K Number
K103410
Device Name
NEOPAP SYSTEM
Date Cleared
2011-03-30

(128 days)

Product Code
Regulation Number
868.5905
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The NeoPAP System is intended to provide continuous positive airway pressure (CPAP) for use in hospitals to treat newborns and infants, weighing less than 5 kg, with RDS or who are recovering from RDS. (Respiratory Distress Syndrome).
Device Description
The Children's Medical Ventures LLC NeoPAP System provides nasal CPAP (continuous positive airway pressure) as a non-invasive method of breathing support for infants and newborns with compromised respiratory systems who weigh less than 5 kg. The NeoPAP System provides CPAP therapy delivered to patient through a nasal cannula or mask. The system has a feature called Baby-Trak which actively regulates pressure at all times by monitoring pressure at patient's nose and adjusting flow from oxygen and air gas supplies using proportional solenoids. The Baby-Trak feature eliminates the need for a closely-fitted patient interface that can damage fragile tissues. In addition to CPAP mode, the NeoPAP System offers a constant flow mode where the device delivers a fixed flow of humidified air/oxygen mixture through the nasal prongs or mask. There is also a resuscitation mode where the device delivers a fixed flow of unhumidified air/oxygen through the resuscitation port for use with a resuscitation or hyperinflation bag. The percentage of oxygen can be adjusted while in any mode by using a rotary knob on the user interface. The NeoPAP System alarms when pressure levels or oxygen concentration are above or below alarm limits. The system also detects occlusion, disconnect, and other alarm conditions. The system displays a real-time bar graph that indicates delivered pressure.
More Information

No
The description details a system that uses sensors and proportional solenoids to regulate pressure, which is a form of automated control, but does not mention or imply the use of AI or ML algorithms for learning or decision-making beyond pre-programmed logic. The "Baby-Trak" feature describes active regulation based on monitoring, not learning or adaptive algorithms typically associated with AI/ML.

Yes
The device is intended to provide continuous positive airway pressure (CPAP) for treating newborns and infants with respiratory distress syndrome (RDS), which clearly indicates a therapeutic purpose.

No

The device is described as providing continuous positive airway pressure (CPAP) for treatment, not for diagnosing a condition. Its features like pressure monitoring and alarm systems are for maintaining therapy, not for diagnostic assessment.

No

The device description clearly outlines hardware components like gas supplies, solenoids, nasal cannula/mask, and a user interface with a rotary knob. The performance studies also mention bench verification testing and biocompatibility testing, which are typically associated with hardware components.

Based on the provided information, the NeoPAP System is not an IVD (In Vitro Diagnostic) device.

Here's why:

  • Intended Use: The intended use clearly states that the NeoPAP System is for providing continuous positive airway pressure (CPAP) to newborns and infants with respiratory distress. This is a therapeutic intervention, not a diagnostic test performed on samples taken from the body.
  • Device Description: The description details a system that delivers air/oxygen mixtures and regulates pressure to support breathing. This is consistent with a respiratory support device, not a device used for analyzing biological samples.
  • Lack of IVD Characteristics: The document does not mention any of the typical characteristics of an IVD device, such as:
    • Analyzing biological samples (blood, urine, tissue, etc.)
    • Detecting or measuring substances in samples
    • Providing information for diagnosis, monitoring, or screening based on sample analysis

The NeoPAP System is a medical device used for treatment and support of respiratory function, not for diagnosis based on in vitro analysis.

N/A

Intended Use / Indications for Use

The NeoPAP System is intended to provide continuous positive airway pressure (CPAP) for use in hospitals to treat newborns and infants, weighing less than 5 kg, with RDS or who are recovering from RDS. (Respiratory Distress Syndrome).

Product codes (comma separated list FDA assigned to the subject device)

BZD

Device Description

The Children's Medical Ventures LLC NeoPAP System provides nasal CPAP (continuous positive airway pressure) as a non-invasive method of breathing support for infants and newborns with compromised respiratory systems who weigh less than 5 kg.

The NeoPAP System provides CPAP therapy delivered to patient through a nasal cannula or mask. The system has a feature called Baby-Trak which actively regulates pressure at all times by monitoring pressure at patient's nose and adjusting flow from oxygen and air gas supplies using proportional solenoids. The Baby-Trak feature eliminates the need for a closely-fitted patient interface that can damage fragile tissues.

In addition to CPAP mode, the NeoPAP System offers a constant flow mode where the device delivers a fixed flow of humidified air/oxygen mixture through the nasal prongs or mask. There is also a resuscitation mode where the device delivers a fixed flow of unhumidified air/oxygen through the resuscitation port for use with a resuscitation or hyperinflation bag. The percentage of oxygen can be adjusted while in any mode by using a rotary knob on the user interface.

The NeoPAP System alarms when pressure levels or oxygen concentration are above or below alarm limits. The system also detects occlusion, disconnect, and other alarm conditions. The system displays a real-time bar graph that indicates delivered pressure.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Newborns and infants

Intended User / Care Setting

Hospitals

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Design verification tests were performed on the Children's Medical Ventures LLC NeoPAP System as a result of the risk analysis and product requirements.
Formal test protocols were written and executed to verify and validate the modified NeoPAP System. All tests that were created for the modified NeoPAP System had passing results with acceptance criteria successfully met, which demonstrates the safety & effectiveness of the system.
Testing included software code reviews, software unit testing, software integration testing, bench verification testing, biocompatibility testing, environmental testing, user manual/labeling inspection, drawing inspections, and a clinical simulation (usability testing confirmed that the NeoPAP System performs equivalently to the predicate device (Guardian Neonate CPAP / Humidification System, K040862). All tests had acceptable results or discrepancies that do not impact safety or effectiveness.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K040862

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Fisher & Paykel MR850 Humidifier (K073706)

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 868.5905 Noncontinuous ventilator (IPPB).

(a)
Identification. A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.(b)
Classification. Class II (performance standards).

0

TAB 5

K103410
NeoPAP System

MAR 3 0 2011

510(K) SUMMARY

| Official Contact | Scott Wright
Senior Regulatory Affairs Engineer
Children's Medical Ventures LLC
191 Wyngate Drive
Monroeville, PA 15146
Scott.Wright1@philips.com |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 412-380-8812 t
724-681-2407 c
412-380-8850 f |
| Date of Submission | November 22, 2010 |
| Classification Reference | 21 CFR 868.5905 |
| Product Code | BZD - ventilator, non-continuous (respirator) |
| Common/Usual Name | CPAP System |
| Proprietary Name | NeoPAP System |
| Predicate Device(s) | Guardian Neonate CPAP / Humidification System (K040862) |
| Reason for submission | Modified Device |

Substantial Equivalence

The modified device has the following similarities to the previously cleared predicate device:

  • Similar intended use. 0
  • Same operating principle. ם
  • Same technology. 0
  • ם Same manufacturing process.

Design verification tests were performed on the Children's Medical Ventures LLC NeoPAP System as a result of the risk analysis and product requirements. Children's Medical Ventures LLC has

1

determined that the modifications have no impact on the safety and effectiveness of the device. In summary, the device described in this submission is substantially equivalent to the predicate device.

The modified device complies with the applicable standards referenced in the Guidance for FDA Reviewers and Industry "Guidance for the Content of Pre-market Submissions for Software Contained in Medical Devices," May 2006.

Formal test protocols were written and executed to verify and validate the modified NeoPAP System. All tests that were created for the modified NeoPAP System had passing results with acceptance criteria successfully met, which demonstrates the safety & effectiveness of the system.

Testing included software code reviews, software unit testing, software integration testing, bench verification testing, biocompatibility testing, environmental testing, user manual/labeling inspection, drawing inspections, and a clinical simulation (usability testing confirmed that the NeoPAP System performs equivalently to the predicate device (Guardian Neonate CPAP / Humidification System, K040862). All tests had acceptable results or discrepancies that do not impact safety or effectiveness.

Intended Use

The NeoPAP System is intended to provide continuous positive airway pressure (CPAP) for use in hospitals to treat newborns and infants, weighing less than 5 kg, with RDS or who are recovering from RDS. (Respiratory Distress Syndrome). The intended use of the NeoPAP System is unchanged from the predicate device (Guardian Neonate CPAP / Humidification System, commonly referred to as NeoPAP System, K040862).

Device Description

The Children's Medical Ventures LLC NeoPAP System provides nasal CPAP (continuous positive airway pressure) as a non-invasive method of breathing support for infants and newborns with compromised respiratory systems who weigh less than 5 kg.

The NeoPAP System provides CPAP therapy delivered to patient through a nasal cannula or mask. The system has a feature called Baby-Trak which actively regulates pressure at all times by monitoring pressure at patient's nose and adjusting flow from oxygen and air gas supplies using proportional solenoids. The Baby-Trak feature eliminates the need for a closely-fitted patient interface that can damage fragile tissues.

In addition to CPAP mode, the NeoPAP System offers a constant flow mode where the device delivers a fixed flow of humidified air/oxygen mixture through the nasal prongs or mask. There is also

2

a resuscitation mode where the device delivers a fixed flow of unhumidified air/oxygen through the resuscitation port for use with a resuscitation or hyperinflation bag. The percentage of oxygen can be adjusted while in any mode by using a rotary knob on the user interface.

The NeoPAP System alarms when pressure levels or oxygen concentration are above or below alarm limits. The system also detects occlusion, disconnect, and other alarm conditions. The system displays a real-time bar graph that indicates delivered pressure.

Predicate Comparison

The fundamental scientific technology of the NeoPAP System is unchanged from the predicate device (K040862). Children's Medical Ventures LLC has made the following changes to the previously cleared NeoPAP System to be considered for this submission:

  • Flow Mode -- Flow Mode for the previously cleared system only delivered flow into an oxygen . hood, whereas the modified system enables the device to deliver a constant flow of oxygenated gas through a nasal cannula or mask. In the modified system, the process for adjusting flow settings was modified to match the process of adjusting pressure settings in CPAP Mode. Both processes now utilize a rotary knob.
  • Resuscitation Mode The previously cleared system could only deliver 100% oxygen to the . patient while in Resuscitation Mode. Since research has shown that high levels of oxygen for an extended period of time can lead to eye damage, the modified system is capable of delivering a range of oxygen concentration between 21 – 100% while in Resuscitation Mode.
  • Aesthetics The external design of the device has been modified to better align with the . Philips brand image.
  • User Interface The user interface (UI) of the modified device was changed to emphasize . ease of use and to be more intuitive than the original design. Key changes are the addition of a flashing alarm bar, removal of the flow segment display, replacement of the majority of membrane keys with capacitive touch keys, and the addition of pressure, FiQ2, and battery status LEDs changing from white to red when a specific parameter causes an alarm. These changes allow for a less cluttered appearance and more visible feedback for the user to determine what parameter is in an alarm state.
  • Occlusion Alarms Software was modified so that occlusion alarms only cause gas delivery . to stop in the case of high manifold or patient pressure. It is no longer halted in the case of flow readings being lower than expected. The software was also modified so that all occlusion alarms auto-reset. Once the trigger for the alarm is no longer present, the occlusion flag is clear and gas delivery will automatically restart if it had been halted.

3

  • User Confirmation After Turning Rotary Knobs Software was modified so that the user must . confirm that settings should be changed after turning the O2 or pressure/flow setting control knobs. This will prevent inadvertent changes to settings.
  • Humidification The original NeoPAP System utilized a Vapotherm humidification cartridge . which was internal to the device and a delivery circuit that utilized a heated water jacket to minimize rainout within the delivery line. However, due to Vapotherm's recall of the cartridge for potential of bacterial growth, the NeoPAP System was voluntarily recalled after less than a year on the market. For this reason, the modified NeoPAP System does not contain an internal humidifier, but is instead compatible with an external humidifier that is currently being demonstrated to be compatible with the NeoPAP System is the Fisher & Paykel MR850 Humidifier (K073706).
  • Battery Life— The stand-alone battery life of the modified system is 2-hours instead of 1-hour t as in the original NeoPAP System. The longer battery life will be useful in the event that power is lost and a back-up generator is not available or that the device is used during intrahospital transport with no AC outlet available.
  • Standby Mode The original NeoPAP System did not have a Standby Mode. It was added . to the modified NeoPAP System to accommodate system set-up prior to patient receiving therapy and care procedures where the patient is taken off of the device for a period of time.
  • Configuration Menu For both the original and modified NeoPAP System it is necessary to . power off the device in order to enter the configuration menu. In order to allow the clinician to more easily adjust alarm parameters, several items were moved from the configuration menu to the delivery mode menu. The items that were moved are leak delay, high pressure delay, low pressure delay, pressure alarm band, and FiO2 alarm band.
  • Device Packaging -- The original device was packaged using foam inserts. The modified . device utilizes suspension packaging which suspends the device in the airspace of the shipping container between two layers of highly resilient, low-slip film.
  • Patient Interface Circuit The original NeoPAP System patient interface consisted of a . "generator" circuit into which the clinician would plug nasal prongs or nasal masks of an appropriate size for their patient. The tubing came over the top of the patient's head and between the eyes. The modified patient interface (commonly referred to as the PI2 System) consists of a one-piece cannula and one-piece mask. The tubing comes from both sides of the patient's head similar to an adult patient on oxygen. Additionally, a pressure tap was added to the most proximal part of the patient interface (at the patient) to allow the NeoPAP

4

System to better respond to changes in patient pressure to maintain the pressure level set by the clinician.

  • Patient Interface Sizes/Shape Masks are still available in 2 sizes and are unchanged from . the original NeoPAP System in terms of overall footprint/outline. Cannulas are still available in 5 sizes and are unchanged from the original NeoPAP System in terms of prong spacing and prong diameter. The shape of the nasal prongs have been modified from a flare shape to a tapered shape for ease of manufacturing.
  • Patient Interface Tubing and Connections –The pressure and delivery tubing in the original . NeoPAP patient interface was surrounded by a water jacket. It was connected via a slip fit to the patient tubing. The PI2 System utilizes corrugated delivery tubing packaged standard for Fisher & Paykel RT324 humidification circuits with a heated wire to minimize condensation so that a water jacket is not necessary. The pressure line is permanently bonded to the exhalation body to prevent misconnection. The patient tube is permanently bonded to a cork that plugs into the exhalation body.
  • Patient Interface Exhalation The original NeoPAP patient interface had a series of holes in . the tubing located about 7 inches from the patient. Instead of holes in the tubing, the P12 has an exhalation body located about 15 inches from the patient with three large openings on different faces of the body so that the exhalation is not easily blocked regardless of orientation.
  • Patient Interface Pop Off Valve The original NeoPAP patient interface had an umbrella .
  • valve located directly at the patient in the generator which held the prongs or mask. The Pl2 . System utilizes the exact same valve made by the same supplier, but it is located in the exhalation body delivery air path.
  • Patient Interface Bonnet The original NeoPAP patient interface included a cotton bonnet . available in 12 sizes to secure the "generator" circuit. The Pl2 System includes a bonnet available in 5 sizes with a hook and loop material to adjust the fit. It has a unique foam material on the inside lining to minimize shifting of the bonnet and adjustable bonnet clips to hold the PI2 tubing in place.
  • Patient Interface Usage Type: The device manual of the origina! NeoPAP System specified . that the delivery tube, patient tube, and bonnet were single patient use while the nasal prongs and mask were single use. The device manual for the modified NeoPAP System indicates that entire patient interface is single patient use and lists acceptable agents/procedure for cleaning the exterior of the patient interface.

5

and the country of the country of the county of the county of the county of the county of

Traditional 510(k) Tab 5 – 510(k) Summary

Table 1 (follows) compares the NeoPAP System with the predicate device.

Table 1 – Device Comparison
Guardian Neonate CPAP /
Humidification System (K040862)NeoPAP System
Intended UseIntended to provide CPAP for use
in hospitals to treat newborns and
infants with RDS or are recovering
from RDS. (Respiratory Distress
Syndrome) May or may not include
humidification capabilitiesUnchanged from K040862, other than
rewording.
The NeoPAP System is intended to
provide continuous positive airway
pressure (CPAP) for use in hospitals
to treat newborns and infants,
weighing less than 5 kg, with RDS or
who are recovering from RDS
(Respiratory Distress Syndrome).
Intended Environment of UseHospitalUnchanged from K040862.
Patient PopulationInfants with or recovering from RDS
weighing less than 5 kg.Unchanged from K040862.
Product CodeBZDUnchanged from K040862.

Operating Mode Descriptions and Ranges

Gas DeliveredAir / Oxygen MixtureUnchanged from K040862.
CPAP Mode - DescriptionDevice actively regulates pressure
at all times by monitoring pressure
and adjusting flow from air and O2
supplies using proportional
solenoids.Unchanged from K040862.
CPAP Mode - Delivery
LocationNasal prongs / mask.Unchanged from K040862.
CPAP Mode - Range of
Pressure2 to 10 cmH2OUnchanged from K040862.
CPAP Mode - %O2 Range21 – 100%Unchanged from K040862.
Flow Mode - DescriptionDevice delivers fixed flow of
humidified air/oxygen mixture.Unchanged from K040862.

스 """

. ---

:

6

の 2017年 10:00 PM 2017 10:

Table 1 - Device Comparison
Guardian Neonate CPAP /
Humidification System (K040862)NeoPAP System
Flow Mode - Delivery
LocationOxygen hood.Nasal cannula / mask.
Flow Mode – Range of Flow5 to 15 L/min5 to 10 L/min
Flow Mode - %O2 Range21 - 100%Unchanged from K040862.
Resuscitation Mode -
DescriptionResuscitation Mode provides un-
humidified gas through the
resuscitation port for use with a
resuscitation or hyperinflation bag.Unchanged from K040862.
Resuscitation Mode - Fixed
Flow10 L/minUnchanged from K040862.
Resuscitation Mode - %O2
Range100%21 - 100%
Standby ModeNoYes
Humidification MethodVapotherm Microporous MembraneThe Fisher & Paykel MR850 Humidifier
(K073706) has been demonstrated to
be compatible with the NeoPAP
System in this submission.
Range of Temperature of
Gas Delivered33 to 41 °CThe F&P MR850 humidifier does not
have communications with the NeoPAP
device so the temperature is set on the
humidifier instead of the NeoPAP
device.

User Interface / Display of Measured Data

| Circuit Pressure (bar graph

display) Range$0-12 cmH_2O$Unchanged from K040862.
%O₂ (window display) Range21 – 100%Unchanged from K040862.
Gas Temperature Range10 to 50 °CThe F&P MR850 humidifier does not
have communication with the NeoPAP
device so the measured gas

. . . . . . . . .

7

. . . . . .

Tab 5 - 510(k) Summary
Table 1 – Device Comparison
Guardian Neonate CPAP /
Humidification System (K040862)NeoPAP System
temperature will only be displayed on
the humidifier and not on the NeoPAP
device.
Flow (bar graph)Indicator only – indicates that
patient is inhaling / exhaling but
does not give actual flow reading.Flow bar graph removed to prevent
confusion and provide a less cluttered
appearance.
AlarmsHigh / Low Air Supply Pressure
High / Low O₂ Supply Pressure
High / Low FiO2
High / Low Flow (Flow mode only)
High / Low Pressure (CPAP mode
only)
High / Low Temperature
High / Low Water Pressure
No Water Supply
No AC Power
No O₂ Sensor
Battery Not Present
Low Battery
Replace Battery
Pressure Unstable (CPAP mode
only)
Partial Occlusion
Occlusion
Leak
Unexpected Restart
Miscellaneous System Error AlarmsUnchanged from K040862 with the
exception that the NeoPAP device does
not have communication with the F&F
MR850 humidifier so the NeoPAP will
not alarm for high / low gas
temperature, high / low water pressure
or no water supply. However, the
humidifier has its own alarming system
for high / low gas temperature.
Alarm Visual IndicatorBacklit outline of a red bell.Flashing alarm bar. Additionally,
pressure and FiO2 readings that are
typically displayed using white LEDs
change to red so that the user can
more easily determine which of the
Table 1 – Device Comparison
Guardian Neonate CPAP /
Humidification System (K040862)NeoPAP System
System Halt ConditionsMains power disconnected
Air or O2 supply pressure over 93
psi
Internal battery voltage depleted
High pressure detected during
startup
Delivery tube disconnected during
operation
Ambient temperature out of range
Tube occluded during system start
up
Other miscellaneous halt conditionsUnchanged from K040862.
Input TypesRotary knobs and membrane keys.Rotary knobs, membrane keys, and
capacitive keys.
Configuration Menu AccessTurn off device and power on while
holding the Alarm Reset key.Unchanged from K040862.
Configuration Menu ItemsAlarm Sound Gap
Auto-Reset Setting
Patm
Language
Leak Delay
High Pressure Delay
Low Pressure Delay
Pressure Alarm Band
Temperature Alarm Band
FiO₂ Alarm BandAlarm Sound Gap, Auto-Reset Setting,
Patm, and Language remain in
Configuration Menu, and Leak Delay,
High Pressure Delay, Low Pressure
Delay, Pressure Alarm Band, and FiO₂
Alarm Band where moved to the
Delivery Mode Menu so they could be
accessed without powering off the
device. Temperature Alarm Band item
does not exist in current software since
the F&P MR850 humidifier does not
have communication with the NeoPAP
device so it will not alarm for high / low
gas temperature.
Table 1 – Device Comparison
Guardian Neonate CPAP /
Humidification System (K040862)NeoPAP System
PowerAC and BatteryUnchanged from K040862.
Stand-alone Battery Life1 hour2 hours
Oxygen SensorVentrex Oxygen Sensor (K963415)Maxtec Oxygen Sensor (K972992).
Oxygen Sensor LocationMixing block upstream of vapor
transfer cartridgeMoved to pressure manifold block to
allow FiO2 to also be measured in
Resuscitation Mode.
Patient Interface
Usage TypeDelivery tube, patient tube, bonnet,
are single patient use. Nasal
prongs / mask are single use.All components are single patient use.
The entire patient interface can be
cleaned and reused on the same
patient.
Sterility ConditionNon-sterileUnchanged from K040862.
Interface DesignNasal prongs/mask is snapped into
"generator" which is connected to
dual lumen tubing that is routed
above the nose (secured by a clip
integral to the hat) to over the top of
the head.Nasal cannula/mask is one-piece
molding that is connected to dual lume
pressure/delivery tube and single
lumen exhalation tube that are routed
to the sides of the head (secured by
bonnet clips) and then over the head.
Exhalation ManagementSeries of holes in tubing
approximately 7 inches from
patient. Tubing must be oriented
correctly so that holes are not
blocked.Exhalation body approximately 15
inches from patient with 3 large
openings on different facets of the
exhalation body so that exhalation is
not easily blocked regardless of
orientation.
Patient Pressure
Measurement LocationAt connection between
delivery/pressure tubing and patient
tube (approximately 10 inches from
patient).At most proximal point of patient
interface (closest to patient) to provide
most accurate readings.
Pressure Relief Valve
LocationLocated directly at the patient in the
"generator" which holds nasalLocated in the exhalation body delivery
air path approximately 15 inches
Table 1 - Device Comparison
Guardian Neonate CPAP /
Humidification System (K040862)NeoPAP System
prongs or maskupstream from patient.
Relief Valve Opening
Pressure (Measured at
Patient)18 cmH2O15 cmH2O
Nasal Cannula SizesExtra large, large, medium, small,
and extra small.Unchanged from K040862.
Nasal Mask SizesLarge and small.Unchanged from K040862.
Bonnet DesignCotton bonnet without mechanism
to adjust fit of bonnet. Laces
thread through bonnet eyelets to
secure patient interface.Foam bonnet with hook and loop
mechanism to adjust fit of bonnet.
Adjustable bonnet clips to secure
patient interface.
Bonnet SizesSizes 0 to 11.Extra large, large, medium, small, and
extra small.

and the comments of the production of the comments of the comments of the

.

:- -

8

. .

and the country of the country of the county


  • --- · · · · ·

:

9

. . . . . . . .


'

·

10

.

:

. .

:

.

. . . . . .


.

:

:

·

11

Image /page/11/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three tail feathers and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circle around the eagle. The eagle is black, and the text is also black.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002

Children's Medical Ventures, LLC C/O Mr. Scott Wright Senior Regulatory Affairs Engineer 191 Wyngate Drive Monroeville, Pennsylvania 15146

MAR 3 0 2011

Re: K103410

Trade/Device Name: NeoPAP System Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: BZD Dated: March 18, 2011 Received: March 23, 2011

Dear Mr. Wright:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

12

Page 2 - Mr. Wright

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutlFDA/CentersOffices/CDRH/CDRHOffices/ucm] 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Nh for

Anthony Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

13

Indications for Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: _ _ _ NeoPAP System_

Indications for Use:

・・

The NeoPAP System is intended to provide continuous positive airway pressure (CPAP) for use in hospitals to treat newborns and infants, weighing less than 5 kg, with RDS or who are recovering from RDS. (Respiratory Distress Syndrome).

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

L. Schulte

on of Anesthesioner. General Infection Control and De