K Number
K111391
Date Cleared
2012-09-26

(497 days)

Product Code
Regulation Number
880.2800
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The STERRAD® 100NX® DUO Cycle Test Pack is used for routine monitoring of the STERRAD® 100NX® DUO Sterilization Cycle and is also used for the periodic testing of a STERRAD® 100NX® System DUO Cycle, using hospital-defined loads containing devices that do not exceed claims of the cycle. The STERRAD® 100NX® DUO Cycle Test Pack consists of a STERRAD® CYCLESURE® 24 Biological Indicator, vial and cap to hold the BI.

Device Description

The STERRAD® 100NX® DUO Test Pack consists of a CYCLESURE® 24 Biological Indicator (BI) and a STERRAD® NX® Test Vial with cap for holding the BI during the sterilization cycle.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and the study that proves the device meets those criteria:

Acceptance Criteria and Device Performance

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Resistive Equivalence to Biological Model: The Test Pack must be at least as resistant to the sterilization process as the biological model developed for the DUO Cycle.Passed. Based on both survival curves and fraction negative data as a function of dose, the Test Pack configuration was found to be at least as resistant as the biological model.
Functionality of Chemical Indicator Disc: The chemical indicator in a DUO Test Pack configuration must show appropriate response in half-cycle parameters of the DUO Cycle.Passed. The response was determined to be appropriate for a chemical indicator.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: "DUO Test Packs assembled from three lots of CYCLESURE® 24 Biological Indicator (BI)" were used for the resistance evaluation.
  • Data Provenance: Not explicitly stated, but given that the applicant is Advanced Sterilization Products, a division of Ethicon, Inc. located in Irvine, CA, it is highly likely the testing was conducted prospectively and in the United States. The study description implies a laboratory or controlled setting.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

There is no mention of human experts being used to establish the "ground truth" for this device. The ground truth is intrinsically defined by the biological model and the chemical indicator's expected behavior within the sterilization process. The "ground truth" here is the scientific measurement of biological indicator survival and chemical indicator color change.

4. Adjudication Method for the Test Set

Not applicable. There was no human adjudication as the assessment was based on objective biological and chemical responses.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is typically used for diagnostic or imaging devices where human interpretation is a key component, and the AI's impact on human performance is being measured. This device is a biological indicator for sterilization monitoring, not a diagnostic tool requiring human interpretation for its primary function.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, in essence. The entire evaluation focuses on the performance of the STERRAD® 100NX® DUO Cycle Test Pack product itself, which functions as a standalone indicator of sterilization effectiveness. Its performance is measured directly against established biological models and chemical indicator specifications, without human interaction influencing the core function it is designed to evaluate (i.e., whether sterilization occurred). The "algorithm" here isn't a complex AI, but rather the intrinsic biological and chemical response mechanisms of the test pack.

7. The Type of Ground Truth Used

The ground truth used is based on:

  • Biological Model: The established and validated "biological model developed for the DUO Cycle" represents the benchmark for resistance to the sterilization process. This model likely involves characterized microorganisms and their known inactivation kinetics.
  • Expected Chemical Indicator Response: The predetermined, appropriate response of the chemical indicator to the sterilization conditions (specifically half-cycle parameters).

8. The Sample Size for the Training Set

There is no explicit mention of a "training set" in the context of an AI/ML algorithm. This device is not an AI/ML product. The testing involves physical devices (biological indicators) rather than training data for an algorithm. The "training" in this context would be the historical data and scientific principles used to develop the biological model itself, but that's not described here.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no explicitly defined "training set" for an AI/ML algorithm. The "ground truth" for the device's functional evaluation (as described in point 7) is established through scientific validation of the biological model and the chemical indicator's properties, likely using established standards and lab procedures.

{0}------------------------------------------------

K111391 1 of 3

510(k) Summary

Applicant's Name, Address, Telephone, FAX, Contact Person

Advanced Sterilization Products, Division of Ethicon, Inc. 33 Technology Drive Irvine, CA 92618

Contact Person

Nancy Chu Manager, Regulatory Affairs Tel: (949) 453-6435 Fax: (949) 789-3900

Summary Date: September 10, 2012

1. CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME

Classification Name: Common/Usual Name: Product Classification: Classification Regulation: Proprietary Name:

Biological Sterilization Process Indicator Biological Indicator (Test Pack) II 21 CFR 880.2800 STERRAD® 100NX® DUO Cycle Test Pack

2. PREDICATE DEVICES

STERRAD® 100NX® Test Pack, K071537 cleared on December 18, 2007

3. INDICATIONS FOR USE

The STERRAD® 100NX® DUO Cycle Test Pack is used for routine monitoring of the STERRAD® 100NX® DUO Sterilization Cycle and is also used for the periodic testing of a STERRAD® 100NX® System DUO Cycle, using hospital-defined loads containing devices that do not exceed claims of the cycle. The STERRAD® 100NX® DUO Cycle Test Pack consists of a STERRAD® CYCLESURE® 24 Biological Indicator, vial and cap to hold the BI.

SEP 26 2012

{1}------------------------------------------------

4. DESCRIPTION OF DEVICE

The STERRAD® 100NX® DUO Test Pack consists of a CYCLESURE® 24 Biological Indicator (BI) and a STERRAD® NX® Test Vial with cap for holding the BI during the sterilization cycle.

5. SUMMARY OF NONCLINICAL TESTS

The STERRAD® 100NX® DUO Test Pack has been evaluated for its resistance to the DUO Cycle in the STERRAD® 100NX® Sterilizer.

A comparison of the DUO Test Pack to the biological model developed for the DUO Cycle indicates that the Test Pack is at least as resistant to the sterilization process as the biological model. This is based on both survival curves and fraction negative data as a function of dose.

DUO Test Packs assembled from three lots of CYCLESURE® 24 Biological Indicator (BI) were exposed to several doses of hydrogen peroxide in a DUO Cycle. These survivor curves were compared to the survivor curves for the biological model developed for the DUO Cycle. The test data showed that the Test Pack configuration was at least as resistant as the biological model.

Additionally, fraction negative data were collected using Test Packs assembled from three lots of CYCLESURE® 24 BI and exposed to increasing volumes of hydrogen peroxide in a DUO Cycle. The results indicated that the Test Pack configuration was at least as resistant as the biological model.

Indicative functionality of the chemical indicator in a DUO Test Pack configuration was evaluated using half-cycle parameters of the DUO Cycle and the response was determined to be appropriate for a chemical indicator.

The subject device and its predicate device have the same intended use which is for routine monitoring of the sterilizer cycle. Additionally, they have the same technological characteristics, the same operating principles and are subjected to the same sterilant (hydrogen peroxide) and therefore, the subject device is substantially equivalent to the predicate.

The Table below lists the tests performed to demonstrate that the DUO Test Pack functions as intended in the STERRAD® 100NX® sterilizer using a DUO Cycle.

{2}------------------------------------------------

Studies PerformedResults
Design Evaluation and Performance Qualification ofSTERRAD® 100 NX® DUO Test PackPassed
Functionality Study of the Chemical Indicator Disc ofCYCLESURE® 24 BI in Test PackPassed

6. DESCRIPTION OF CHANGE:

  • Indications for Use Statement has been revised to incorporate the STERRAD® . 100NX® DUO Test Pack.
  • Instructions for Use have been revised to include STERRAD® 100NX® DUO Test . Pack information.

7. OVERALL PERFORMANCE CONCLUSIONS

The performance data shows that the STERRAD® 100NX® DUO Test Pack has the necessary resistance relative to the biological model to be an appropriate challenge for testing the DUO Cycle of the STERRAD® 100NX® Sterilizer and it is substantially equivalent to predicate device.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its wing and tail. The eagle is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ms. Nancy Chu Manager, Regulatory Affairs Advanced Sterilization Products 33 Technology Drive Irvine, California 92618

SEP 2 6 2012

Re: K111391

Trade/Device Name: STERRAD® 100NX® DUO Cycle Test Pack Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: FRC Dated: September 10, 2012 Received: September 12, 2012

Dear Ms. Chu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2- Ms. Chu

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

For

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K111391

STERRAD® 100NX® DUO Cycle Test Pack Device Name:

Indication for Use:

The STERRAD® 100NX® DUO Cycle Test Pack is used for routine monitoring of the STERRAD® 100NX® DUO Sterilization Cycle and is also used for the periodic testing of a STERRAD® 100NX® System DUO Cycle, using hospital-defined loads containing devices that do not exceed claims of the cycle. The STERRAD® 100NX® DUO Cycle Test Pack consists of a STERRAD® CYCLESURE® 24 Biological Indicator, vial and cap to hold the BI.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use __ V (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

fu uld.

Page 1 of 1

(Division Sign-Off) (Division Sign-Only)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices i

K 11 510(k) Number:_

§ 880.2800 Sterilization process indicator.

(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).