Intended Use

The device is indicated for acute and chronic hemodialysis, apheresis, and infusion. It may be inserted percutaneously or by cut down. Catheters with greater than 40 cm implant length are indicated for femoral placement.

Device Description

The Tal Palindrome™ Symmetric Tip Dual Lumen Catheter has a radiopaque shaft with two large inner lumens designed in an opposing "double D" configuration. The proximal end of the catheter features two color-coded luer adapters. The luer adapters are connected to clear extension tubes. Each extension tube contains a clamp and is connected to the catheter hub which contains suture wings. The distal end of the catheter hub is connected to the double lumen catheter shaft contains a cuff and extends to a symmetrical distal tip configuration.

The Mahurkar® Maxid™ Dual Lumen Catheter has a 14.5 Fr. radiopaque shaft with two large inner lumens designed in an opposing "double D" configuration. The proximal end of the catheter features two color-coded luer adapters. The luer adapters are connected to clear extension tubes. Each extension tube contains a clamp and is connected to the catheter hub which contains suture wings. The distal end of the catheter hub is connected to the double lumen catheter shaft. The shaft contains a cuff and extends to a staggered distal tip configuration which is available with or without side holes. Configurations with side holes contain five side holes on the arterial lumen and two side holes on the venous lumen.

The Palindrome™ RT Reverse-Tunneled Catheter will be supplied with a detached "proximal end" allowing for the catheter tip to be positioned in the vein first before the catheter shaft is pulled through the patient's subcutaneous tunnel tract in a retrograde fashion. The fully assembled Palindrome™ RT Reverse-Tunneled Catheter has a 15.0 Fr. radiopaque shaft with two large inner lumens designed in an opposing "double D" configuration. The proximal end of the catherer features two color-coded luer adapters. The luer adapters are connected to clear extension tubes. Each extension tube contains a clamp and is connected to the hub assembly which contains suture wings. The distal end of the catheter hub is connected to the double lumen catheter shaft. The shaft contains a cuff and extends to a symmetrical distal tip configuration.

AI/ML Overview

The provided text is a 510(k) summary for several hemodialysis catheters, submitted to the FDA. It does not describe an AI or software as a medical device (SaMD), but rather physical medical devices. Therefore, the questions related to AI/SaMD performance, such as acceptance criteria for algorithms, sample sizes for test/training sets, expert adjudication, MRMC studies, or standalone algorithm performance, are not applicable to this document.

The document focuses on demonstrating substantial equivalence of modified catheter designs to previously cleared predicate devices through non-clinical benchtop functional testing and biocompatibility testing.

Here's an analysis of the provided text in the context of the questions asked, highlighting why most are not applicable:

  1. A table of acceptance criteria and the reported device performance

    • Not Applicable in the requested format for SaMD.
    • The document states: "Bench top functional testing was completed to support substantial equivalence between the modified device and the current device. The test regimen evaluated the devices' resistance to kink, leak, burst, catheter collapse, fatigue, and the tensile strength at various points of the results of the performance testing show that the modified devices continue to meet the relevant product specifications."
    • It also states: "Biocompatibility testing per ISO 10993: Biological Evaluation of Medical Devices was completed... The results... show that the modified devices continue to be biocompatible for its intended use."
    • While acceptance criteria for these physical properties (e.g., maximum kink resistance, minimum tensile strength, pass/fail for biocompatibility) would have been used internally by the manufacturer, they are not explicitly detailed in this 510(k) summary. The summary only reports that the devices "continue to meet the relevant product specifications" and "continue to be biocompatible."
  2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable. This document describes testing for physical medical devices (catheters), not an AI/SaMD. There isn't a "test set" in the context of data used for algorithm evaluation. The testing involved physical samples of the catheters. The document does not specify the number of catheters tested for each characteristic.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. Ground truth in the context of expert consensus for data labeling is irrelevant here, as this is not an AI/SaMD.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. Adjudication methods are relevant for resolving discrepancies in expert labeling or diagnoses of imaging/data, which is not what this document is about.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. MRMC studies are for evaluating the performance of diagnostic devices, often medical imaging AI, with human readers. This is a physical catheter.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is not an algorithm or software.
  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable in the requested context for SaMD ground truth. For the physical devices, the "ground truth" would be the physically measured performance against established engineering or biocompatibility standards.
  8. The sample size for the training set

    • Not Applicable. There is no training set as this is not an AI/SaMD.
  9. How the ground truth for the training set was established

    • Not Applicable. There is no training set and therefore no ground truth to establish for it in the context of AI/SaMD.

In summary, the provided 510(k) summary for hemodialysis catheters describes the process of demonstrating substantial equivalence for physical medical devices through benchtop functional testing and biocompatibility testing, not through the evaluation of an AI or software algorithm. Therefore, the specific questions posed regarding acceptance criteria and study details for AI/SaMD are not answered by this document.

{0}------------------------------------------------

K11372

Page 1 of 2

AUG - 4 2011

Image /page/0/Picture/3 description: The image shows the word "COVIDIEN" in all capital letters. To the left of the word is a black and white symbol that resembles a plus sign. The text and symbol are in black and white, and the background is white.

510(K) SUMMARY

1. 510(k) Owner:

Covidien 15 Hampshire Street Mansfield. MA 02048 Telephone: (508) 261-6596 Fax: (508) 261 - 8149

Contact: Mr. Wing Ng Title: Manager, Regulatory Affairs Date Prepared: May 6, 2011

2. Device:

Trade Names: Tal Palindrome™ Symmetric Tip Dual Lumen Catheter Mahurkar® Maxid™ Dual Lumen Catheter Palindrome™ RT Reverse-Tunneled Catheter Common Name: Catheter Classification Name: Implanted Hemodialysis Catheter Regulation Number: 21 CFR 876.5540 Product Code: MSD Classification: Class III

3. Predicate Devices:

Tal Palindrome™ Symmetric Tip Dual Lumen Catheter (K043272) Mahurkar® Maxid™ Dual Lumen Catheter (K002901) Palindrome™ RT Reverse-Tunneled Catheter (K092205)

4. Device Description:

The Tal Palindrome™ Symmetric Tip Dual Lumen Catheter has a radiopaque shaft with two large inner lumens designed in an opposing "double D" configuration. The proximal end of the catheter features two color-coded luer adapters. The luer adapters are connected to clear extension tubes. Each extension tube contains a clamp and is connected to the catheter hub which contains suture wings. The distal end of the catheter hub is connected to the double lumen catheter shaft contains a cuff and extends to a symmetrical distal tip configuration.

The Mahurkar® Maxid™ Dual Lumen Catheter has a 14.5 Fr. radiopaque shaft with two large inner lumens designed in an opposing "double D" configuration. The proximal end of the catheter features two color-coded luer adapters. The luer adapters are connected to clear extension tubes. Each extension tube contains a clamp and is connected to the catheter hub which contains suture wings. The distal end of the catheter hub is connected to the double lumen catheter shaft. The shaft contains a cuff and extends to a staggered distal tip configuration which is available with or without side holes. Configurations with side holes contain five side holes on the arterial lumen and two side holes on the venous lumen.

The Palindrome™ RT Reverse-Tunneled Catheter will be supplied with a detached "proximal end" allowing for the catheter tip to be positioned in the vein first before the catheter shaft is pulled through the patient's subcutaneous tunnel tract in a retrograde fashion. The fully assembled Palindrome™ RT Reverse-

{1}------------------------------------------------

K111372 Page 2 of 2

Tunneled Catheter has a 15.0 Fr. radiopaque shaft with two large inner lumens designed in an opposing "double D" configuration. The proximal end of the catherer features two color-coded luer adapters. The luer adapters are connected to clear extension tubes. Each extension tube contains a clamp and is connected to the hub assembly which contains suture wings. The distal end of the catheter hub is connected to the double lumen catheter shaft. The shaft contains a cuff and extends to a symmetrical distal tip configuration.

5. Intended Use:

The Tal Palindrome™ Symmetric Tip Dual Lumen Catheter is indicated for acute and chronic hemodialysis, apheresis, and infusion. It may be inserted percutaneously or by cut down. Catheters with greater than 40 cm implant length are indicated for femoral placement.

The Mahurkar® Maxid™ Dual Lumen Catheter is indicated for acute and chronic hemodialysis, apheresis, and infusion. It may be inserted percutaneously or by cut down.

The Palindrome™ RT Reverse-Tunneled Catheter is indicated for acute and chronic hemodialysis, apheresis, and infusion. It may be inserted either perculaneously or by cut down. Catherers with greater than 40 cm implant length are indicated for femoral insertion.

6. Technological Characteristics:

The modified devices have the same technological characteristics as compared to their respective predicate devices.

7. Performance Data:

Bench top functional testing was completed to support substantial equivalence between the modified device and the current device. The test regimen evaluated the devices' resistance to kink, leak, burst, catheter collapse, fatigue, and the tensile strength at various points of the results of the performance testing show that the modified devices continue to meet the relevant product specifications.

Biocompatibility testing per ISO 10993: Biological Evaluation of Medical Devices was completed to support biocompatibility between the modified device and the current device. Material characterization testing was included to show material equivalence where applicable. The results of the biocompatibility testing show that the modified devices continue to be biocompatible for its intended use.

The results of functional testing, biocompatibility testing, and material analytical testing supports the determination of substantial equivalence. .

8. Conclusion:

Based on non-clinical testing results, Covidien has demonstrated that the modified chronic catheters are substantially equivalent to their respective existing chronic catheters.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is an abstract image of an eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MID 20993-0002

Mr. Wing Ng Manager, Regulatory Affairs Covidien Vascular Therapies 15 Hampshire Street MANSFIELD MA 02048

AUG - 4 2011

Re: K111372

Trade/Device Name: Tal Palindrome™ Symmetric Tip Dual Lumen Catheter Mahurkar® Maxid™ Dual Lumen Catheter Palindrome™ RT Reverse-Tunneled Catheter Regulation Number: 21 CFR §876.5540 Regulation Name: Blood access device and accessories Regulatory Class: III Product Code: MSD Dated: July 29, 2011 Received: August 1, 2011

Dear Mr. Ng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

ff you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Helena Lewis

Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

K111372

Section 4 Indications for Use Statement

510(k) Number (if known): To-Be Determined K111372

Device Name: Tal Palindrome™ Symmetric Tip Dual Lumen Catheter Mahurkar® Maxid™ Dual Lumen Catheter Palindrome™ RT Reverse-Tunneled Catheter

Indications for Use:

The device is indicated for acute and chronic hemodialysis, apheresis, and infusion. It may be inserted percutaneously or by cut down. Catheters with greater than 40 cm implant length are indicated for femoral placement.

Prescription Use X . (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Hecho Reme

ision Signi-C sion of Reproductive, Gastro-Rena ological Devices 510(k) Number

§ 876.5540 Blood access device and accessories.

(a)
Identification. A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device.
(2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle).
(3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
(b)
Classification. (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
1 ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
2 ) Proper care and maintenance of the device and device exit site;(
3 ) Removal of the device;(
4 ) Anticoagulation;(
5 ) Management of obstruction and thrombus formation; and(
6 ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.