(169 days)
The Fujifilm Digital Mammography System, Aspire HD (FDR MS-1000), generates full-field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and is intended for use in the same clinical applications as traditional screen-film mammography systems.
The Aspire HD is an integrated FFDM system combining an X-ray system made by Siemens with Fujifilm's a-Se detector and Acquisition Workstation (AWS). The Aspire HD creates digital mammography images by direct capture of x-ray energy using the a-Se detector is a Fujifilm design utilizing an exclusive dual layer a-Se scintillator with Direct Optical switching circuitry to acquire image data and transfer images to the AWS for automated post processing, technologist preview and QC, and subsequent transmission to hard copy printers, diagnostic workstations and archiving systems. The Aspire HD provides automated compression and one AEC mode.
The Aspire HD Acquisition Workstation (FDR 1000AWS) includes an off the shelf personal computer, the application software, either Microsoft Vista or Windows 7 Operating System, a Smegapixel portrait type monitor, and a hub. The hub transmits signals between the personal computer and control cabinet, and between the personal computer and exposure stand.
The AWS display primarily consists of three windows:
- Patient Information Input window .
- Exposure Menu Selection window t
- . Study window.
The user may switch between these windows depending on the operation being performed. The X-ray control panel, which controls and observes the exposure stand, is always displayed in the lower part of each window. This allows setting the exposure conditions and confirming the radiation conditions on a single view.
Here's an analysis of the provided text regarding the FUJIFILM Aspire HD Full-Field Digital Mammography X-ray System, focusing on the acceptance criteria and the study that proves the device meets them.
It's important to note that the provided documents are a 510(k) Summary and the FDA's Substantial Equivalence letter. These documents typically focus on demonstrating equivalence to a predicate device rather than presenting a detailed clinical study with explicit acceptance criteria and corresponding performance metrics for an AI algorithm. Since this device is a digital mammography system itself and not an AI-based diagnostic tool, the interpretation below will reflect the information provided for such a system.
Acceptance Criteria and Device Performance
The provided text does not contain a typical "acceptance criteria table" with specific quantitative metrics (e.g., sensitivity, specificity, AUC) and corresponding reported performance values that would be expected for an AI diagnostic device. Instead, for a digital mammography system, the "acceptance criteria" are implied by demonstrating substantial equivalence to predicate devices through technical comparisons and a qualitative clinical image attribute review.
Implicit Acceptance Criteria and Reported Performance (for a digital mammography system):
| Acceptance Criteria (Implied by 510(k)) | Reported Device Performance |
|---|---|
| Technical Equivalence: Similar imaging characteristics to predicate devices (Siemens Mammomat Novation, Siemens Inspiration). | "The technological characteristics of the device are similar as demonstrated by the comparison of imaging characteristics such as MTF, Noise Analysis, DQE, and phantom testing." |
| Image Quality for Mammographic Use: Sufficeintly acceptable image quality for screening and diagnosis of breast cancer, as determined by independent expert review. | "A clinical image attribute review was conducted by independent mammographic radiologists in accordance with the FFDM 510(k) Guidance document. The mammographic attributes of six (6) image sets of screening and diagnostic cases were reviewed concluding that the Aspire HD provides sufficiently acceptable quality for mammographic use." |
| Safety and Efficacy Equivalence: No new safety or efficacy issues compared to predicate devices; suitable for the same clinical applications as traditional screen-film mammography systems. | "The Aspire HD introduces no new safety or efficacy issues other than those already identified with the predicate devices." "This 510(k) premarket notification submission has demonstrated Substantial Equivalence... We conclude the subject device to be as safe and effective as the predicate devices." "Used for screening and diagnosis of breast cancer and is intended for use in the same clinical applications as traditional screen-film mammography systems." |
Study Details
-
Sample size used for the test set and the data provenance:
- Sample Size: Six (6) image sets of screening and diagnostic cases.
- Data Provenance: Not explicitly stated, but clinical cases are implied. The country of origin and whether the data was retrospective or prospective are not mentioned.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of Experts: "independent mammographic radiologists" (number not specified, but plural indicating more than one).
- Qualifications: "independent mammographic radiologists." Specific experience (e.g., 10 years) is not detailed.
-
Adjudication method for the test set:
- The document states an "image attribute review was conducted by independent mammographic radiologists." The specific adjudication method (e.g., 2+1, 3+1) is not provided. It seems to have been a collective review or consensus among the independent radiologists.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study involving AI assistance was not done. This document pertains to a device (mammography system), not an AI algorithm intended to assist human readers. The clinical review was to assess the image quality of the Aspire HD system itself, not its impact on human reader performance with or without AI.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- No, this is not applicable. The Aspire HD is a full-field digital mammography X-ray system, not a standalone algorithm. The "performance" being assessed is the system's ability to produce diagnostically acceptable mammographic images.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The "ground truth" for the image attribute review was based on the expert opinion/consensus of independent mammographic radiologists regarding the "sufficiently acceptable quality for mammographic use" of the images. It was not based on pathology or outcomes data in the typical sense of a diagnostic accuracy study.
-
The sample size for the training set:
- Not applicable. This document describes a medical imaging device (hardware and integrated software for image acquisition), not a machine learning model that relies on a "training set" in the context of AI. The system itself undergoes design and engineering, but doesn't have a data-driven "training set" like an AI algorithm would.
-
How the ground truth for the training set was established:
- Not applicable, as there is no "training set" for this device in the context of AI development.
{0}------------------------------------------------
419 WEST AVENUE STAMFORD, CT 06902 203-324-2000 PHONE: TOLL-FREE: 800-431-1850 203-353-0926 FAX:
KI10729
510(k) Summary
March 10, 2011 Date Prepared:
Submitter's Information
FUJIFILM Medical Systems U.S.A., Inc. 419 West Avenue Stamford, Connecticut 06902 Telephone: (203) 602-3576 Facsimile: (203) 353-0296 Contact: Peter Altman
Device Name and Classification:
| Product Name: | Aspire HD Full-Field Digital Mammography X-ray System |
|---|---|
| Model Number: | FDR MS-1000 |
| Classification Name: | Full-Field Digital Mammography X-ray System |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR 892.1715 |
| Device Class: | Class II |
| Product Code: | MUE |
Substantial Equivalence/Predicate Devices:
Siemens Mammomat Novation (P030010)
Siemens Inspiration (P030010/S006)
Although the wording changes slightly from device to device, the Aspire HD and the Siemens Mammomat Novation on both have the same basic Indication For Use (IFU). The IFU for the Siemens Mammomat Inspiration is not publically available, although it is expected to be very similar to the Novation and the Aspire HD. All three devices generate digital mammographic images that are intended for screening and diagnosis of breast cancer and are intended for use in the same clinical applications as traditional screen-film mammography systems.
All systems employ digital amorphous selenium x-ray detectors integrated into gantry (stand) based x-ray systems. The x-ray stand, tube and generator of the Aspire HD and Inspiration are identical. Note that the Aspire HD's X-ray system is manufactured by Siemens and is the same as that used in the recently approved (November 5, 2010) Siemens Mammomat Inspiration (P030010/S006). The Inspiration is Siemens' successor to the Mammomat Novation 148 (P030010). The technological characteristics of the device are similar as demonstrated by the comparison of imaging characteristics such as MTF, Noise Analysis, DQE, and phantom testing.
A clinical image attribute review was conducted by independent mammographic radiologists in accordance with the FFDM 510(k) Guidance document. The mammographic attributes of six (6) image sets of screening and diagnostic cases were reviewed concluding that the Aspire HD provides sufficiently acceptable quality for mammographic use.
page lotz
Page 25 of 1560
{1}------------------------------------------------
FUJ!FILM
FUJIFILM MEDICAL SYSTEMS USA, INC.
419 WEST AVENUE STAMFORD, CT 06902 203-324-2000 800-431-1850 203-353-0926
Description of the Device:
The Aspire HD is an integrated FFDM system combining an X-ray system made by Siemens with Fujifilm's a-Se detector and Acquisition Workstation (AWS). The Aspire HD creates digital mammography images by direct capture of x-ray energy using the a-Se detector is a Fujifilm design utilizing an exclusive dual layer a-Se scintillator with Direct Optical switching circuitry to acquire image data and transfer images to the AWS for automated post processing, technologist preview and QC, and subsequent transmission to hard copy printers, diagnostic workstations and archiving systems. The Aspire HD provides automated compression and one AEC mode.
The Aspire HD Acquisition Workstation (FDR 1000AWS) includes an off the shelf personal computer, the application software, either Microsoft Vista or Windows 7 Operating System, a Smegapixel portrait type monitor, and a hub. The hub transmits signals between the personal computer and control cabinet, and between the personal computer and exposure stand.
The AWS display primarily consists of three windows:
- Patient Information Input window .
- Exposure Menu Selection window t
- . Study window.
The user may switch between these windows depending on the operation being performed. The X-ray control panel, which controls and observes the exposure stand, is always displayed in the lower part of each window. This allows setting the exposure conditions and confirming the radiation conditions on a single view.
Intended Use:
The Fujifilm Digital Mammography System, Aspire HD (FDR MS-1000), generates full-field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and is intended for use in the same clinical applications as traditional screen-film mammography systems.
Safety Information:
The Aspire HD introduces no new safety or efficacy issues other than those already identified with the predicate devices. The results of the Hazard Analysis combined with the appropriate preventive measures taken indicate that the device is of moderate concern as per the May 11. 2005 issue of the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." and is consistent with the level of concern indicated in the "Class II Special Controls Guidance Document: Full-Field Digital Mammography System" document issued on: November 5, 2010.
Conclusion:
This 510/k) premarket notification submission has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health. We conclude the subject device to be as safe and effective as the predicate devices.
page 2 of 2
Page 26 of 1560
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo of the Department of Health & Human Services. The logo consists of a symbol on the left and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" on the right. The symbol appears to be a stylized representation of a human figure.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. Peter Altman Regulatory Consultant FUJIFILM Medical Systems USA. Inc. 419 West Avenue STAMFORD CT 06902
SEP - 1 2011
K110729 Trade Name: Aspire HD (FDR MS-1000) Regulation Number: 21 CFR § 892.1715 Regulation Name: Full-field digital mammography system Regulatory Class: II Product Code: MUE Dated: June 22, 2011 Received: June 23, 2011
Dear Mr. Altman:
Re:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{3}------------------------------------------------
Page 2
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours.
Mary S. Postel
Mary Pastel, ScD. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known): _ K (10729
Device Name: Aspire HD (FDR MS-1000)
Indications for Use:
The Fujifilm Digital Mammography System, Aspire HD (FDR MS-1000), generates full-field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and is intended for use in the same clinical applications as traditional screen-film mammography systems.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Mary S Postel
(Division Sign-Off)
Radiolonio Office of In vial Diagnostic Device Evaluation and Safety
510K K110729
§ 892.1715 Full-field digital mammography system.
(a)
Identification. A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”See § 892.1(e) for the availability of this guidance document.