(53 days)
The StaXx® XD System is a vertebral body replacement device intended for use in the thoracolumbar spine (T1-L5) to replace and restore height of a collapsed, damaged, or unstable vertebral body or portion thereof, due to tumor or trauma (i.e., fracture). The system is to be placed bilaterally and used with autograft or allograft and supplemental spinal fixation. The supplemental fixation system that is intended to be used with the StaXx® XD System is the CapSure® PS Spine System.
The StaXx® XD System is composed of wafers that are stacked into an expandable implant to adjust the height of the implant. The implant components are manufactured from PEEK-OPTIMA with 6% Barium Sulfate and may contain tantalum markers for additional visualization under fluoroscopy.
The provided text is related to a 510(k) submission for a medical device called the StaXx® XD System, which is a vertebral body replacement device. The purpose of this specific submission (K111469) is stated as a modification to the labeling and "minor modifications have been made to the surgical gun used to deliver the StaXx® XD Expandable Device."
Crucially, the document explicitly states: "No testing is required for an update to the labeling." This means that the submission does not detail specific acceptance criteria or a study proving the device meets those criteria, as it's primarily a labeling update and minor modification to a delivery tool, not a new device or significant design change requiring extensive performance validation.
Therefore, many of the requested details about acceptance criteria and a study cannot be extracted directly from this document.
However, I can provide what is available or inferable from the document:
1. A table of acceptance criteria and the reported device performance:
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Substantial Equivalence to Predicate Devices (for the modified labeling and minor surgical gun changes) | The device is "substantially equivalent" to the predicate devices (K052670, K090315, K101288, K102682) as determined by the FDA. |
| Safety and Effectiveness (implied from substantial equivalence) | No new issues of safety or effectiveness are presented by the modifications. |
| Same Intended Use and Indications for Use as Predicates | The Intended Use and Indications for Use are identical to the predicate StaXx® XD System. |
| Same Materials as Predicates | The device is manufactured from the same materials (PEEK-OPTIMA with 6% Barium Sulfate, optional tantalum markers) as the predicate. |
| Same Technological Characteristics as Predicates | The device maintains the same technological characteristics as the predicate. |
2. Sample size used for the test set and the data provenance:
- Sample Size: Not applicable. The document states "Non-clinical testing was performed to demonstrate that the subject StaXx® XD System is substantially equivalent to the listed predicate device. No testing is required for an update to the labeling." This indicates that specific performance testing for this submission, which would include a "test set," was not deemed necessary due to the nature of the changes (labeling and minor surgical gun modifications).
- Data Provenance: Not applicable for this specific submission as detailed test data is not provided. The previous 510(k) submissions for the predicate devices would contain such information.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. As noted above, specific performance testing with a "test set" and corresponding ground truth establishment was not performed for this submission.
4. Adjudication method for the test set:
- Not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is a surgical implant, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not applicable. This device is a surgical implant.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not applicable for this submission. The "ground truth" for demonstrating substantial equivalence primarily relied on comparison to legally marketed predicate devices based on their established performance data and the lack of new safety or effectiveness issues with the minor changes.
8. The sample size for the training set:
- Not applicable. This submission does not involve machine learning or AI models that require training sets.
9. How the ground truth for the training set was established:
- Not applicable.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the words "Spine Wave" in a bold, sans-serif font. The words are arranged horizontally, with "Spine" on the left and "Wave" on the right. A curved line runs beneath the words, resembling a wave.
JUL 19 2011
510(k) Summary StaXx® XD System Surgical Gun
1. Submitter Information
| Submitter: | Spine Wave, Inc. |
|---|---|
| Address: | Three Enterprise DriveSuite 210Shelton, CT 06484 |
| Telephone: | 203-712-1847 |
| Telefax: | 203-944-9493 |
| Contact: | Denise Duchene |
| Date Prepared: | June 30, 2011 |
2. Device Information
| Trade Name: | S |
|---|---|
| Common Name: | V |
| Classification Name: | S |
| Classification/Code: | C |
StaXx® XD System /ertebral Body Replacement Spinal Intervertebral Body Fixation Orthosis Class II per 21 CFR 888.3060; MQP
3. Purpose of Submission
The purpose of this submission is to describe a modification to the labeling for the StaXx® XD System. In addition minor modifications have been made to the surgical gun used to deliver the StaXx® XD Expandable Device.
4. Predicate Device Information
The StaXx® XD System described in this submission is substantially equivalent to StaXx® XD System included in the following 510(k)s:
| Predicate Device | Manufacturer | 510(k) No. |
|---|---|---|
| StaXx® XD System | Spine Wave, Inc. | K052670 |
| StaXx® XD System | Spine Wave, Inc. | K090315 |
| StaXx® XD System | Spine Wave, Inc. | K101288 |
| StaXx® XD System | Spine Wave, Inc. | K102682 |
{1}------------------------------------------------
5. Device Description
The StaXx® XD System is composed of wafers that are stacked into an expandable implant to adjust the height of the implant. The implant components are manufactured from PEEK-OPTIMA with 6% Barium Sulfate and may contain tantalum markers for additional visualization under fluoroscopy.
6. Intended Use
The StaXx® XD System is a vertebral body replacement device intended for use in the thoracolumbar spine (T1-L5) to replace and restore height of a collapsed, damaged, or unstable vertebral body or portion thereof, due to tumor or trauma (i.e., fracture). The system is to be placed bilaterally and used with autograft or allograft and supplemental spinal fixation. The supplemental fixation system that is intended to be used with the StaXx® XD System is the CapSure® PS Spine System.
7. Comparison of Technological Characteristics
The substantial equivalence of the subject StaXx® XD System is shown by similarity in intended use, indications for use, materials and performance to the cited predicate device. The StaXx® XD System and the predicate devices are manufactured from the same materials and have the same indications and technological characteristics.
8. Performance Data
Non-clinical testing was performed to demonstrate that the subject StaXx® XD System is substantially equivalent to the listed predicate device. No testing is required for an update to the labeling.
9. Conclusion
Based on the indications for use, technological characteristics, performance testing and comparison to predicates, the subject StaXx® XD System has been shown to be substantially equivalent to the predicate device identified in this submission, and does not present any new issues of safety or effectiveness.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wavy lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Spine Wave, Inc. % Ms. Denise Duchene Sr. Manager, Regulatory Affairs Three Enterprise Drive, Suite 210 Shelton, Connecticut 06484
JUL 19 2011
Re: K111469
Trade/Device Name: StaXx® XD System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis -Regulatory Class: Class II Product Code: MOP Dated: June 30, 2011 Received: July 01, 2011
Dear Ms. Duchene:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
Page 2 - Ms. Denise Duchene
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
erely yours,
Mark N. Melkerson // Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known):
StaXx® XD System Device Name:
Indications for Use:
The StaXx® XD System is a vertebral body replacement device intended for use in the thoracolumbar spine (T1-L5) to replace and restore height of a collapsed, damaged, or unstable vertebral body or portion thereof, due to tumor or trauma (i.e., fracture). The system is to be placed bilaterally and used with autograft or allograft and supplemental spinal fixation. The supplemental fixation system that is intended to be used with the StaXx® XD System is the CapSure® PS Spine System.
Prescription Use __ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign Off)
Page 1 of 1
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number_________________________________________________________________________________________________________________________________________________________________
§ 888.3060 Spinal intervertebral body fixation orthosis.
(a)
Identification. A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.(b)
Classification. Class II.