K Number
K103298
Device Name
ACAM 5 AUDIOMETER SYSTEM
Manufacturer
Date Cleared
2011-04-08

(151 days)

Product Code
Regulation Number
874.1050
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
ACAM® 5 is intended as an aid in diagnosing hearing loss and fitting of hearing aids. The device consists of an audiometer that can be used individually or with the other ACAM modules for performing Real Ear Measurements (REM), Hearing Instrument Test (HIT) and loudness scaling. ACAM 5 should be used by persons trained in audiology on patients of any age who are able to understand the basic principles of the test method when explained by the operator.
Device Description
ACAM® 5 is an audiometer with optional components for performing Real Ear Measurements (REM) Hearing Instrument Testing (HIT) and loudness scaling. It is intended to aid trained audiologists and hearing aid professionals in the diagnosis of hearing loss and the fitting of hearing aids. As a part of the REM and HIT modules ACAM performs several kinds of signal analysis including root mean squared (RMS), fast Fourier transformation (FFT) and percentile analysis (PA). ACAM® 5 gives an audiologist the ability to use one or any combination of the diagnostic tools recommended by the National Acoustics Laboratory (NAL) and Dr. Richard Seewald (DSL-IO). Available ACAM® 5 hardware consists of an audiometer with a modular control unit for peripheral components, which include a large or small hearing instrument test box and an in sity ear probe for real ear measurement. Headphones for audiometry, bone conduction and loudness scaling. loudspeakers, a microphone and an interrupter switch are available for use with the audiometer. ACAM®'s operational software meets the international software standards established by the Hearing Instruments Software Manufacturers Association (HIMSA) and the speech recognition standards (ISTS, International Speech Test Signals) established by the-European Hearing Instrument Manufacturers Association (EHIMA). The ACAM® 5 system meets or exceeds the following international standards: IEC 60645, ISO 8253, ISO 16832, ANSI S3.6, ANSI S3.22, EN 60118-0, EN 60118-7, DIN IEC 60118-15, ISO 12124, ANSI S3.43 ACAM® 5 should be operated by individuals with education and training in the field of audiology such as audiologists, hearing aid dispensers, health care and school nurses and ear nose and throat specialists.
More Information

Not Found

No
The document describes standard audiometry and hearing aid fitting tools with signal analysis techniques like RMS, FFT, and percentile analysis, which are not indicative of AI/ML. There is no mention of AI, ML, or related terms.

No.
The device is intended as an aid in diagnosing hearing loss and fitting hearing aids, not for treating hearing loss.

Yes

The "Intended Use / Indications for Use" section explicitly states that "ACAM® 5 is intended as an aid in diagnosing hearing loss and fitting of hearing aids." The "Device Description" also reiterates its intention to "aid trained audiologists and hearing aid professionals in the diagnosis of hearing loss."

No

The device description explicitly lists available hardware components, including an audiometer, control unit, test boxes, ear probe, headphones, loudspeakers, microphone, and interrupter switch. This indicates it is a system with both hardware and software, not a software-only device.

Based on the provided information, the ACAM® 5 device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
  • ACAM® 5 Function: The ACAM® 5 is an audiometer and related equipment used for in vivo testing of hearing. It directly interacts with the patient's ear to measure their hearing ability and aid in fitting hearing aids. It does not analyze samples taken from the body.
  • Intended Use: The intended use clearly states it's for "diagnosing hearing loss and fitting of hearing aids," which are procedures performed directly on the patient.
  • Device Description: The description details hardware like headphones, bone conduction transducers, loudspeakers, and probes that are used in direct contact with or in close proximity to the patient's ear.
  • Lack of Sample Analysis: There is no mention of analyzing biological samples or using reagents, which are characteristic of IVDs.

Therefore, the ACAM® 5 falls under the category of a medical device used for diagnosis and treatment support, but not an in vitro diagnostic device.

N/A

Intended Use / Indications for Use

ACAM® 5 is intended as an aid in diagnosing hearing loss and fitting of hearing aids. The device consists of an audiometer that can be used individually or with the other ACAM modules for performing Real Ear Measurements (REM), Hearing Instrument Test (HIT) and loudness scaling. ACAM 5 should be used by persons trained in audiology on patients of any age who are able to understand the basic principles of the test method when explained by the operator.

Product codes (comma separated list FDA assigned to the subject device)

EWO, ETW, ETS

Device Description

ACAM® 5 is an audiometer with optional components for performing Real Ear Measurements (REM) Hearing Instrument Testing (HIT) and loudness scaling. It is intended to aid trained audiologists and hearing aid professionals in the diagnosis of hearing loss and the fitting of hearing aids. As a part of the REM and HIT modules ACAM performs several kinds of signal analysis including root mean squared (RMS), fast Fourier transformation (FFT) and percentile analysis (PA). ACAM® 5 gives an audiologist the ability to use one or any combination of the diagnostic tools recommended by the National Acoustics Laboratory (NAL) and Dr. Richard Seewald (DSL-IO).

Available ACAM® 5 hardware consists of an audiometer with a modular control unit for peripheral components, which include a large or small hearing instrument test box and an in sity ear probe for real ear measurement. Headphones for audiometry, bone conduction and loudness scaling. loudspeakers, a microphone and an interrupter switch are available for use with the audiometer.

ACAM®'s operational software meets the international software standards established by the Hearing Instruments Software Manufacturers Association (HIMSA) and the speech recognition standards (ISTS, International Speech Test Signals) established by the-European Hearing Instrument Manufacturers Association (EHIMA).

The ACAM® 5 system meets or exceeds the following international standards: IEC 60645, ISO 8253, ISO 16832, ANSI S3.6, ANSI S3.22, EN 60118-0, EN 60118-7, DIN IEC 60118-15, ISO 12124, ANSI S3.43

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Ear

Indicated Patient Age Range

patients of any age who are able to understand the basic principles of the test method when explained by the operator.

Intended User / Care Setting

ACAM® 5 should be operated by individuals with education and training in the field of audiology such as audiologists, hearing aid dispensers, health care and school nurses and ear nose and throat specialists.

Clinics and on-site applications

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

As is the case with the predicate device, the ACAM system was subjected to electrical safety and electromagnetic compatibility testing and performance specifications and was found to be in compliance with the requirements of the international consensus standards shown below. Conformity with the requirements of these standards demonstrates that the ACAM System is safe and effective and is substantially equivalent to the predicate device.

Safety:
The ACAM System has been designed, constructed and tested to meet the safety and performance requirements of various national safety codes and standards. The ACAM System complies with the following international safety standards:

  • . EN 60601-1 1996-03 Medical electrical equipment: Part 1: General requirements for safety: (FDA Recognition # 5-4)
  • EN 60601-1-1:2001 Medical electrical equipment; Part 1-1: General requirements for safety; . 1. Collateral standard: Safety requirements for medical electrical systems (FDA Recognition # 5-27)
  • EN 60601-1-2:2001 Medical electrical equipment; Part 1: General requirements for safety; . 2. Collateral standard: Electromagnetic compatibility - Requirements and tests (FDA Recognition # 5-30)

Effectiveness:
The effectiveness of the ACAM System has been shown through testing in accordance with the following international standards:

  • ANSI S3.6-2004. Specification for Audiometers. (FDA Recognition # 4-123) .
  • ANSI S3.46-1997(R2007). Methods of Measurement of Real-Ear Performance . Characteristics of Hearing Aids (FDA Recognition # 4-175)
  • ANSI S3.22-2003. Specification of Hearing Aid Characteristics (FDA Recognition # 4-124) .

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K071462

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 874.1050 Audiometer.

(a)
Identification. An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.(b)
Classification. Class II. Except for the otoacoustic emission device, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, if it is in compliance with American National Standard Institute S3.6-1996, “Specification for Audiometers,” and subject to the limitations in § 874.9.

0

APR - 8 2011

l

: 510(k) Summary, ACAM 5

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

| Submitted By: | Acousticon GmbH
Hirschbachstrasse 48
D-64354 Reinheim
Germany
Telephone: +49 6162 93240
Fax: +49 6162 932449
Email: harald.bonsel@acousticon.de |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Maureen N. Garner
New World Regulatory Solutions, Inc.
Telephone: (732) 779 7422
Fax: (732) 270 4829
Email: NWRSinc@gmail.com |
| Date Prepared: | March 3, 2011 |
| Proprietary Name: | ACAM® 5 Audiometer System |
| Common Name: | Audiometer |
| Classification Status: | Class II exempt, 21 CFR 874.1050, Audiometer |
| Subsequent, Classifications: | Class II, 21 CFR 874.3310, Hearing Aid Calibrator and
Analysis System
Class II, 21 CFR 874.1120, Electronic noise generator for
audiometric testing |
| Regulation Name: | Audiometer |
| Product Code: | EWO |
| Subsequent Product Codes: | ETW, ETS |

Predicate Device Information:

.

  • Siemens Unity 2 (K# unknown) and Siemens Unity 2 HIA and Probe, K071462 .
    .

DC
4//////

1

510(k) SUMMARY, continued

Device Description

ACAM® 5 is an audiometer with optional components for performing Real Ear Measurements (REM) Hearing Instrument Testing (HIT) and loudness scaling. It is intended to aid trained audiologists and hearing aid professionals in the diagnosis of hearing loss and the fitting of hearing aids. As a part of the REM and HIT modules ACAM performs several kinds of signal analysis including root mean squared (RMS), fast Fourier transformation (FFT) and percentile analysis (PA). ACAM® 5 gives an audiologist the ability to use one or any combination of the diagnostic tools recommended by the National Acoustics Laboratory (NAL) and Dr. Richard Seewald (DSL-IO).

Available ACAM® 5 hardware consists of an audiometer with a modular control unit for peripheral components, which include a large or small hearing instrument test box and an in sity ear probe for real ear measurement. Headphones for audiometry, bone conduction and loudness scaling. loudspeakers, a microphone and an interrupter switch are available for use with the audiometer.

ACAM®'s operational software meets the international software standards established by the Hearing Instruments Software Manufacturers Association (HIMSA) and the speech recognition standards (ISTS, International Speech Test Signals) established by the-European Hearing Instrument Manufacturers Association (EHIMA).

The ACAM® 5 system meets or exceeds the following international standards: IEC 60645, ISO 8253, ISO 16832, ANSI S3.6, ANSI S3.22, EN 60118-0, EN 60118-7, DIN IEC 60118-15, ISO 12124, ANSI S3.43

ACAM® 5 should be operated by individuals with education and training in the field of audiology such as audiologists, hearing aid dispensers, health care and school nurses and ear nose and throat specialists. ·

Intended Use/Indications for Use

ACAM® 5 is intended as an aid in diagnosing hearing loss and fitting of hearing aids. The device consists of an audiometer that can be used individually or with the other ACAM modules for performing Real Ear Measurements (REM). Hearing Instrument Test (HIT) and loudness scaling. ACAM 5° should be used by persons trained in audiology on patients of any age who are able to understand the basic principles of the test method when explained by the operator.

2

510(k) Summary, continued

Technological Characteristics

The ACAM® 5 System has substantially the same technological characteristics as the predicate, Siemens Unity 2 audiometer (K# unknown), with the HIA and Probe for use with Unity 2, cleared on June 5, 2007 under K071462.

Similarities

Both the Acousticon ACAM® 5 and the Siemens Unity 2 are modular systems intended for the diagnosis of hearing loss and optimization of the acoustical parameters of hearing aid fitting. Both the ACAM® 5 (applicant) and Siemens (predicate) systems provide modular components that can be used individually or in tandem to support processes from hearing aid fitting. Both the predicate and the applicant systems are computer controlled and include software specific for their applications, including links to the NOAH™ software applications provided by the Hearing Instrument Manufacturers' Software Association (HIMSA™). Both the applicant and the predicate systems utilize Real Ear Measurement (REM) fitting protocols that include percentile measurement to analyze the dynamic feature of amplified speech signals. Both the applicant and the predicate systems allow for the use of fitting protocols provided by the National Acoustic Laboratory (NAL-NL1) and by Dr Richard Seewald (DSL-IO). Both the predicate and applicant systems utilize optional visual stimuli that accompany ambient background sounds. Both the predicate and applicant systems offer ambient sound modules to optimize hearing aid performance in real life situations. Both Systems have an audiometer module to measure hearing losses with pure tone and speech following the requirements of ANSI S3.6.

Differences

The predicate system uses single tone frequency scaling over a range of 125 - 8,000 Hz (normal mode) and 125 - 16.000 Hz (high frequency mode), using mean background noise levels and a fixed signal to noise ratio (SNR), while the applicant system uses single tone frequency scaling over a range of 125 - 8.000 Hz and does not have a high frequency mode. The predicate system offers ambient sounds accompanied by still pictograms (SPL O gram) of environmental surroundings and two speakers, while the applicant system employs video sequences (real life fitting, or RLF sequence) with up to 12 speakers. The applicant system provides a broader menu of ambient sound selections than the predicate system. The predicate integrated system includes hearing aid programming with a Siemens hearing aid. The applicant does not include hearing aid programming.

Comparison to Predicate Device

The ACAM 5 Audiometer is equivalent to the predicate device in intended use its technological characteristics, design and function, as demonstrated in the comparison table below:

3

| Device

SpecificationsSiemens Unity 2, with HIA and ProbeACAM 5
K NumberK# unknown, K071462 (HIA and Probe)K103298
Intended useSee indications for useSee indications for use
Indications for useThe Unity 2 is indicated for use in
conducting diagnostic hearing
evaluations and assisting in the diagnosis
of possible otologic disorders. The Unity
2 HIA and Probe are intended to be used
as a system to: (1) perform Real Ear
Measurement; (2) get an objective
indication of the characteristics of a
hearing aid; and (3) assist in the
adjustment of hearing aids while in use
by the patient. It is used by ENT
professionals and in clinics for hearing
aid fitting. This device can either be sold
individually or together with other Unity 2
modules.ACAM 5 is intended as an aid in
diagnosing hearing loss and fitting of
hearing aids. The device consists of an
audiometer that can be used individually
or with the other ACAM modules for
performing Real Ear Measurements
(REM), Hearing Instrument Test (HIT)
and loudness scaling. ACAM 5 should
be used by persons trained in audiology
on patients of any age who are able to
understand the basic principles of the
test method when explained by the
operator.
Target populationIndividuals of any age who have
demonstrated hearing loss, and who are
able to understand and respond to the
basic principles of the test method when
explained by the operator.Individuals of any age who have
demonstrated hearing loss, and who are
able to understand and respond to the
basic principles of the test method when
explained by the operator
Anatomical sitesEarEar
Where usedClinics and on-site applicationsClinics and on-site applications
Energy used
and/or deliveredHzdB HLHzdB HL
Air line
Headphones125100 / 9012590
250110 / 110250110
500115 / 120500120
750120 / 120750120
1000120 / 1201000120
1500115 / 1201500120
2000115 / 1202000120
3000115 / 1203000120
4000115 / 1204000120
6000105 / 1206000120
8000110 / 1058000110
10000/ 100
12500/ 90
16000/ 60
Bone line
Headphones1254012545
2505025060
5007050070
7507075070
100070100070
150070150070
200070200075
300070300080
400060400080
600060600060
Table of Similarities and Differences, continued
Device
SpecificationsSiemens Unity 2 with HIA and ProbeACAM 5
Energy used
and/or deliveredHzdB HLHzdB HL
Air Line Speakers125
250
500
750
1000
1500
2000
3000
4000
6000
800080
90
90
90
90
90
90
90
90
80
65125
250
500
750
1000
1500
2000
3000
4000
6000
800068
85
95
95
95
95
95
100
100
95
75
Human factorsSubjective loudness responseSubjective loudness response
DesignModular components consisting of:
Audiometer and headphones, external
speakers, hearing instrument test box
and microphone, softwareModular components consisting of:
Audiometer and headphones, external
speakers, hearing instrument test box
and microphone, software
Testing Features• Real Ear measurement
• Pure tone and speech audiometry
• Loudness scaling and noise impulse
audiometry
• ISTS. International Speech Test
Signal
• Pediatric audiometry, if requested
• Automated fitting report
• Percentile analysis• Real Ear measurement
• Pure tone and speech audiometry
• Loudness scaling and noise impulse
audiometry
• ISTS. International Speech Test
Signal
• Pediatric audiometry, if requested
• Automated fitting report
• Percentile Analysis
Available TestsAudiometer:
• Air conduction
• Bone conduction
• Free field (10 W integrated 2-channel
amplifier)Audiometer:
• Air conduction
• Bone conduction
• Free field 4 Channel Standard
(optional up to 12 Channel) with 20W
Power
Tests conducted
above hearing
threshold• SISI test
• ABLB test (Fowler)
• DLI test (Luescher)
• Stenger test
• Tone decay test (Carhart)• ABLB test (Fowler)
• Stenger test
• Tone decay test (Carhart)
• Speech Audiometry in quiet and in
Noise
• ANL(acceptable Noise Level)
Tests conducted
above hearing
threshold• SISI test
• ABLB test (Fowler)
• DLI test (Luescher)
• Stenger test
• Tone decay test (Carhart)• ABLB test (Fowler)
• Stenger test
• Tone decay test (Carhart)
• Speech Audiometry in quiet and in
Noise
• ANL (acceptable Noise Level)

Table of Similarities and Differences

4

Table of Similarities and Differences, continued

: .

5

| Device

SpecificationsSiemens Unity 2 with HIA and ProbeACAM 5
Additional Tests
Offered• REUR (Real Ear Unaided Response),
• OEG (Open Ear Gain)
• REOR (Real Ear Occluded Response)
• REAR (Real Ear Aided Response,
SPL level at the eardrum)• REUR (Real Ear Unaided Response),
• OEG (Open Ear Gain)
• REAR (Real Ear Aided Response,
SPL level at the eardrum)
• REIG (Real Ear Insertion Gain)
• RECD (Real Ear-to-Coupler
Difference)
• LI/LO (Input-/Output curve)
• Sound Mapping
• On top mode
• Percentile analysis• REIG (Real Ear Insertion Gain)
• RECD (Real Ear-to-Coupler
Difference)
• LI/LO (Input-/Output curve)
• Percentile analysis
Technical
specificationsCommon Data for all Applications:
• Frequency Accuracy: ± 1 %
• FFT: Resolution 1024 points,
Averaging: 10-100
• Intensity Accuracy: ± 1.5 dB
• Sweep Speed: 1.5 - 12 sec./decade at
25 Hz warble frequency (depending
on frequency resolution)
• Frequency Resolution: 1/3, 1/6, 1/12
and 1/24 octave
• Stimulus Distortion: Less than 1 %
THDCommon Data for all Applications:
• Frequency Accuracy: 140 dB SPL
Intensity Range
Loudspeaker• Max. 6 W into 8 ohms• Max. 20 W into 4 ohms
Output• Max. 10 W into 4 ohms
Hearing Instrument• Single measurement (configurable):• Single measurement (configurable):
Analyzer• OSPL90• OSPL90
• Full On Gain• Full On Gain
• Input/Output• Input/Output
• Attack/Release Time• Attack/Release Time
• Reference Test Gain• Reference Test Gain
• Frequency Response• Frequency Response
• Equivalent Input Noise• Equivalent Input Noise
• Harmonic Distortion• Harmonic Distortion
• Battery Current Drain• Battery Current Drain
• Automatic test sequence acc. to EN
60118 or ANSI 3.22 (customizable)• Automatic test sequence acc. to EN
60118 or ANSI 3.22 (customizable)
• Percentile analysis• Delay Time (Group Delay)
• Percentile analysis
• Dynamic Compression Ratio
• Long Term Test
• Sonogram
• Amplitude statistics
• Loudness
• Roughness
• Spectrum of envelope
• Intermodulation
• Reverberation time
• Induction
Loudspeaker:integrated test box speakerintegrated test box speaker
Hearing Instrument test box
Technical specifications
Telecoil drive
in test box10 - 100 mA/Meter0.01mA/meter - 1 A/meter
Total Harmonic
DistortionRanges 0 - 55%, Resolution 0.1 %Ranges 0 - 100 %, Resolution 0.1 %
Battery-Simulator:Standard types are selectable, Custom
types within 1.1 - 1.6 V,0 - 25 Ohm
rangeVoltage adjustable 0 to 3V (resolution
0.01Volt)
Battery Current
Test:• Ranges: 0 - 50 mA
• Resolution: 0.1 mA
• Accuracy: ± 5 %• Ranges: 0 - >20 mA
• Resolution: 0.01 mA
• Accuracy: ± 5 %
Computer
Communication:Built-in USB1.1 computer interface.
Windows software availableNetwork Protocol TCP/IP over Twisted
Pair (Standard Network Interface)
Supported
Operating systems:Windows 98 SE and higher (except
Windows NT)Windows XP SP3
Device
SpecificationsSiemens Unity 2 with HIA and ProbeACAM 5
ConstructionAnodized all-aluminium caseIndustry 19" Rack
PowerAC 50/60 Hz, 100 - 240V, 5 x 20 mm,AC 50/60 Hz, 100 - 240V, 2 X 1 AT
Fuses2.15 AT
Consumption75 W110W
Dimensions (cm)• Control Unit: 31.9L x 31.7W x 7.5H
• Docking Station: 42.8L x 23.0X x 22.2H• Control Unit: 36L x 36.5W x 14H
• Open Test Box, Small: 20L x 20X x 12H
• Closed Test Box, Large: 40L x 33 x 30H
Standards MetAudiometerAudiometer
EN 60645-1EN 60645-1
EN 60645-2EN 60645-2
EN 60645-4ISO 8253 - 1,2,3
ANSI S3.6ISO 16832
ANSI S3.22ANSI S3.6
ANSI S3.22
Hearing Instrument AnalyzerHearing Instrument Analyzer
EN 60118-0EN 60118-0
EN 60118-7EN 60118-7
ANSI S3.43ISO 12124
ANSI S3.43
MaterialsNANA
BiocompatibilityNANA
Compatibility with the
environment and
other devicesNANA
SterilityNANA
Electrical safetyEN 60601-1, Class 1, Type B
EN 60601-1-2EN 60601-1, Class 1, Type B
EN 60601-1-2
Mechanical safetyNANA
Chemical safetyNANA
Thermal safetyNANA
Radiation safetyNANA

Table of Similarities and Differences, continued

6

Table of Similarities and Differences, continued

.

7

Table of Similarities and Differences, continued

8

510(k) Summary, Performance Test Data and Conclusions

Performance Test Data and Conclusions

As is the case with the predicate device, the ACAM system was subjected to electrical safety and electromagnetic compatibility testing and performance specifications and was found to be in compliance with the requirements of the international consensus standards shown below. Conformity with the requirements of these standards demonstrates that the ACAM System is safe and effective and is substantially equivalent to the predicate device.

Safetv:

The ACAM System has been designed, constructed and tested to meet the safety and performance requirements of various national safety codes and standards. The ACAM System complies with the following international safety standards:

  • . EN 60601-1 1996-03 Medical electrical equipment: Part 1: General requirements for safety: (FDA Recognition # 5-4)
  • EN 60601-1-1:2001 Medical electrical equipment; Part 1-1: General requirements for safety; . 1. Collateral standard: Safety requirements for medical electrical systems (FDA Recognition # 5-27)
  • EN 60601-1-2:2001 Medical electrical equipment; Part 1: General requirements for safety; . 2. Collateral standard: Electromagnetic compatibility - Requirements and tests (FDA Recognition # 5-30)

Effectiveness:

The effectiveness of the ACAM System has been shown through testing in accordance with the following international standards:

  • ANSI S3.6-2004. Specification for Audiometers. (FDA Recognition # 4-123) .
  • ANSI S3.46-1997(R2007). Methods of Measurement of Real-Ear Performance . Characteristics of Hearing Aids (FDA Recognition # 4-175)
  • ANSI S3.22-2003. Specification of Hearing Aid Characteristics (FDA Recognition # 4-124) .

9

Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Acousticon GmbH c/o Ms. Maureen N. Garner President New World Regulatory Solutions, Inc. P.O. Box 5374 Toms River, NJ 08754

APR = 8 2011

Re: K103298

Trade/Device Name: ACAM® 5 Audiometer System Regulation Number: 21 CFR 874.1050 Regulation Name: Audiometer Regulatory Class: Class II Product Code: EWO Dated: March 25, 2011 Received: March 29, 2011

Dear Ms. Garner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

10

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Melvin R. Felder, M.D.

Malvina B. Evdelman. M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

11

103298

INDICATIONS FOR USE

510(k) Number K103298

Device Name: ACAM® 5

Prescription

Indications for Use: ACAM® 5 is intended as an aid in diagnosing hearing loss and fitting of hearing aids. The device consists of an audiometer that can be used individually or with the other ACAM modules for performing Real Ear Measurements (REM), Hearing Instrument Test (HIT) and loudness scaling. ACAM 5 should be used by persons trained in audiology on patients of any age who are able to understand the basic principles of the test method when explained by the operator.

  • AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)

Over-The-Counter

Use

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Use

Signature

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number K103298