K Number
K103573
Date Cleared
2011-12-28

(387 days)

Product Code
Regulation Number
888.3070
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ART2 Spinal Fixation System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

As a pedicle screw system ART2 Spinal Fixation System is indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

Device Description

The ART2 Spinal Fixation System has been developed with the objective of providing the surgeon with an adaptable pedicle screw system in order to carry out posterior pedicle screw fixation of the spine simply, quickly and effectively. It is a modification of the ART Pedicle Screw system (K033150).

The system consists of a variety of color coded top loading pedicle screws. The pedicle screws are available in various lengths and diameters. The screw is connected to the rod connector. Two sizes of connectors are available, short and long. The long is used in cases of spondylolisthesis where the short connector would not be able to engage the rod. The rods are available in multiple lengths.

The components are manufactured from titanium alloy (Ti 6Al 4V ELI) per ASTM F136.

AI/ML Overview

The provided document describes the ART2 Spinal Fixation System, a pedicle screw system. The acceptance criteria and the study proving it meets these criteria are outlined in the "Summary of Nonclinical Tests" section.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria StudyAcceptance Criteria (from ASTM Standards)Reported Device Performance
Static and dynamic compressionMet criteria laid out in ASTM F1717"The acceptance criteria was/were met." (Implies the ART2 system performed comparably or better than the predicate devices and passed the ASTM F1717 requirements)
Static torsionMet criteria laid out in ASTM F1717"The acceptance criteria was/were met." (Implies the ART2 system performed comparably or better than the predicate devices and passed the ASTM F1717 requirements)
Static Axial SlippageMet criteria laid out in ASTM F1798"The acceptance criteria was/were met." (Implies the ART2 system performed comparably or better than the predicate devices and passed the ASTM F1798 requirements)
Static Torsional SlippageMet criteria laid out in ASTM F1798"The acceptance criteria was/were met." (Implies the ART2 system performed comparably or better than the predicate devices and passed the ASTM F1798 requirements)

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size (number of devices or tests performed) for each specific test or the data provenance (country of origin, retrospective/prospective). It only mentions what tests were conducted and that the acceptance criteria were met. For non-clinical, mechanical testing like this, the "test set" would refer to the number of devices or components subjected to each test type. The data provenance would be from laboratory testing.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

Not applicable. This is a submission for a spinal fixation system, which involves mechanical safety and performance testing, not diagnostic or interpretive tasks where expert ground truth would be established from patient data. The "ground truth" here is compliance with established ASTM (American Society for Testing and Materials) standards for medical devices.

4. Adjudication Method for the Test Set

Not applicable. As this is mechanical testing against an engineering standard, there is no human adjudication process involved. The results are quantitative measurements compared against predefined thresholds in the ASTM standards.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. The document explicitly states: "No clinical studies were performed." Therefore, no MRMC study with human readers was conducted.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

No. This is a medical device (spinal fixation system) for surgical implantation, not a software algorithm that operates without human intervention. The testing described is mechanical, material, and functional testing of the physical device.

7. Type of Ground Truth Used

The ground truth used for performance validation is based on established engineering and material standards, specifically:

  • ASTM F1717 (Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model)
  • ASTM F1798 (Standard Test Method for Static and Dynamic Axial Compression-Shear, Torsion, and Combined Loading Fatigue Tests of Spinal Intervertebral Body Fusion Devices)

These standards define methods and acceptance criteria for mechanical properties of spinal implants.

8. Sample Size for the Training Set

Not applicable. This is a physical medical device, not an AI or machine learning model. Therefore, there is no "training set" in the context of data for model development. The design and manufacturing process would involve extensive engineering and material science, but not in the way of a data-driven training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for an AI model. The "ground truth" for the device's design and functionality would be based on biomechanical principles, material science, clinical requirements for spinal stabilization, and previous generations of similar devices (the predicate devices).

{0}------------------------------------------------

510(k) Summary

for the ART2 Spinal Fixation System

DEC 2 8 2011

In accordance with 21 CFR 807.92 of the Federal Code of Regulations the following 510(k) summary is submitted for the ART2 Spinal Fixation System

Date Prepared: November 30, 2010

  • Contact Person: 1. Owner: J.D. Webb Advanced Medical Technologies The OrthoMedix Group, Inc. Kasteler Strasse 11 1001 Oakwood Blvd 66620 Nonnweiler-Braunhausen Germany Round Rock, TX 78681 Telephone: 49-6873-66 88 111 Telephone: 512-388-0199 Fax: 49-6873-66 88 232 Fax: 512-692-369
    1. Trade name: ART2 Spinal Fixation System Common Name: orthosis, pedicle screw system Classification Name: orthosis, spondylolisthesis spinal fixation orthosis, spinal pedicle fixation 21 CFR Sec. 888.3070 MNH MNI Class II

Predicate or legally marketed devices which are substantially equivalent: ന്

ART Posterior Spinal System - K033150 (AMT) ISOBAR - K990118/K013444 (Scient'x)

4. Description of the device:

The ART2 Spinal Fixation System has been developed with the objective of providing the surgeon with an adaptable pedicle screw system in order to carry out posterior pedicle screw fixation of the spine simply, quickly and effectively. It is a modification of the ART Pedicle Screw system (K033150).

The system consists of a variety of color coded top loading pedicle screws. The pedicle screws are available in various lengths and diameters. The screw is connected to the rod connector. Two sizes of connectors are available, short and long. The long is used in cases of spondylolisthesis where the short connector would not be able to engage the rod. The rods are available in multiple lengths.

Materials:

The components are manufactured from titanium alloy (Ti 6Al 4V ELI) per ASTM F136.

Function:

The ART2 Spinal Fixation System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments until fusion takes place.

Changes in this submission:

  • Modification to pedicle screw design ●
  • . Addition of Ø8.0mm screws
  • � Addition of cannulated screws and reposition screws
  • � Addition of Ø5.5mm straight and curved rods

ഗ് Intended Use:

The ART2 Spinal Fixation System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine:

{1}------------------------------------------------

November 30, 2010 Advanced Medical Technologies 510(k) Summary ART2 Spinal Fixation System K103573

degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

As a pedicle screw system ART2 Spinal Fixation System is indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (13 to sacrum) with removal of the implants after the attainment of a solid fusion.

6. Technological characteristics compared to the predicate device(s):

ART2 Spinal Fixation System is substantially equivalent to the ART Posterior Spinal System and ISOBAR Spinal System in terms of technological characteristics, intended use, design, materials used, and performance. The information provided within this premarket notification supports substantial equivalence to the predicate device(s).

Summary of Nonclincal Tests: 7.

The following tests were conducted:

  • . Static and dynamic compression per ASTM F1717. The acceptance criteria was/were met.
  • Static torsion per ASTM F1717. The acceptance criteria was/were met. .
  • Static Axial Slippage per ASTM F1798. The acceptance criteria was/were met. .
  • . Static Torsional Slippage per ASTM F1798. The acceptance criteria was/were met.

The results of this testing indicate that the ART2 Spinal Fixation System is equivalent to predicate device(s).

8. Clinical Test Summary:

No clinical studies were performed

ರಿ. Conclusions Nonclinical and Clinical:

Conclusions drawn from the nonclinical and clinical tests demonstrate that the ART2 Spinal Fixation System is as safe, as effective, and performs as well as or better than the predicate device(s).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing health, human services, and well-being. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

DEC 2 8 2011

Advanced Medical Technologies % The OrthoMedix Group, Inc. Mr. J.D. Webb 1001 Oakwood Boulevard Round Rock. Texas 78681

Re: K103573

Trade/Device Name: ART2 Spinal Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNH, MNI Dated: December 15, 2011 Received: December 20, 2011

Dear Mr. Webb:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{3}------------------------------------------------

Page 2 - Mr. J.D. Webb

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark A. Millenson

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

KID3273 510(k) Number (if known):

ART2 Spinal Fixation System Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications for Use:

The ART2 Spinal Fixation System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis).

As a pedicle screw system ART2 Spinal Fixation System is indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Lirdell Castle

Division of Surgica! Orthopedic, and Restorative Devices

510(k) Number K103573

N/A