K Number
K111179

Validate with FDA (Live)

Date Cleared
2011-07-01

(65 days)

Product Code
Regulation Number
872.5570
Panel
Dental
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Siesta Medical, Inc. ENCORE Tongue Suspension System is intended to be used for anterior advancement of the tongue base by means of a bone screw threaded with suture. It is indicated for the treatment of mild or moderate obstructive sleep apnea (OSA) and /or snoring.

Device Description

The ENCORE Tongue Suspension System is designed for anterior tongue base suspension by fixation of the soft tissue of the tongue base to the mandible bone using a bone screw and suture. The ENCORE Tongue Suspension System consists of an integrated suture passer pre-loaded with size #2-0 braided polyester suture, a titanium bone screw that is pre-mounted on an inserter, a bone screw lock tool. and a Threading Tool. In addition, one of the following suspension sutures are provided depending on the model number: 1) a size #1 monofilament polypropylene suspension suture, 2) a size #1 monofilament polypropylene suspension suture with a radiopaque marker, 3) a size #2 monofilament polypropylene suspension suture with a radiopaque marker, or 4) a size #2 braided polyester suspension suture.

Similar to the PRELUDE III System, the ENCORE Tongue Suspension System 1) uses a bone screw, 2) uses a bone screw inserter to place a bone screw into the bone, and 3) uses a needle to pass a suspension suture submucosally through the base of the tongue. The ENCORE uses a Suture Passer and working suture loops to pass a suture from the posterior position in the tongue. The ENCORE advances the tongue in the anterior direction with a suspension suture. A Threading Tool is provided to ease passage of the suspension suture through the bone screw.

AI/ML Overview

The provided text describes the 510(k) submission for the Siesta Medical, Inc. ENCORE Tongue Suspension System. The submission primarily focuses on demonstrating substantial equivalence to a predicate device (PRELUDE III Tongue Suspension System) through performance testing, rather than a clinical study evaluating diagnostic or treatment effectiveness in humans.

Here's an analysis based on your requested information:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Suture Endurance"the suture was able to endure 10.5 million cycles"
Bone Screw to Suture Attachment Strengths"the bone screw to suture attachment strengths were no worse than the written nominal ratings" and "the performance characteristics of the ENCORE Tongue Suspension System compare favorably to the predicate device." (Specific quantitative acceptance criteria for "nominal ratings" are not provided in the excerpt, but the favorable comparison to the predicate device implies meeting equivalence standards).

2. Sample size used for the test set and the data provenance

The document describes device performance testing (Suture Endurance Test and Bone Screw Fixation Strength Test), not a clinical study on human subjects or a diagnostic algorithm. Therefore, "test set" in the context of clinical data or AI evaluation is not directly applicable. The "samples" would refer to the physical devices or components tested. The exact number of samples for each engineering test is not specified in the provided text, nor is their "country of origin" as they are manufactured devices. These were likely prospective engineering tests performed in a lab setting.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable. The study described is not a clinical study involving human patients or an AI device requiring expert-established ground truth. The "ground truth" here is determined by the physical properties and performance benchmarks of the medical device components themselves.

4. Adjudication method for the test set

This question is not applicable as it refers to a clinical or AI study with human interpretation, not a bench test for device performance.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not performed, nor is it applicable. This is a 510(k) submission for a physical medical device, not an AI or diagnostic tool that assists human readers.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

No, this question is not applicable. The ENCORE Tongue Suspension System is a physical medical device, not a standalone algorithm.

7. The type of ground truth used

The "ground truth" for the device performance tests was based on engineering specifications and established performance benchmarks derived from the predicate device and potentially industry standards for medical device materials and components. For instance:

  • Suture Endurance: The acceptance criterion was enduring 10.5 million cycles, which represents a physical performance threshold.
  • Bone Screw to Suture Attachment Strengths: The ground truth was described as "written nominal ratings" and "favorable comparison to the predicate device," implying a quantitative or comparative engineering standard.

8. The sample size for the training set

This question is not applicable. As previously stated, this is a 510(k) submission for a physical medical device, not an AI or machine learning algorithm that requires a training set.

9. How the ground truth for the training set was established

This question is not applicable. There was no training set for an AI algorithm.

{0}------------------------------------------------

Image /page/0/Picture/11 description: The image shows a logo with the word "SIESTA" in large, bold letters. Below "SIESTA" is the word "MEDICAL" in smaller, but still bold, letters. The logo is black and white and appears to be a cropped portion of a larger document or advertisement.

510(k) for the Siesta Medical, Inc ENCORE Tongue Suspension System April 26, 2011

510(k) Summary (per 21CFR 807.92) Section 5

JUL - 1 2011

General Company Information
Name:Siesta Medical, Inc.
Contact:Michael KolberVice President, Regulatory Affairs
Address:101 Church Street, Suite 3Los Gatos, CA 95030
Telephone:408-505-6626
Fax:408-399-7600
Date Prepared:April 26, 2011
General Device Information
Product Name:ENCORETM Tongue Suspension System
Common Name:Bone Screw System
Classification:21CFR872.5570Intraoral devices for snoring and intraoral devices for snoringand obstructive sleep apnea.
Device Class:Class II
Product Code:ORY
Predicate Device
ManufacturerDevice Name510(k) Number
Siesta Medical, Inc.PRELUDE III TongueSuspension SystemK110127

Description - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - The ENCORE Tongue Suspension System is designed for anterior tongue base suspension by fixation of the soft tissue of the tongue base to the mandible bone using a bone screw and suture. The ENCORE Tongue Suspension System consists of an integrated suture passer pre-loaded with size #2-0 braided polyester suture, a titanium bone screw that is pre-mounted on an inserter, a bone screw lock tool. and a Threading Tool. In addition, one of the following suspension sutures are provided depending on the model number: 1) a size #1 monofilament polypropylene suspension suture, 2) a size #1 monofilament polypropylene suspension suture with a radiopaque marker, 3) a size #2 monofilament polypropylene suspension suture with a radiopaque marker, or 4) a size #2 braided polyester suspension suture.

Similar to the PRELUDE III System, the ENCORE Tongue Suspension System 1) uses a bone screw, 2) uses a bone screw inserter to place a bone screw into the bone, and 3) uses a needle to pass a suspension suture submucosally through the base of the tongue. The ENCORE uses a Suture Passer and working suture loops to pass a suture from the posterior position in the tongue. The ENCORE advances the tongue in the anterior direction with a suspension suture. A Threading Tool is provided to ease passage of the suspension suture through the bone screw.

Intended Use (Indications) , The Siesta Medical, Inc. ENCORE Tongue Suspension System is intended to be used for anterior advancement of the tongue base by means of a bone screw threaded with suture. It is indicated for the treatment of mild or moderate obstructive sleep apnea (OSA) and /or snoring.

Substantial Equivalence

This submission supports the position that the Siesta Medical, Inc. ENCORE Tongue Suspension System is substantially equivalent to the PRELUDE III Tongue Suspension System (K110127). The 510(k) notice contains a summary of the Suture Endurance Test and the Bone Screw Fixation

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Siesta Medical. The word "SIESTA" is in large, bold, sans-serif font with a distressed or slightly eroded appearance. A curved line is positioned above the word "SIESTA", adding a design element to the logo. Below "SIESTA", the word "MEDICAL" is printed in a smaller, sans-serif font, also in bold.

Strength Test, which were used to evaluate the performance characteristics of the ENCORE Tongue Suspension System. This data demonstrate that the suture was able to endure 10.5 million cycles Suspension System. This the bone screw to suture attachment strengths were no worse than the writtion ordaings and that the performance characteristics of the ENCORE Tongue Suspension System compare favorably to the predicate device.

Conclusions

Siesta Medical, Inc. believes that the information provided demonstrates that the proposed device is Substantially equivalent to the predicate device and does not raise any new issues of safety or efficacy. Based on the indications for use, technological characteristics, and comparison to the entially. Based the ENCORE Tongue Suspension System has been shown to be substantially equivalent to predicate device as described under the Federal Food, Drug and Cosmetic Act.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002

Mr. Michael Kolber Vice President, Regulatory Affairs Siesta Medical, Incorporated 101 Church Street, Suite 3 Los Gatos, California 95030

JUL - 1 2011

Re: K111179

Trade/Device Name: ENCORETM Tongue Suspension System Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices for Snoring and Intraoral Devices for Snoring and Obstructive Sleep Apnea Regulatory Class: II Product Code: ORY Dated: June 1, 2011 Received: June 2, 2011 .. . . . . . . . . . . . .

Dear Mr. Kolber:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Mr. Kolber

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requrements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Anthony D. Watson

Clostformy D. Only .. Oninton
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the words "SIESTA MEDICAL" in a bold, sans-serif font. The word "SIESTA" is on the top line, and the word "MEDICAL" is on the bottom line. There is a line above the word "SIESTA". The text is black and the background is white.

Indications for Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Siesta Medical, Inc. ENCORE™ Tongue Suspension System

Indications for Use: The Siesta Medical, Inc. ENCORE Tongue Suspension System is intended to be used for anterior advancement of the tongue base by means of a bone screw threaded with suture. It is indicated for the treatment of mild or moderate obstructive sleep apnea (OSA) and /or snoring.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Venner

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

-111179 510(k) Number:

ﺑﺎﻟﺘﻲ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪ

§ 872.5570 Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea.

(a)
Identification. Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.(b)
Classification. Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”