K Number
K112337

Validate with FDA (Live)

Date Cleared
2011-12-01

(108 days)

Product Code
Regulation Number
862.1660
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Randox Liquid Immunoassay Premium Controls Level 1, Level 2, Level 3 and Tri level are liquid controls developed for use in the quality control of the quantitative assays stated on the package insert. This in vitro diagnostic device is intended for prescription use only.

Device Description

Randox Liquid Immunoassay Premium Controls are manufactured at three levels. Level 1. Level 2 and Level 3. as well as a Tri Level product. The analyte concentrations in each of the three levels have been chosen to span a range that includes the chemically significant or medical decision level(s). The analyte concentrations have been reviewed by a panel of experts to ensure that the concentrations are clinically relevant for use in routine hospital laboratories.

AI/ML Overview

This document, K112337, describes the Randox Liquid Immunoassay Premium Controls. This is a quality control material and as such, the acceptance criteria and study proving it meets these criteria are focused on its stability and its equivalence to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Stability (Shelf Life)12 months shelf life claim
Equivalence to Predicate DeviceTesting results indicate the proposed device is substantially equivalent to the Bio-Rad Liquichek Immunoassay Plus Controls.

2. Sample Size Used for the Test Set and Data Provenance

The document states: "To assess the shelf life stability claims, real time stability studies of the device have been carried out. A shelf life stability claim of 12 months has been noted for the device, with 2 lots having achieved at least this claim, and tests ongoing to obtain real time data for 24 months."

  • Sample Size for Test Set: 2 lots were used to achieve the 12-month stability claim.
  • Data Provenance: The studies were "real time stability studies" conducted by Randox Laboratories Limited, based in the United Kingdom. This indicates the data is prospective and originates from the United Kingdom.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document mentions: "The analyte concentrations have been reviewed by a panel of experts to ensure that the concentrations are clinically relevant for use in routine hospital laboratories."

  • Number of Experts: A "panel of experts" is mentioned, implying more than one, but a specific number is not provided.
  • Qualifications of Experts: The document states their role was to ensure "clinically relevant" concentrations for "routine hospital laboratories." While their specific qualifications (e.g., radiologist with 10 years of experience) are not listed, their expertise is in clinical relevance for laboratory settings, likely implying clinical chemists, pathologists, or similar professionals with experience in laboratory medicine.

4. Adjudication Method for the Test Set

The document does not explicitly state an adjudication method for the "panel of experts" that reviewed the analyte concentrations. It simply says they "reviewed" them. For the stability studies and the substantial equivalence claim, the "adjudication" is based on meeting the predefined stability parameters and demonstrating comparable performance to the predicate, likely through quantitative comparisons rather than expert consensus on individual "cases."

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, an MRMC comparative effectiveness study was not done. This device is a quality control material, not an AI-powered diagnostic tool. The concept of "human readers improving with AI" is not applicable here.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Not applicable. This is a quality control material, not an algorithm. Performance is assessed on its inherent stability and accuracy in providing expected values for various analytes, which is its standalone function.

7. The Type of Ground Truth Used

  • For Analyte Concentrations: The ground truth for the target analyte concentrations in the control materials was established through a "panel of experts" who determined clinically relevant levels. This aligns with expert consensus regarding appropriate control ranges.
  • For Stability: The ground truth for stability is the ability of the control material to maintain its expected values over time under specified storage conditions. This is assessed through direct measurement and comparison to initial values or recognized standards over time.
  • For Substantial Equivalence: The ground truth for proving substantial equivalence to the predicate device is a direct comparison of performance characteristics (e.g., analyte values, stability profile) to the established performance of the legally marketed predicate device (Bio-Rad Liquichek Immunoassay Plus Controls).

8. The Sample Size for the Training Set

  • Not applicable. This device is a quality control material, not a machine learning model, so there is no "training set" in the conventional sense. The "training" or development of the product would involve formulating the controls to achieve desired analyte concentrations and stability, which isn't described in terms of a data set size like an algorithm.

9. How the Ground Truth for the Training Set was Established

  • Not applicable. As there is no training set for an algorithm, the concept of establishing ground truth for it does not apply. The development process focused on achieving specific, clinically relevant analyte concentrations and ensuring their stability.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "RANDO" in a bold, sans-serif font. The letters are black and have a slightly textured appearance. The "O" in "RANDO" is represented by two concentric circles.

Image /page/0/Picture/2 description: The image shows a close-up of a dark, textured surface that fills most of the frame. The texture appears granular and uneven, with variations in shading that suggest depth and complexity. A curved, dark line defines the edge of the surface, creating a sense of containment within the frame. The overall impression is one of density and roughness, with a focus on the tactile qualities of the material.

K112337

510(k) Summary

Safety and Effectiveness as Required by 21 CFR 807.92

Manufacturer and Submitter

Name: Randox Laboratories Limited

Address: 55 Diamond Road, Crumlin, County Antrim, BT29 4QY, United Kingdom.

Telephone: +44 (0) 28 9442 2413 Fax: +44 (0) 28 9445 2912 E-mail: marketing@randox.com

Device Name

Trade Name: Randox Liquid Immunoassay Premium Controls Levels 1, 2 and 3, and Tri Level

Common Name: Liquid Immunoassay Premium Controls Levels 1, 2 and 3, and Tri Level

Classification: Multianalyte Controls, All kinds (Assayed and Unassayed)

Product Code: JJY

Date of Summary Preparation 22nd November 2011

Predicate Devices

Bio-Rad Liquichek Immunoassay Plus Controls Levels 1, 2 and 3

Image /page/0/Picture/18 description: The image shows three different logos. The first logo has a crown on top of a cursive letter "e" with a bird-like figure in the middle, and the number "2007" below. The second logo is a circle with a checkmark inside, with the letters "SGS" below. The third logo is a square with a crown on top of a checkmark, with the letters "URAS" below.

Randox Laboratories Limited, 55 Diamond Road, Crumlin, County Antrim, BT29 4QY, United Kingdom T +44 (0) 28 9442 2413 F +44 (0) 28 9445 2912 E marketing@randox.com | www.randox.com

...

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "RANDOXX" in a bold, sans-serif font. The letters are filled with a textured pattern, giving them a slightly rough appearance. The two O's in the word are designed as concentric circles, adding a unique visual element to the text.

Image /page/1/Picture/1 description: The image shows a close-up of a dark, textured surface that fills most of the frame. The texture appears granular or speckled, with small, light-colored dots scattered across a darker background. The surface curves along the right and top edges, suggesting it might be part of a larger, rounded object. The lower-left corner of the image is a solid white, contrasting sharply with the textured area.

Device Description

Randox Liquid Immunoassay Premium Controls are manufactured at three levels. Level 1. Level 2 and Level 3. as well as a Tri Level product. The analyte concentrations in each of the three levels have been chosen to span a range that includes the chemically significant or medical decision level(s). The analyte concentrations have been reviewed by a panel of experts to ensure that the concentrations are clinically relevant for use in routine hospital laboratories.

Intended Use

The Randox Liquid Immunoassay Premium Controls, Level 1, Level 2, Level 3 and Tri Level are liquid controls developed for use in the quality control of quantitative assays stated in the package insert.

Similarity to Predicate Device

  • Both are assayed quality control serums. �
  • Both are intended to monitor the precision of laboratory testing procedures 0 for the analytes named in the product insert.
  • Both are in vitro diagnostic medical devices. .
  • Both devices are in liquid format and manufactured from human serum. .

Stability

Serum is stable for 7 days (when thawed) at +2 - +8°C. To assess the shelf life stability claims, real time stability studies of the device have been carried out. A shelf life stability claim of 12 months has been noted for the device, with 2 lots having achieved at least this claim, and tests ongoing to obtain real time data for 24 months.

Conclusion

Testing results indicate that the proposed device is substantially equivalent to the predicate device.

Randox Laboratories Limited, 55 Diamond Road, Crumlin, County Antrim, BT29 4QY: United Kingdom +44 (0) 28 9442 2413 F +44 (0) 28 9445 2912 E marketing@randox.com I www.randox.com

Image /page/1/Picture/16 description: The image contains three logos. The first logo has a crown on top of a cursive letter "e" and the number 2007 below it. The second logo is a circle with a checkmark inside and the letters "SGS" below it. The third logo has a crown on top of the letters "ULAL" and some lines below it.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with outstretched wings.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

Randox Laboratories Limited c/o Pauline Armstrong 55 Diamond Road Crumlin Antrim United Kingdom BT29 4QY

DEC - 1 2011

Re: K112337

Trade/Device Name: Randox Liquid Immunoassay Premium Controls, Level 1, level 3 and Tri Level Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material Regulatory Class: Class I, reserved Product Code: JJY Dated: October 4, 2011 Received: October 6, 2011

Dear Dr. Armstrong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{3}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical

Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance...

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm

Sincerely yours.

signature

Counney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): Unknown

Device Name: Randox Liquid Immunoassay Premium Controls Level 1, Level 2, Level 3 and Tri-Level

Indication For Use:

The Randox Liquid Immunoassay Premium Controls Level 1, Level 2, Level 3 and Tri level are liquid controls developed for use in the quality control of the quantitative assays stated on the package insert. This in vitro diagnostic device is intended for prescription use only.

Prescription Use > (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Rute Chale

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k)_112337

§ 862.1660 Quality control material (assayed and unassayed).

(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.