(192 days)
The Willow™ channels airflow noninvasively to a patient from a positive airway pressure (PAP) device or a bilevel system The Willow ™ is:
(1) to be used by adult patients (> 66lb / 30 kg) for the treatment of sleep disordered breathing, such as obstructive sleep apnea (OSA), for whom positive airway pressure has been prescribed.
(2) to be used for single-patient reuse in the home environment.
The 3B Willow™ is a mask interface, of the nasal pillow variety, that directs airflow from a positive pressure device to the patient's nose. The mask is held in place with adjustable headgear that straps the mask to the face.
The 3B WillowTM is a prescription device supplied non-sterile.
This document describes the 3B Willow™ Nasal Mask and its substantial equivalence to predicate devices, focusing on technological characteristics and performance bench testing.
1. Table of Acceptance Criteria and Reported Device Performance:
The document primarily focuses on demonstrating substantial equivalence to a predicate device (ResMed Mirage Swift II), rather than establishing novel acceptance criteria for a new device type. The "acceptance criteria" are implicitly defined by the reported performance of the predicate device.
| Performance Characteristic | Acceptance Criteria (Predicate: ResMed Mirage Swift II - K042403) | Reported Device Performance (3B Willow™ Nasal Mask) |
|---|---|---|
| Therapy Pressure | 4-20 cmH2O | 4-20 cmH2O |
| Intentional Leak (at 4 cmH2O) | 20 L/min | 20 L/min |
| Intentional Leak (at 10-12 cmH2O) | 37 L/min (at 12 cmH2O) | 31.5 L/min (at 10 cmH2O) |
| Intentional Leak (at 20 cmH2O) | 49 L/min | 45 L/min |
| Resistance (at 50 L/min) | 0.4 cmH2O | 1.0 cmH2O |
| Resistance (at 100 L/min) | 1.6 cmH2O | 3.1 cmH2O |
| Dead Space | 91 ml | 96 ml |
2. Sample Size Used for the Test Set and Data Provenance:
The document explicitly states: "To verify the substantial equivalence claim, the Willow™ was performance bench tested against the ResMed Mirage Swift (K042043)."
- Sample Size for Test Set: The document does not specify a numerical sample size for the bench testing. It refers to general "performance bench testing." For this type of device (nasal mask), this typically involves testing a small number of production units.
- Data Provenance: The data is generated from bench testing performed on the 3B Willow™ Nasal Mask and compared against the specifications of the ResMed Mirage Swift II. There is no indication of human subject data, country of origin, or retrospective/prospective nature as this is a physical device performance comparison.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
Not applicable. This is a bench test comparing physical performance characteristics of a medical device against a predicate device's specifications. It does not involve expert review or ground truth establishment in the clinical diagnostic sense. The "ground truth" here is the established and publicly available performance specifications of the predicate device.
4. Adjudication Method for the Test Set:
Not applicable. As this is bench testing of physical properties against technical specifications, there is no need for adjudication by experts. The measurements are objective.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This is a physical medical device (nasal mask) and not an AI/software as a medical device (SaMD) that would involve human readers or AI in diagnostic interpretation.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
Not applicable. This is a physical medical device.
7. The Type of Ground Truth Used:
The ground truth used for comparison is the published performance specifications and characteristics of the predicate device, ResMed Mirage Swift II (K042403), derived from its 510(k) submission and/or product documentation.
8. The Sample Size for the Training Set:
Not applicable. This is a physical medical device. There is no "training set" in the context of machine learning.
9. How the Ground Truth for the Training Set Was Established:
Not applicable. As there is no training set for an AI/ML model, there is no ground truth establishment for a training set.
{0}------------------------------------------------
Section 5: 510(k) Summary
3B Willow™ Nasal Mask
FEB 1 6 2012
Section 5:
510(k) SUMMARY [As required by 21 CFR §807.92(c)]
Date Prepared:
July 27, 2011
Official Contact:
Alex Lucio Managing Partner 3B Products, LLC 1142 N. Scenic Highway Lake Wales, FL 33853 Tel: (863) 676-5948 Email: alucio@3Bproducts.com
| Device Trade Name: | 3B Willow™ Nasal Mask | |
|---|---|---|
| Device Common Name/Classification Name: | Vented Nasal Mask | |
| Classification: | 21 CFR §868.5905, 73 BZD (CLASS II) | |
| Predicate Devices: | Manufacturer: | Resmed |
| Trade Name: | Mirage Swift II | |
| 510(k) Number: | K042403 |
Manufacturer: Respironics ComfortLite 2 Trade Name: 510(k) Number: K082558
Fisher & Paykel Manufacturer: Opus 360 Trade Name: K063036 510(k) Number:
Device Description:
The 3B Willow™ is a mask interface, of the nasal pillow variety, that directs airflow from a positive pressure device to the patient's nose. The mask is held in place with adjustable headgear that straps the mask to the face.
The 3B Willow™ is safe when used under the conditions and purposes intended as indicated the labeling provided with the product.
{1}------------------------------------------------
Section 5: 510(k) Summary
3B Willow™ Nasal Mask
The 3B WillowTM is a prescription device supplied non-sterile.
Intended Use:
The Willow™ channels airflow noninvasively to a patient from a positive
airway pressure (PAP) device or a bilevel system. The Willow™ is:
- (1) to be used by adult patients (> 66lb / 30 kg) for the treatment of sleep disordered breathing, such as obstructive sleep apnea (OSA), for whom positive airway pressure has been prescribed.
- (2) to be used for single-patient reuse in the home environment.
Contraindications:
None
Technological Characteristics Comparison:
A comparative table of the Willow TM alongside the three identified predicate devices indicates that the Willow Nasal Pillows is substantially equivalent with the other three with respect to the form, performance, materials, structures and the ventilation characteristics.
| Manufacturer | BMC | ResMed | Respironics | Fisher&Paykel |
|---|---|---|---|---|
| Model | WillowTMNasal PillowsSystem | Mirage SwiftII NasalPillowsSystem(K042403) | ComfortLiteTM 2System(K082558) | OpusTM 360 NasalPillows Mask(K063036) |
| Pic | Image: BMC Willow Nasal Pillows System | Image: ResMed Mirage Swift II Nasal Pillows System | Image: Respironics ComfortLite 2 System | Image: Fisher&Paykel Opus 360 Nasal Pillows Mask |
| Nasal pillows | Yes | Yes | Yes | Yes |
| With headgear | Yes | Yes | Yes | Yes |
| Latex free | Yes | Yes | Yes | Yes |
| Multi Size | Yes | Yes | Yes | Yes |
| Connector | 22mm | 22mm | 22mm | 22mm |
| TherapyPressure | 4-20cmH20 | 4-20cmH20 | 4-30cmH20 | 3-25cmH20 |
| Intentional | 4cmH20=20L/ | 4cmH20=20L | 4cmH20=22L/min | 5cmH20=22L/min |
{2}------------------------------------------------
| leak | min10cmH20=31.5L/min20cmH20=45L/min | /min12cmH20=37/min20cmH20=49L/min | 12cmH20=34L/min20cmH20=42L/min | 11cmH20=33L/min21cmH20=48L/min25cmH20=52L/min |
|---|---|---|---|---|
| Resistance | 1.0 cmH20 at 50L/min3.1 cmH20 at 100L/min | 0.4 cmH20 at 50L/min1.6 cmH20 at 100L/min | NA | 1.2 cmH20 at 50L/min5.4 cmH20 at 100L/min |
| Dead Space | 96ml | 91ml | 19-36ml(mask only) | 22ml(mask only) |
To verify the substantial equivalence claim, the Willow™ was performance bench tested against the ResMed Mirage Swift (K042043). The results of the performance bench testing, along with a description of the testing protocol are found at Appendix A. In terms of performance characteristics (i.e. passive exhalation port flow, resistance to flow, and dead space), the two devices are substantially equivalent.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
3B Products, LLC C/O Mr. Alex Lucio Managing Partner 1142 N. Scenic Highway Lake Wales, Florida 33853
FEB 1 6 2012
Re: K112271
Trade/Device Name: 3B Willow™ Nasal Mask Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator Regulatory Class: II Product Code: BZD Dated: February 4, 2012 Received: February 9, 2012
Dear Mr. Lucio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device · Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Page 2 - Mr. Lucio
{4}------------------------------------------------
Page 2 - Mr. Lucio
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to Sempilations ification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the I bailing) of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours,
h for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
3B Willow™ Nasal Mask
Section 4: Indications for Use
K1112271
Indications for Use
510(k) Number: K112271 Device Name: 3B Willow™ Nasal Mask
Indications For Use:
The Willow™ channels airflow noninvasively to a patient from a positive airway pressure (PAP) device or a bilevel system The Willow ™ is:
- (1) to be used by adult patients (> 66lb / 30 kg) for the treatment of sleep disordered breathing, such as obstructive sleep apnea (OSA), for whom positive airway pressure has been prescribed.
- (2) to be used for single-patient reuse in the home environment.
AND/OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________ × Prescription Use _ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
y. Schulther
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: 以上1227/
§ 868.5905 Noncontinuous ventilator (IPPB).
(a)
Identification. A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.(b)
Classification. Class II (performance standards).