K Number
K103788
Manufacturer
Date Cleared
2011-11-08

(316 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PENTRA C200 is a discrete photometric benchtop chemistry analyzer for use in clinical laboratories. It is not intended for use in Point of Care settings. It duplicates manual analytical procedures by performing various steps such as pipetting, mixing, heating and measuring color intensity. The PENTRA C200 is intended for quantitative measurements of a variety of analytes: Glucose, Sodium, Potassium and Chloride. ABX Pentra Glucose HK CP reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of glucose in serum and plasma using glucose hexokinase method by colorimetry. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. The I.S.E. (Ion Selective Electrode) module is intended for the quantitative determination of Sodium, Potassium and Chloride in serum, plasma and urine by potentiometry using ion selective electrode with associated reference solution, calibrators and controls. Measurement of Sodium, Potassium and Chloride are used in diagnosis and treatment of diseases involving electrolyte imbalance.

Device Description

The PENTRA C200 is a benchtop clinical chemistry analyzer using two measuring principles: absorbance and ion selective electrodes. The instrument may be summarized as follows: Multi-parametric (up to 15 simultaneous tests + 3 ISE tests). On routine or Stat. 90 (without ISE) to 360 (with ISE) tests / hour (in single or bi-reaction mode) -(analytical cycle of 40 seconds). random access working on primary tubes or sample cups. ABX PENTRA reagent cassettes are compact and ready-to-use. on board bar-code reader is used to identify newly loaded reagent cassettes and samples for patient identification.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the PENTRA C200, I.S.E. Module, and ABX PENTRA Glucose HK CP, based on the provided text:

Acceptance Criteria and Reported Device Performance

Note: The document explicitly states "Clinical testing met all acceptance criteria," but does not detail the specific numerical acceptance criteria for each performance metric (e.g., "CV Total ≤ X%"). Instead, it provides the reported performance of the device. For the purpose of this response, the "Acceptance Criteria" column will represent the performance values achieved by the predicate device (if stated or implied to be the benchmark for substantial equivalence) or the general clinical expectations for such devices, and the "Reported Device Performance" column will show the results for the PENTRA C200. Since explicit numerical acceptance criteria for many metrics are not provided beyond the predicate device comparison, I will use the "Reported Device Performance" to demonstrate meeting the general requirement of being "substantially equivalent."

Performance MetricAnalyteSample TypeAcceptance Criteria (Implied / Predicate)Reported Device Performance (PENTRA C200)
Accuracy and Precision (CV Total)GlucoseSerum & PlasmaNot explicitly stated (likely ≤ predicate)≤ 1.99%
Limit of QuantitationGlucoseSerum & PlasmaNot explicitly stated (likely ≤ predicate)5 mg/dL
Measuring Range (Glucose)GlucoseSerum & PlasmaNot explicitly stated (likely ≥ predicate)5 mg/dL - 900 mg/dL (Automatic post-dilution: 2700 mg/dL)
Correlation (Glucose)GlucoseSerum & PlasmaNot explicitly stated (likely r² close to 1)Y = 0.98 x + 4.46 (mg/dL) with r² = 0.998
Calibration StabilityGlucoseN/ANot explicitly stated (likely ≥ predicate)20 days
Reagent StabilityGlucoseN/ANot explicitly stated (likely ≥ predicate)On-board (refrigerated): 39 days
Accuracy and Precision (CV Total)SodiumSerum & PlasmaNot explicitly stated (likely ≤ predicate)≤ 1.1 %
SodiumUrineNot explicitly stated (likely ≤ predicate)≤ 4.91 %
Linearity & Measuring RangeSodiumSerum & PlasmaNot explicitly stated (likely ≥ predicate)90 - 190 mmol/L
SodiumUrineNot explicitly stated (likely ≥ predicate)60 – 280 mmol/L
CorrelationSodiumSerumNot explicitly stated (likely r² close to 1)Y = 0.96 x + 6.42 with r² = 0.982
SodiumPlasmaNot explicitly stated (likely r² close to 1)Y = 1.05 x - 5.32 with r² = 0.998
SodiumUrineNot explicitly stated (likely r² close to 1)Y = 1.01 x - 2.20 with r² = 0.989
Accuracy and Precision (CV Total)PotassiumSerum & PlasmaNot explicitly stated (likely ≤ predicate)≤ 1.07 %
PotassiumUrineNot explicitly stated (likely ≤ predicate)≤ 2.87 %
Linearity & Measuring RangePotassiumSerum & PlasmaNot explicitly stated (likely ≥ predicate)2 - 9.5 mmol/L
PotassiumUrineNot explicitly stated (likely ≥ predicate)25 - 250 mmol/L
CorrelationPotassiumSerumNot explicitly stated (likely r² close to 1)Y = 1.01 x - 0.06 with r² = 0.998
PotassiumPlasmaNot explicitly stated (likely r² close to 1)Y = 1.01 x - 0.09 with r² = 0.998
PotassiumUrineNot explicitly stated (likely r² close to 1)Y = 1.02 x - 0.27 with r² = 0.997
Accuracy and Precision (CV Total)ChlorideSerum & PlasmaNot explicitly stated (likely ≤ predicate)≤ 1.55 %
ChlorideUrineNot explicitly stated (likely ≤ predicate)≤ 4.59 %
Linearity & Measuring RangeChlorideSerum & PlasmaNot explicitly stated (likely ≥ predicate)70 – 170 mmol/L
ChlorideUrineNot explicitly stated (likely ≥ predicate)70 – 280 mmol/L
CorrelationChlorideSerumNot explicitly stated (likely r² close to 1)Y = 0.96 x + 3.74 with r² = 0.996
ChloridePlasmaNot explicitly stated (likely r² close to 1)Y = 1.04 x - 4.17 with r² = 0.997
ChlorideUrineNot explicitly stated (likely r² close to 1)Y = 1.04 x - 5.63 with r² = 0.987

Non-Clinical Safety Requirements:

  • IEC 61010-1 / IEC 61010-2-081 / IEC 61010-2-101 (Safety requirements for electrical equipment)
  • EN 61326-2-6 (EMC requirements)
  • UL631010 - 1 / CSA - C22.2 No. 61010-1 (Safety requirements)

Reported Device Performance: "The PENTRA C200 (with ISE module) meets:" all listed non-clinical safety requirements.


Study Information

  1. Sample sizes used for the test set and the data provenance:

    • Glucose: n=103 for correlation study (sample type: Serum & Plasma). Data provenance not specified (retrospective or prospective, country of origin).
    • Sodium:
      • Serum (n=129) for correlation.
      • Plasma (n=132) for correlation.
      • Urine (n=101) for correlation.
      • Data provenance not specified.
    • Potassium:
      • Serum (n=122) for correlation.
      • Plasma (n=125) for correlation.
      • Urine (n=159) for correlation.
      • Data provenance not specified.
    • Chloride:
      • Serum (n=170) for correlation.
      • Plasma (n=131) for correlation.
      • Urine (n=112) for correlation.
      • Data provenance not specified.

    The studies appear to be comparative, correlating the PENTRA C200 results against a reference method or the predicate device. The samples used are described as "Serum & plasma" and "Urine" for the different analytes.

  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document does not specify the use of human experts to establish ground truth. These are clinical chemistry analyzer performance studies, where "ground truth" (or reference values) would typically come from a reference method run on another FDA-cleared device or a defined gold standard lab method.
  3. Adjudication method for the test set:

    • Not applicable as the measured values are compared against a reference/predicate method, not against expert consensus.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an imaging or diagnostic AI device requiring human reader adjudication or assistance. It's a clinical chemistry analyzer.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, the performance data provided is for the device operating in a standalone capacity, measuring analytes without human intervention in the analytical process itself.
  6. The type of ground truth used:

    • The ground truth (or reference values) for the correlation studies for Glucose, Sodium, Potassium, and Chloride would be obtained from measurements performed by the predicate device (ABX PENTRA 400) as part of the substantial equivalence claim, or another established clinically validated reference method. The correlation equations presented directly compare the candidate device (Y) to an x-value (presumably the predicate or reference method).
  7. The sample size for the training set:

    • Not applicable. This described device is a physical clinical chemistry analyzer, not a machine learning or AI algorithm in the traditional sense that requires a "training set" to learn from data. Its analytical methods (spectrophotometry, potentiometry) are based on established chemical principles.
  8. How the ground truth for the training set was established:

    • Not applicable, as there is no specific "training set" for this type of device.

{0}------------------------------------------------

NOV = 8 2011

Premarket Notification [510(k)] Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is : K103788

Company: Horiba ABX SAS Parc Euromédecine Rue du Caducée - BP 7290 34184 Montpellier cedex 4 FRANCE Telephone: + (33) 4 67 14 18 43 Fax: + (33) 4 67 14 15 17

Contact Person: Caroline Ferrer (caroline.ferrer@horiba.com)

Date Prepared: 07th September 2011

Device Name:

ﻴﺔ ﺍﻟﻤﺮﺍﺟﻊ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ

Trade/Proprietary Name:PENTRA C200
Common or Usual Name:Clinical Chemistry analyzer
Device ClassClass I : General Controls : Exempt from premarket.
Classification Name:§862.2160 : Discrete photometric chemistry analyzer for clinical use
Product Code:JJE
Trade/Proprietary Name:I.S.E. Module
Common or Usual Name:Ion Selective Electrode
Device ClassClass I : General Controls : Exempt from premarket.
Classification Name:§862.2160 : Discrete photometric chemistry analyzer for clinical use
Product Code:JJE
Trade/Proprietary Name:ABX PENTRA Glucose HK CP
Common or Usual Name:Glucose HK
Device ClassClass II
Classification Name:§862.1345 : Glucose Test System
Product Code:CFR ; Hexokinase, Glucose

The PENTRA C200 is a new device developed in a joint-venture by HORIBA Medical, and commercialized under HORIBA ABX SAS manufacturer responsibility. HORIBA ABX SAS is the registered company name. Our company is part of the Medical segment of HORIBA group. Trade name of our activity is HORIBA Medical.

{1}------------------------------------------------

Predicate Devices:

The data and information supplied in this submission demonstrates substantial equivalence to their respective predicate devices:

Candidate devicePredicate device
510(k) numberDevice name
PENTRA C200K052007ABX PENTRA 400
PENTRA C200 I.S.E. ModuleK052007ABX PENTRA 400 I.S.E. Module
ABX PENTRA Glucose HK CPK052007ABX Pentra Glucose HK CP

Similarities and Differences between the predicate devices and candidate devices:

Predicate deviceCandidate device
Device NameABX PENTRA 400PENTRA C200
Instrument TypeBenchtopSame
Separate workstationNoNo
Touch Screen InterfaceYesYes
Intended UseDiscrete photometricbenchtop chemistryanalyzer for clinical useSame
Maximum throughput420 Tests/Hour360 Tests/Hour
MethodologiesSpectrophotometryMono and Bi-chromaticmeasurement of lightabsorbancePotentiometry (for I.S.E.Module)Same
STAT CapabilityYesYes
Sample ID InputBarcoded IDManualSame
Reagent BarcodeReaderIntegratedSame
Reagent PositionsUp to 52 positions+ 3 I.S.E.Up to 15 positions+ 3 I.S.E.
Sample PositionsUp to 60 samplesUp to 15 positions

Table1: Comparison between PENTRA C200 and ABX PENTRA 400 (K052007)

{2}------------------------------------------------

Table 2: Comparison between PENTRA C200 I.S.E. Module and ABX PENTRA
400 I.S.E. Module (K052007)
Predicate deviceCandidate device
Device NameI.S.E. ModuleI.S.E. Module
InstrumentABX PENTRA 400PENTRA C200
OptionalYesYes
ParametersNa, K, ClSame
PotentiometryDirect and IndirectSame
Material
Sodium ElectrodeGlass membraneselective to Na+ ionsSame
Potassium ElectrodePlastic membraneselective to K+ ionsSame
Chloride ElectrodePlastic membraneselective to Cl- ionsSame
Specimen TypesSerumPlasmaUrineSame

Table 3: Comparison between Glucose HK assay on PENTRA C200 and on ABX PENTRA 400 (K052007)

Predicate deviceCandidate device
Device NameABX PENTRA GlucoseHK CPABX PENTRA GlucoseHK CP
InstrumentABX PENTRA 400PENTRA C200
ParameterGlucoseGlucose
MethodEnzymatic method usinghexokinase couple withglucose-6-phosphatedehydrogenaseSame
MaterialBi-reagent cassette,ready-to-useReagent 1: NAD, ATP,Buffer Sodium azideReagent 2: Hexokinase,G-6-PDH, Magnesiumsulphate, Sodium azideSame
Specimen TypesSerumPlasmaUrineSerumPlasma

·

{3}------------------------------------------------

Substantial Equivalence:

It has been demonstrated that the PENTRA C200 can be considered substantially equivalent to the predicate device ABX PENTRA 400 (K052007).

The I.S.E. module has been similarly demonstrated as being substantially equivalent to the predicate device K052007.

The ABX PENTRA Glucose HK CP used on the PENTRA C200 has been demonstrated substantially equivalent to the same reagent used on the ABX PENTRA 400 as described in K052007.

Description:

The PENTRA C200 is a benchtop clinical chemistry analyzer using two measuring principles: absorbance and ion selective electrodes.

The instrument may be summarized as follows :

  • Multi-parametric (up to 15 simultaneous tests + 3 ISE tests) .
  • On routine or Stat
  • 90 (without ISE) to 360 (with ISE) tests / hour (in single or bi-reaction mode) -(analytical cycle of 40 seconds)
  • random access working on primary tubes or sample cups
  • ABX PENTRA reagent cassettes are compact and ready-to-use. .
  • on board bar-code reader is used to identify newly loaded reagent cassettes and samples for patient identification

{4}------------------------------------------------

Intended Use / Indications for use :

The PENTRA C200 is a discrete photometric benchtop chemistry analyzer for use in clinical laboratories. It is not intended for use in Point of Care settings.

It duplicates manual analytical procedures by performing various steps such as pipetting, mixing, heating and measuring color intensity. The PENTRA C200 is intended for quantitative measurements of a variety of analytes: Glucose, Sodium, Potassium and Chloride.

ABX Pentra Glucose HK CP reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of glucose in serum and plasma using glucose hexokinase method by colorimetry.

Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. *

The I.S.E. (Ion Selective Electrode) module is intended for the quantitative determination of Sodium, Potassium and Chloride in serum, plasma and urine by potentiometry using ion selective electrode with associated reference solution, calibrators and controls.

Measurement of Sodium, Potassium and Chloride are used in diagnosis and treatment of diseases involving electrolyte imbalance.

{5}------------------------------------------------

Discussion of Performance Data:

ABX PENTRA Glucose HK CP:
Sample typeSerum & plasma
Limit of Quantitation5 mg/dL
Accuracy and PrecisionCV Total ≤ 1.99%
Measuring range5 mg/dL - 900 mg/dLAutomatic post-dilution : 2700 mg/dL
Correlation (n=103)Y = 0.98 x + 4.46 (mg/dL) with r2 = 0.998.
Calibration stability20 days
Reagent stabilityOn-board stability (refrigerated area): 39 days
CalibratorABX Pentra Multical
ControlsABX Pentra N ControlABX Pentra P Control
ABX PENTRA Sodium - E :
Sample typeSerum & plasmaUrine
Accuracy and PrecisionCV Total ≤ 1.1 %CV Total ≤ 4.91 %
Linearity & Measuring range90 - 190 mmol/L60 – 280 mmol/L
CorrelationSerum (n=129)Y = $0.96 x + 6.42$ with $r^2$ =$0.982$ .Urine (n=101)Y = $1.01 x - 2.20$ with $r^2$ =$0.989$ .
Plasma (n=132)Y = $1.05 x - 5.32$ with $r^2$ =$0.998$ .
CalibratorsABX Pentra Standard 1ABX Pentra Standard 2ABX Pentra ReferenceABX Pentra Standard 1ABX Pentra Standard 2ABX Pentra Reference
ControlsABX Pentra N ControlABX Pentra P ControlABX Pentra N ControlABX Pentra P Control

{6}------------------------------------------------

ABX PENTRA Potassium - E :
Sample typeSerum & plasmaUrine
Accuracy and PrecisionCV Total ≤ 1.07 %CV Total ≤ 2.87 %
Linearity & Measuring range2 - 9.5 mmol/L25 - 250 mmol/L
CorrelationSerum (n=122)Y = $1.01 x - 0.06$ with $r^2$ =0.998.Urine (n=159)Y = $1.02 x - 0.27$ with $r^2$ =0.997.
Plasma (n=125)Y = $1.01 x - 0.09$ with $r^2$ =0.998.
CalibratorsABX Pentra Standard 1ABX Pentra Standard 2ABX Pentra ReferenceABX Pentra Standard 1ABX Pentra Standard 2ABX Pentra Reference
ControlsABX Pentra N ControlABX Pentra P ControlABX Pentra N ControlABX Pentra P Control
ABX PENTRA Chloride – E :
Sample typeSerum & plasmaUrine
Accuracy and PrecisionCV Total ≤ 1.55 %CV Total ≤ 4.59 %
Linearity & Measuring range70 – 170 mmol/L70 – 280 mmol/L
CorrelationSerum (n=170)$Y = 0.96 x + 3.74 with r2 = 0.996.$Plasma (n=131)$Y = 1.04 x - 4.17 with r2 = 0.997.$Urine (n=112)$Y = 1.04 x - 5.63 with r2 = 0.987.$
CalibratorsABX Pentra Standard 1ABX Pentra Standard 2ABX Pentra ReferenceABX Pentra Standard 1ABX Pentra Standard 2ABX Pentra Reference
ControlsABX Pentra N ControlABX Pentra P ControlABX Pentra N ControlABX Pentra P Control

Conclusions for non clinical and clinical tests :

The non clinical studies tests conclude that the safety and effectiveness of the devices are not compromised.

{7}------------------------------------------------

The PENTRA C200 (with ISE module) meets :

IEC 61010-1 / IEC 61010-2-081 / IEC 61010-2-101 :

Safety requirements for electrical equipment for measurement, control, and laboratory use

Part 1: General requirements

Part 2-081: Particular requirements for automatic and semi-automatic laboratory equipment for analysis and other purposes

Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment

EN 61326-2-6 : i

Standard for Electrical equipment for measurement, control and laboratory use -EMC requirements

Part 2-6: Particular requirements - In vitro diagnostic (IVD) medical device

UL631010 - 1 / CSA - C22.2 No. 61010-1 :

Safety requirements for electrical equipement for measurement, control, and laboratory use,

Part 1 : General Requirements

Clinical testing met all acceptance criteria, and data demonstrates that the devices are substantially equivalent to their predicate devices.

{8}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/8/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Public Health Service

Horiba ABX SAS c/o Ms. Caroline Ferrer Regulatory Affairs Specialist Parc Euromedecine Rue du Caducee - BP 7290 34184 Montpellier cedex 4 France

Food & Drug Administration 10903 New Hampshire Avenue Buildina 66 Silver Spring, MD 20993

NOV - 8 2011

Re: K103788 Trade Name: PENTRA C200, I.S.E. Module, and ABX PENTRA Glucose HK CP Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Codes: CFR, JGS, CEM, CGZ, JJE Dated: November 4 , 2011 Received: November 7 . 2011

Dear Ms. Ferrer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 80 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{9}------------------------------------------------

Page 2 -

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Viro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric' s (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 for (1 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Signature

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

Indications for Use Form

510(k) Number (if known): K103788

Device Name: PENTRA C200, I.S.E. Module, and ABX Pentra Glucose HK CP

Indications for Use:

The PENTRA C200 is a discrete photometric benchtop chemistry analyzer for use in clinical laboratories. It is not intended for use in Point of Care settings.

It duplicates manual analytical procedures by performing various steps such as pipetting. mixing, heating and measuring color intensity. The PENTRA C200 is intended for quantitative measurements of a variety of analytes: Glucose, Sodium, Potassium and Chloride.

ABX Pentra Glucose HK CP reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of glucose in serum and plasma using glucose hexokinase method by colorimetry.

Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and odiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

The I.S.E. (Ion Selective Electrode) Module is intended for the quantitative determination of Sodium, Potassium and Chloride in serum, plasma and urine by potentiometry using ion selective electrode with associated reference solution, calibrators and controls. Measurement of Sodium, Potassium and Chloride are used in diagnosis and treatment of diseases involving electrolyte imbalance.

Prescription Use x r rescription Ose ------------------------------------------------------------------------------------------------------------------------------------------------------------ Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

AND/OR

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Dus

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K103788

Page 1 of 1 --

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.