K Number
K110635
Date Cleared
2011-06-23

(111 days)

Product Code
Regulation Number
868.5165
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The INOmax DS delivery system delivers INOmax® (nitric oxide of inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

The INOmax DS provides continuous integrated monitoring of inspired O2, NO3, and NO, and a comprehensive alarm system.

The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup.

The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

Device Description

The INOmax DSIR uses a "dual-channel" design to ensure the safe delivery of INOmax. The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2, and O2 cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOmax delivery independent of monitoring but also allows the monitoring system to shutdown INOmax delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm.

AI/ML Overview

The provided document describes the INOmax DSIR delivery system for nitric oxide and its compatibility with additional ventilators. The study discussed is a non-clinical test to establish substantial equivalence for the updated device rather than an AI-powered diagnostic tool, therefore many of the requested criteria are not applicable.

Here's an analysis of the provided information, focusing on the relevant sections and explicitly stating when information is not applicable (N/A) for this type of device and study:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Requirements for operation and interface to selected ventilators)Reported Device Performance
O2 dilutionPerformed within published specifications
Effect on delivered pressuresPerformed within published specifications
INOmax DSIR delivery accuracyPerformed within published specifications
NO2 generationPerformed within published specifications
Backup DeliveryPerformed within published specifications

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set:
    • Five INOmax DSIR settings were used: 0 (baseline), 1, 5, 20, 40, and 80 ppm.
    • These settings were applied for each setting and mode of ventilation/patient cannula for the five additional respiratory care devices.
    • The "Backup mode" was also tested for each.
    • (Note: The document doesn't specify the number of times each setting/mode combination was tested or the duration of each test).
  • Data Provenance: Not specified, but implied to be from laboratory testing related to the submitter (INO Therapeutics/Ikaria). The study is prospective in that it was conducted specifically for this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

N/A. This was a non-clinical engineering and performance study of a medical device, not a diagnostic study requiring expert human interpretation of data for ground truth. The "ground truth" was the physical performance of the device against predefined technical specifications.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

N/A. As this was a non-clinical performance study, there was no need for human adjudication of results in the way it's applied in diagnostic studies. The results were measured and recorded against established device specifications.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

N/A. This study is not an MRMC comparative effectiveness study, nor does it involve AI assistance for human readers. It's a non-clinical compatibility and performance test of a medical device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

N/A. This device does not have an "algorithm only" component in the diagnostic sense that would be tested for standalone performance. The device itself is the "system" being tested.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this non-clinical study was the published specifications of the INOmax DSIR and the compatibility requirements for its operation with the selected ventilators. The testing aimed to demonstrate that the device performed within these established specifications.

8. The sample size for the training set

N/A. There is no mention of a training set as this is a non-clinical device performance study, not an AI or machine learning model development.

9. How the ground truth for the training set was established

N/A. There is no training set for this type of non-clinical device performance study.

{0}------------------------------------------------

INO Therapeutics/Ikaria Premarket Notification

INOmax DS1R with additional Ventilators

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided: Date: February 28, 2011

  • INO Therapeutics/Ikaria Submitter: 2902 Dairy Drive Madison, Wisconsin 53718
    Primary Contact Person: Larry Lepley Associate Director, Regulatory Affairs INO Therapeutics/Ikaria T: 608-226-3415 F: 608-226-3402

  • Secondary Contact Person: David Trueblood Director, Regulatory Affairs INO Therapeutics/Ikaria T: 608-395-3910 F: 608-226-3402

    • INOmax® DS (Delivery System) Device: Trade Name:

Common/Usual Name: Nitric Oxide Administration Apparatus (primary) Nitric Oxide Administration Apparatus, Back-up System Nitric Oxide Analyzer Nitrogen Dioxide Analyzer

  • Classification Names: Apparatus, Nitric Oxide Delivery, or Apparatus, Nitric Oxide Backup Delivery, Class II - 21 CFR 868.5165
    • MRN (Primary), MRQ, MRP Product Code:

K061901, K070867, K071516, K080484, K081691, K090958, Predicate Device(s): K092545. K093922

Device Description: The INOmax DSIR uses a "dual-channel" design to ensure the safe delivery of INOmax. The first channel has the delivery CPU, the flow controller and the injector module to ensure the accurate delivery of NO. The second channel is the monitoring system, which includes a separate monitor CPU, the gas cells (NO, NO2, and O2 cells) and the user interface including the display and alarms. The dual-channel approach to delivery and monitoring permits INOmax delivery independent of monitoring but also allows the monitoring system to shutdown INOmax delivery if it detects a fault in the delivery system such that the NO concentration could become greater than 100 ppm.

{1}------------------------------------------------

Intended Use:

The INOmax DS delivery system delivers INOmax® (nitric oxide of inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

The INOmax DS provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system. The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup.

The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

Technology: All revisions of INOmax DSIR utilize component technology to deliver Nitric Oxide gas to the patient. The components consist of the Delivery System unit, the blender, a stand/cart and the NO gas tanks. In this proposed revision the INOmax DSIR, technological characteristic of design has not changed with the inclusion of five additional respiratory care devices that Ikaria has tested to be compatible for use with the INOmax DSJR system. Also there are two duration tables added to the INOmax DSJR Operation Manual. These were added as a courtesy to the user. By providing a table, the user can quickly determine how long a cylinder will last given the set dose and ventilator flow rate. This is especially helpful during medical transport. The five additional respiratory care devices include:

(K101803) Newport E360TM

(K091238) Impact Instrumentation EMV+ Ventilator®

(K061736) Teleflex Comfort FloTM Humidification System and Nasal Cannula

(K093632) Dräger Babylog® VN500

(K093633) Dräger Evita Infinity® V500

{2}------------------------------------------------

Determination ofSummary of Non-Clinical Tests:
Substantial Equivalence:The testing concluded all requirements necessary for the operation of the INOmax DSIR and interface to the selected

ventilators to be compatible:

All five respiratory care devices were set up and calibrated according to the manufacturer's recommendations, and tested using the settings established for each respiratory care device test. The INOmax DSR was set up and calibrated according to the manufacturer's recommendations.

Five INOmax DSIR settings were used [0 (baseline), 1, 5, 20, 40, and 80 ppm] for each setting and mode of ventilation/patient cannula, as well as the Backup mode.

The measured values on the INOmax DSJR were also recorded along with any anomalies found.

The testing concluded five requirements necessary for the operation of the INOmax DSIR and the five respiratory care devices to be compatible:

  • O2 dilution
  • Effect on delivered pressures
  • INOmax DSIR delivery accuracy
  • NO2 generation
  • Backup Delivery

Testing Conclusion:

The INOmax DSJR performed within published specifications when used with each of the ventilators.

Summary of Clinical Tests:

The subject of this premarket submission, INOmax DSIR, interfaced to each of the selected respiratory care devices, did not require clinical studies to support substantial equivalence.

INO Therapeutics/Ikaria considers the INOmax DSIR to be as safe Conclusion: and as effective as the predicate device, with performance substantially equivalent to the predicate device.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Larry Lepley Associate Director, Regulatory Affairs INO Therapeutics LLC/DBA Ikaria 2902 Dairy Drive Madison, Wisconsin 53718

JUN 2 3 2011

Re: K110635

Trade/Device Name: INOmax® DS Delivery System Regulation Number: 21 CFR 868.5165 Regulation Name: Nitric Oxide Administration Apparatus Regulatory Class: II Product Code: MRN Dated: May 24, 2011 Received: May 25, 2011

Dear Mr. Lepley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2- Mr. Lepley

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Susan Quare

Anthony D. Watson, B.S., M.S., M.B.A. Director . Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

INO Therapeutics/Ikaria Premarket Notification

INOmax DSJR with additional Ventilators

510(k) Number (if known):

Device Name: INOmax DS

Indications for Use:

The INOmax DS delivery system delivers INOmax® (nitric oxide of inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.

The INOmax DS provides continuous integrated monitoring of inspired O2, NO3, and NO, and a comprehensive alarm system.

The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source.

The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup.

The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

Prescription Use_X____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Anesthesiology, General Hospital .nfection Control, Dental Devices

510(k) Number: K110635

§ 868.5165 Nitric oxide administration apparatus.

(a)
Identification. The nitric oxide administration apparatus is a device used to add nitric oxide to gases that are to be breathed by a patient. The nitric oxide administration apparatus is to be used in conjunction with a ventilator or other breathing gas administration system.(b)
Classification. Class II. The special control for this device is FDA's “Guidance Document for Premarket Notification Submissions for Nitric Oxide Administration Apparatus, Nitric Oxide Analyzer, and Nitrogen Dioxide Analyzer.”