K Number
K111178
Device Name
JUELL SOFT CURE
Manufacturer
Date Cleared
2011-09-12

(138 days)

Product Code
Regulation Number
872.3760
Panel
DE
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

JUELLCure Soft is intended for use as a permanent soft relining for total and partial dentures

  • . To relieve pressure from pressure points
  • To dam the palatal vibrating line ●
  • . In cases of flabby ridge and/or insufficient adhesion
  • . To cushion sharp-edged alveolar processes
Device Description

JuellCure Soft is a self-curing, silicone-based and permanently soft relining material for removable dentures.

AI/ML Overview

The provided text describes the 510(k) summary for the JUELLCure Soft device, which is a soft impression material for dentures. The submission aims to demonstrate substantial equivalence to a predicate device, Ufi Gel SC.

Here's an analysis based on the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for JUELLCure Soft are based on demonstrating equivalence to the predicate device, Ufi Gel SC, across chemical composition and key physical properties.

Test ItemAcceptance Criteria (Equivalent to Ufi Gel SC)JUELLCure Soft Performance
Monomer Matrix (Base)70-80 %70-80 %
Rheologic Modifier (Base)15-30 %15-30 %
Catalyst (Base)0.1-0.5 %0.1-0.5 %
Colorant (Base)

§ 872.3760 Denture relining, repairing, or rebasing resin.

(a)
Identification. A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.(b)
Classification. Class II.