(166 days)
K061741 Mac-Lab, CardioLab, ComboLab, and SpecialsLab System
No
The document explicitly states that optional accessories include "research tools to be used exclusively outside active patient care settings" with the purpose "to assist researchers or clinicians in developing algorithms." This indicates that the core device itself does not incorporate AI/ML for clinical use, but provides tools for external algorithm development. There is no mention of AI/ML being integrated into the device's primary functions of data acquisition, processing, display, or recording for patient care.
No.
The "Intended Use" section for both Mac-Lab and CardioLab explicitly states: "The Mac-Lab System does not have alarms, does not generate energy delivered to the patient, does not administer drugs and does not perform any life-supporting or life-sustaining functions." and "The CardioLab System does not have alarms, does not generate energy delivered to the patient, does not administer drugs and does not perform any life-supporting or life-sustaining functions." These statements indicate that the device does not provide any therapeutic intervention.
Yes
The text indicates that the device is intended for "acquiring, filtering, digitizing, amplifying, measuring and calculating, displaying, recording and monitoring of clinical data" which is then "interpreted" by "licensed healthcare practitioners." This process of collecting and displaying clinical data for interpretation by a healthcare professional falls under the definition of a diagnostic device, as it provides information used to aid in diagnosis. While it explicitly states it is "not intended for use...in situations where diagnostic arrhythmia detection is required," the overall function of collecting and presenting data for interpretation by clinicians in a diagnostic setting (like interventional laboratories) supports its role as a diagnostic device.
No
The device description explicitly states that the systems are "hemodynamic and electrophysiology (EP) recording systems" and mentions running on "the same hardware," indicating the presence of hardware components beyond just software.
Based on the provided information, these devices (Mac-Lab, CardioLab, ComboLab, and SpecialsLab Recording Systems) are not IVDs (In Vitro Diagnostics).
Here's why:
- IVD Definition: In Vitro Diagnostic devices are used to examine specimens derived from the human body (like blood, urine, tissue) to provide information for diagnosis, monitoring, or screening.
- Device Function: The descriptions clearly state that these systems are intended for acquiring, filtering, digitizing, amplifying, measuring and calculating, displaying, recording and monitoring of clinical data from adult and pediatric patients. This data includes things like ECG waveforms, blood pressure, heart rate, etc., which are acquired directly from the patient, not from a specimen in vitro.
- Lack of Specimen Analysis: There is no mention of analyzing biological specimens or performing tests on samples outside of the body.
- Focus on Physiological Signals: The data they process are physiological signals and parameters derived from the patient's body.
Therefore, these devices fall under the category of physiological monitoring and recording systems, not In Vitro Diagnostic devices.
N/A
Intended Use / Indications for Use
Mac-Lab
The Mac-Lab System is intended for acquiring, filtering, digitizing, amplifying, measuring and calculating, displaying, recording and monitoring of clinical data from adult and pediatric patients. The Mac-Lab System is configurable. Clinical data includes: ECG waveforms, heart rate, pulse oximetry (SpO2), respiration rate, CO2 (EtCO2), temperature, hemodynamic measures [e.g. valve gradients and areas, cardiac output, shunts, Fractional Flow Reserve (FFR), invasive and noninvasive blood pressure] Physiological parameters such as diastolic, systolic, and mean pressures, and heart rate are derived from the signal data. displayed and recorded. The data is entered manually or acquired via interfaced devices and/or information systems and may be used for report generation.
Procedural information and optional anatomical and physiological imaging and data devices may be interfaced (e.g. X-ray, ultrasound, patient monitors and information systems). The Mac-Lab System can display, store and annotate images previously acquired and stored by other systems. Data may be provided to other systems via multiple formats (e.g. HL7. DICOM, Analog outputs). Data may be received from other devices via multiple formats (e.g. DICOM, Analog inputs).
Optional accessories for hardware and software include research tools to be used exclusively outside active patient care settings. The purpose of the research tools is to assist researchers or clinicians in developing algorithms.
The Mac-Lab System does not have alarms, does not generate energy delivered to the patient, does not administer drugs and does not perform any life-supporting or life-sustaining functions. The Mac-Lab System is not intended for use on unattended patients, or in situations where diagnostic arrhythmia detection is required.
The Mac-Lab System provides the ability to transmit patient data for storage, analysis and viewing at distributed locations within a clinical facility via network connectivity. The Mac-Lab System also functions as a stand-alone device. The Mac-Lab System is used in a variety of hospital and clinical settings including interventional laboratories (e.g. cardiac catheterization and radiology), operating room environments, and pre and post areas all under the direct supervision of licensed healthcare practitioners who are responsible for interpreting the data.
CardioLab
The CardioLab System is intended for acquiring, filtering, digitizing, amplifying, measuring and calculating, displaying, recording and monitoring of clinical data from adult and pediatric patients. The CardioLab System is configurable. Clinical data includes: ECG waveforms, intracardiac signals, stimulus data, ablation data, pulse oximetry (SpO2), respiration rate, CO2 (EtCO2), temperature, and invasive and noninvasive blood pressure. Physiological parameters such as diastolic, systolic, mean pressures, heart rate, and cycle length are derived from the signal data, displayed and recorded. The data is entered manually or acquired via interfaced devices and/or information systems and may be used for report generation.
Procedural information and optional anatomical and physiological imaging and data devices may be interfaced [e.g. X-ray, ultrasound, mapping systems, ablation generators (e.g. RF and cryogenic)], stimulators, patient monitors and information systems. The CardioLab System can display, store and annotate images previously acquired and stored by other systems. Data may be provided to other systems via multiple formats (e.g. HL7. DICOM, Analog outputs). Data may be received from other devices via multiple formats (e.g. DICOM, Analog inputs).
Optional accessories for hardware and software include research tools to be used exclusively outside active patient care settings. The purpose of the research tools is to assist researchers or clinicians in developing algorithms.
Optional accessories for hardware and software includes a waveform simulator to be used exclusively outside active patient care settings. The waveform simulator may be used for training, demonstration without a patient attached, and as a troubleshooting tool on the CardioLab System.
The CardioLab System does not have alarms, does not generate energy delivered to the patient, does not administer drugs and does not perform any life-supporting or life-sustaining functions. The CardioLab System is not intended for use on unattended patients, or in situations where diagnostic arrhythmia detection is required.
The CardioLab System provides the ability to transmit patient data for storage, analysis and viewing at distributed locations within a clinical facility via network connectivity. The CardioLab System also functions as a stand-alone device. The CardioLab System is used in a variety of hospital and clinical settings including interventional laboratories (e.g. electrophysiology and cardiac catheterization), operating room environments, and pre and post areas all under the direct supervision of licensed healthcare practitioners who are responsible for interpreting the data.
SpecialsLab
The SpecialsLab System executes the same software and runs on the same hardware in the same environments as the Mac-Lab System. Products designated as a SpecialsLab System support fewer options than the Mac-Lab system.
ComboLab
The ComboLab System is the combination of both the Mac-Lab and CardioLab Systems. The ComboLab System allows the user to run either the Mac-Lab System or the CardioLab System, although only one system may be used at a time. The ComboLab System executes the same software and runs on the same hardware in the same environments as the Mac-Lab and CardioLab Systems.
Product codes (comma separated list FDA assigned to the subject device)
DOK
Device Description
Mac-Lab, CardioLab, ComboLab, and SpecialsLab Recording Systems are hemodynamic and electrophysiology (EP) recording systems.
The product will be available in the following configurations: Mac-Lab application, CardioLab application, SpecialsLab application, or a combination of both CardioLab and Mac-Lab applications marketed as ComboLab. The product designated as SpecialsLab is the same as the Mac-Lab System with the exception that it will support fewer options. The SpecialsLab System performs the same intended use as the Mac-Lab, executes the same software, and runs on the same hardware.
The Mac-Lab, CardioLab, and ComboLab Recording Systems are each available in several configurations ranging from basic to full advanced functionality.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
X-ray, ultrasound
Anatomical Site
Not Found
Indicated Patient Age Range
adult and pediatric patients
Intended User / Care Setting
licensed healthcare practitioners in a variety of hospital and clinical settings including interventional laboratories (e.g. cardiac catheterization and radiology, electrophysiology), operating room environments, and pre and post areas.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The subject of this premarket submission, Mac-Lab, CardioLab, ComboLab, and SpecialsLab Recording Systems, did not require clinical studies to support substantial equivalence.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
K061741 Mac-Lab, CardioLab, ComboLab, and SpecialsLab System
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.1425 Programmable diagnostic computer.
(a)
Identification. A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.(b)
Classification. Class II (performance standards).
0
111200
Image /page/0/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circular border. The letters are stylized with curved lines, and the border has a textured appearance.
OCT 1 2 2011
GE Healthcare 510(k) Premarket Notification Submission
ે.
જે તેમ
510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided: Date: April 27, 2011 Submitter: GE Healthcare,
(GE Medical Systems Information Technologies, Inc.) 8200 West Tower Avenue Milwaukee, WI 53223
Primary Contact Person: Ms. Carol Alloian Regulatory Leader
GE Healthcare, (GE Medical Systems Information Technologies, Inc.) Telephone: 224 280-1008 Fax: 847 589 8524
Secondary Contact Person: Mr. Philip Malca Regulatory Affairs Director
GE Healthcare, (GE Medical Systems Information Technologies, Inc.) Telephone: 33(0) 1 3070 4207 Fax: 33(0) 1 3070 4399
- Device: Trade Name: Mac-Lab, CardioLab, ComboLab, and SpecialsLab Recording Systems
- Common/Usual Name: Hemodynamic and Electrophysiology (EP) Recording Systems
- Classification Names: 21 CFR 870.1425 Computer, Diagnostic Programmable Product Code: DOK
- Predicate Device(s): K061741 Mac-Lab, CardioLab, ComboLab, and SpecialsLab System
1
Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The intertwined letters are enclosed within a circular border, which has a decorative, wave-like pattern around its edge. The logo is black and white.
1/5
Device Description: Mac-Lab, CardioLab, ComboLab, and SpecialsLab Recording Systems are hemodynamic and electrophysiology (EP) recording systems.
The product will be available in the following configurations: Mac-Lab application, CardioLab application, SpecialsLab application, or a combination of both CardioLab and Mac-Lab applications marketed as ComboLab. The product designated as SpecialsLab is the same as the Mac-Lab System with the exception that it will support fewer options. The SpecialsLab System performs the same intended use as the Mac-Lab, executes the same software, and runs on the same hardware.
The Mac-Lab, CardioLab, and ComboLab Recording Systems are each available in several configurations ranging from basic to full advanced functionality.
Intended Use/ Mac-Lab
Indications for Use:
The Mac-Lab System is intended for acquiring, filtering, digitizing, amplifying, measuring and calculating, displaying, recording and monitoring of clinical data from adult and pediatric patients. The Mac-Lab System is configurable. Clinical data includes: ECG waveforms, heart rate, pulse oximetry (SpO2), respiration rate, CO2 (EtCO2), temperature, hemodynamic measures [e.g. valve gradients and areas, cardiac output, shunts, Fractional Flow Reserve (FFR), invasive and noninvasive blood pressure] Physiological parameters such as diastolic, systolic, and mean pressures, and heart rate are derived from the signal data. displayed and recorded. The data is entered manually or acquired via interfaced devices and/or information systems and may be used for report generation.
Procedural information and optional anatomical and physiological imaging and data devices may be interfaced (e.g. X-ray, ultrasound, patient monitors and information systems). The Mac-Lab System can display, store and annotate images previously acquired and stored by other systems. Data may be provided to other systems via multiple formats (e.g. HL7. DICOM, Analog outputs). Data may be received from other devices via multiple formats (e.g. DICOM, Analog inputs).
Optional accessories for hardware and software include research
2
Image /page/2/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circular border. The design is simple and recognizable, representing the company's brand identity.
3/c
Intended Use/ Indications for Use: tools to be used exclusively outside active patient care settings. The purpose of the research tools is to assist researchers or clinicians in developing algorithms.
The Mac-Lab System does not have alarms, does not generate energy delivered to the patient, does not administer drugs and does not perform any life-supporting or life-sustaining functions. The Mac-Lab System is not intended for use on unattended patients, or in situations where diagnostic arrhythmia detection is required.
The Mac-Lab System provides the ability to transmit patient data for storage, analysis and viewing at distributed locations within a clinical facility via network connectivity. The Mac-Lab System also functions as a stand-alone device. The Mac-Lab System is used in a variety of hospital and clinical settings including interventional laboratories (e.g. cardiac catheterization and radiology), operating room environments, and pre and post areas all under the direct supervision of licensed healthcare practitioners who are responsible for interpreting the data.
CardioLab
The CardioLab System is intended for acquiring, filtering, digitizing, amplifying, measuring and calculating, displaying, recording and monitoring of clinical data from adult and pediatric patients. The CardioLab System is configurable. Clinical data includes: ECG waveforms, intracardiac signals, stimulus data, ablation data, pulse oximetry (SpO2), respiration rate, CO2 (EtCO2), temperature, and invasive and noninvasive blood pressure. Physiological parameters such as diastolic, systolic, mean pressures, heart rate, and cycle length are derived from the signal data, displayed and recorded. The data is entered manually or acquired via interfaced devices and/or information systems and may be used for report generation.
Procedural information and optional anatomical and physiological imaging and data devices may be interfaced [e.g. X-ray, ultrasound, mapping systems, ablation generators (e.g. RF and cryogenic)], stimulators, patient monitors and information systems. The CardioLab System can display, store and annotate images previously acquired and stored by other systems. Data may be provided to other systems via multiple formats (e.g. HL7.
3
Image /page/3/Picture/0 description: The image shows a black and white logo of General Electric (GE). The logo consists of the letters 'GE' intertwined in a stylized, circular design. The letters are surrounded by a circular border with a decorative pattern.
4/2-
Intended Use/ Indications for Use:
DICOM, Analog outputs). Data may be received from other devices via multiple formats (e.g. DICOM, Analog inputs).
Optional accessories for hardware and software include research tools to be used exclusively outside active patient care settings. The purpose of the research tools is to assist researchers or clinicians in developing algorithms.
Optional accessories for hardware and software includes a waveform simulator to be used exclusively outside active patient care settings. The waveform simulator may be used for training, demonstration without a patient attached, and as a troubleshooting tool on the CardioLab System.
The CardioLab System does not have alarms, does not generate energy delivered to the patient, does not administer drugs and does not perform any life-supporting or life-sustaining functions. The CardioLab System is not intended for use on unattended patients, or in situations where diagnostic arrhythmia detection is required.
The CardioLab System provides the ability to transmit patient data for storage, analysis and viewing at distributed locations within a clinical facility via network connectivity. The CardioLab System also functions as a stand-alone device. The CardioLab System is used in a variety of hospital and clinical settings including interventional laboratories (e.g. electrophysiology and cardiac catheterization), operating room environments, and pre and post areas all under the direct supervision of licensed healthcare practitioners who are responsible for interpreting the data.
SpecialsLab
The SpecialsLab System executes the same software and runs on the same hardware in the same environments as the Mac-Lab System. Products designated as a SpecialsLab System support fewer options than the Mac-Lab system.
ComboLab
The ComboLab System is the combination of both the Mac-Lab and CardioLab Systems. The ComboLab System allows the user to run either the Mac-Lab System or the CardioLab System,
4
Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The design is simple and iconic, representing the company's brand identity.
75
although only one system may be used at a time. The ComboLab System executes the same software and runs on the same hardware in the same environments as the Mac-Lab and CardioLab Systems.
The proposed Mac-Lab, CardioLab, ComboLab, and SpecialsLab Technology: Recording Systems employ the same fundamental scientific technology as their predicate devices.
Determination of Summary of Non-Clinical Tests:
Substantial Equivalence: The Mac-Lab, CardioLab, ComboLab, and SpecialsLab Recording Systems and their applications comply with voluntary standards as detailed in Section 9, 11 and 17 of this premarket submission. The following quality assurance measures were applied to the development of the system:
- Risk Analysis .
- Requirements Reviews .
- . Design Reviews
- Testing on unit level (Module verification) .
- Integration testing (System verification) .
- . Performance testing (Verification)
- Safety testing (Verification) .
- Simulated use and Usability testing (Validation) .
Summary of Clinical Tests:
The subject of this premarket submission, Mac-Lab, CardioLab, ComboLab, and SpecialsLab Recording Systems, did not require clinical studies to support substantial equivalence.
Conclusion: GE Healthcare considers the Mac-Lab, CardioLab, ComboLab, and SpecialsLab Recording Systems to be as safe and effective as, and to have performance substantially equivalent to, the predicate devices.
6
5
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular in shape, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter of the circle. In the center of the circle is an abstract symbol that resembles an eagle or other bird in flight. The symbol is composed of three curved lines that converge at the bottom.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20995-0002
OCT 1 2 2011
GE Healthcare c/o Ms. Carol Alloian. RAC Regulatory Leader, Interventional Systems 9900 W Innovation Drive Wauwatosa, WI 53266-4856
Re: K111200
Trade/Device Names: Mac-Lab, CardioLab, ComboLab and SpecialsLab Recording Systems, Version 6.9 Regulatory Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II (Two) Product Code: DOK Dated: September 19, 2011 Received: September 21, 2011
Dear Ms. Alloian:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act . The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your devil (1111) be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
: .
6
Page 2 - Ms. Carol Alloian. RAC
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. -0-Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
7
111200
Image /page/7/Picture/1 description: The image shows a circular object with a textured surface. The surface appears to have a pattern of concentric circles or spirals, possibly indicating a layered or structured composition. The object is surrounded by a scattering of small, dark spots, which could be imperfections, shadows, or part of the object's design.
GE Healthcare 510(k) Premarket Notification Submission
KIII Zoo 510(k) Number (if known): To Be Assigned
Device Name: Mac-Lab, CardioLab, ComboLab, and SpecialsLab Recording Systems
Indications for Use: Mac-Lab
The Mac-Lab System is intended for acquiring, filtering, amplifying, measuring and calculating, displaying, recording and monitoring of clinical data from adult and pediatric patients. The Mac-Lab System is configurable. Clinical data includes: ECG waveforms, heart rate, pulse oximerry (SpO2), respiration rate, CO2 (EtCO2), temperature, hemodynamic measures |e.g. valve gradients and areas, cardiac output, shunts, Fractional Flow Reserve (FFR), invasive and noninvasive blood pressure) Physiological parameters such as diastolic, systolic, and heart rate are derived from the signal data, displayed and recorded. The data is entered manually or acquired via interfaced devices and/or information systems and mav be used for report generation.
Procedural information and optional analomical and physiological imaging and data devices may be interfaced (e.g. X -rav, ultrasound, patient monitors and information systems). The Mac-Lab System can display, store and annotate images previously acquired and stored by other systems. Data may be provided to other systems via multiple formats (e.g. HL7, DICOM. Analog outputs). Data may be received from other devices via multiple formats (e.g. DICOM, Analog inputs).
Optional accessories for hardware and software include research tools to be used exclusively outside active patient care settings. The purpose of the research tools is to assist researchers or developing algorithms.
The Mac-Lab System does not have alarms, does not generate energy delivered to the patient, does not administer drugs and does not perform any life-supporting or life-sustaining functions. The Mac-Lab System is not intended for use on unattended patients, or in situations where diagnostic archythmia detection is required.
The Mac-Lab System provides the ability to transmit patient data for storage, analysis and viewing at distributed locations within a clinical facility via network connectivity. The Mac-Lab System also functions as a stand-alone device. The Mac-Lab System is used in a variety of hospital and clinical settings including interventional laboratories (e.g. cardiac catheterization and radiology), operating room environments, and pre and post areas all under the direct supervision of licensed healthcare practitioners who are responsible for interpreting the data.
Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
(Division Sign-Off) | |
---|---|
--------------------- | -- |
Division of Cardiovascular Devices
510(k) Number | K117200 |
---|---|
--------------- | --------- |
R-C 000
8
KII200 510(k) Number (if known): To Be Assigned
Device Name: Mac-Lab, CardioLab, ComboL.ab, and SpecialsL.ab Recording Systems
Indications for Use: CardioLab
The CardioLab System is intended for acquiring, filtering, amplitying, measuring and calculating, displaying, recording and monitoring of clinical data from adult and pediatric patients. The CardioLab System is configurable. Clinical data includes: ECG waveforms, intracardiac signals, stimulus data, ablation data, pulse oximetry (SpO2), respiration rate, CO2 (EiCO2), temperature, and invasive and noninvasive blood pressure. Physiological parameters such as diastolic, systolic, mean pressures, heart rate, and cycle length are derived from the signal data, displayed and recorded. The data is entered manually or acquired via interfaced devices and/or information systems and may be used for report generation.
Procedural information and optional anatomical and physiological imaging and data devices may be interfaced [e.g. X-ray, ultrasound, mapping systems, ablation generators (e.g. RF and cryogenic)], stimulators, patient monitors and information systems. The CardioLab System can display, store and annotate images previously acquired and stored by other systems. Data may be provided to other systems via multiple formats (e.g. HL7, DICOM, Analog outputs). Data may be received from other devices via multiple formats (e.g. DICOM, Analog inputs).
Optional accessories for hardware and software include research tools to be used exclusively outside active patient care settings. The purpose of the research tools is to assist researchers or clinicians in developing algorithms.
Optional accessories for hardware and software include a waveform simulator to be used exclusively outside active patient care settings. The waveform simulator may be used for training, demonstration without a patient attached, and as a troubleshooting tool on the CardioLab System.
The CardioLab System does not have alarms, does not generate energy delivered to the patient, does not administer drugs and does not perform any life-supporting or life-sustaining functions. The CardioLab System is not intended for use on unattended patients, or in situations where diagnostic arrhythmia detection is required.
The CardioLab System provides the ability to transmit patient data for storage, analysis and viewing at distributed locations within a clinical facility via network connectivity. The CardioLab System also functions as a stand-alone device. The CardioLab System is used in a variety of hospital and clinical settings including interventional laboratories (e.g. electrophysiology and cardiac catheterization), operating room environments, and post areas all under the direct supervision of licensed healthcare practitioners who are responsible for interpreting the data.
Prescription Use x (Part 21 CFR 801 Subpart D)
510(k) Numb
AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
(Division Sign-Off) |
---|
Division of Cardiovascular Devices |
R-C 000003
9
Image /page/9/Picture/1 description: The image shows a circular logo with the letters 'GE' intertwined in the center. The letters are stylized and appear to be connected. The entire logo is enclosed within a circular border, giving it a classic and recognizable appearance. The logo is black and white.
510(k) Number (if known): To Be Assigned
Device Name: Mac-Lab, CardioLab, ComboLab, and SpecialsLab Recording Systems
Indications for Use:
SpecialsLab
The SpecialsLab System executes the same software and runs on the same hardware in the same environments as the Mac-Lab System. Products designated as a SpecialsLab System support fewer options than the Mac-Lab system:
ComboLab
The ComboLab System is the combination of both the Mac-Lab and CardioLab Systems. The ComboLab System allows the user to run either the Mac-Lab System or the CardioLab System, although only one system may be used at a time. The Combol.ab System executes the same software and runs on the same hardware in the same environments as the Mac-Lab and CardioLab Systems.
Prescription Use_X | |
---|---|
(Part 21 CFR 801 Subpart D) |
AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
(Division Sign-Off) | |
---|---|
Division of Cardiovascular Devices | |
510(k) Number | (111200 |
4
R-C 000004