K Number
K112342
Manufacturer
Date Cleared
2012-04-26

(255 days)

Product Code
Regulation Number
876.5980
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For use to manage fecal incontinence through the collection of liquid to semi-liquid stool and to provide access to administer medications.

Device Description

The Flexi-Seal® SIGNAL™ Fecal Management System is comprised of a soft catheter tube assembly, a Luer Syringe, a collection bag with filter and a cinch clamp to pinch off flow in the catheter when required for medication retention. The components are contained in a snap closed tray.

The catheter main drain tube is fabricated from collapsible silicone rubber. The drain tube has a low-pressure polyurethane or silicone retention balloon at the distal end and a connector for attaching a collection bag (provided with the device and separately) at the proximal end. There is a recess under the balloon for the clinician's finger allowing the device to be positioned digitally.

Two ports are attached to the side of the catheter. One port is used to inflate the retention balloon with water or saline after the device has been inserted into the patient's rectum. This port also provides a visual and tactile signal of when the low pressure retention balloon is filled to its optimal volume. The other port is used to flush the device if needed and administer medication, if prescribed.

A syringe (provided with the device) is used to fill and evacuate the retention balloon for insertion and removal.

The device, collection bay and syringe are intended for single us, are provided nonsterile, and have no components made with animal products, natural rubber latex or DEHP.

This 510(k) concerns modifications to the indications for use, materials and design featured with the predecessor device, the ConvaTec Fecal Management System (ref. K032734).

AI/ML Overview

The provided text outlines the performance summary of the Flexi-Seal® SIGNAL™ Fecal Management System. However, it does not include detailed acceptance criteria or a study design with specific metrics, sample sizes, expert qualifications, or ground truth establishment as would be typically found for AI/ML device evaluations.

Instead, the document focuses on demonstrating substantial equivalence to predicate devices through non-clinical testing. This means the device is compared against existing, legally marketed devices to show it performs similarly and is equally safe and effective.

Here's an analysis based on the provided text, addressing your questions where possible:


Acceptance Criteria and Device Performance

The document describes non-clinical testing performed to establish substantial equivalence.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria TypeReported Device Performance
Functional Equivalence"device's performance has been found to be substantially equivalent to the aforementioned predicate devices both functionally..." (implies performance related to managing fecal incontinence, collecting stool, and providing medication access, similar to predicates)
Structural Equivalence"...and structurally (material strength, catheter size, balloon size, etc.)" (implies device components meet comparable strength and dimensional specifications as predicates)
Biocompatibility"The device has also been evaluated for biocompatibility in accordance with the US Food and Drug Administration's guidance entitled Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing', issued May 1, 1995, and has been found safe in such respect for its intended use." (implies meeting ISO 10993 standards for biological safety)

Note: The document does not provide specific quantitative acceptance thresholds (e.g., "must collect X mL of stool per hour") or quantitative performance metrics (e.g., "collected Y mL of stool per hour"). The criteria are implicitly met by demonstrating "substantial equivalence" to predicate devices through non-clinical testing.

2. Sample size used for the test set and the data provenance
The document refers to "non-clinical testing" and states that "the device's performance has been found to be substantially equivalent." This implies laboratory or bench testing rather than clinical data from a "test set" of patient cases. Therefore, the concepts of data provenance (e.g., country of origin, retrospective/prospective) and a "test set" in the context of clinical data are not applicable here.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. There is no mention of a "test set" requiring expert ground truth, as the evaluation was based on non-clinical, functional, and structural testing.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. There is no "test set" or human review for this type of non-clinical evaluation.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a fecal management system, not an AI/ML diagnostic or prognostic tool that would involve human readers or AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This device is not an algorithm or AI.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
The "ground truth" for this device's evaluation is based on engineering and materials standards, functional performance specifications, and biocompatibility standards applied to the device and its predicate devices. For instance, material strength would be validated against established material properties, and biocompatibility against ISO 10993 standards.

8. The sample size for the training set
Not applicable. This is not an AI/ML device that requires a training set.

9. How the ground truth for the training set was established
Not applicable. This is not an AI/ML device that requires a training set.


In summary: The provided 510(k) summary focuses on demonstrating substantial equivalence to predicate devices through non-clinical functional, structural, and biocompatibility testing. It is not an AI/ML device, and therefore, many of the typical questions related to AI/ML device evaluation (such as test set sample sizes, expert review, training data, MRMC studies) are not applicable to this submission. The "study" proving the device meets acceptance criteria is the aggregate of the described non-clinical tests that showed comparable performance to the legally marketed predicate devices.

{0}------------------------------------------------

APR 2 6 2012

510(K) SUMMARY

.

. · · ·

Subject Device:Flexi-Seal® SIGNAL™ Fecal Management System
Date Prepared:4/24/2012
Applicant:ConvaTec Inc.200 Headquarters Park DriveSkillman, New Jersey 08558
Contact:Katrina FiedlerAssociate Director, US Regulatory AffairsConvaTec Inc.Tel: 908-904-2541Fax: 908-904-2235
Device Trade Name:Flexi-Seal® SIGNAL™ Fecal Management System
Classification Name:Gastrointestinal Tube and Accessories (ref. 21 CFR 876.5980; Product Code KNT)
Device Class:Class II

Predicate Devices:

Trade Name:Flexi-Seal® Fecal Management System - ConvaTec Inc.
Classification Name:Gastrointestinal Tube and Accessories (ref. 21 CFR 876.5980; Product Code KNT)
Device Class:Class II
510(k) Substantial Equivalence:K032734 – determined substantially equivalent on April 8, 2004
Trade Name:Indwelling Fecal Management System - Bowel Management Systems, LLC
Classification Name:Gastrointestinal Tube and Accessories (ref. 21 CFR 876.5980; Product Code KNT)
Device Class:Class II
510(k) Substantial Equivalence:K012113 – determined substantially equivalent on May 3, 2002

{1}------------------------------------------------

Device Description:

The Flexi-Seal® SIGNAL™ Fecal Management System is comprised of a soft catheter tube assembly, a Luer Syringe, a collection bag with filter and a cinch clamp to pinch off flow in the catheter when required for medication retention. The components are contained in a snap closed tray.

The catheter main drain tube is fabricated from collapsible silicone rubber. The drain tube has a low-pressure polyurethane or silicone retention balloon at the distal end and a connector for attaching a collection bag (provided with the device and separately) at the proximal end. There is a recess under the balloon for the clinician's finger allowing the device to be positioned digitally.

Two ports are attached to the side of the catheter. One port is used to inflate the retention balloon with water or saline after the device has been inserted into the patient's rectum. This port also provides a visual and tactile signal of when the low pressure retention balloon is filled to its optimal volume. The other port is used to flush the device if needed and administer medication, if prescribed.

A syringe (provided with the device) is used to fill and evacuate the retention balloon for insertion and removal.

The device, collection bay and syringe are intended for single us, are provided nonsterile, and have no components made with animal products, natural rubber latex or DEHP.

This 510(k) concerns modifications to the indications for use, materials and design featured with the predecessor device, the ConvaTec Fecal Management System (ref. K032734).

Intended use:

The Flexi-Seal® SIGNAL™ Fecal Management System is an indwelling fecal management catheter intended for use to manage fecal incontinence through the collection of liquid to semi-liquid stool and to provide access to administer medications.

Summary of Technological Characteristics:

The Flexi-Seal® SIGNAL™ Fecal Management System has the same intended use and indications for use but with the addition of the access to administer medication. The approximate geometry of the new device is the same as the predicate devices. The materials and construction are the same as the predicate devices with the exception of a polyurethane balloon version additional to the original silicone retention balloon. This just provides an alternate substitute material for the retention balloon. The only differences in operation between the Flexi-Seal® SIGNAL™ Fecal Management System and the predicate devices are the use of a port to gather samples rather than the bay coupling and the use of an external separate cinch clamp to retard waste flow when desired. One of the predicate devices Indwelling Fecal Management System (ref. K012113) uses an internal inflatable balloon to block flow and has a sampling port.

Summary of Performance (Non-Clinical Testing) Data:

{2}------------------------------------------------

K 112342
page 3 of 3

Non-clinical testing of the subject device for functional and structural parameters has been performed. In this testing, the device's performance has been found to be substantially equivalent to the aforcmentioned predicate devices both functionally and structurally (material strength, catheter size, balloon size, etc.). The device has also been evaluated for biocompatibility in accordance with the US Food and Drug Administration's guidance entitled Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing', issued May 1, 1995, and has been found safe in such respect for its intended use.

In conclusion, the subject device has been demonstrated as safe and effective and substantially equivalent to the predicate devices.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Katrina Fiedler Associate Director. US Regulatory Affairs ConvaTec Inc. 200 Headquarters Park Drive SKILLMAN NJ 08558

APR 2 6 2012

Re: K112342

Trade/Device Name: Flexi-Seal® SIGNALTM Fecal Management System (FMS) Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: KNT Dated: March 27, 2012 Received: March 29, 2012

Dear Ms. Fiedler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I vathar sthates that stequirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{4}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Bariano K. torkin

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications For Use

510(K) Number (if known): Not yet assigned

Device Name: Flexi-Seal® SIGNALTM Fecal Management System (FMS)

Indication for Use:

For use to manage fecal incontinence through the collection of liquid to semi-liquid stool and to provide access to administer medications.

Prescription Use XX (21 CFR 801 Subpart D) AND/OR

Over the Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Hubert Reum

Division Sign-Off vision of Reproductive, Gastro-Renal, Urological Devices 510(k) Number

§ 876.5980 Gastrointestinal tube and accessories.

(a)
Identification. A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).(b)
Classification. (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.