(199 days)
The NaviAid™ ICVI device is an accessory to an endoscope and is intended to ensure positioning of a standard endoscope in the small intestine (i.e., an endoscope that is used for standard intestinal endoscopic visualization and has a compatible instrument channel as defined in the instructions for use)
The modified NaviAid™ ICVI is an on-demand disposable that is inserted through the instrument channel of the endoscope in order to enable advancement and positioning of a standard endoscope in the small intestine. The modified NaviAid™ ICVI Device comprises a disposable balloon system and an air supply control unit for inflating and deflating the balloon system. The role of the ICVI disposable is to facilitate advancement of a standard endoscope into the small intestine. The modified NaviAid™ ICVI disposable includes the ICVI Balloon and the ICVI inflation tube. The balloon is inflated by ambient air. Either the Air Supply Unit ("ASU") or Single Balloon Air Supply Unit ("SB ASU") may operate and control the inflation and deflation of the balloon through a foot-pedal. The balloon is connected to a dedicated inflation tube that runs inside the instrument channel of the endoscope, and is connected at its proximal (user) end to the ASU or SB ASU.
The ICVI balloon can be advanced ahead of the endoscope tip or pulled back through pushing/pulling action on the ICVI inflation tube at its proximal side, outside the patient's body. When the ICVI balloon is advanced and then inflated, it functions as a distal anchor, to which the endoscope tip is advanced, and the ICVI inflation tube serves as a rail that leads the endoscope as it is pushed towards the anchoring ICVI balloon.
The balloon and tube negligibly compromise the endoscope's flexibility, or its field of view. Additionally, the ICVI disposable negligibly compromise the maneuverability of the endoscope's tip and does not limit the usage of any standard endoscopy tools, such as biopsy forceps, snare, needle etc. The ICVI disposable can be pulled back at any time during the procedure in order to allow use of therapy tools.
The NaviAid™ ICVI disposable is intended for-single use, while the ASU and SB ASU are re-usable.
The provided text describes a 510(k) submission for the NaviAid™ ICVI device, focusing on its substantial equivalence to a predicate device and non-clinical performance data. It does not contain information about acceptance criteria or a study proving the device meets those criteria, as typically understood for AI/ML performance evaluation. The document confirms that clinical performance data was "Not Applicable" for this submission.
Therefore, many of the requested details cannot be extracted from the provided text.
Here is a breakdown of what can be extracted or inferred based on the document's content:
1. A table of acceptance criteria and the reported device performance
The document does not specify quantitative acceptance criteria or a table of reported device performance in the way a typical AI/ML study would. Instead, it refers to "design and performance specifications" being met through non-clinical tests.
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Sample size for test set: Not applicable, as clinical performance data was "Not Applicable" and only non-clinical (in vitro, mechanical) tests were performed.
- Data provenance: Not applicable.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
- Number of experts: Not applicable.
- Qualifications of experts: Not applicable.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- Adjudication method: Not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC study: No. The document explicitly states "Clinical Performance Data: Not Applicable." The device is an accessory to an endoscope, not an AI/ML diagnostic tool for human readers.
- Effect size: Not applicable.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Standalone study: No. The device is a physical accessory to an endoscope, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Type of ground truth: Not applicable in the context of clinical studies. For non-clinical tests, the "ground truth" would be established engineering specifications and physical measurements (e.g., balloon inflation, material properties, compatibility with endoscope channels).
8. The sample size for the training set
- Sample size for training set: Not applicable, as this is a physical medical device, not an AI/ML algorithm that requires a training set.
9. How the ground truth for the training set was established
- How ground truth for training set was established: Not applicable.
Summary of available information:
The document describes the NaviAid™ ICVI device's non-clinical performance testing. These tests focused on ensuring the device meets engineering specifications and functions as intended, rather than evaluating performance against a clinical ground truth or human reader performance.
-
Acceptance Criteria & Device Performance: The document generally states that "The non-clinical tests demonstrated that the modified NaviAid™ ICVI device meets its design and performance specifications." Specific criteria are implied by the types of tests conducted:
- ICVI-In Vitro Validation
- PU Balloon (material properties/functionality)
- AB Tip (material properties/functionality)
- Irrigation Cap (functionality)
- AB Compatibility with 3.2mm Working Channel using a Handle (fit and function)
- ICVI Tube Length (dimensional accuracy)
- Set Point Pressure Tolerance (for the SB ASU)
- Inflation/Deflation Cycles (durability/functionality)
- Electrical & Mechanical Safety, Electromagnetic Compatibility, Software Validation, Biocompatibility (compliance with recognized standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, FDA Guidelines, ISO 10993)
-
Study Proving Acceptance Criteria: The study described is a series of non-clinical performance tests conducted on the modified NaviAid™ ICVI device. These tests focused on the device's physical and mechanical properties, compatibility, and safety.
- The ASU and SB ASU (Air Supply Units) components' performance tests were referenced from previous 510(k) submissions (K060923 and K110291, respectively), indicating reliance on prior established performance data for shared components.
- The ICVI disposable's performance tests were also referenced from K101191.
- New tests for the modified device specifically addressed the ICVI-R model's rotation capability, new tip material, irrigation cap, and modified working length.
-
Clinical Performance Data: Explicitly stated as "Not Applicable," meaning no human clinical trials were conducted to establish safety or effectiveness for this 510(k) submission. The FDA cleared the device based on substantial equivalence to a predicate device and satisfactory non-clinical testing.
{0}------------------------------------------------
SECTION 5 - 510(K) SUMMARY OF SAFETY & EFFECTIVENESS
NAVIAID™ ICVI DEVICE
510(k) Number K_11Q1000
Applicant's Name:
| Company name: | Smart Medical Systems Ltd. |
|---|---|
| Address: | 10 Hayetsira Str. |
| Ra'anana 43663 | |
| ISRAEL | |
| Tel.: | +972-9-7444321 |
| Fax: | +972-9-7444543 |
| e-mail: | ayoselzon@smartmedsys.com |
Contact Person:
| Official Correspondent: | Adva Yoselzon |
|---|---|
| Company name: | Smart Medical Systems Ltd. |
| Address: | 10 Hayetsira Str.Ra'anana 43663ISRAEL |
| Tel.: | +972-9-7444321 |
| Fax: | +972-9-7444543 |
| e-mail: | avoselzon@smartmedsys.com |
Name of the device:
NaviAid™ ICVI device
Trade or proprietary name, if applicable:
NaviAid™ ICVI device
Common or usual name:
NaviAid™ ICVI Device
Establishment Registration No.:
3005261802
!
{1}------------------------------------------------
Classification Name:
Endoscope and accessories
Classification:
FDA has classified Endoscope and accessories devices as a Class II medical device, with product code ODC and 21 CFR classification code 876.1500. Review by the Gastroenterology/Urology Devices Panel.
Predicate Device:
The modified NaviAid™ ICVI device is substantially equivalent to the NaviAid™ ICVI device (manufactured by Smart Medical Systems Ltd. and the subject of 510(k) document no. K110291). A comparison table and detailed discussion are presented in Section 12 of this application.
Device Description:
The modified NaviAid™ ICVI is an on-demand disposable that is inserted through the instrument channel of the endoscope in order to enable advancement and positioning of a standard endoscope in the small intestine. The modified NaviAid™ ICVI Device comprises a disposable balloon system and an air supply control unit for inflating and deflating the balloon system. The role of the ICVI disposable is to facilitate advancement of a standard endoscope into the small intestine. The modified NaviAid™ ICVI disposable includes the ICVI Balloon and the ICVI inflation tube. The balloon is inflated by ambient air. Either the Air Supply Unit ("ASU") or Single Balloon Air Supply Unit ("SB ASU") may operate and control the inflation and deflation of the balloon through a foot-pedal. The balloon is connected to a dedicated inflation tube that runs inside the instrument channel of the endoscope, and is connected at its proximal (user) end to the ASU or SB ASU.
The ICVI balloon can be advanced ahead of the endoscope tip or pulled back through pushing/pulling action on the ICVI inflation tube at its proximal side, outside the patient's body. When the ICVI balloon is advanced and then inflated, it functions as a distal anchor, to which the endoscope tip is advanced, and the ICVI inflation tube serves as a rail that leads the endoscope as it is pushed towards the anchoring ICVI balloon.
{2}------------------------------------------------
The balloon and tube negligibly compromise the endoscope's flexibility, or its field of view. Additionally, the ICVI disposable negligibly compromise the maneuverability of the endoscope's tip and does not limit the usage of any standard endoscopy tools, such as biopsy forceps, snare, needle etc. The ICVI disposable can be pulled back at any time during the procedure in order to allow use of therapy tools.
The NaviAid™ ICVI disposable is intended for-single use, while the ASU and SB ASU are re-usable.
Intended Use / Indication for Use:
The NaviAid™ ICVI device is an accessory to an endoscope and is intended to ensure positioning of a standard endoscope in the small intestine (i.e., an endoscope that is used for standard intestinal endoscopic visualization and has a compatible instrument channel as defined in the instructions for use)
Comparison of Technological Characteristics with the predicate device:
The modified NaviAid™ ICVI device is similar to the NaviAid™ ICVI device (also manufactured by Smart Medical Systems Ltd. and the subject of 510(k) number K110291) regarding the intended use, technological characteristics, principle of operation, specifications, and safety requirements.
Both the NaviAid™ ICVI device and the modified NaviAid™ ICVI device are intended to facilitate penetration of standard endoscopes into the small intestine, while maintaining all the advantages of an endoscopic procedure, such as back-and-forth navigation, stopping propagation if needed, real-time operation, video imaging, instrument channel including biopsy and treatment. The ICVI disposable can be pulled back at any time during the procedure in order to allow use of therapy tools.
Both devices have the same basic technology and principle of operation, the utilization of a single balloon added to an endoscopy system that is alternately inflated and deflated in order to progress the endoscope and ensure positioning of the endoscope in the small intestine. Both devices are inserted on-demand through the instrument channel of the endoscope.
The NaviAid™ ICVI device includes 3 models (ICVI-F, ICVI-S and ICVI-B) that are compatible for different anatomies of the intestine. These 3 models are compatible with a 3.7mm instrument channel diameter. The modified ICVI Device includes an additional
{3}------------------------------------------------
model, the ICVI-R. The balloon of the ICVI-R may be rotated to fit a 3.2mm instrument channel diameter.
Balloon specifications of all models are the same.
The working length of the modified ICVI is slightly different; the working length of the NaviAid™ ICVI is 3000mm whereas the working length of the modified ICVI is 3500mm. The modified NaviAid™ ICVI device also includes an irrigation cap.
Both devices are supplied non-sterile, for single use only.
The NaviAid™ ICVI Tip is made out of PU whereas the Tip of the modified NaviAid™ ICVI is made out of either the same PU or Pebax as an alternative material. All other patient contact materials are the same.
Both the NaviAid™ ICVI and the modified NaviAid™ ICVI may be controlled by either the Air Supply Unit ("ASU") or Single Balloon Air Supply Unit ("SB ASU").
The ASU is identical between the NaviAid™ ICVI and the modified NaviAid™ ICVI. The set point pressure tolerance of the SB ASU included in the modified NaviAid™ ICVI device is slightly different.
All other components and parameters are the same with no change in function or user interface.
Both the ASU and SB ASU that control the inflation of the balloons comply with the electrical and mechanical safety testing requirements and the electromagnetic compatibility testing requirements for electronic medical devices.
Detailed comparison between the NaviAid™ ICVI device and the modified NaviAid™ ICVI device is provided in section12 of this application
Non-Clinical Performance Data
The performance tests include testing of the modified NaviAid™ ICVI Device.
The ASU of the modified NavAid™ ICVI is the same as the ASU of the NaviAid™ ICVI device and the NaviAid™ BGE (also manufactured by Smart Medical Systems), therefore performance tests for the ASU are described in detail in 510(k) number K060923 of the NaviAid™ BGE device.
The SB ASU of the NaviAid™ ICVI is similar to the SB ASU of the modified NaviAid™ ICVI device. The difference between them is the set point pressure tolerance. The set point pressure test is detailed in the performance testing summary. All other tests ર-4
{4}------------------------------------------------
performed for the SB ASU of the NaviAid™ ICVI device are also relevant for the SB ASU of the modified NaviAid™ ICVI device and are described in detail in 510(k) number K110291 of the NaviAid™ ICVI device.
The performance testing of the NaviAid™ ICVI disposable are described in 510(k) number K101191 of the NaviAid™ ICVI device
The following performance tests were conducted on the modified NaviAid™ ICVI device:
1.ICVI-In Vitro Validation
-
PU Balloon
-
AB Tip
-
Irrigation Cap
-
AB Compatibility with 3.2mm Working Channel using a Handle
-
ICVI Tube Length
-
Set Point Pressure Tolerance
-
Inflation/Deflation Cycles
Clinical Performance Data
Not Applicable
Conclusions Drawn from Non-Clinical and Clinical Tests:
The non-clinical tests demonstrated that the modified NaviAid™ ICVI device meets its design and performance specifications. Furthermore, the tests showed that the modified NaviAid™ ICVI device is easy to use and user friendly and does not cause damage to the intestine or to the endoscope.
The modified NaviAid™ ICVI device may be safely and effectively used in procedures in order to ensure the positioning of a standard endoscope in endoscopy of the small intestine and reach depths of the intestine that may not otherwise be accessible with a standard endoscope device.
{5}------------------------------------------------
Substantial Equivalence:
In summary, the modified NaviAid™ ICVI device is a direct derivative of the NaviAid™ ICVI device. The intended use, basic technology, principle of operation specifications and safety requirements are similar. The minor difference do not raise new questions of safety and effectiveness.
Consequently, the modified NaviAid™ ICVI device is substantially equivalent to the NaviAid™ ICVI device.
Performance Standards:
The NaviAid™ ICVI device complies with the voluntary recognized standards:
-
- Electrical & Mechanical Safety Testing (IEC 60601-1)
-
- Electromagnetic Compatibility Testing (IEC 60601-1-2)
-
- Software Validation (IEC 60601-1-4 & FDA Guidelines)
-
- Biocompatibility Testing (ISO 10993)
Note: For the ASU, performance tests relating to standards 1-3 are described in detail in 510(k) number K060923 of the NaviAidTM BGE device. For the SB ASU, performance tests relating to standards 1-3 are described in detail in 510(k) number K110291 of the NaviAid™ ICVI Device.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular fashion around the eagle.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Adva Yoselzon QA Manager and Director of Clinical & Regulatory Affairs Smart Medical Systems, Ltd. 10 Hayetsira Street RA'ANANA 43663 ISRAEL
FEB - 6 2012
Re: K112106
Trade/Device Name: NaviAid™ ICVI Device Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: ODC Dated: January 18, 2012 Received: January 23, 2012
Dear Ms. Yoselzon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{7}------------------------------------------------
Page 2 -
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Sincerely yours,
Benjamin K. Evanko
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
. Enclosure
{8}------------------------------------------------
INDICATIONS FOR USE STATEMENT SECTION 4
510(k) Number (if known): K112106/
Device Name:
NaviAid™ ICVI Device
Indications for use:
The NaviAid™ ICVI device is an accessory to an endoscope and is intended to ensure positioning of a standard endoscope in the small intestine (i.e., an endoscope that is used for standard intestinal endoscopic visualization and has a compatible instrument channel as defined in the instructions for use)
ﺮ ﻣﺴﺎ
| Prescription Use | |
|---|---|
| Use | OR |
| (Per 21 C.F.R. 801 Subpart D) |
Over-The-Counter
(Optional Format Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Aonm Wham
al, and
§ 876.1500 Endoscope and accessories.
(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.