(105 days)
The AIS SIBD XP Spinal System is a stand-alone device intended to be used with the four supplied bone screws if no supplemental fixation is used.
As an intervertebral body fusion device designed for use with autograft, the SIBD XP Spinal System is intended for spinal fusion procedures at one or two contiguous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients may have had previous non-fusion spinal surgery at the involved spinal level(s).
Patients should be skeletally mature and must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the Aesculap® Implant Systems device.
The AIS SIBD XP Spinal System is an implantable spinal device manufactured from PEEK-OPTIMA® LT (Polyetheretherketone) per ASTM F2026, with a titanium laver and a vacuum plasma spray coating (Plasmapore®). The device will have Tantalum markers per ASTM F-560.The implant is secured to vertebral bodies by four titanium screws inserted through the anterior screw holes. The implants are offered in a variety of shapes and sizes to meet the requirements of the individual patient anatomy.
This is a medical device submission, not an AI/ML device, so many of the requested fields are not applicable. Here's what can be extracted:
Acceptance Criteria and Device Performance Study
The submission describes the Aesculap® Implant Systems (AIS) - SIBD XP Spinal System as a substantially equivalent device to its predicate. The "acceptance criteria" are implied by the performance of the predicate devices and existing ASTM standards. The study performed is a series of non-clinical tests to demonstrate that the new device meets or exceeds these established performance benchmarks.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Implied by Predicate/Standards) | Reported Device Performance (AIS SIBD XP Spinal System) |
|---|---|
| Static and dynamic torsion per ASTM F2077 | Meets or exceeds performance of predicate devices |
| Static and dynamic axial compression per ASTM F2077 | Meets or exceeds performance of predicate devices |
| Static and dynamic shear compression testing per ASTM F2077 | Meets or exceeds performance of predicate devices |
| Subsidence per ASTM F2267 | Meets or exceeds performance of predicate devices |
| Wear Debris per ASTM F2077 & ASTM F1877 | Meets or exceeds performance of predicate devices |
| Expulsion per ASTM Draft Standard F-04.25.02.02 | Meets or exceeds performance of predicate devices |
| Microstructure of the coating per ASTM F1854 | Demonstrated |
| Static Tensile Strength per ASTM F1147 | Demonstrated |
| Static Shear Strength per ASTM F1044 | Demonstrated |
| Shear Fatigue Test per ASTM F1160 | Demonstrated |
| Abrasion Resistance per ASTM F1978 | Demonstrated |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not specified. Standard practice for these types of mechanical tests would typically involve a small number of samples (e.g., 5-10 per test condition) to establish statistical significance within the test's scope.
- Data Provenance: Not explicitly stated, but these are non-clinical tests, meaning they were conducted in a laboratory setting. Not applicable to country of origin or retrospective/prospective in the conventional sense.
3. Number of Experts Used to Establish Ground Truth and Qualifications
- Not Applicable. This is for a physical implant device and involves mechanical and material testing, not expert review of diagnostic interpretations. The "ground truth" is defined by established engineering and material science standards (ASTM).
4. Adjudication Method for the Test Set
- Not Applicable. Mechanical tests are evaluated against predefined acceptance limits set by ASTM standards and comparison to predicate devices, not by human adjudication of ambiguous results.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Not Applicable. This is a mechanical device, not an AI diagnostic tool.
6. Standalone (Algorithm Only) Performance
- Not Applicable. This is a mechanical device, not an AI algorithm.
7. Type of Ground Truth Used
- Engineering and Material Standards: The ground truth is established by recognized industry standards, specifically ASTM (American Society for Testing and Materials) standards for spinal implants and coating characteristics. The performance is also benchmarked against legally marketed predicate devices.
8. Sample Size for the Training Set
- Not Applicable. This is a physical device, and the concept of a "training set" is not relevant here. The design of the device is based on established engineering principles and prior validated devices.
9. How the Ground Truth for the Training Set Was Established
- Not Applicable. As there is no training set, this question is not relevant. The device's design is informed by existing knowledge, predicate devices, and regulatory requirements rather than a data-driven training process.
{0}------------------------------------------------
510(k) SUMMARY (as required by 21 CFR 807.92) B.
AUG - 4 2011
Aescular® Implant Systems(AIS) - SIBD XP Spinal System August 1, 2011
| COMPANY: | Aesculap®Implant Systems, Inc.3773 Corporate ParkwayCenter Valley, PA 18034Establishment Registration Number: 3005673311 |
|---|---|
| ---------- | -------------------------------------------------------------------------------------------------------------------------------------- |
CONTACT: Lisa M. Boyle 610-984-9274 (phone) 610-791-6882 (fax)
TRADE NAME: AIS SIBD XP Spinal System
COMMON NAME. Intervertebral Body Fusion Device
CLASSIFICATION NAME: Intervertebral Fusion Device with Integrated Fixation, Lumbar
REGULATION NUMBER: 888.3080
PRODUCT CODE: OVD
PURPOSE FOR PREMARKET NOTIFICATION
The purpose for this submission is to gain marketing clearance for the AIS SIBD XP Spinal System.
SUBSTANTIAL EQUIVALENCE
Aesculap® Implant Systems, Inc. believes that the AIS SIBD XP Spinal System is substantially equivalent to the design of the AIS SIBD Spinal System (K100802). The Plasmapore® coating used for the subject device is similar to the the Plasmapore® coating of the AIS Hydrolift VBR System (K083186). Furthermore, the coating has been used and cleared in a number of legally marketed systems in the US and Europe for many years.
DEVICE DESCRIPTION
The AIS SIBD XP Spinal System is an implantable spinal device manufactured from PEEK-OPTIMA® LT (Polyetheretherketone) per ASTM F2026, with a titanium laver and a vacuum plasma spray coating (Plasmapore®). The device will have Tantalum markers per ASTM F-560.The implant is secured to vertebral bodies by four titanium screws inserted through the anterior screw holes. The implants are offered in a variety of shapes and sizes to meet the requirements of the individual patient anatomy.
page 1 of 3
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the word "killing" written in a handwritten style. The letters are connected, and the word appears to be written with a dark ink or marker on a white background. The handwriting is somewhat stylized, with variations in the thickness of the strokes.
INDICATIONS FOR USE
The AIS SIBD XP Spinal System is a stand-alone device intended to be used with the four supplied bone screws if no supplemental fixation is used.
As an intervertebral body fusion device designed for use with autograft, the SIBD XP Spinal System is intended for spinal fusion procedures at one or two contiquous levels in the lumbar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondvlolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients may have had previous non-fusion spinal surgery at the involved spinal level(s).
Patients should be skeletally mature and must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the Aescular® Implant Systems device.
TECHNOLOGICAL CHARACTERISTICS(compared to Predicate(s))
The components of the SIBD XP Spinal System are offered in the same range of shapes and sizes as the predicate device. The material used for the Aescular® Implant Systems device is the same as that used to manufacture the predicate devices. The only difference between the predicate device and the subject device is the titanium laver and a vacuum plasma spray coating (Plasmapore®).
PERFORMANCE DATA
As recommended by the FDA Guidance for Spinal System 510(k)'s, non-clinical testing was performed to demonstrate that the AIS SIBD XP Spinal System is substantially equivalent to other predicate devices. The following testing was performed:
- Static and dynamic torsion per ASTM F2077 .
- Static and dynamic axial compression per ASTM F2077 .
- Static and dynamic shear compression testing per ASTM F2077 .
- Subsidence per ASTM F2267 .
- . Wear Debris per ASTM F2077 & ASTM F1877
- . Expulsion per ASTM Draft Standard F-04.25.02.02
In addition to FDA's Spine Guidance, Aesculap has also completed non-clinical testing recommended in the "Guidance for Industry on the Testing of Metallic Plasma Spraved Coatings on Orthopedic Implants to Support Reconsideration of Postmarket Surveillance Requirements." The following tests were performed:
- Microstructure of the coating per ASTM F1854 .
- Static Tensile Strength per ASTM F1147 .
- · Static Shear Strength per ASTM F1044
page 2 of 3
{2}------------------------------------------------
illzz
- Shear Fatigue Test per ASTM F1160
- · Abrasion Resistance per ASTM F1978
The results of these tests showed that the AIS SIBD XP Spinal System meets or exceeds the performance of the predicate devices, and the device is therefore found to be substantially equivalent.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol on the right side. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
- 4 2011 പ്രിട്ര
Aesculap Implant Systems, Inc. % Ms. Lisa M. Boyle 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K111122
Trade/Device Name: Aesculap® Implant Systems (AIS) - SIBD XP Spinal System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD Dated: July 08, 2011 Received: July 11, 2011
Dear Ms. Boyle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{4}------------------------------------------------
Page 2 - Ms. Lisa M. Boyle
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/About/DA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark A. Mulhern
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
INDICATIONS FOR USE STATEMENT A.
510(k) Number: k11/22
Device Name: Aesculap® Implant Systems (AIS) - SIBD XP Spinal System
Indications for Use:
The AIS SIBD XP Spinal System is a stand-alone device intended to be used with the four supplied bone screws if no supplemental fixation is used.
As an intervertebral body fusion device designed for use with autograft, the SIBD XP Spinal System is intended for spinal fusion procedures at one or two contiguous levels in the Uyotar spine from L2 to S1 in patients with degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis at the involved level(s). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients may have had previous non-fusion spinal surgery at the involved spinal level(s).
Patients should be skeletally mature and must have undergone a regimen of at least six ({) months of non-operative treatment prior to being treated with the Aesculap® Implant Systems device.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (per 21 CFR 801.109)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Auckth for MXM 8/3/11
(Division Sign-Off) Division of Surgical. Orthopedic, and Restorative Devices
510(k) Number K111222
N/A