K Number
K103788
Device Name
PENTRA C200, ISE MODULE AND ABX PENTRA GLUCOSE HK CP REAGENT
Manufacturer
Date Cleared
2011-11-08

(316 days)

Product Code
Regulation Number
862.1345
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The PENTRA C200 is a discrete photometric benchtop chemistry analyzer for use in clinical laboratories. It is not intended for use in Point of Care settings. It duplicates manual analytical procedures by performing various steps such as pipetting, mixing, heating and measuring color intensity. The PENTRA C200 is intended for quantitative measurements of a variety of analytes: Glucose, Sodium, Potassium and Chloride. ABX Pentra Glucose HK CP reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of glucose in serum and plasma using glucose hexokinase method by colorimetry. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. The I.S.E. (Ion Selective Electrode) module is intended for the quantitative determination of Sodium, Potassium and Chloride in serum, plasma and urine by potentiometry using ion selective electrode with associated reference solution, calibrators and controls. Measurement of Sodium, Potassium and Chloride are used in diagnosis and treatment of diseases involving electrolyte imbalance.
Device Description
The PENTRA C200 is a benchtop clinical chemistry analyzer using two measuring principles: absorbance and ion selective electrodes. The instrument may be summarized as follows: Multi-parametric (up to 15 simultaneous tests + 3 ISE tests). On routine or Stat. 90 (without ISE) to 360 (with ISE) tests / hour (in single or bi-reaction mode) -(analytical cycle of 40 seconds). random access working on primary tubes or sample cups. ABX PENTRA reagent cassettes are compact and ready-to-use. on board bar-code reader is used to identify newly loaded reagent cassettes and samples for patient identification.
More Information

Not Found

No
The summary describes a standard benchtop clinical chemistry analyzer performing automated photometric and potentiometric measurements. There is no mention of AI, ML, image processing, or any other technology typically associated with AI/ML applications in medical devices. The performance studies focus on standard analytical metrics like accuracy, precision, linearity, and correlation.

No
The device is a diagnostic chemistry analyzer used for quantitative measurements of analytes in clinical laboratories, not for treating patients.

Yes
The device is described as a "discrete photometric benchtop chemistry analyzer for use in clinical laboratories" and is intended for "quantitative measurements of a variety of analytes" such as Glucose, Sodium, Potassium, and Chloride. It also states that these measurements are "used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia" and "in diagnosis and treatment of diseases involving electrolyte imbalance," which are all diagnostic purposes.

No

The device description clearly states it is a "benchtop clinical chemistry analyzer" that performs physical actions like pipetting, mixing, heating, and measuring color intensity, indicating it is a hardware device with integrated software.

Yes, this device is an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The "Intended Use / Indications for Use" section explicitly states that the PENTRA C200 is for "quantitative in vitro diagnostic determination of glucose in serum and plasma" and "quantitative determination of Sodium, Potassium and Chloride in serum, plasma and urine". The term "in vitro diagnostic" is used directly.
  • Purpose: The device is used to measure analytes (Glucose, Sodium, Potassium, Chloride) in biological samples (serum, plasma, urine) to aid in the diagnosis and treatment of various conditions (carbohydrate metabolism disorders, electrolyte imbalance). This is the core function of an IVD.
  • Reagents and Calibrators: The description mentions the use of "ABX Pentra Glucose HK CP reagent with associated calibrators and controls" and "associated reference solution, calibrators and controls" for the ISE module. These are essential components of an IVD system.
  • Clinical Laboratories: The intended user/care setting is "clinical laboratories," which is a typical environment for IVD testing.

The information provided clearly indicates that the PENTRA C200 is designed and intended for use in performing diagnostic tests on samples taken from the human body, which is the definition of an In Vitro Diagnostic device.

N/A

Intended Use / Indications for Use

The PENTRA C200 is a discrete photometric benchtop chemistry analyzer for use in clinical laboratories. It is not intended for use in Point of Care settings.

It duplicates manual analytical procedures by performing various steps such as pipetting, mixing, heating and measuring color intensity. The PENTRA C200 is intended for quantitative measurements of a variety of analytes: Glucose, Sodium, Potassium and Chloride.

ABX Pentra Glucose HK CP reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of glucose in serum and plasma using glucose hexokinase method by colorimetry.

Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

The I.S.E. (Ion Selective Electrode) module is intended for the quantitative determination of Sodium, Potassium and Chloride in serum, plasma and urine by potentiometry using ion selective electrode with associated reference solution, calibrators and controls.

Measurement of Sodium, Potassium and Chloride are used in diagnosis and treatment of diseases involving electrolyte imbalance.

Product codes (comma separated list FDA assigned to the subject device)

CFR, JGS, CEM, CGZ, JJE

Device Description

The PENTRA C200 is a benchtop clinical chemistry analyzer using two measuring principles: absorbance and ion selective electrodes.

The instrument may be summarized as follows :

  • Multi-parametric (up to 15 simultaneous tests + 3 ISE tests) .
  • On routine or Stat
  • 90 (without ISE) to 360 (with ISE) tests / hour (in single or bi-reaction mode) -(analytical cycle of 40 seconds)
  • random access working on primary tubes or sample cups
  • ABX PENTRA reagent cassettes are compact and ready-to-use. .
  • on board bar-code reader is used to identify newly loaded reagent cassettes and samples for patient identification

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Clinical laboratories.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

ABX PENTRA Glucose HK CP:
Sample type: Serum & plasma
Limit of Quantitation: 5 mg/dL
Accuracy and Precision: CV Total ≤ 1.99%
Measuring range: 5 mg/dL - 900 mg/dL; Automatic post-dilution : 2700 mg/dL
Correlation (n=103): Y = 0.98 x + 4.46 (mg/dL) with r2 = 0.998.
Calibration stability: 20 days
Reagent stability: On-board stability (refrigerated area): 39 days
Calibrator: ABX Pentra Multical
Controls: ABX Pentra N Control, ABX Pentra P Control

ABX PENTRA Sodium - E:
Sample type: Serum & plasma; Urine
Accuracy and Precision: Serum & plasma: CV Total ≤ 1.1 %; Urine: CV Total ≤ 4.91 %
Linearity & Measuring range: Serum & plasma: 90 - 190 mmol/L; Urine: 60 – 280 mmol/L
Correlation:
Serum (n=129): Y = 0.96 x + 6.42 with r2 = 0.982.
Plasma (n=132): Y = 1.05 x - 5.32 with r2 = 0.998.
Urine (n=101): Y = 1.01 x - 2.20 with r2 = 0.989.
Calibrators: ABX Pentra Standard 1, ABX Pentra Standard 2, ABX Pentra Reference
Controls: ABX Pentra N Control, ABX Pentra P Control

ABX PENTRA Potassium - E:
Sample type: Serum & plasma; Urine
Accuracy and Precision: Serum & plasma: CV Total ≤ 1.07 %; Urine: CV Total ≤ 2.87 %
Linearity & Measuring range: Serum & plasma: 2 - 9.5 mmol/L; Urine: 25 - 250 mmol/L
Correlation:
Serum (n=122): Y = 1.01 x - 0.06 with r2 = 0.998.
Plasma (n=125): Y = 1.01 x - 0.09 with r2 = 0.998.
Urine (n=159): Y = 1.02 x - 0.27 with r2 = 0.997.
Calibrators: ABX Pentra Standard 1, ABX Pentra Standard 2, ABX Pentra Reference
Controls: ABX Pentra N Control, ABX Pentra P Control

ABX PENTRA Chloride – E:
Sample type: Serum & plasma; Urine
Accuracy and Precision: Serum & plasma: CV Total ≤ 1.55 %; Urine: CV Total ≤ 4.59 %
Linearity & Measuring range: Serum & plasma: 70 – 170 mmol/L; Urine: 70 – 280 mmol/L
Correlation:
Serum (n=170): Y = 0.96 x + 3.74 with r2 = 0.996.
Plasma (n=131): Y = 1.04 x - 4.17 with r2 = 0.997.
Urine (n=112): Y = 1.04 x - 5.63 with r2 = 0.987.
Calibrators: ABX Pentra Standard 1, ABX Pentra Standard 2, ABX Pentra Reference
Controls: ABX Pentra N Control, ABX Pentra P Control

Conclusions for non clinical and clinical tests:
The non clinical studies tests conclude that the safety and effectiveness of the devices are not compromised. Clinical testing met all acceptance criteria, and data demonstrates that the devices are substantially equivalent to their predicate devices.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K052007

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

0

NOV = 8 2011

Premarket Notification [510(k)] Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is : K103788

Company: Horiba ABX SAS Parc Euromédecine Rue du Caducée - BP 7290 34184 Montpellier cedex 4 FRANCE Telephone: + (33) 4 67 14 18 43 Fax: + (33) 4 67 14 15 17

Contact Person: Caroline Ferrer (caroline.ferrer@horiba.com)

Date Prepared: 07th September 2011

Device Name:

ﻴﺔ ﺍﻟﻤﺮﺍﺟﻊ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ

Trade/Proprietary Name:PENTRA C200
Common or Usual Name:Clinical Chemistry analyzer
Device ClassClass I : General Controls : Exempt from premarket.
Classification Name:§862.2160 : Discrete photometric chemistry analyzer for clinical use
Product Code:JJE
Trade/Proprietary Name:I.S.E. Module
Common or Usual Name:Ion Selective Electrode
Device ClassClass I : General Controls : Exempt from premarket.
Classification Name:§862.2160 : Discrete photometric chemistry analyzer for clinical use
Product Code:JJE
Trade/Proprietary Name:ABX PENTRA Glucose HK CP
Common or Usual Name:Glucose HK
Device ClassClass II
Classification Name:§862.1345 : Glucose Test System
Product Code:CFR ; Hexokinase, Glucose

The PENTRA C200 is a new device developed in a joint-venture by HORIBA Medical, and commercialized under HORIBA ABX SAS manufacturer responsibility. HORIBA ABX SAS is the registered company name. Our company is part of the Medical segment of HORIBA group. Trade name of our activity is HORIBA Medical.

1

Predicate Devices:

The data and information supplied in this submission demonstrates substantial equivalence to their respective predicate devices:

Candidate devicePredicate device
510(k) numberDevice name
PENTRA C200K052007ABX PENTRA 400
PENTRA C200 I.S.E. ModuleK052007ABX PENTRA 400 I.S.E. Module
ABX PENTRA Glucose HK CPK052007ABX Pentra Glucose HK CP

Similarities and Differences between the predicate devices and candidate devices:

Predicate deviceCandidate device
Device NameABX PENTRA 400PENTRA C200
Instrument TypeBenchtopSame
Separate workstationNoNo
Touch Screen InterfaceYesYes
Intended UseDiscrete photometric
benchtop chemistry
analyzer for clinical useSame
Maximum throughput420 Tests/Hour360 Tests/Hour
MethodologiesSpectrophotometry
Mono and Bi-chromatic
measurement of light
absorbance
Potentiometry (for I.S.E.
Module)Same
STAT CapabilityYesYes
Sample ID InputBarcoded ID
ManualSame
Reagent Barcode
ReaderIntegratedSame
Reagent PositionsUp to 52 positions
  • 3 I.S.E. | Up to 15 positions
  • 3 I.S.E. |
    | Sample Positions | Up to 60 samples | Up to 15 positions |

Table1: Comparison between PENTRA C200 and ABX PENTRA 400 (K052007)

2

Table 2: Comparison between PENTRA C200 I.S.E. Module and ABX PENTRA
400 I.S.E. Module (K052007)
Predicate deviceCandidate device
Device NameI.S.E. ModuleI.S.E. Module
InstrumentABX PENTRA 400PENTRA C200
OptionalYesYes
ParametersNa, K, ClSame
PotentiometryDirect and IndirectSame
Material
Sodium ElectrodeGlass membrane
selective to Na+ ionsSame
Potassium ElectrodePlastic membrane
selective to K+ ionsSame
Chloride ElectrodePlastic membrane
selective to Cl- ionsSame
Specimen TypesSerum
Plasma
UrineSame

Table 3: Comparison between Glucose HK assay on PENTRA C200 and on ABX PENTRA 400 (K052007)

Predicate deviceCandidate device
Device NameABX PENTRA Glucose
HK CPABX PENTRA Glucose
HK CP
InstrumentABX PENTRA 400PENTRA C200
ParameterGlucoseGlucose
MethodEnzymatic method using
hexokinase couple with
glucose-6-phosphate
dehydrogenaseSame
MaterialBi-reagent cassette,
ready-to-use
Reagent 1: NAD, ATP,
Buffer Sodium azide
Reagent 2: Hexokinase,
G-6-PDH, Magnesium
sulphate, Sodium azideSame
Specimen TypesSerum
Plasma
UrineSerum
Plasma

·

3

Substantial Equivalence:

It has been demonstrated that the PENTRA C200 can be considered substantially equivalent to the predicate device ABX PENTRA 400 (K052007).

The I.S.E. module has been similarly demonstrated as being substantially equivalent to the predicate device K052007.

The ABX PENTRA Glucose HK CP used on the PENTRA C200 has been demonstrated substantially equivalent to the same reagent used on the ABX PENTRA 400 as described in K052007.

Description:

The PENTRA C200 is a benchtop clinical chemistry analyzer using two measuring principles: absorbance and ion selective electrodes.

The instrument may be summarized as follows :

  • Multi-parametric (up to 15 simultaneous tests + 3 ISE tests) .
  • On routine or Stat
  • 90 (without ISE) to 360 (with ISE) tests / hour (in single or bi-reaction mode) -(analytical cycle of 40 seconds)
  • random access working on primary tubes or sample cups
  • ABX PENTRA reagent cassettes are compact and ready-to-use. .
  • on board bar-code reader is used to identify newly loaded reagent cassettes and samples for patient identification

4

Intended Use / Indications for use :

The PENTRA C200 is a discrete photometric benchtop chemistry analyzer for use in clinical laboratories. It is not intended for use in Point of Care settings.

It duplicates manual analytical procedures by performing various steps such as pipetting, mixing, heating and measuring color intensity. The PENTRA C200 is intended for quantitative measurements of a variety of analytes: Glucose, Sodium, Potassium and Chloride.

ABX Pentra Glucose HK CP reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of glucose in serum and plasma using glucose hexokinase method by colorimetry.

Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. *

The I.S.E. (Ion Selective Electrode) module is intended for the quantitative determination of Sodium, Potassium and Chloride in serum, plasma and urine by potentiometry using ion selective electrode with associated reference solution, calibrators and controls.

Measurement of Sodium, Potassium and Chloride are used in diagnosis and treatment of diseases involving electrolyte imbalance.

5

Discussion of Performance Data:

ABX PENTRA Glucose HK CP:
Sample typeSerum & plasma
Limit of Quantitation5 mg/dL
Accuracy and PrecisionCV Total ≤ 1.99%
Measuring range5 mg/dL - 900 mg/dL
Automatic post-dilution : 2700 mg/dL
Correlation (n=103)Y = 0.98 x + 4.46 (mg/dL) with r2 = 0.998.
Calibration stability20 days
Reagent stabilityOn-board stability (refrigerated area): 39 days
CalibratorABX Pentra Multical
ControlsABX Pentra N Control
ABX Pentra P Control
ABX PENTRA Sodium - E :
Sample typeSerum & plasmaUrine
Accuracy and PrecisionCV Total ≤ 1.1 %CV Total ≤ 4.91 %
Linearity & Measuring range90 - 190 mmol/L60 – 280 mmol/L
CorrelationSerum (n=129)
Y = $0.96 x + 6.42$ with $r^2$ =
$0.982$ .Urine (n=101)
Y = $1.01 x - 2.20$ with $r^2$ =
$0.989$ .
Plasma (n=132)
Y = $1.05 x - 5.32$ with $r^2$ =
$0.998$ .
CalibratorsABX Pentra Standard 1
ABX Pentra Standard 2
ABX Pentra ReferenceABX Pentra Standard 1
ABX Pentra Standard 2
ABX Pentra Reference
ControlsABX Pentra N Control
ABX Pentra P ControlABX Pentra N Control
ABX Pentra P Control

6

ABX PENTRA Potassium - E :
Sample typeSerum & plasmaUrine
Accuracy and PrecisionCV Total ≤ 1.07 %CV Total ≤ 2.87 %
Linearity & Measuring range2 - 9.5 mmol/L25 - 250 mmol/L
CorrelationSerum (n=122)
Y = $1.01 x - 0.06$ with $r^2$ =
0.998.Urine (n=159)
Y = $1.02 x - 0.27$ with $r^2$ =
0.997.
Plasma (n=125)
Y = $1.01 x - 0.09$ with $r^2$ =
0.998.
CalibratorsABX Pentra Standard 1
ABX Pentra Standard 2
ABX Pentra ReferenceABX Pentra Standard 1
ABX Pentra Standard 2
ABX Pentra Reference
ControlsABX Pentra N Control
ABX Pentra P ControlABX Pentra N Control
ABX Pentra P Control
ABX PENTRA Chloride – E :
Sample typeSerum & plasmaUrine
Accuracy and PrecisionCV Total ≤ 1.55 %CV Total ≤ 4.59 %
Linearity & Measuring range70 – 170 mmol/L70 – 280 mmol/L
CorrelationSerum (n=170)
$Y = 0.96 x + 3.74 with r2 = 0.996.$
Plasma (n=131)
$Y = 1.04 x - 4.17 with r2 = 0.997.$Urine (n=112)
$Y = 1.04 x - 5.63 with r2 = 0.987.$
CalibratorsABX Pentra Standard 1
ABX Pentra Standard 2
ABX Pentra ReferenceABX Pentra Standard 1
ABX Pentra Standard 2
ABX Pentra Reference
ControlsABX Pentra N Control
ABX Pentra P ControlABX Pentra N Control
ABX Pentra P Control

Conclusions for non clinical and clinical tests :

The non clinical studies tests conclude that the safety and effectiveness of the devices are not compromised.

7

The PENTRA C200 (with ISE module) meets :

IEC 61010-1 / IEC 61010-2-081 / IEC 61010-2-101 :

Safety requirements for electrical equipment for measurement, control, and laboratory use

Part 1: General requirements

Part 2-081: Particular requirements for automatic and semi-automatic laboratory equipment for analysis and other purposes

Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment

EN 61326-2-6 : i

Standard for Electrical equipment for measurement, control and laboratory use -EMC requirements

Part 2-6: Particular requirements - In vitro diagnostic (IVD) medical device

UL631010 - 1 / CSA - C22.2 No. 61010-1 :

Safety requirements for electrical equipement for measurement, control, and laboratory use,

Part 1 : General Requirements

Clinical testing met all acceptance criteria, and data demonstrates that the devices are substantially equivalent to their predicate devices.

8

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/8/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Public Health Service

Horiba ABX SAS c/o Ms. Caroline Ferrer Regulatory Affairs Specialist Parc Euromedecine Rue du Caducee - BP 7290 34184 Montpellier cedex 4 France

Food & Drug Administration 10903 New Hampshire Avenue Buildina 66 Silver Spring, MD 20993

NOV - 8 2011

Re: K103788 Trade Name: PENTRA C200, I.S.E. Module, and ABX PENTRA Glucose HK CP Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Codes: CFR, JGS, CEM, CGZ, JJE Dated: November 4 , 2011 Received: November 7 . 2011

Dear Ms. Ferrer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 80 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

9

Page 2 -

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Viro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric' s (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 for (1 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Signature

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

10

Indications for Use Form

510(k) Number (if known): K103788

Device Name: PENTRA C200, I.S.E. Module, and ABX Pentra Glucose HK CP

Indications for Use:

The PENTRA C200 is a discrete photometric benchtop chemistry analyzer for use in clinical laboratories. It is not intended for use in Point of Care settings.

It duplicates manual analytical procedures by performing various steps such as pipetting. mixing, heating and measuring color intensity. The PENTRA C200 is intended for quantitative measurements of a variety of analytes: Glucose, Sodium, Potassium and Chloride.

ABX Pentra Glucose HK CP reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of glucose in serum and plasma using glucose hexokinase method by colorimetry.

Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and odiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

The I.S.E. (Ion Selective Electrode) Module is intended for the quantitative determination of Sodium, Potassium and Chloride in serum, plasma and urine by potentiometry using ion selective electrode with associated reference solution, calibrators and controls. Measurement of Sodium, Potassium and Chloride are used in diagnosis and treatment of diseases involving electrolyte imbalance.

Prescription Use x r rescription Ose ------------------------------------------------------------------------------------------------------------------------------------------------------------ Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

AND/OR

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Dus

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K103788

Page 1 of 1 --