(162 days)
Powder-free yellow synthetic vinyl patient examination gloves is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Powder-free yellow synthetic vinyl patient examination gloves that meets all of the requirements of ASTM standard D 5250-06.
Here's an analysis of the provided 510(k) summary, focusing on the acceptance criteria and study details:
Acceptance Criteria and Device Performance
The device in question is a "Powder-free yellow synthetic vinyl patient examination glove." The summary indicates that the device's performance was evaluated against established standards for patient examination gloves.
1. Table of Acceptance Criteria and Reported Device Performance:
Characteristics | Acceptance Criteria (Standard) | Reported Device Performance |
---|---|---|
Dimension | ASTM standard D 5250-06 | Meets |
Physical Properties | ASTM standard D 5250-06 | Meets |
Freedom from pinholes | 21 CFR 800.20 | Meets |
Powder Residual | ASTM standard D 5250-06 and D6124-01 |
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.