(210 days)
The Disposable Pulse Oximeter Sensor is indicated for non-invasive spot checking and/or continuous monitoring of arterial oxygen saturation and pulse rate of the adult or neonatal patients with oximeter equipment in hospitals, hospital-type facilities, and home care environments.
The Reusable Pulse Oximeter Sensors are indicated for non-invasive spot checking and/or continuous monitoring of arterial oxygen saturation and pulse rate of the adult or pediatric patients with oximeter equipment in hospitals, hospital-type facilities, and home care environments.
The proposed devices of the Pulse Oximeter Sensors (Disposable and Reusable), are used in conjunction with oximeter equipment to measure and display the arterial oxygen saturation of blood.
The provided document is limited in detail regarding the acceptance criteria and study that proves the device meets them. It primarily focuses on substantiating equivalence to predicate devices based on safety and effectiveness, rather than providing specific performance metrics against defined acceptance criteria from a standalone study.
Here's an analysis based on the available information:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly define quantitative acceptance criteria for device performance (e.g., accuracy ranges for SpO2 or pulse rate) nor does it present a table of reported device performance against such criteria. The "Test Conclusion" section mentions "Bench tests were conducted to verify that the proposed device met all design specifications," but these specifications are not detailed or quantified in the provided text.
The closest equivalent to "acceptance criteria" discussed are the standards the device complies with, which are safety and electrical performance standards, not clinical accuracy standards.
Acceptance Criteria (Implied/Standard) | Reported Device Performance |
---|---|
Compliance with IEC 60601-1 (Medical Electrical Equipment - General Requirements for Safety) | Stated that "Bench tests were conducted to verify that the proposed device met all design specifications" and listed compliance with IEC 60601-1. |
Compliance with IEC 60601-1-2 (Collateral Standard: Electrical equipment - Electromagnetic compatibility) | Stated compliance with IEC 60601-1-2. |
Compliance with ISO 9919:2005 (Medical electrical equipment - Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use) | Stated compliance with ISO 9919:2005. This standard would typically include accuracy requirements for SpO2, but the specific performance results against these requirements are not detailed in this document. |
Substantial Equivalence to Predicate Devices | Concluded "Substantially Equivalent (SE) to the predicate devices... in respect of safety and effectiveness." This implies that the device is expected to perform comparably to the predicate without explicitly stating numerical performance. |
2. Sample Size Used for the Test Set and Data Provenance
The document mentions "Bench tests were conducted" but does not provide any information regarding the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective). Without details on a clinical accuracy study, this information is absent.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided in the document as there is no mention of a clinical accuracy study that would involve expert assessment for establishing ground truth. The submission focuses on technical equivalence and compliance with general safety/performance standards.
4. Adjudication Method for the Test Set
This information is not provided as there is no mention of a human-involved study or test set requiring adjudication.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size
No, an MRMC comparative effectiveness study was not mentioned or conducted according to this document. The submission is for a pulse oximeter sensor, which is a diagnostic device that outputs numerical values (SpO2 and pulse rate), not an imaging or diagnostic aid that would typically involve human readers interpreting cases with and without AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
The document describes "Bench tests" and compliance with standards, which can be considered standalone testing of the device's performance. However, specific details of these tests (e.g., specific accuracy values for SpO2 and pulse rate, details of controlled hypoxia studies) are not provided in this summary. For pulse oximeters, standalone performance typically involves testing against a reference oximeter during induced hypoxia studies, but the specifics of this particular device's results from such a study are not in this summary. The mention of ISO 9919:2005 implies such testing was likely performed.
7. The Type of Ground Truth Used
The document does not explicitly state the type of ground truth used for performance evaluation. For pulse oximeters, the ground truth for SpO2 measurements in clinical accuracy studies is typically established using co-oximetry of arterial blood samples. However, as a detailed clinical study is not summarized here, the method for establishing ground truth is not specified. For the bench tests, the ground truth would typically be established by calibrated reference equipment.
8. The Sample Size for the Training Set
This information is not applicable as the device is a pulse oximeter sensor, not an AI/ML-based algorithm that requires a "training set" in the conventional sense. The device's underlying technology and calibration would be based on physiological models and empirical data, but not referred to as an "AI training set."
9. How the Ground Truth for the Training Set Was Established
This information is not applicable for the reasons stated above (not an AI/ML device requiring a training set).
In summary: The provided document is a 510(k) summary that focuses on establishing substantial equivalence through regulatory compliance and bench testing without detailing specific clinical performance acceptance criteria or the results of a dedicated clinical accuracy study. It primarily asserts that the device meets safety and basic performance criteria common to pulse oximeters and is comparable to existing predicate devices.
§ 870.2700 Oximeter.
(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).