K Number
K111804
Device Name
INTELLISPACE PACS 4.X
Date Cleared
2011-07-28

(31 days)

Product Code
Regulation Number
892.2050
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
IntelliSpace PACS 4.x is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians. The system is a software package used with general purpose computing hardware to acquire, store, distribute, process and display images and associated data throughout a clinical environment. The software performs digital image processing, measurement, communication and storage. Lossy compressed mammographic images and digitized screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.
Device Description
IntelliSpace PACS 4.x is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians. The system is a software package used with general purpose computing hardware to acquire, store, distribute, process and display images and associated data throughout a clinical environment. The software performs digital image processing, measurement, communication and storage. IntelliSpace PACS 4.x supports receiving, sending, printing, storing and displaying studies received from the following modality types via DICOM: CT, MR, NM, US, XA, PET, DX, DR, RF, RT, MG, SC, VL, as well as hospital/radiology information systems.
More Information

No
The document describes a standard PACS system for image management and processing, with no mention of AI or ML capabilities. The performance studies focus on functional requirements, interoperability, and equivalence to a predicate device, not on AI/ML performance metrics.

No
This device is an image management system that acquires, stores, processes, and displays medical images. It does not directly treat or diagnose a disease or condition.

No

This device is an image management system (PACS) that acquires, stores, distributes, processes, and displays medical images. Its intended use is to manage images, not to provide a diagnosis independently. While it supports the interpretation process by displaying images, the primary interpretation is explicitly stated to require an FDA approved monitor (for mammography), implying the tool itself does not perform the diagnostic interpretation.

Yes

The device is explicitly described as a "software package" that is used "with general purpose computing hardware." It performs functions like image processing, measurement, communication, and storage, all of which are software-based operations. While it interacts with hardware (general purpose computing hardware and imaging modalities), the device itself is the software component.

Based on the provided text, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections. They are used to provide information for diagnosis, monitoring, or screening.
  • Device Function: The description clearly states that IntelliSpace PACS 4.x is an image management system. Its primary functions are to acquire, store, distribute, process, and display medical images from various imaging modalities (CT, MR, etc.). It works with images generated by other medical devices, not on biological samples.
  • Lack of Biological Sample Analysis: There is no mention of the device analyzing biological samples or performing any tests on bodily fluids or tissues.

The device is a Picture Archiving and Communication System (PACS), which is a type of medical imaging software used for managing and viewing medical images. This falls under the category of medical devices, but not specifically In Vitro Diagnostics.

N/A

Intended Use / Indications for Use

IntelliSpace PACS 4.x is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians.

The system is a software package used with general purpose computing hardware to acquire, store, distribute, process and display images and associated data throughout a clinical environment. The software performs digital image processing, measurement, communication and storage.

Lossy compressed mammographic images and digitized screen images must not be reviewed for primary image interpretations.

Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.

Product codes

LLZ

Device Description

IntelliSpace PACS 4.x is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians.

The system is a software package used with general purpose computing hardware to acquire, store, distribute, process and display images and associated data throughout a clinical environment. The software performs digital image processing, measurement, communication and storage.

IntelliSpace PACS 4.x supports receiving, sending, printing, storing and displaying studies received from the following modality types via DICOM: CT, MR, NM, US, XA, PET, DX, DR, RF, RT, MG, SC, VL, as well as hospital/radiology information systems.

Mentions image processing

Yes

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

CT, MR, NM, US, XA, PET, DX, DR, RF, RT, MG, SC, VL

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

trained professionals, including but not limited to physicians, nurses and medical technicians.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

The software design was verified to ensure design output met functional and operational requirements. The verification results showed that the proposed device met the expectations in safety, performance and reliability. Additionally, performance was evaluated and interoperability and compatibility testing was conducted to confirm conformance to DICOM, HL7 and IHE standards.

The validation testing was performed at the system level with production ready software under an actual or simulated environment in which the device is expected to be used. The validation testing covered software and risk management validation. The validation testing results confirmed that the proposed device met user needs and intended use.

Testing included regression testing to legacy features and demonstrated substantial equivalence to earlier versions of this product.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K063267

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

0

K111804
Page 1 of 3

510(k) Summary

JUL 2 8 2011

6.0 510(k) Summary

In accordance with the provisions of the Safe Medical Device Act of 1990, Philips Healthcare Informatics, Inc. is providing a summary of Safety and Effectiveness information regarding the IntelliSpace PACS 4.x software.

6.1 Company Identification

Philips Healthcare Informatics, Inc.

4100 East Third Avenue, Suite 101

Foster City, CA 94404

Registration Number: 2954704

6.2 Contact Person

Victoria S. Mackinnon Director of Quality and Regulatory Telephone: 650-293-2373

Fax: 650-293-2301

6.3 Preparation Date

April 25, 2011

6.4 ldentification of Product and Classification

Device Trade Name: IntelliSpace PACS 4.x

Classification Name: Picture Archiving and Communications System

Classification Panel: Radiology

CRF Section: 892.2050

Device Class: II

Product Code: LLZ

6.5 Substantial Equivalence

Manufacturer: Stentor, Inc., a Philips Medical Systems Company

Model: iSite PACS v4.x

510(k) Number: K063267

Philips Healthcare Informatics 510(k) submission IntelliSpace PACS 4.x

1

K111804
Page 2 of 3
510(k) Summary

6.6 Device Description and Intended Use

IntelliSpace PACS 4.x is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians.

The system is a software package used with general purpose computing hardware to acquire, store, distribute, process and display images and associated data throughout a clinical environment. The software performs digital image processing, measurement, communication and storage.

IntelliSpace PACS 4.x supports receiving, sending, printing, storing and displaying studies received from the following modality types via DICOM: CT, MR, NM, US, XA, PET, DX, DR, RF, RT, MG, SC, VL, as well as hospital/radiology information systems.

6.7 Software Development

Philips Healthcare Informatics certifies that the IntelliSpace PACS 4.x software is designed, developed, tested and validated according to written procedures. These procedures identify individuals within the organization responsible for developing and approving product specifications, coding and testing, validation testing and field maintenance. The software developed for this product is used to provide diagnostic-quality images and associated information to the intended users.

6.8 Performance Testing

The software design was verified to ensure design output met functional and operational requirements. The verification results showed that the proposed device met the expectations in safety, performance and reliability. Additionally, performance was evaluated and interoperability and compatibility testing was conducted to confirm conformance to DICOM, HL7 and IHE standards.

The validation testing was performed at the system level with production ready software under an actual or simulated environment in which the device is expected to be used. The validation testing covered software and risk management validation. The validation testing results confirmed that the proposed device met user needs and intended use.

Testing included regression testing to legacy features and demonstrated substantial equivalence to earlier versions of this product.

6.9 Safety and Effectiveness

General Safety and Effectiveness Concerns:

The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device. It is the user's responsibility to ensure that Philips Healthcare Informatics Page 6-2 of 6-3

510(k) submission IntelliSpace PACS 4.x

2

K111804
Page 3 of 3
510(k) Summary

display quality, environmental lighting and other possible distractions are consistent with the clinical environment. The hardware components specified are all "off the shelf" computer components.

Conclusion:

IntelliSpace PACS 4.x is substantially equivalent in design and intended use to the predicate device, iSite PACS v4.x, which includes diagnostic radiological workstations, PACS and image management. Any differences between the IntelliSpace PACS 4.x software and the predicate device have no significant influence on safety or effectiveness. Therefore, IntelliSpace PACS 4.x raises no new issues of safety or effectiveness from the predicate device.

3

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Phillips Healthcare Informatics % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313

JUL 28 2011

Re: K111804

Trade/Device Name: IntelliSpace PACS 4.x Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: June 24, 2011 Received: June 27, 2011

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

4

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours.

Mary Pastel

Mary S. Pastel. Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

5

Indications for Use Form . JA CDRH DMC

JUN 2 7 2011

Received

Indications for Use Form 5.0

(111804 510(k) Number (if known): Unknown

Device Name: IntelliSpace PACS 4.x

Indications for Use:

IntelliSpace PACS 4.x is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians.

The system is a software package used with general purpose computing hardware to acquire, store, distribute, process and display images and associated data throughout a clinical environment. The software performs digital image processing, measurement, communication and storage.

Lossy compressed mammographic images and digitized screen images must not be reviewed for primary image interpretations.

Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K111804

Page 1 of

Philips Healthcare Informatics 510(k) submission IntelliSpace PACS 4.x

Page 5-1 of 5-1