 |
Linearity
Specificity (Inhibition) Studies
Specificity of each allergen was verified through competitive inhibition testing using a single serum sample or pool of sera. A negative sample was used to measure the background response.
To initiate the inhibition experiment, 70 uL of undiluted and 3-4 levels of 5-fold serially diluted inhibitor extract were mixed with 250 uL of sample or pool to achieve final inhibitor concentrations of 218.75, 43.75, 1.75, 0.35, 0.08, 0.07, 0.02, 0.01, 0.003 µg/mL. This mixture was incubated at room temperature (15-28 °C) for 1 hour allowing the immunological reaction to occur. Each sample mixture containing the inhibitor extract and the appropriate controls was assayed with 1 lot of each allergen. The percent (%) inhibition was calculated according to the following formula:
Image /page/10/Figure/7 description: This image shows a mathematical equation. The equation is a fraction with "(Response of pos. control (pos. sample – neg. sample) - sample response with inhibitor extract)" as the numerator and "(Response of pos. control (pos. sample - neg. sample))" as the denominator. The fraction is then multiplied by 100.
The inhibition plots demonstrate that the allergens tested are inhibited by the relevant inhibitor extract in a concentration dependent fashion. Also, the target % inhibition of 50% for the highest inhibitor concentration tested was met. These results indicate specificity of Native Ole e 1, Native Art v 1, Dog Serum Albumin, (Can f 3), Native Bet v 1 and Cat Serum Albumin, (Fel d 2) allergens.
{11}------------------------------------------------
Inhibition Using Negative Controls
Additional inhibitions studies were conducted to show that the specific allergens are not cross-reacting to unrelated allergens.
Procedures were followed according to CLSI ILA20-A2. Testing was performed using one positive sample with three unrelated allergen extracts at 1 mg/ml. A negative sample was used to measure the background response. Inhibition for each allergen with three unrelated extracts did not generate any significant inhibition. These results indicate specificity of nOle e 1, nArt v 1, Dog Serum Albumin (Can f 3), nBet v 1, and Cat Serum Albumin (Fel d 2) allergens.
Clinical Performance Studies
Clinical performance was demonstrated by testing serum samples against specific allergens from clinically diagnosed atopic and non-atopic individuals. Allergen-specific testing was obtained using the IMMULITE® 2000 3gAllergy™ assay.
| Clinical Diagnosis Data | |||
|---|---|---|---|
| IMMULITE® 2000 Analyzer | Clinical | Normal | Total |
| Positive | 148 | 21 | 169 |
| Negative | 45 | 479 | 524 |
| Total | 193 | 500 | 693 |
| 76.7% | 95.8% | 90.5% | |
| Sensitivity | Specificity | Agreement | |
| Lower Confidence Interval | 71% | 94% | 88% |
| Upper Confidence Interval | 83% | 98% | 93% |
| Allergens included: nOle e 1, nArt v 1, Dog Serum Albumin (Can f 3), nBet v 1, andCat Serum Albumin (Fel d 2). |
Data summary agreement of the IMMULITE® 2000 3gAllergy™ results to clinical data is presented in the table below.
IMMULITE® 2000 3gAllergy™ assay results for all allergens compare well with clinical documentation of presence or absence of signs, symptoms and other diagnostic evidence of allergen sensitivity
{12}------------------------------------------------
Conclusions for all Studies
Allergens including nOle e 1, nArt v 1, Dog Serum Albumin (Can f 3), nBet v 1 and Cat Serum Albumin (Fel d 2) for use with the IMMULITE® 2000 3gAllergy Specific IgE assay demonstrate acceptable analytical performance including precision, linearity and specificity. IMMULITE® 2000 assay results compare well with clinical documentation of presence or absence of signs, symptoms and other diagnostic evidence of allergen sensitivity. Substantial equivalence was demonstrated to clinical data, supporting the following intended use of the IMMULITE® 2000 3gAllergys Specific IgE assay and the five previously listed allergens:
For in vitro diagnostic use with the IMMULITE® 2000 Analyzer -- for the quantitative measurement of allergen-specific IgE in human serum, as an aid in the clinical diagnosis of IgE-mediated allergic disorders. The test results are to be used in conjunction with clinical findings and other laboratory tests.
As per 21 CFR 807.92(b)(3) the tests conducted for the IMMULITE® 2000 3gAllergy Specific IgE assay demonstrates that the device is substantially equivalent and safe and effective and performs as well as the predicate device.
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird, with its wings forming a flowing, abstract design.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Siemens Healthcare Diagnostics, Inc. c/o Mr. Garo Mimaryan Technical Regulatory Affairs Specialist III 511 Benedict Avenue Tarrytown, NY 10591
JAN 0 6 2012
Re: K112523
Trade/Device Name: IMMULITE® 2000 3gAllergy™ Specific IgE Assay Kit Regulation Number: 21 CFR §866.5750 Regulation Name: Radioallergosorbent (RAST) immunological test system Regulatory Class: Class II Product Code: DHB Dated: January 4, 2012 Received: January 5, 2012
Dear Mr. Mimaryan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed
{14}------------------------------------------------
Page 2 – Mr. Garo Mimaryan
predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
ia mchar
Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{15}------------------------------------------------
Indication for Use
510(k) Number (if known): K112523
Device Name: IMMULITE® 2000 3gAllergy™ Specific IgE Assay
Indication For Use:
For in vitro diagnostic use with the IMMULITE 2000 Analyzer - for the quantitative measurement of allergen-specific IgE in human serum, as an aid in the clinical diagnosis of IgE-mediated allergic disorders. The test results are to be used in conjunction with clinical findings and other laboratory tests.
Prescription Use V (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use ___ : (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Maria M Chan
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K112523
§ 866.5750 Radioallergosorbent (RAST) immunological test system.
(a)
Identification. A radioallergosorbent immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the allergen antibodies (antibodies which cause an allergic reaction) specific for a given allergen. Measurement of specific allergen antibodies may aid in the diagnosis of asthma, allergies, and other pulmonary disorders.(b)
Classification. Class II (special controls). The device, when intended to detect any of the allergens included in Table 1 in this paragraph, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 866.9. Table 1—Class II Exempt Allergens Under § 866.5750—Radioallergosorbent (RAST) Immunological Test Systems
Allergen code Allergen product Source (taxonomical name) Grass Pollens g1 Sweet vernal grass Anthoxanthum odoratum. g3 Cocksfoot grass, Orchard grass Dactylis glomerata. g4 Meadow fescue Festuca elatior. g5 Rye-grass (perennial rye grass) Lolium perenne. g7 Common reed (common reed grass) Phragmites communis. g8 Meadow grass, Kentucky blue (June grass) Poa pratensis. g9 Redtop, Bentgrass Agrostis stolonifera ,Agrostis gigantea (Agrostis alba ). g11 Brome grass Bromus inermis. g12 Cultivated rye (cultivated rye grass) Secale cereale. g13 Velvet grass Holcus lanatus. g14 Cultivated oat (cultivated oat grass) Avena sativa. g15 Cultivated wheat (cultivated wheat grass) Triticum aestivum (Triticum spp.). g16 Meadow foxtail (meadow foxtail grass) Alopecurus pratensis. g17 Bahia grass Paspalum notatum. g24 Wheat grass, Western Agropyron smithii (Elymus smithii ). g30 Bluegrass, annual Poa annua. g70 Wild rye grass Elymus triticoides Elymus condensatus. g71 Canary grass Phalaris arundinacea. g201 Barley, cultivated Hordeum vulgare. g202 Maize, corn (cultivated corn) Zea mays. g203 Salt grass Distichlis spicata. g204 False oat-grass Arrhenatherum elatius. g216 Cyn d 1 Cynodon dactylon. g701 Phl p 1.0102, Phl p 5.0101 Phleum pratense. g702 Phl p 7.0101 Phleum pratense. g703 Phl p 12.0101 Phleum pratense. Weed Pollens w2 Western ragweed Ambrosia psilostachya. w4 False ragweed Ambrosia acanthicarpa (Franseria acanthicarpa ). w5 Wormwood Artemisia absinthium Artemisia annua. w6 Mugwort Artemisia vulgaris. w7 Marguerite, ox-eye daisy Chrysanthemum leucanthemum. w8 Dandelion Taraxacum vulgare ,Taraxacum officinale. w9 Plantain (English), Ribwort Plantago lanceolata. w10 Goosefoot, lamb's quarters Chenopodium album. w11 Saltwort (prickly), Russian thistle Salsola kali (Salsola pestifer ). w12 Goldenrod Solidago virgaurea (Solidago spp.). w13 Cocklebur, common Xanthium commune. w14 Common pigweed (rough pigweed) Amaranthus retroflexus. w15 Scale, Lenscale Atriplex lentiformis. w16 Rough marsh elder Iva ciliate ,Iva annua. w17 Firebush (Kochia) Kochia scoparia. w18 Sheep sorrel Rumex acetosella. w19 Wall pellitory Parietaria officinalis. w20 Nettle (Common stinging nettle) Urtica dioica. w21 Wall pellitory Parietaria judaica. w22 Japanese hop (careless weed) Humulus japonicas (Humulus scandens ). w23 Yellow dock, Yellow dockweed Rumex crispus. w24 Spiny pigweed Amaranthus spinosus. w27 Carnation Dianthus spp. w28 Rose Rosa rugosa. w33 Clover Trifolium pratense. w35 Mexican tea Chenopodium ambrosioides. w36 Rabbit bush Ambrosia deltoidea (Franseria deltoides ). w37 Salt bush, annual Atriplex wrightii. w39 Water hemp, Western Amaranthus rudis (Acnida tamariscina ). w41 Burrobrush Hymenoclea salsola. w42 Poverty weed Baccharis neglecta. w43 Common sagebrush Artemisia tridentata. w45 Alfalfa Medicago sativa. w46 Dog fennel Eupatorium capillifolium. w53 Geranium Geranium spp. w67 Groundsel bush Baccharis halimifolia. w69 Iodine bush Allenrolfea occidentalis. w70 Ragweed, slender Ambrosia confertiflora. w75 Wing scale (wingscale) Atriplex canescens. w82 Careless weed Amaranthus palmeri ,Amaranthus hybridus. w90 Japanese hop Humulus japonicas (Humulus scandens ). w203 Rape (rape pollen) Brassica napus. w204 Sunflower Helianthus annuus. w206 Camomile Matricaria chamomilla. w207 Lupin Lupinus spp. w210 Sugar-beet Beta vulgaris. w211 Par j 2.0101 Parietaria judaica. w231 Art v 1 Artemisia vulgaris (Mugwort). w232 Sal k 1 Salsola kali. w233 Art v 3 Artemisa vulgaris (LTP, Mugwort). w234 Pla l 1 Plantago lanceolata. w235 Che a 1.0101 Chenopodium album. w236 Mer a 1.0101 Mercurialis annua. a753 Art v 1 Artemisia vulgaris (Mugwort weed). Tree Pollens t1 Box-elder (Maple) Acer negundo ,Acer saccharum. t2 Gray alder, speckled alder (alder) Alnus incana. t4 Hazel, hazelnut Corylus avellana ,Corylus americana. t5 American beech (beech) Fagus grandifolia (Fagus americana ). t6 Mountain juniper, Mountain cedar Juniperus ashei (Juniperus sabinoides ). t8 Elm Ulmus americana. t9 Olive Olea europaea. t10 Walnut Juglans californica ,Juglans nigra. t11 Maple leaf sycamore, London plane, Plane tree Platanus acerifolia. t61 Sycamore Platanus occidentalis. t12 Willow Salix caprea ,Salix nigra. t14 Cottonwood (Eastern Cottonwood/Black Cottonwood) Populus deltoides. t15 White ash Fraxinus americana. t16 White pine Pinus strobus. t18 Eucalyptus, gum-tree Eucalyptus globulus (Eucalyptus spp.). t19/t26 Acacia Acacia longifolia (Acacia spp.). t20 Mesquite Prosopis glandulosa/ Prosopis juliflora. t21 Melaleuca, cajeput tree Melaleuca quinquenervia (Melaleuca leucadendron ). t22 Pecan, hickory Carya illinoinensis (Carya pecan ). t23 Italian/Mediterranean/funeral cypress Cupressus sempervirens. t24 Japanese cypress Chamaecyparis obtusa (Chamaecyparis spp.). t25 Ash Fraxinus excelsior. t27 Maple, red Acer rubrum. t29 Acacia Acacia spp. t30 Birch, white Betula populifolia. t32 Willow, black Salix nigra. t33 Ash, Arizona Fraxinus velutina. t35 Cedar, salt Tamarix gallica. t37 Bald cypress (white bald cypress) Taxodium distichum. t38 Elm, Chinese/Siberian Ulmus pumila. t40 Hazelnut tree Corylus americana. t41 White hickory Carya alba (Carya tomentosa ). t42 Oak, red Quercus rubra. t43 Loblolly pine Pinus taeda. t44 Hackberry Celtis occidentalis. t45 Cedar elm Ulmus crassifolia. t47 Juniper, one seed Juniperus monosperma. t48 Pine, lodgepole Pinus contorta. t49 Pine, ponderosa Pinus ponderosa. t50 Beech, European Fagus sylvatica. t51 Tree of Heaven Ailanthus altissima. t52 Western white pine Pinus monticola. t54 Russian olive Elaeagnus angustifolia. t55 Scotch broom Cytisus scoparius. t56 Bayberry Myrica cerifera. t57 Red cedar Juniperus virginiana. t60 Western juniper Juniperus occidentalis. t61 Sycamore Platanus occidentalis. t70 Mulberry (white mulberry) Morus alba. t71 Red mulberry Morus rubra. t72 Queen palm Arecastrum romanzoffiamon. t73 Australian pine Casuarina equisetifolia. t77 Oak mix (red, white, black) Quercus spp. t80 Japanese cypress Chamaecyparis obtusa. t81 Japanese alder Alnus japonica. t83 Mango tree Mangifera indica. t90 Walnut, black Juglans nigra. t96 Poplar, white (poplar) Populus alba. t103/t218 Virginia live oak (live oak) Quercus virginiana. t105 Pepper tree Schinus molle. t110 Orange tree Citrus sinensis. t201 Spruce, Norway spruce Picea abies (Picea excelsa ). t202 Alder, smooth Alnus incana spp.Rugosa (Alnus rugosa ). t203 Horse chestnut Aesculus hippocastanum. t205 Elder Sambucus nigra. t206 Chestnut Castanea sativa. t207 Douglas fir Pseudotsuga menziesii (Pseudotsuga taxifolia ). t208 Linden Tilia cordata. t209 Horn beam Carpinus betulus. t210 Privet Ligustrum vulgare. t211 Sweet gum Liquidambar styraciflua. t212 Cedar Libocedrus decurrens. t213 Pine Pinus radiata. t214 Date palm Phoenix canariensis. t215 Lilac Syringa vulgaris. t217 Pepper tree Schinus molle. t217 Red alder Alnus rubra. t218 Virginia live oak Quercus virginiana. t218 Bayberry (bayberry/sweet gale) Myrica gale. t219 Palo verde Cercidium floridum. t219 Red cedar Juniperus virginiana. t220 Bet v 4 Betula verrucosa (Birch). t221 Bet v 2.0101, Bet v 4 Betula verrucosa (Birch). t222 Cypress (Arizona cypress) Cupressus arizonica. t223 Oil palm Elaeis guineensis. t224 Ole e 1 Olea europaea. t225 Bet v 6 Betula verrucosa (Birch). t226 Cup a 1 Cupressus arizonica. t227 Ole e 7 Olea Europaea. t228 Aspen, quaking Populus tremuloides. t229 Eastern hemlock Tsuga canadensis. t230 Redwood (sequoia) Sequoia sempervirens. t232 Pussy willow Salix discolor. t240 Ole e 9.0101 Olea Europaea. t241 Pla a 1.0101 Platanus acerifolia. t242 Pla a 2 Platanus acerifolia. t243 Pla a 3.0101 Platanus acerifolia. t244 Cor a 1.0103 Corylus avellana. t245 Aln g 1.0101 Alnus glutinosa. t246 Cry j 1 Cryptomeria japonica. t280 Locust tree Robinia pseudoacacia. t401 Brazilian peppertree Schinus terebinthifolius. t402 Mastic tree Pistacia lentiscus. t404 Tree of heaven Ailanthus altissima. t406 Date palm Phoenix dactylifera. a482 Ole e 1 Olea europaea (Olive Oil). Mites d207 Blo t 5.0101 Blomia tropicalis. d208 Lep d 2.0101 Lepidoglyphus destructor. Microorganisms, Molds, Yeast m1 Penicillium chrysogenum (Penicillium notatum ) Penicillium chrysogenum (Penicillium notatum ). m2 Cladosporium herbarum (Hormodendrum ) Cladosporium herbarum (Hormodendrum ). m3 Aspergillus fumigatus Aspergillus fumigatus. m4 Mucor racemosus Mucor racemosus. m5 Candida albicans Candida albicans. m7 Botrytis cinerea Botrytis cinerea. m8 Drechslera halodes (Setomelanomma rostrata, Helminthosporium halodes, Helminthosporium interseminatum ) Drechslera halodes (Setomelanomma rostrata, Helminthosporium halodes. m9 Fusarium moniliforme (Fusarium proliferatum ) Fusarium moniliforme (Fusarium proliferatum ). m10 Stemphylium botryosum Stemphylium herbarum (Stemphylium botryosum ). m11 Rhizopus nigricans Rhizopus nigricans. m12 Aureobasidium pullulans Aureobasidium pullulans. m13 Phoma betae Phoma betae. m14 Epicoccum purpurascens Epicoccum purpurascens (Epicoccum nigrum ). m15 Trichoderma viride Trichoderma viride. m16 Curvularia lunata Curvularia lunata ,Curvularia specifera ([K923044](https://510k.innolitics.com/device/[K923044](https://510k.innolitics.com/device/K923044)) ). m17 Cladosporium fulvum Cladosporium fulvum. m18 Fusarium culmorum Fusarium culmorum. m19 Aspergillus versicolor Aspergillus versicolor. m20 Mucor mucedo Mucor mucedo. m21 Aspergillus clavatus Aspergillus clavatus. m22 Micropolyspora faeni Saccharopolyspora rectivirgula (Micropolyspora faeni ). m23 Thermoactinomyces vulgaris Thermoactinomyces vulgaris. m24 Stachybotrys atra Stachybotrys chartarum (Stachybotrys atra ). m24 Paecilomyces spp Paecilomyces spp. m25 Aspergillus versicolor Aspergillus versicolor. m25 Penicillium brevicompactum Penicillium brevicompactum. m26 Cladosporium cladosporioides Cladosporium cladosporioides. m26 Penicillium citrinum Penicillium citrinum. m27 Penicillium spp Penicillium spp. m29 Aspergillus repens Aspergillus repens. m30 Penicillium roqueforti Penicillium roqueforti. m32 Cladosporium cladosporioides Cladosporium cladosporioides. m34 Serpula lacrymans Serpula lacrymans. m36 Aspergillus terreus Aspergillus terreus. m37 Trichophyton mentagrophytes Trichophyton mentagrophytes. m40 Aspergillus amstelodami Aspergillus amstelodami. m43 Saccharomyces Carlsberg Saccharomyces carlsbergensis. m44 Saccharomyces cerevisiae Saccharomyces cerevisiae. m45 Hormodendrum hordei Hormodendrum hordei. m46 Bipolaris spicifera Bipolaris spicifera. m47 Aspergillus nidulans Aspergillus nidulans. m48 Aspergillus oryzae Aspergillus oryzae. m49 Fusarium oxysporum Fusarium oxysporum. m50 Micropolyspora faeni Saccharopolyspora rectivirgula (Micropolyspora faeni ). m51 Thermoactinomyces vulgaris Thermoactinomyces vulgaris. m53 Microspora canis Microsporum canis (Microspora canis ). m54 Aspergillus flavus Aspergillus flavus. m63 Helminthosporium intersemin Helminthosporium intersemin. m66 Mucor plumbeus Mucor plumbeus. m67 Mycogone Mycogone perniciosa. m68 Nigrospora oryzae Nigrospora oryzae. m69 Rhodotorula Rhodotorula rubra (Rhodotorula mucilaginosa ). m70 Malassezia furfur (Pityrosporum orbiculare ) Malassezia furfur (Pityrosporum orbiculare ). m71 Spondylocladium Spondylocladium spp. m72 Epidermophyton Epidermophyton floccosum. m73 Epicoccum nigrum Epicoccum purpurascens (Epicoccum nigrum ). m80 Staphylococcal enterotoxin A (Sta a SEA ) Staphylococcus aureus. m80 Helminthosporium spp Helminthosporium spp. m81 Staphylococcal enterotoxin B (Sta a SEB ) Staphylococcus aureus. m88 Stemphylium solani Stemphylium solani. m93 Gliocladium fimbriatum Gliocladium fimbriatum. m94 Phycomyces blakesleeanus Phycomyces blakesleeanus. m201 Tilletia tritici (Ustilago nuda, Ustilago tritici ) (Barley smut) Tilletia tritici (Ustilago nuda, Ustilago tritici ). m202 Acremonium kiliense (Cephalosporium acremonium ) Acremonium kiliense (Cephalosporium acremonium ). m203 Trichosporon pullulans Trichosporon pullulans. m204 Ulocladium chartarum Ulocladium chartarum. m205 Trichophyton rubrum Trichophyton rubrum. m207 Aspergillus niger Aspergillus niger. m208 Chaetomium globosum Chaetomium globosum. m209 Penicillium frequentans Penicillium glabrum (Penicillium frequentans ). m209 Stachybotrys chartarum Stachybotrys chartarum (Stachybotrys atra ). m210 Trichophyton mentagrophytes var. goetzii Trichophyton mentagrophytes var. goetzii. m211 Trichophyton mentagrophytes var. interdigitale Trichophyton mentagrophytes var. interdigitale. m211 Oat smut Ustilago avenae. m212 Micropolyspora faeni Saccharopolyspora rectivirgula (Micropolyspora faeni ). m212 Geotrichum candidum Geotrichum candidum. m213 Bermuda grass smut Ustilago cynodontis. m214 Johnson grass smut Sphacelotheca cruenta. m215 Corn smut Ustilago maydis. m218 Asp f 1.0101 Aspergillus fumigatus. a3050 Asp r 1 Aspergillus restrictus. m219 Asp f 2 Aspergillus fumigatus. m220 Asp f 3.0101 Aspergillus fumigatus. m221 Asp f 4 Aspergillus fumigatus. m222 Asp f 6.0101 Aspergillus fumigatus. m223 Staphylococcal enterotoxin C (Sta a SEC ) Staphylococcus aureus. m224 Staphylococcal enterotoxin D (Sta a SED ) Staphylococcus aureus. m226 Staphylococcal enterotoxin TSST (Sta a TSST ) Staphylococcus aureus. m227 Malassezia spp. (Pityrosporum spp.) Malassezia spp. (Pityrosporum spp.). m228 Aspergillus flavus. m229 Alt a 1.0101 Alternaria alternata (Alternaria tenuis ). m230 Alt a 6.0101 Alternaria alternata (Alternaria tenuis ). m231 Cla h 8.0101 Cladosporium herbarum (Hormodendrum ). m300 Eurotium spp Eurotium spp. m304 Aspergillus oryzae Aspergillus oryzae. m305 Penicillium brevicompactum Penicillium brevicompactum. m309 Aspergillus terreus Aspergillus terreus. m310 Aspergillus nidulans Aspergillus nidulans. m311 Aspergillus flavus Aspergillus flavus. m312 Aspergillus clavatus Aspergillus clavatus. Epidermal & Animal e6 Guinea pig epithelium Cavia porcellus. e7 Pigeon droppings Columba palumbus ,Columba livia. e25 Chicken serum Gallus domesticus (Gallus gallus domesticus; Gallus spp.). e26 Parrot serum Psittacoidea spp. e62 Camel Camelus dromedaries. e70 Goose feathers Anser anser. e71 Mouse epithelium Mus musculus (Mus spp.). e73 Rat epithelium Rattus norvegicus. e74 Rat urine proteins Rattus norvegicus ,Rattus rattus. e75 Rat serum proteins Rattus norvegicus ,Rattus rattus. e76 Mouse serum proteins Mus musculus (Mus spp.). e77 Budgerigar droppings Melopsittacus undulatus. e78 Budgerigar feathers Melopsittacus undulatus. e79 Budgerigar serum proteins Melopsittacus undulatus. e80 Goat epithelium Capra hircus. e81 Sheep epithelium Ovis aries (Ovis spp.). e82 Rabbit epithelium Oryctolagus cuniculus (Oryctolagus spp.). e83 Swine epithelium Sus scrofa (Sus scrofa domesticus; Sus spp.). e84 Hamster epithelium Cricetus cricetus, Mesocricetus auratus, and Phodopus sungorus. e85 Chicken feathers Gallus domesticus (Gallus gallus domesticus; Gallus spp.). e86 Duck feathers Anas platyrhynchos. e87 Rat epithelium, serum proteins, and urine proteins Rattus norvegicus Rattus rattus. e88 Mouse epithelium, serum proteins, and urine proteins (mouse) Mus musculus (Mus spp.). e89 Turkey feathers Meleagris gallopavo. e90 Budgerigar serum proteins, feathers, and droppings Melopsittacus undulatus. e91 Pigeon serum proteins, feathers, and droppings Streptopelia roseogrisea ,Psittacidae spp. e92 Parrot serum proteins, feathers, and droppings Ara spp. e93 Pigeon serum proteins Streptopelia roseogrisea. e94 Fel d 1.0101 Felis domesticus. a345 Fel d 1 Felis domesticus. e98 Parrot droppings Psittacoidea spp. e101 Can f 1.0101 Canis familiaris (Canis domesticus ). a174 Can f 1 Canis familiaris (Canis domesticus ). e102 Can f 2.0101 Canis familiaris (Canis domesticus ). e196 Parakeet feathers Nymphicus hollandicus. e197 Parakeet droppings Nymphicus hollandicus. e198 Parakeet serum Nymphicus hollandicus. e199 Canary bird serum Serinus canarius. e200 Canary bird droppings Serinus canarius. e201 Canary bird feathers (Canary feathers) Serinus canarius. e202 Reindeer epithelium Rangifer tarandus. e203 Mink epithelium Mustela spp. e204 Bos d 6 Bos domesticus (Bos taurus; Bos spp.). e205 Horse, serum proteins Equus caballus (Equus spp.). e206 Rabbit, serum proteins Oryctolagus cuniculus (Oryctolagus spp.). e208 Chinchilla epithelium Chinchilla laniger. e209 Gerbil epithelium Meriones unguiculatus. e210 Fox epithelium Vulpes vulpes. e211 Rabbit, urine proteins Oryctolagus cuniculus (Oryctolagus spp.). e212 Swine, urine proteins Sus scrofa (Sus scrofa domesticus; Sus spp.). e213 Parrot feathers Ara spp. e214 Finch feathers Lonchura domestica. e215 Pigeon feathers Streptopelia roseogrisea (Streptopelia spp.),Columbia spp. e216 Deer epithelium Dama dama. e217 Ferret epithelium Mustela putorius. e218 Chicken droppings Gallus domesticus (Gallus gallus domesticus; Gallus spp.). e219 Chicken, serum proteins Gallus domesticus (Gallus gallus domesticus; Gallus spp.). e220 Fel d 2, Cat serum albumin Felis domesticus. e221 Can f 3 Canis familiaris (Canis domesticus ) (Dog serum albumin ). e222 Swine serum albumin (Sus s PSA) Sus scrofa (Sus scrofa domesticus; Sus spp.). e225 Lovebird feathers Psittacoidea agapomis. e226 Can f 5.0101 Canis familiaris. e227 Equ c 1.0101 Equus caballus. e228 Fel d 4.0101 Felis domesticus. e230 Equ c 3 Equus caballus. e231 Mus m 1 Mus musculus. Food f9 Rice Oryza sativa. f12 Pea (green pea) Pisum sativum. f15 White bean Phaseolus vulgaris. f19 Cayenne pepper Capsicum frutescens (Capsicum annum ). f21 Sugar cane Saccharum officinarum. f22 Raspberry Rubus idaeus. f26 Pork Sus scrofa (Sus scrofa domesticus; Sus spp.). f29 Watermelon Citrullus lanatus (Citrullus vulgaris ). f31 Carrot Daucus carota. f32 Oyster mushroom Pleurotus ostreatus. f33 Orange Citrus sinensis. f35 Potato Solanum tuberosum. f43 Mother's milk Homo sapiens. f44 Strawberry Fragaria vesca (Fragaria spp.). f45 Yeast, baker's Saccharomyces cerevisiae. f46 Pepper, Red Capsicum annuum. f47 Garlic Allium sativum. f48 Onion Allium cepa. f49 Apple Malus x domestica (Malus spp.). f51 Bamboo shoot Phyllostachys pubescens. f52 Cacao/chocolate Theobroma cacao. f54 Sweet potato Ipomoea batatas. f55 Common millet Panicum miliaceum. f56 Foxtail millet Setaria italica. f57 Japanese millet Echinochloa crus-galli. f58 Pacific squid Todarodes pacificus. f59 Octopus Octopus vulgaris (Octopus spp.). f63 Kefir NA. f67 Parmesan cheese NA. f81 Cheese, cheddar type NA. f82 Cheese, mold type NA. f83 Chicken Gallus domesticus (Gallus gallus domesticus; Gallus spp.). f86 Parsley Petroselinum crispum. f87 Melon Cucumis melo Cucumis melo + Citrullus lanatus. f88 Mutton (lamb) Ovis aries (Ovis spp.). f90 Malt Hordeum vulgare. f92 Banana Musa spp. f93 Cacao Theobroma cacao. f94 Pear Pyrus communis (Pyrus spp.). f97 Yam Dioscorea spp.Dioscorea opposita. f97 Chamomile tea Matricaria chamomilla. f98 Gliadin Triticum aestivum (Triticum spp.). f102 Cantaloupe Cucumis melo var. cantalupensis. f105 Chocolate Theobroma cacao. f109 Cottonseed Gossypium hirsutum. f110 Giant radish Raphanus sativus. f118 Zucchini Cucurbita pepo. f119 Radish Raphanus sativus. f120 Venison Capreolus capeolus. f121 Pinto bean Phaseolus vulgaris. f122 Cheese, American NA. f127 Black-eyed pea Vigna unguiculata. f131 Black Olive Olea europaea. f136 Red beet Beta vulgaris var. conditiva. f139 Goat's Cheese Capra aegagrus. f140 Bran NA. f141 Corn (vegetables) Zea mays. f152 Green bell pepper Capsicum annuum. f155 Brewer's yeast Saccharomyces carlsbergensis. f157 Duck Anas domesticus. f158 Goose Anser anser. f160 Camembert cheese NA. f162 Nectarine Prunus persica var. nucipersica. f163 Kohlrabi Brassica oleracea var. gongylodes. f65 Perch f166 Leek Allium porrum. f170 Cheese (Switzerland) (Swiss cheese) NA. f174 Fig Ficus carica. f177 Cranberry Vaccinium macrocarpon. f179 Raisin Vitis spp. f182 Lima bean Phaseolus lunatus. f198 Flaxseed (bruised grain) Linum usitatissimum. f199 Untreated native milk Bos domesticus (Bos taurus; Bos spp.). f208 Lemon Citrus limon. f209 Grapefruit Citrus paradisi. f210 Pineapple Ananas comosus. f211 Blackberry Rubus fruticosus. f212 Mushroom (champignon) Agaricus hortensis (Agaricus spp.). f213 Rabbit Oryctolagus cuniculus (Oryctolagus spp.). f214 Spinach Spinacia oleracea. f215 Lettuce Lactuca sativa. f216 Cabbage Brassica oleracea var. capitata. f217 Brussels sprouts Brassica oleracea var. gem. f218 Paprika, sweet pepper Capsicum annuum. f219 Fennel seed Foeniculum vulgare. f219 Sage Salvia officinalis. f220 Cinnamon Cinnamomum spp. f221 Coffee Coffea spp. f222 Tea Camellia sinensis. f223 Green olive Olea europaea. f225 Summer squash, pumpkin Cucurbita pepo. f225 Pumpkin Cucurbita maxima. f226 Pumpkin seed Cucurbita pepo. f227 Sugar-beet seed Beta vulgaris. f229 Safflower Seed Carthamus tinctorius. f231 Milk, boiled Bos domesticus (Bos taurus; Bos spp.). f234 Vanilla Vanilla planifolia. f237 Apricot Prunus armeniaca. f241 Gouda cheese NA. f242 Cherry Prunus avium. f244 Cucumber Cucumis sativus. f246 Guar, guar gum Cyamopsis tetragonoloba. f247 Honey NA. f248 Rosemary Rosmarinus officinalis. f254 Plaice Pleuronectes platessa. f255 Plum Prunus domestica ,Prunus americana. f258 Squid Loligo spp. f259 Grape Vitis vinifera (Vitis spp.). f260 Broccoli Brassica oleracea var. italica (Brassica oleracea var. cultivar ). f261 Asparagus Asparagus officinalis. f262 Aubergine, eggplant Solanum melongena. f263 Green pepper Piper nigrum ,Capsicum annuum. f264 Eel Anguilla anguilla. f265 Caraway Carum carvi. f265 Cumin Cuminum cyminum. f266 Mace Myristica fragrans. f267 Cardamon Elettaria cardamomum. f268 Clove Syzygium aromaticum. f269 Basil Ocimum basilicum. f270 Ginger Zingiber officinale. f271 Anise Pimpinella anisum. f272 Tarragon Artemisia dracunculus. f273 Thyme Thymus vulgaris. f274 Marjoram Origanum majorana. f275 Lovage Levisticum officinale. f276 Fennel, fresh Foeniculum vulgare. f277 Dill Anethum graveolens. f278 Bay leaf Laurus nobilis. f279 Chili pepper Capsicum frutescens. f280 Black pepper Piper nigrum. f281 Curry (Santa Maria) NA. f282 Nutmeg Myristica fragrans. f283 Oregano Origanum vulgare. f284 Turkey meat Meleagris gallopavo. f285 Elk/moose meat Alces spp. f286 Mare's milk Equus caballus (Equus spp.). f287 Red kidney bean Phaseolus vulgaris. f288 Blueberry Vaccinium myrtillus (Vaccinium spp.). f289 Date Phoenix dactylifera. f291 Cauliflower Brassica oleracea var. botrytis. f292 Guava Psidium guajava. f293 Papaya Carica papaya. f294 Passion fruit, Maracuja Passiflora edulis (Passiflora spp.). f295 Carambola Averrhoa carambola. f296 Carob Ceratonia siliqua. f297 Gum Arabic Acacia senegal (Acacia spp.). f298 Tragacanth Astragalus spp. f299 Sweet chestnut (chestnut) Castanea sativa. f300 Pinto bean Phaseolus spp. f301 Persimmon (kaki fruit, sharon) Diospyros kaki. f302 Mandarin (tangerine, clementine, satsumas) Citrus reticulata. f305 Fenugreek Trigonella foenum-graecum. f306 Lime Citrus aurantifolia. f307 Hake Merluccius merluccius. f308 Sardine (pilchard) Sardina pilchardus. f310 Blue vetch Lathyrus sativus. f311 Megrim Lepidorhombus whiffiagonis. f315 Green bean Phaseolus vulgaris. f316 Rape seed Brassica napus. f317 Coriander Coriandrum sativum. f318 Jack fruit Artocarpus heterophyllus. f319 Beetroot Beta vulgaris. f320 Crayfish Astacus astacus. f321 Horse meat Equus caballus (Equus spp.). f322 Red currant Ribes sylvestre. f324 Hop (fruit cone) Humulus lupulus. f325 Saffron Colchicum autumnale. f328 Fig Ficus carica. f329 Watermelon Citrullus lanatus. f330 Rose hip Rosa spp. f331 Saffron Crocus sativus. f332 Mint Mentha piperita. f333 Linseed Linum usitatissimum. f336 Jujube Ziziphus jujuba. f336 Wine vinegar Vitis vinifera (Vitis spp.). f337 Sole Solea solea. f337 English sole Parophrys vetulus. f338 Wine, white Vitis vinifera (Vitis spp.). f339 Allspice Pimenta dioica. f339 Wine, red Vitis vinifera (Vitis spp.). f341 Cranberry Vaccinium oxycoccus ,Vaccinium macrocarpon. f342 Olive (black, fresh) Olea europaea. f343 Raspberry Rubus idaeus. f344 Sage Salvia officinalis. f346 Chives Allium schoenoprasum. f347 Quinoa Chenopodium quinoa. f348 Litchi Litchi chinensis. f349 Chum salmon roe Oncorhynchus keta. f358 Artichoke Cynara scolymus. f360 Yogurt NA. f368 Black bass Micropterus dolomieu (Micropterus dolomieui ). f374 Karaya gum Sterculia urens. f375 Horseradish Armoracia rusticana. f377 Maple syrup NA. f379 Okra Abelmoschus esculentus. f382 Beet, sugar Beta vulgaris var. altissima. f401 Loquat Eriobotrya japonica. f402 Fig Ficus carica. f403 Brewer's yeast Saccharomyces cerevisiae. f405 Mint Mentha spp. f406 Arugula Eruca vesicaria. House Dust h1 Greer Labs., Inc NA. h2 Hollister-Stier Labs NA. h6 Japan NA. Venoms & Insects i7 Midge Chironomus yoshimatsui. i8 Moth Bombyx mori ,Heterocera spp. i47 Water flea Daphnia spp. i49 Deer fly Chrysops spp. i51 Black ant Camponotus pennsylvanicus. i54 Flea mix (dog/cat), common flea Ctenocephalides spp. i71 Mosquito Aedes communis ,Aedes spp.and Culex spp. i72 Green nimitti Cladotanytarsus lewisi. i73 Blood worm Chironomus thummi ,Chironomusri parius ,Chironomus spp. i75 European hornet Vespa crabro. i76 Berlin beetle Trogoderma angustum. i77 European paper wasp Polistes dominulus. i78 Fly Musca domestica. i80 Bumblebee Bombus pennsylvanicus. i201 Horse bot fly Gasterophilus intestinalis. i202 Grain weevil Sitophilus granarius. i203 Mediterranean flour moth Ephestia kuehniella (Anagasta kuehniella ). i204 Horse fly Tabanus spp. i205 Bumblebee Bombus terrestris. i208 Api m 1.0101 Apis mellifera. a45 Api m 1 Apis mellifera. i209 Ves v 5.0101 Vespula vulgaris. a670 Ves v 5 Vespula vulgaris. i210 Pol d 5.0101 Polistes dominulus. i211 Ves v 1.0101 Vespula vulgaris. i213 Api m 4 Apis mellifera. i214 Api m 2 Apis mellifera. i215 Api m 3 Apis mellifera. i216 Api m 5 Apis mellifera. i217 Api m 10 Apis mellifera. i220 Bla g 1.0101 Blattella germanica. i221 Bla g 2.0101 Blattella germanica. i222 Bla g 5.0101 Blattella germanica. i223 Bla g 7 Blattella germanica. a46 Api m 2 Apis mellifera. Miscellaneous o1 Cotton, crude fibers Gossypium spp. o3 Cotton (treated) Gossypium spp. o70 Seminal fluid Homo sapiens. o71 Staphylococcus aureus Staphylococcus aureus. o72 Pichia pastoris crude extract customer specific Pichia pastoris. o72 Sperm-sediment Homo sapiens. o73 Pichia pastoris crude extr. vector customer specific Pichia pastoris. o74 Pichia pastoris with vector customer specific Pichia pastoris. o201 Tobacco leaf, tobacco dust Nicotiana tabacum. o202 Artemia salina, fish feed Artemia salina. o203 Tetramin, fish feed NA. o207 Daphnia, fish feed Daphnia spp. o211 Mealworm Tenebrio molitor. o212 Streptavidin Streptomyces avidini. o213 MBP (maltose binding protein) Escherichia coli. o214 CCD; MUXF3 from bromelain Ananas comosus. o72 Enterotoxin A (Sta a SEA) Staphylococcus aureus. o73 Enterotoxin B (Sta a SEB) Staphylococcus aureus. Parasites p1 Ascaris Ascaris suum. p2 Echinococcus Echinococcus granulosus. p3 Schistosoma Schistosoma mansoni. p4 Anisakis (Herring Worm) Anisakis simplex (Anisakis spp.). p5 Toxocara canis Toxocara canis. p10 Ani s 3.0101 Anisakis simplex (Anisakis spp.). p11 Ani s 1 Anisakis simplex (Anisakis spp.). Occupational k4 Threshing dust NA. k5 Flax NA. k7 Hay Dust NA. k8 Hop (hops) Humulus lupulus. k12 Grain mill dust NA. k14 Kapok NA. k20 Sheep's wool (treated) (wool) Ovis aries (Ovis spp.). k21 Sheep's wool (Untreated) Ovis aries (Ovis spp.). k23 Straw Dust NA. k33 Oak NA. k70 Green coffee bean Coffea spp. k71 Castor bean Ricinus communis. k72 Ispaghula Plantago psyllium/Plantago ovata. k73 Silk waste NA. k74 Silk Bombyx mori. k75 Isocyanate TDI (Toluene diisocyanate) NA. k76 Isocyanate MDI (Diphenylmethane diisocyanate) NA. k77 Isocyanate HDI (Hexamethylen diisocyanate) NA. k78 Ethylene oxide NA. k79 Phthalic anhydride NA. k80 Formaldehyde/Formalin NA. k81 Ficus Ficus benjamina (Ficus spp.). k83 Cotton seed Gossypium hirsutum. k84 Sunflower seed Helianthus annuus. k85 Chloramin T NA. k86 Trimellitic anhydride, TMA NA. k87 Asp o 21, alpha-amylase Aspergillus oryzae. k89 Orris root Iris florentina. k99 HSA (Human Serum Albumin) (Hom s HSA) Homo sapiens. k201 Car p 1, Papain Carica papaya. k202 Ana c 2, Bromelain Ananas comosus. k204 Maxatase Bacillus licheniformis. k205 Alcalase Bacillus spp. k206 Savinase, Protease 1 (Bac l Subtilisin) Bacillus spp. k208 Gal d 4, Lysozyme Gallus domesticus (Gallus gallus domesticus; Gallus spp.). k209 Hexahydrophtalic anhydrid NA. k210 Maleic anhydride NA. k211 Methyltetrahydrophtalic anhydrid NA. k212 Abachi wood dust Triplochiton scleroxylon. k213 Pepsin (Sus s Pepsin) Sus scrofa (Sus scrofa domesticus; Sus spp.). k213 TCPA NA. k214 Bougainvillea Bougainvillea spp. k225 Horse radish peroxidase (Arm r HRP) Armoracia rusticana. k226 Ascorbate oxidase (Cuc p ascorbate oxidase) Cucurbita pepo. k301 Flour dust Triticum spp. k501 Savinase customer specific Proprietary knowledge of customer. k502 Lipolase customer specific Proprietary knowledge of customer. k503 Termamyl customer specific Proprietary knowledge of customer. k504 Clazinase customer specific Proprietary knowledge of customer.