K Number
K111989
Device Name
CORONIS FUSION 4MP DL
Manufacturer
Date Cleared
2011-09-22

(71 days)

Product Code
Regulation Number
892.2050
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The Coronis Fusion 4MP DL (MDCC-4130) is intended to be used in displaying and viewing digital images, for review and analysis by trained medical practitioners. These devices must not be used in primary image diagnosis in mammography.
Device Description
Not Found
More Information

Not Found

Not Found

No
The summary describes a display device for viewing digital images and does not mention any AI or ML capabilities.

No.
The device is for displaying and viewing digital images for review and analysis, not for treating or preventing diseases or conditions.

No
The device is described as being for "displaying and viewing digital images, for review and analysis," but explicitly states, "These devices must not be used in primary image diagnosis in mammography." While it is used in a medical context for review, its exclusion from primary diagnosis indicates it's not a diagnostic device itself.

No

The device is described as a display device (Coronis Fusion 4MP DL (MDCC-4130)), which is a hardware component used for viewing images. The summary does not indicate it is solely software.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood or tissue, to detect diseases, conditions, or infections.
  • Device Function: The description states the device is for "displaying and viewing digital images" for review and analysis. This is a display device, not a test that analyzes biological samples.
  • Intended Use: The intended use is for viewing medical images, not for performing diagnostic tests on patient samples.

Therefore, the Coronis Fusion 4MP DL (MDCC-4130) is a medical image display device, not an in vitro diagnostic device.

N/A

Intended Use / Indications for Use

The Coronis Fusion 4MP DL (MDCC-4130) is intended to be used in displaying and viewing digital images, for review and analysis by trained medical practitioners. These devices must not be used in primary image diagnosis in mammography.

Product codes

LLZ

Device Description

Not Found

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

trained medical practitioners

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

Not Found

Key Metrics

Not Found

Predicate Device(s)

Not Found

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

0

Image /page/0/Picture/0 description: The image shows a circular seal with text around the perimeter and a stylized bird-like symbol in the center. The text around the perimeter is small and difficult to read, but it appears to be the name of an organization or agency. The bird-like symbol in the center is composed of three curved lines that converge at a point, with a small circle at the bottom representing the bird's head. The overall design is simple and clean, with a focus on conveying a sense of authority and professionalism.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Mr. Lieven De Wandel Officer Regulatory & Environmental Affairs Barco n.v. President Kennedypark 35 8500 Kortrijk BELGIUM

SEP ? 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2

Re: K111989

Trade/Device Name: 4MP DL (MDCC-4130) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: H Product Code: LLZ Dated: September 9. 2011 Received: September 12, 2011

Dear Mr. De Wandel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

1

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Pari 820). This letter will allow you to begin marketing your device as described in vour Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Viro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely Yours,

Mary Patel

Mary S. Pastel, Sc.D. · Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

2

510(k) Number (if known):

Device Name: Coronis Fusion 4MP DL (MDCC-4130)

Indications For Use:

.

The Coronis Fusion 4MP DL (MDCC-4130) is intended to be used in displaying and viewing digital images, for review and analysis by trained medical practitioners. These devices must not be used in primary image diagnosis in mammography.

Prescription Use × (Part 21 CFR 801 Subpart D)

OR

Over-The-counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Mary Pastel

Division Sigr Division of Radiological Device Division of Statuation and Safety >f In

K111989
.10K

Special 510(k) Submission Coronis Fusion 4MP DL - MDCC-4130