K Number
K111996

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-09-14

(63 days)

Product Code
Regulation Number
862.1660
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ELITech Clinical Systems ISE CONTROL I and ELITech Clinical Systems ISE CONTROL II are control sera for in vitro diagnostic use in quality control of ELITech Clinical Systems ISE Na, K, Cl, Total CO2 on ELITech Clinical Systems Selectra analyzers equipped with ISE Module.

Device Description

ELITech Clinical Systems ISE CONTROL I and ELITech Clinical Systems ISE CONTROL I are two level quality control products consisting of lyophilized human serum with added constituents of purified biochemicals chemicals, therapeutic drugs , preservatives and stabilizers. ISE CONTROL I and ISE CONTROL II are prepared exclusively from the blood of donors tested individually and found to be negative for HbsAg and to antibodies to HCV and HIV according to FDA-approved methods.

AI/ML Overview

The provided document is a 510(k) summary for the ELITech Clinical Systems ISE CONTROL I and CONTROL II. It details the device's characteristics and compares it to a predicate device, concluding with a statement of substantial equivalence. However, the document does not explicitly present acceptance criteria in a table format with reported device performance data, nor does it describe a study to prove those criteria were met in the comprehensive manner requested.

The document is a regulatory submission demonstrating substantial equivalence to a predicate device, which is a different type of evaluation than an independent study focused on predefined acceptance criteria and performance metrics.

Based on the information available in the provided text, I can infer some aspects and highlight what is missing:

Inferred Information:

The core "acceptance criterion" in this context is "substantial equivalence" to the predicate device, the Biorad-Lyphochek Assayed Chemistry Control Level 1 and 2 (K040273). The conclusion explicitly states: "The performance data and other information demonstrate that the safety and effectiveness of these devices versus the predicate that the salety compromised, and that it met all acceptance criteria, demonstrating that the device is substantially equivalent to its respective predicate device."

This means the acceptance criteria are implicitly met if the new device performs similarly to the predicate device in terms of its intended use, format, handling, and stability for quality control of ISE Na, K, Cl, Total CO2 measurements.

Missing Information (which would be present in a dedicated study as requested):

  1. Table of Acceptance Criteria and Reported Device Performance: This would typically list specific quantitative or qualitative thresholds (e.g., "bias must be within X%", "CV must be less than Y%") and then present the ELITech device's actual performance against those thresholds. The current document only shows a side-by-side comparison of device characteristics with the predicate, not performance data.
  2. Sample Size Used for the Test Set and Data Provenance: The document does not describe a "test set" in the context of performance testing for acceptance criteria. The comparison is based on the characteristics and intended use.
  3. Number of Experts Used to Establish Ground Truth and Qualifications: Not applicable as this is a quality control material and not a diagnostic device requiring expert interpretation for ground truth.
  4. Adjudication Method: Not applicable for the same reason as above.
  5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: Not applicable. This is a quality control material, not a diagnostic device requiring human-in-the-loop performance evaluation.
  6. Standalone (Algorithm Only) Performance: Not applicable. This is a chemical control material.
  7. Type of Ground Truth Used: For quality control materials, the "ground truth" would be the assayed values provided by the manufacturer. The document implies that the device's performance is assessed by its ability to provide stable and reliable control values.
  8. Sample Size for the Training Set: Not applicable as this is not an AI/ML device requiring a training set.
  9. How the Ground Truth for the Training Set was Established: Not applicable.

Summary of available information based on the prompt's categories:

  1. A table of acceptance criteria and the reported device performance:

    • Acceptance Criteria (Inferred): Substantial equivalence to the predicate device (Biorad-Lyphochek Assayed Chemistry Control Level 1 and 2), meaning similar intended use, format, handling, and stability for quality control of ISE Na, K, Cl, Total CO2.
    • Reported Device Performance: The document provides a comparative table of characteristics, but not specific quantitative performance data (e.g., accuracy, precision) against predefined numerical acceptance criteria. The conclusion states that "performance data and other information demonstrate that...it met all acceptance criteria," but these specific data are not presented in the provided text.
    CharacteristicELITech Clinical Systems Device (ISE CONTROL I & ISE CONTROL II)Predicate Device (Biorad-Lyphochek Assayed Chemistry Control Level 1 and Level 2)
    Intended UseControl sera for in vitro diagnostic use in quality control of ELITech Clinical Systems ISE Na, K, Cl, Total CO2 on ELITech Clinical Systems Selectra analyzers equipped with ISE module.Assayed quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in the package insert. (More general, but includes Na, K, Cl, Total CO2 likely)
    FormatLyophilized human serumLyophilized human serum
    LevelsTwo levelsTwo levels
    Handling/RxReconstitute with 5 mL distilled/deionized water, dissolve contents completely by occasional gentle swirling within 30 minutes, avoid foam.Reconstitute with 5.0 mL distilled or deionized water. Replace stopper and allow to stand for 20 minutes, swirling occasionally.
    Stability (Unreconstituted)When stored at 2-8 °C and protected from light, stable until expiry date.Stable until expiration date at 2-8°C.
    Stability (Reconstituted)Between 2-8 °C: 7 days. Between -20 and -10 °C: 28 days (frozen once). Store tightly capped and protected from light.7 days when stored tightly capped at 2-8°C. 30 days at -10 to -20°C. Once thawed, do not refreeze.
    Device DescriptionLyophilized human serum with added constituents (purified biochemicals, therapeutic drugs, preservatives, stabilizers). Prepared from blood of donors tested negative for HbsAg, HCV, HIV by FDA-approved methods.(General description of predicate implies similar composition as assayed chemistry control)
  2. Sample size used for the test set and the data provenance: Not explicitly stated as a separate "test set" for performance metrics. The underlying data for supporting substantial equivalence would have been generated internally by the manufacturer, likely from a lab setting, but details are not provided.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable for a quality control material.

  4. Adjudication method: Not applicable.

  5. If a multi reader multi case (MRMC) comparative effectiveness study was done: No. This is not a diagnostic imaging device.

  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a QC material, not an AI algorithm.

  7. The type of ground truth used: For a quality control material, the ground truth would be the assigned values or reference values established by the manufacturer through rigorous analytical testing against reference methods or calibrators. The document mentions "performance data," which implies analytical studies were performed to establish these values and stability, but the details are not provided.

  8. The sample size for the training set: Not applicable (not an AI/ML device).

  9. How the ground truth for the training set was established: Not applicable.

In conclusion, the provided 510(k) summary focuses on demonstrating substantial equivalence based on device characteristics and intended use, rather than presenting a detailed "acceptance criteria" study with specific performance metrics against those criteria. The acceptance criterion is implicitly meeting the performance expectations of a similar, legally marketed predicate device.

{0}------------------------------------------------

..............................................................................................................................................................................

ﻤﺎ ﻫﺎ ﺳ

. कु

SECTION 7510(k) SummarySEP 14 2ELITech Clinical Systems Device(ISE CONTROL I&ISE CONTROL II)Predicate device(Biorad-Lyphochek Assayed Chemistry Control Level 1and Level 2 (K040273))
IntroductionAccording to the requirements of 21 CFR 807.92, the following informationprovides sufficient detail to understand the basis for a determination of sub-stantial equivalence.Intended useELITech Clinical Systems ISE CONTROL I and ISE CONTROL II are control sera for in vitro diagnostic use in quality control of ELITech Clinical Systems ISE Na, K, Cl, Total CO2 on ELITech Clinical Systems Selectra analyzers equipped with ISE module.Lyphochek Assayed Chemistry Control is intended for use as an assayed quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in the package insert.
The assigned 510(k) number is: K111996FormatLyophilized human serumLyophilized human serum
SubmitterAddressPhone numberFax numberELITechGroup SEPPIM S.A.S.Zone Industrielle, 61500 SEES, FRANCE+ 33 (0)2 33 81 21 00+ 33 (0)2 33 28 77 51LevelsTwo levelsTwo levels
ContactDate of PreparationValérie GOURDON (Email: v.gourdon@elitechgroup.com)July 8th, 2011HandlingCarefully open the vial, avoiding the loss of lyophilizate, and pipette in exactly 5 mL of distilled/deionized water. Carefully close the vial and dissolve the contents completely by occasional gentle swirling within 30 minutes avoiding the formation of foam.Using a volumetric pipet, reconstitute each vial with 5.0 mL of distilled or deionized water. Replace the stopper and allow the control to stand for 20 minutes, swirling occasionally.
Device namesStabilityPrior to reconstitution, when stored at 2-8 °C and protected from light, the controls are stable until the expiry date stated on the label.After reconstitution :Between 2-8 °C: 7 daysBetween -20 and -10 °C: 28 days (when frozen once)Note :Store control sera tightly capped and protected from light after reconstitution.When stored unopened the product is stable at until the expiration date at 2-8°C.Once the control is reconstituted all the analytes are stable 7 days when stored tightly capped at 2-8°C and 30 days at -10 to -20°C.Once thawed, do not refreeze.
REAGENTTrade/proprietary Name:Common or Usual Name:Device ClassClassification nameProduct codeELITech Clinical Systems ISE CONTROL I and ISE CONTROL II"ISE CONTROL I & ISE CONTROL II"Class ICFR 862.1660 - Quality control material (assayed and unassayed)JJY - multi-analyte controls, all kinds (assayed)
Predicate deviceBiorad-Lyphochek Assayed Chemistry Control Level 1 and 2.(K040273)
Device descriptionELITech Clinical Systems ISE CONTROL I and ISE CONTROL I are two levelquality control products consisting of lyophilized human serum with addedconstituents of purified biochemicals chemicals, therapeutic drugs , preserva-tives and stabilizers.ISE CONTROL I and ISE CONTROL II are prepared exclusively from theblood of donors tested individually and found to be negative for HbsAg and toantibodies to HCV and HIV according to FDA-approved methods.
Intended UseELITech Clinical Systems ISE CONTROL I and ISE CONTROL II are controlsera for in vitro diagnostic use in quality control of ELITech Clinical SystemsISE Na, K, Cl, Total CO2 on ELITech Clinical Systems Selectra analyzersequipped with ISE module.

.

{1}------------------------------------------------

ексендерінен жасының қалықты жасының тұрған қаласының құрақтық басырып тұрған басырып ба

1988

Comparison to Predicate device

1

ﺎ ﺗﻢ

Conclusion

The performance data and other information demonstrate that the safety and effectiveness of these devices versus the predicate that the salety compromised, and that it met all acceptance criteria, demonstrating that the device is substantially equivalent to its respective predicate device.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

ELITech Group e/o Debra Hutson 21720 231d Dr SE, Suite 150 Bothell, WA 98021

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

SEP 1 4 2011

Re: K111996 Trade Name: ELITech Clinical Systems ISE Control I and Control II Regulation Number: 21 CFR 862.1660 Regulation Name: Quality Control Material (Assayed and Unassayed) Regulatory Class: Class I, Reserved Product Codes: JJY Dated: September 6, 2011 Received: September 7, 2011

Dear Ms. Hutson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition. FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{3}------------------------------------------------

Page 2 -

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Viro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please one the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 800, 97) For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the responsible of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll the rich the Drission of Shirts 5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor ou/ndustry/dffail.htm.

Sincerely yours,

Signature

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Form

510(k) Number (if known): K ||| 99\0

ELITech Clinical Systems ISE CONTROL I Device Name: ELITech Clinical Systems ISE CONTROL II

Indications for Use:

ELITech Clinical Systems ISE CONTROL I and ISE CONTROL II are control sera for in vitro diagnostic use in quality control of ELITech Clinical Systems ISE Na, K, Cl, Total CO2 on ELITech Clinical Systems Selectra analyzers equipped with ISE Module.

Prescription Use X (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

AND/OR

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Richard Klinger

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K111996

Page 1 of

§ 862.1660 Quality control material (assayed and unassayed).

(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.