K Number
K110821
Device Name
BARF BAND
Date Cleared
2012-02-27

(340 days)

Product Code
Regulation Number
N/A
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Barf Band is indicated for the relief of nausea. Nausea is a symptom that may be experienced due to a variety of causes, for example: pregnancy (morning sickness), motion sickness, anesthesia and chemotherapy.

Device Description

The Barf Band is a wristband with adjustable properties that operates by putting pressure on Nei Kuan acupressure point (also known as the P6 point) on each wrist by means of a plastic dome/stud. Barf Bands are made from PVC free polyurethane, ABS plastic covered with Santoprene and 100% cotton. The band itself is composed of polyurethane. The ABS plastic covered with Santoprene is used to construct the plastic dome/stud that applies the pressure to the P6 point. The cotton is used inside the band to make the band comfortable to the end user.

AI/ML Overview

The provided text is a 510(k) summary for the "Barf Band" and a subsequent FDA letter of substantial equivalence. This type of document, particularly for an acupressure device, focuses primarily on establishing substantial equivalence to predicate devices rather than presenting detailed clinical study data with specific acceptance criteria, statistical analyses for performance, or ground truth establishment as would be seen for a more complex diagnostic or therapeutic device.

Therefore, many of the requested elements for a study proving acceptance criteria are not applicable or not present in this document. The submission relies on demonstrating similarity in intended use, technological characteristics, and principle of operation to existing approved devices (PSI Band and Sea Band), along with performance testing to show the applied force falls within the range of the predicates.

Here's an attempt to address your request based on the available information, indicating when information is not present:


Description of Acceptance Criteria and Study Proving Device Meets Criteria

Overview: The Barf Band's acceptance criteria and proof of meeting them are established primarily through demonstrating substantial equivalence to predicate devices (PSI Band and Sea Band). The key performance metric evaluated was the force applied by the wristband, which was shown to fall within the range established by the predicate devices. This approach is common for lower-risk devices where direct head-to-head clinical efficacy studies against specific criteria are not typically required for 510(k) clearance if substantial equivalence can be shown.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criterion (Implicit)Reported Device Performance
Applied force on Nei Kuan (P6) acupressure point falls within the range of predicate devices."The force applied by the Barf Band wristband falls within the range established by the predicate devices." (Specific numerical range not provided in the summary.)
Intended Use/Indications for Use are the same as predicate devices."The Barf Band has the same intended uses and indications for use... for the relief of nausea due to pregnancy (morning sickness), motion sickness, anesthesia, and chemotherapy." (Verified by comparison to predicate device labeling)
Similar technological characteristics.Barf Band: wristband with adjustable properties, plastic dome/stud for pressure on P6. Predicates are described as having similar characteristics.
Same principles of operation.Achieves pressure on the Nei Kuan (P6) acupressure point on each wrist. This is the same as the predicate devices.
No new issues of safety or effectiveness."The minor technological differences... raise no new issues of safety or effectiveness."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated for performance testing. The document refers to "Performance testing conducted on the Barf Band and the predicate devices," implying a comparison. It does not appear to be a clinical trial with a "test set" in the traditional sense of patient data.
  • Data Provenance: Not applicable. The testing described appears to be bench testing of the device's physical properties (force application) and a comparison of those properties to predicate devices, rather than clinical patient data.
  • Retrospective/Prospective: Not applicable based on the nature of the described testing.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: Not applicable. The "ground truth" for this submission revolves around the characteristics of the device itself and its comparison to predicate devices, not expert interpretation of clinical data.
  • Qualifications of Experts: Not applicable.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. This was not a study involving expert adjudication of patient cases.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • MRMC Study: No, an MRMC comparative effectiveness study was not done. The submission is for a physical acupressure device, not an AI-assisted diagnostic or therapeutic tool for which such a study would be relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Standalone Performance: Not applicable. The Barf Band is a physical device, designed for human use, not an algorithm.

7. The Type of Ground Truth Used

  • Type of Ground Truth: The "ground truth" in this context is the physical properties and performance characteristics of the predicate devices (PSI Band and Sea Band) regarding the force applied to the acupressure point. The Barf Band's performance was compared to this established range.

8. The Sample Size for the Training Set

  • Training Set Sample Size: Not applicable. This device is not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set was Established

  • Training Set Ground Truth Establishment: Not applicable, as there is no training set for this type of device submission.

{0}------------------------------------------------

K110821

FEB 2 7 2012

510(k) SUMMARY

Doodlebug Products's Barf Band

Date Prepared:

1/30/2012

Name of Device and Name/Address of Sponsor

Doodlebug Barf Band

Doodlebug Products 1809 Rockmoor Ave Austin, Tx 78703 Phone: 615-400-0032

Common or Usual Name:

Barf Band

Classification Name:

Acupressure Device

Predicate Devices

PSI Band by Psi Health Solutions, Inc.

Sea Band by Sea-Band Limited

Intended Use / Indications for Use

The Barf Band is indicated for use in: The Barf Band is indicated for the relief of nausea. Nausea is a symptom that may be experienced due to a variety of causes, for example: pregnancy (morning sickness), motion sickness, anesthesia and chemotherapy.

Technological Characteristics

The Barf Band is a wristband with adjustable properties that operates by putting pressure on Nei Kuan acupressure point (also known as the P6 point) on each wrist by means of a plastic dome/stud. The P6

{1}------------------------------------------------

Barf Bands are made from PVC free polyurethane, ABS plastic covered with Santoprene and 100% cotton. The band itself is composed of polyurethane. The ABS plastic covered with Santoprene is used to construct the plastic dome/stud that applies the pressure to the P6 point. The cotton is used inside the band to make the band comfortable to the end user.

Substantial Equivalence

The Barf Band is as safe and effective as the PSI Band (K070766) and Sea Band (K033268). The Barf Band has the same intended uses and indications for use, similar technological characteristics, and same principles of operation as its predicate device. The minor technological differences between the Barf Band and its predicate devices (e.g., dimensional and material composition) raise no new issues of safety or effectiveness. Performance testing conducted on the Barf Band and the predicate devices supports this. The force applied by the Barf Band wristband falls within the range established by the predicate devices. Thus, the Barf Band is substantially equivalent.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three curved lines representing its wings, positioned to the right. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.

Doodlebug Products, LLC c/o Barry E. Sands ROMIS Inc., President 5 Hemingway Lane, West Newbury, MA 01985

Re: K110821

Trade/Device Name: Barf Band Regulation Number: none Regulation Name: n/a Regulatory Class: unclassified Product Code: MVV Dated: January 30, 2012 Received: January 31, 2012

Dear Mr. Sands:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

FEB 2 7 2012

{3}------------------------------------------------

Page 2 - Mr. Barry E. Sands

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

for

Sincerely yours,

Keira Alexander

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement

510(k) Number (if known): NA

K110821

Device Name: Barf Band

Indications for Use:

The Barf Band is indicated for the relief of nausea. Nausea is a symptom that may be experienced due to a variety of causes, for example: pregnancy (morning sickness), motion sickness, anesthesia and chemotherapy.

Prescription Use (Part 21 C.F.R. 801 Subpart D) AND/OR

Over-The-Counter Use_X (21 C.F.R. 807 Subpart C)

(Please DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Tesin.

K110821

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

Page / of /

510(k) Number.

N/A