K Number
K103175

Validate with FDA (Live)

Date Cleared
2011-02-17

(112 days)

Product Code
Regulation Number
866.3980
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The FilmArray Respiratory Panel (RP) is a multiplexed nucleic acid test intended for use with the FilmArray instrument for the simultaneous qualitative detection and identification of multiple respiratory viral nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals suspected of respiratory tract infections. The following virus types and subtypes are identified using the FilmArray RP: Adenovirus, Coronavirus HKU1, Coronavirus NL63, Human Metapneumovirus, Influenza A, Influenza A subtype H1, Influenza A subtype H3, Influenza A subtype 2009 H1, Influenza B, Parainfluenza virus 3, Rhinovirus/Enterovirus, and Respiratory Syncytial Virus. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of a respiratory infection aids in the diagnosis of respiratory infection if used in conjunction with other clinical and epidemiological information. Negative results do not preclude respiratory infection and should not be used as the sole basis for diagnosis, treatment or other management decisions. Positive results do not rule out bacterial infection or co-infection with other organisms. The agent detected may not be the definite cause of disease. The use of additional laboratory testing (e.g. bacterial and viral culture, immunofluorescence, and radiography) and clinical presentation must be taken into consideration in order to obtain the final diagnosis of respiratory infection.

Due to seasonal prevalence, performance characteristics for Influenza A/H1, Influenza A/H3, Influenza A/2009 H1, and Influenza B were established primarily with retrospective clinical specimens.

Due to the genetic similarity between human Rhinovirus and Enterovirus, the FilmArray RP cannot reliably differentiate them. A positive FilmArray RP Rhinovirus/Enterovirus result should be followed-up using an alternate method (e.g. cell culture or sequence analysis).

The FilmArray RP detects Adenovirus species C serotype 6 with reduced sensitivity. It is recommended that specimens found to be negative for Adenovirus after examination using FilmArray RP be confirmed by an alternate method (e.g. FDA cleared molecular test or cell culture).

Performance characteristics for influenza A were established when influenza A/2009 H1N1, A/H1, and A/H3 were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary. If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Device Description

The FilmArray RP System is multiplex nucleic acid test system composed of the FilmArray instrument, the FilmArray software (preinstalled on a laptop computer) and the FilmArray RP pouch. The FilmArray RP pouch contains freeze-dried reagents to perform nucleic acid purification, reverse transcription, and nested, multiplex PCR with DNA melt analysis. The Respiratory Panel (RP) pouch identifies 12 common and emerging viral respiratory pathogens (see Table 1).

A test is initiated by loading Hydration Solution and an unprocessed patient nasopharyngeal swab (NPS) specimen (i.e. specimen mixed with Sample Buffer) into the FilmArray RP pouch. The pouch contains all of the reagents required for specimen testing and analysis in a freeze-dried format; the addition of Hydration Solution and Sample/Buffer Mix rehydrates the reagents. After the pouch is prepared, the FilmArrav software guides the user though the steps of placing the pouch into the instrument, scanning the pouch barcode, entering the sample identification and initiating the run.

The FilmArray instrument contains a coordinated system of inflatable bladders and seal points, which act on the pouch to control the movement of liquid between the pouch blisters. When a bladder is inflated over a reagent blister, it forces liquid from the blister into connecting channels. Alternatively, when a seal is placed over a connecting channel it acts as a valve to open or close a channel. In addition, electronically controlled pneumatic pistons are positioned over multiple plungers in order to deliver the rehydrated reagents into the blisters at the appropriate times. Two Peltier devices control heating and cooling of the pouch to drive the reverse transcription reactions, the PCR reactions, and the melting curve analysis.

Nucleic acid extraction occurs within the FilmArray pouch using mechanical lysis and standard magnetic bead technology. After extracting and purifying nucleic acids from the unprocessed sample, the FilmArray performs a nested multiplex PCR that is executed in two stages. During the first stage, the FilmArray performs a single, large volume, highly multiplexed reverse transcription PCR (rt-PCR) reaction. The products from first stage PCR are then diluted and combined with a fresh, primer-free master mix and a fluorescent double stranded DNA binding dye (LC Green@Plus, Idaho Technology). This second master mix solution, is then distributed to each well of the array. Array wells contain sets of primers designed specifically to amplify sequences internal to the PCR products generated during the first stage PCR reaction. The second stage PCR, or nested PCR, is performed in singleplex fashion in each well of the array. At the conclusion of the 200 stage PCR, the array is interrogated by melting curve analysis for the detection of signature amplicons denoting the presence of specific viral or bacterial targets A digital camera placed in front of the second stage PCR captures fluorescent images of the PCR reactions in real time.

The FilmArray software automatically interprets the results of each DNA melting curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism on the panel.

AI/ML Overview

Here's an analysis of the provided text, focusing on acceptance criteria and supporting study details:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined "acceptance criteria" in a numerical target for metrics like sensitivity or specificity. However, it presents the reported clinical sensitivity (Positive Percent Agreement - PPA) and specificity (Negative Percent Agreement - NPA) for each detected organism from two studies: a prospective clinical study and an archived specimen study. The implicit acceptance criteria would be for these values to be high, ideally close to 100%, and for the 95% Confidence Intervals (CI) to be acceptably narrow.

Table: Reported Device Performance (Clinical Studies)

Organism/VirusStudy TypeSensitivity (PPA)95% CISpecificity (NPA)95% CI
AdenovirusProspective Clinical88.9% (24/27)70.8 - 97.7%98.3% (812/826)97.2 - 99.1%
AdenovirusArchived Specimen100.0% (27/27)87.2 - 100%100.0% (28/28)87.7 - 100%
Coronavirus HKU1Prospective Clinical95.8% (23/24)78.9 - 99.9%99.8% (827/829)99.1 - 100%
Coronavirus NL63Prospective Clinical95.8% (23/24)78.9 - 99.9%100.0% (829/829)99.6 - 100%
Human MetapneumovirusProspective Clinical94.6% (88/93)87.9 - 98.2%99.2% (754/760)98.3 - 99.7%
Human Rhinovirus/EnterovirusProspective Clinical92.7% (190/205)88.2 - 95.8%94.6% (613/648)92.6 - 96.2%
EnterovirusArchived Specimen95.7% (22/23)78.0 - 99.9%100.0% (90/90)96.0 - 100%
Influenza AProspective Clinical90.0% (9/10)55.5 - 99.8%99.8% (841/843)99.2 - 100%
Influenza A/H1Prospective Clinical0/0n/a100.0% (853/853)99.6 - 100%
Influenza A/H1Archived Specimen100.0% (32/32)89.1 - 100%100.0% (127/127)97.1 - 100%
Influenza A/H3Prospective Clinical0/0n/a100.0% (853/853)99.6 - 100%
Influenza A/H3Archived Specimen100.0% (54/54)93.4 - 100%100.0% (105/105)96.5 - 100%
Influenza A/2009 H1Prospective Clinical88.9% (8/9)51.8 - 99.7%99.6% (841/844)99.0 - 99.9%
Influenza A/H1-2009Archived Specimen100.0% (34/34)89.7 - 100%100.0% (125/125)97.1 - 100%
Influenza BProspective Clinical0/0n/a100.0% (853/853)99.6 - 100%
Influenza BArchived Specimen100.0% (30/30)88.4 - 100%100.0% (129/129)97.2 - 100%
Parainfluenza Virus 3Prospective Clinical95.8% (23/24)78.9 - 99.9%98.8% (819/829)97.8 - 99.4%
PIV3Archived Specimen100.0% (36/36)90.3 - 100%100.0% (93/93)96.1 - 100%
Respiratory Syncytial VirusProspective Clinical100.0% (52/52)93.2 - 100%89.1% (714/801)86.8 - 91.2%

2. Sample Sizes and Data Provenance

  • Test Set (Clinical Performance):
    • Prospective Clinical Study:
      • Sample Size: 853 subjects (from an initial enrollment of 857).
      • Data Provenance: 3 U.S. clinical sites, prospective data collected from December 2009 through May 2010.
    • Archived Specimen Study:
      • Sample Size: 349 archived clinical NPS specimens were confirmed to contain or lack the analyte of interest (from 400 initially selected).
      • Data Provenance: Retrospective, preselected archived clinical NPS specimens. Age demographics provided (e.g., 89 (25.5%) ≤5 years, 188 (53.9%) unknown age from a pediatric hospital). Country of origin is implied to be U.S. from clinical sites and some strains are from particular US states, but not explicitly stated for all archived samples.

3. Number of Experts and Qualifications for Ground Truth (Test Set)

The document primarily relies on comparator methods for establishing ground truth, rather than a direct count of human expert consensus.

  • For Adenovirus, Influenza A, Influenza B, Parainfluenza virus 3, Respiratory Syncytial Virus (Prospective Study):
    • Ground Truth Method: Viral culture followed by DFA (Fluorescent Antibody) identification. "True" positives/negatives were defined based on these comparator methods.
  • For FluA/H1, FluA/H3, FluA/2009 H1 (Prospective Study):
    • Ground Truth Method: Viral culture followed by one analytically validated PCR assay with bi-directional sequence confirmation. The comparator PCR assays were designed to amplify different sequences than the FilmArray assays. "True" positives/negatives were based on cultural positivity for Influenza A and bi-directional sequencing results matching NCBI GenBank entries.
  • For Human Rhinovirus, Coronavirus NL63, Coronavirus HKU1, Human Metapneumovirus (Prospective Study):
    • Ground Truth Method: A "predetermined algorithm that used composite comparator methods" consisting of two analytically validated PCR assays followed by bi-directional sequencing. "True" positives/negatives were based on bi-directional sequencing data matching NCBI GenBank entries or negative results from both comparator PCR assays.
  • For Archived Specimen Study:
    • Ground Truth Method: Analyte-specific PCR and bi-directional sequencing were used to confirm the presence or absence of the analyte of interest in each specimen prior to FilmArray RP testing.

The document does not specify the number of experts or their qualifications for interpreting the DFA, viral culture, PCR, or sequencing results. It states these are "analytically validated" assays, implying expert input in their validation and interpretation, but provides no direct detail on the human experts involved in establishing the ground truth for this specific study.

4. Adjudication Method for the Test Set

The document describes comparator methods for defining "true" positive/negative results. For the clinical performance sections:

  • For categories like Adenovirus through RSV in the prospective study, ground truth was set by viral culture + DFA. Discrepancies between FilmArray RP and the comparator method were further investigated (e.g., footnoted retesting and bi-directional sequence analysis for false positives/negatives). For example, "Adenoviruses were identified in 13/14 false positive specimens using bi-directional sequence analysis." This implies adjudication by molecular methods (PCR and sequencing) was used to resolve discrepancies or to further confirm findings where FilmArray differed from the initial comparator method.
  • For Influenza A subtypes and other viruses (Rhinovirus, etc.), the ground truth was already a composite of PCR and bi-directional sequencing, suggesting internal resolution of complex cases within that ground truth establishment process.

The "archived specimen study" explicitly states that the presence or absence of the analyte was confirmed using analyte-specific PCR and bi-directional sequencing prior to FilmArray RP testing, which serves as a highly robust ground truth.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. The study is a diagnostic performance evaluation comparing the device (FilmArray RP) against established laboratory methods (viral culture, PCR, sequencing) for detecting respiratory pathogens. It does not involve human readers interpreting results with or without AI assistance, or quantify the effect size of human improvement. This is a standalone diagnostic test evaluation.

6. Standalone Performance Study (Algorithm Only)

Yes, the clinical performance studies (prospective and archived specimen) describe the standalone performance of the FilmArray RP system. The device is automated with "automated test interpretation and report generation" where the "User cannot access raw data" (Table 3), indicating the algorithm's interpretation delivers the final result.

7. Type of Ground Truth Used

The ground truth used is a combination of:

  • Expert Consensus/Reference Methods (Viral Culture + DFA, PCR + Bi-directional Sequencing): This reflects standard laboratory diagnostic methods considered to be gold standards or highly accurate comparators at the time.
    • "True" positives/negatives were determined by these reference methods.
    • For discrepancies, especially for "false positives" or "false negatives" where the FilmArray RP result differed from the initial comparator, further molecular analysis (bi-directional sequence analysis) was often performed, indicating a robust process to confirm ground truth when needed.

8. Sample Size for the Training Set

The document explicitly describes studies for clinical performance evaluation and analytical validation (LoD, inclusivity, specificity, precision, interference). It does not provide details on a specific "training set" or its size, as would be typical for machine learning models. This suggests that the FilmArray RP system's interpretive algorithm (which automatically interprets melting curve analysis results) was developed and validated internally using methods not detailed in this summary (e.g., extensive analytical testing, algorithm tuning, and potentially internal datasets before these formal clinical studies). The "analytical studies" sections (LoD, inclusivity, etc.) would contribute to refining and confirming the algorithm's performance.

9. How the Ground Truth for the Training Set Was Established

Since a "training set" is not explicitly defined or detailed in this summary, the method for establishing its ground truth is also not provided. However, based on the analytical studies described:

  • LoD (Limit of Detection): Established by testing limiting dilutions of live, quantified viruses (or clinical specimens in one case) where the presence and concentration are precisely known.
  • Analytical Reactivity (Inclusivity): Evaluated using known strains/isolates, often quantified. Bi-directional sequencing and bioinformatics (in silico analyses) were used to confirm specific viral identities and predict reactivity.
  • Analytical Specificity (Cross-reactivity and Exclusivity): Tested against high concentrations of known interfering organisms or substances.

These analytical studies established ground truth for known viral identities and concentrations under controlled conditions, which would underpin the fundamental ability of the algorithm to accurately detect and identify targets. While not a "training set" in the modern ML sense, these analytical validations serve a similar purpose in confirming the algorithm's foundational accuracy against known truths.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a handwritten string of characters. The characters appear to be 'K103175'. The handwriting is somewhat stylized, with distinct shapes for each number and letter. The image is in black and white, with the characters appearing dark against a lighter background.

11 510(k) Summary

FFB 17 2011

510(k) Summary Idaho Technology Inc. FilmArray RP Test System

Introduction: According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

Submitted by:

Idaho Technology Inc 390 Wakara Way Salt Lake City, UT 84108

Telephone: 801-736-6354 Facsimile: 801-588-0507

Contact: Beth Lingenfelter, ext. 407

Date Submitted: February 2, 2011

Device Name and Classification:

Trade Name: FilmArray RP System

Regulation Number: 21 CFR 866.3980

Classification Name: Respiratory Viral Panel Multiplex Nucleic Acid Assay

Predicate Device:

K063765, K081483, K091677 - Luminex® xTAG™ Respiratory Viral Panel (RVP).

Intended Use:

The FilmArray Respiratory Panel (RP) is a multiplexed nucleic acid test intended for use with the FilmArray instrument for the simultaneous qualitative detection and identification of multiple respiratory viral nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals suspected of respiratory tract infections. The following virus types and subtypes are identified using the FilmArray RP: Adenovirus, Coronavirus HKU1, Coronavirus NL63, Human Metapneumovirus, Influenza A, Influenza A subtype H1, Influenza A subtype H3, Influenza A subtype 2009 H1, Influenza B, Parainfluenza virus 3, Rhinovirus/Enterovirus, and Respiratory Syncytial Virus. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and

Idaho Technology Inc. 510(k) FilmArray Respiratory Panel System

{1}------------------------------------------------

symptoms of a respiratory infection aids in the diagnosis of respiratory infection if used in conjunction with other clinical and epidemiological information. Negative results do not preclude respiratory infection and should not be used as the sole basis for diagnosis. treatment or other management decisions. Positive results do not rule out bacterial infection or co-infection with other organisms. The agent detected may not be the definite cause of disease. The use of additional laboratory testing (e.g. bacterial and viral culture, immunofluorescence, and radiography) and clinical presentation must be taken into considertation in order to obtain the final diagnosis of respiratory infection.

Due to seasonal prevalence, performance characteristics for Influenza A/H1, Influenza A/H3, Influenza A/2009 H1, and Influenza B were established primarily with retrospective clinical specimens.

Due to the genetic similarity between human Rhinovirus and Enterovirus, the FilmArray RP cannot reliably differentiate them. A positive FilmArray RP Rhinovirus/Enterovirus result should be followed-up using an alternate method (e.g. cell culture or sequence analysis).

The FilmArray RP detects Adenovirus species C serotype 6 with reduced sensitivity. It is recommended that specimens found to be negative for Adenovirus after examination using FilmArray RP be confirmed by an alternate method (e.g. FDA-cleared molecular test or cell culture).

Performance characteristics for influenza A were established when influenza A/2009 H1N1, A/H1, and A/H3 were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary. If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Device Description:

The FilmArray RP System is multiplex nucleic acid test system composed of the FilmArray instrument, the FilmArray software (preinstalled on a laptop computer) and the FilmArray RP pouch. The FilmArray RP pouch contains freeze-dried reagents to perform nucleic acid purification, reverse transcription, and nested, multiplex PCR with DNA melt analysis. The Respiratory Panel (RP) pouch identifies 12 common and emerging viral respiratory pathogens (see Table 1).

{2}------------------------------------------------

Table 1. Viruses Detected by the FilmArray Respiratory Panel

Viral Respiratory Pathogens
Influenza A
H1 subtype
H3 subtype
2009 H1 subtype
Influenza B
Adenovirus
Coronavirus HKU1
Coronavirus NL63
Human Metapneumovirus
Parainfluenza Virus 3
Respiratory Syncytial Virus
Rhinovirus and Enterovirus

A test is initiated by loading Hydration Solution and an unprocessed patient nasopharyngeal swab (NPS) specimen (i.e. specimen mixed with Sample Buffer) into the FilmArray RP pouch. The pouch contains all of the reagents required for specimen testing and analysis in a freeze-dried format; the addition of Hydration Solution and Sample/Buffer Mix rehydrates the reagents. After the pouch is prepared, the FilmArrav software guides the user though the steps of placing the pouch into the instrument, scanning the pouch barcode, entering the sample identification and initiating the run.

The FilmArray instrument contains a coordinated system of inflatable bladders and seal points, which act on the pouch to control the movement of liquid between the pouch blisters. When a bladder is inflated over a reagent blister, it forces liquid from the blister into connecting channels. Alternatively, when a seal is placed over a connecting channel it acts as a valve to open or close a channel. In addition, electronically controlled pneumatic pistons are positioned over multiple plungers in order to deliver the rehydrated reagents into the blisters at the appropriate times. Two Peltier devices control heating and cooling of the pouch to drive the reverse transcription reactions, the PCR reactions, and the melting curve analysis.

Nucleic acid extraction occurs within the FilmArray pouch using mechanical lysis and standard magnetic bead technology. After extracting and purifying nucleic acids from the unprocessed sample, the FilmArray performs a nested multiplex PCR that is executed in two stages. During the first stage, the FilmArray performs a single, large volume, highly multiplexed reverse transcription PCR (rt-PCR) reaction. The products from first stage PCR are then diluted and combined with a fresh, primer-free master mix and a fluorescent double stranded DNA binding dye (LC Green@Plus, Idaho Technology). This second master mix solution, is then distributed to each well of the array. Array wells contain sets of primers designed specifically to amplify sequences internal to the PCR products generated during the first stage PCR reaction. The second stage PCR, or nested PCR, is performed in singleplex fashion in each well of the array. At the conclusion of the 200 stage PCR, the array is interrogated by melting curve analysis for the detection of signature amplicons denoting the presence of specific viral or bacterial targets A digital camera placed in front of the second stage PCR captures fluorescent images of the PCR reactions in real time.

Idaho Technology Inc. 510(k) FilmArray Respiratory Panel System

{3}------------------------------------------------

The FilmArray software automatically interprets the results of each DNA melting curve analysis and combines the data with the results of the internal pouch controls to provide a test result for each organism on the panel.

Substantial Equivalence:

The Luminex® xTAG™ RVP is a PCR-based system for detecting the presence of viral nucleic acid in nasopharyngeal swabs collected from individuals with signs and symptoms of respiratory illness. The similarities and differences between the xTAG RVP and the FilmArray RP are outlined below.

ElementFilmArray Respiratory Panel TestSystemLuminex® xTAG™ RVP
OrganismsDetectedInfluenza A, Influenza A subtype H1,Influenza A subtype H3, Influenza B,Respiratory Syncytial Virus, humanMetapneumovirus, Adenovirus,Parainfluenza 3, andRhinovirus/EnterovirusSameSee below for differences
AnalyteRNA/DNASame
TechnologicalPrinciplesmultiplex nucleic acidSameSee below for differences
SpecimenTypesNasopharyngeal swabsSame

Table 2. Similarities between the xTAG RVP and the FilmArray RP

Table 3. Differences between the xTAG RVP and the FilmArray RP

ElementFilmArray Respiratory Panel TestSystemLuminex® xTAG™ RVP
OrganismsDetectedCan distinguish Influenza A subtype 2009H1 from Influenza A subtype H1.Also detects Coronavirus NL63 andCoronavirus HKU1.Can distinguish Respiratory Syncytial VirusType A from Respiratory Syncytial VirusType B. Detects Parainfluenza virus 1 andParainfluenza virus 2.
TechnologicalPrinciplesNested multiplex RT-PCR followed by highresolution melting analysis to confirmidentity of amplified product.Multiplex RT-PCR and multiplex TSPEfollowed by Fluorescence-activated sortingof labeled beads coupled to streptavidin-conjugated biotinylated products
InstrumentationFilmArray InstrumentPCR ThermocyclerLuminex® 100 IS or 200 system
Time to resultLess than 1 hourApproximately 8 hours
TestInterpretationAutomated test interpretation and reportgeneration. User cannot access raw data.Semi-automated test interpretation. Usermust review all "no call" results todetermine cause and retesting strategy.
SamplePreparationMethodSample Processing is automated in theFilmArray instrument.Up front sample processing is required toextract nucleic acid.

{4}------------------------------------------------

ElementFilmArray Respiratory Panel TestSystemLuminex® xTAG™ RVP
ReagentStorageReagents are stored at room temperature.Reagents stored at 4°C and -20°C.
ControlsTwo controls are included in each reagentpouch to control for sample processing andboth stages of PCR and melt analysis.Internal control added to each sample.External control processed with each batchof samples.
UserComplexityModerate/LowHigh

Summary of Performance Data

Clinical Performance

The clinical performance of the FilmArray RP system was established during a prospective study at 3 U.S. clinical sites over a 6 month time period (December 2009 through May 2010). Subjects with signs and/or symptoms of respiratory infection were invited to participate. Upon obtaining informed consent, NPS samples were collected for FilmArray and comparator testing; a second respiratory sample was collected from each subject for viral culture reference testing. A total of 857 subjects were initially enrolled in the study and four were withdrawn. Table 4, provides a summary of demographic information for the 853 subjects that participated in the prospective study.

SexNumber of Subjects
Male449 (53%)
Female404 (47%)
Age
≤5484 (57%)
6-2195 (11%)
22-49190 (22%)
≥5084 (10%)

Table 4. Demographic Summary for FilmArray RP Prospective Study

{5}------------------------------------------------

Each NPS specimen was tested with the FilmArray RP. The performance of the FilmArray RP was evaluated by comparing the FilmArray RP test result for each member of the panel with the appropriate comparator/reference methods shown in Table 5.

Organism/VirusReference/Comparator Method(s)
Adenovirus
Influenza AViral culture followed by DFAidentification1
Influenza B
Parainfluenza virus 3
Respiratory Syncytial Virus
FluA/H1 subtyping1 PCR test of influenza A positive viral
FluA/H3 subtypingculture with bi-directional sequence
FluA/2009 H1subtypingconfirmation2
Human Rhinovirus
Enterovirus2 PCR tests of patient specimen with bi-
Coronavirus NL63directional sequence confirmation3
Coronavirus HKU1
Human Metapneumovirus

Table 5. Reference/Comparator Methods Used to Assess FilmArray RP Performance

1 Performance of the FilmArray RP system detecting Adenovirus, Flu A. Flu B. PIV3, or RSV. respectively, was compared to viral culture followed by fluorescent antibody identification. "True" Adenovirus, Influenza B, Parainfluenza virus 3, or RSV positively, were considered as any sample that tested positive for Adenovirus, Influenza A, Influenza B, Parainfluenza virus 3, or RSV, respectively, by viral culture followed by DFA testing. "True" Adenovirus, Influenza A. Influenza virus 3, or RSV negatives, respectively, were considered as any sample that tested negative for Adenovirus. Influenza B. Parainfluenza virus 3, or RSV. respectively, by viral culture followed by DFA testing.

2 Performance of the FilmArray RP system detecting Influenza A/H1, A/H3, or A/2009 H1, respectively, was compared to viral culture followed by one analytically validated PCR assay with bi-directional sequence confirmation. The comparator assays were designed to amplify a different sequence from that amplified by the FilmArray assay(s). None of the comparator PCR assays overlapped any FilmArray amplicon sequence even if the same gene was targeted. "True" Influenza A/H1, A/H3, or A/2009 H1 positively, were considered as any sample that tested positive for Influenza A by viral culture, and had bi-directional sequencing pre-defined quality acceptance criteria that matched Influenza A/H1, A/H3, or A/2009 H1 sequences deposited in the National Center for Biotechnology Information (NCBI) GenBank database (www.ncbi.nlm.nih.gov), respectively, with acceptable E-values. "True" Influenza A/H1, A/H3, or A/2009 H1 negatively, were considered as any sample that tested negative for Influenza A by viral culture, or any sample that tested positive for Influenza A virus by viral culture, but tested negative by the respective Influenza A subtype specific PCR assay.

3 Performance of the FilmArray RP system detecting Human Rhinovirus, Coronavirus NL63, Coronavirus HKU1. or Human Metapneumovirus, respectively, was compared to a predetermined algorithm that used composite comparator methods. The methods consist of two analytically validated PCR assays followed by bi-directional sequencing. The comparator assays were designed to amplify a different sequence from that amplified by the FilmArray assay(s). None of the comparator PCR assays overlapped any FilmArray amplicon sequence even if the same gene was targeted. "True" Human Rhinovirus, Enterovirus NL63, Coronavirus HKU1, or Human Metapneumovirus positives, respectively, were considered as any sample that had bi-directional sequencing data meeting pre-defined quality acceptance criteria that matched Human Rhinovirus, Coronavirus NL63, Coronavirus HKUI, or Human Metapneumovirus sequences deposited in the National Center for Biotechnology Information (NCBI) GenBank database (www.ncbi.nlm,nih.gov), respectively, with acceptable E-values. "True" Human Rhinovirus, Enterovirus, Coronavirus NL63, Coronavirus HKU1, or Human Metapneumovirus negatives, respectively, were considered as any sample that tested negative by both of the comparator PCR assays.

{6}------------------------------------------------

A total of 853 specimens were evaluated in this study. Clinical sensitivity or positive percent agreement (PPA) was calculated as 100% x (TP / TP + FN). True positive (TP) indicates that both the FilmArray RP and comparator method had a positive result for this specific analyte and false negative (FN) indicates that the FilmArray result was negative while the comparator result was positive. Specificity was calculated as 100% x (TN / TN + FP). True negative (TN) indicates that both the FilmArray RP and the comparator method had negative results and a false positive (FP) indicates that the FilmArray RP result was positive but the comparator results was negative. The exact binomial two-sided 95% confidence interval was calculated. The results are summarized in Table 6.

Sensitivity95% CISpecificity95% CI
Adenovirus24/27a88.9%70.8 - 97.7%812/826b98.3%97.2 - 99.1%
Influenza A9/1090.0%55.5 - 99.8%841/843d99.8%99.2 - 100%
Influenza A/H10/0n/an/a853/853100%99.6 - 100%
Influenza A/H30/0n/an/a853/853100%99.6 - 100%
Influenza A/2009 H18/988.9%51.8 - 99.7%841/844d99.6%99.0 - 99.9%
Influenza B0/0n/an/a853/853100%99.6 - 100%
Parainfluenza Virus 323/24e95.8%78.9 - 99.9%819/829f98.8%97.8 - 99.4%
Respiratory Syncytial Virus52/52100%93.2 - 100%714/801g89.1%86.8 - 91.2%
PPA95% CINPA95% CI
Coronavirus HKU123/2495.8%78.9 - 99.9%827/829c99.8%99.1 - 100%
Coronavirus NL6323/2495.8%78.9 - 99.9%829/829100%99.6 - 100%
Human Metapneumovirus88/9394.6%87.9 - 98.2%754/76099.2%98.3 - 99.7%
Human Rhinovirus/Enterovirus190/20592.7%88.2 - 95.8%613/64894.6%92.6 - 96.2%

Table 6, Clinical Sensitivity and Specificity for the Film Array RP Prospective Clinical Study

4 The FilmArty RP system detected Adenovirus in 1/3 false negative specimens when retested. The Adenovirus in the retested specimen was identified as species C by bi-directional sequence analysis. The remaining two false negative specimens were identified as species C and species B.

8 Adenoviruses were identified in 13/14 false positive specimens using bi-directional sequence analysis. Ten were identified as species C, two as species B, and one as species E.

CoV-HKUI viruses were identified in 2/2 false positive specimens using bi-directional sequence analysis with an alternate assav.

4 Influenza A viruses (2009 H1 subtype) were identified in 3/3 false positive specimens (2/2 false positive compared to influenza A culture alone) using sequence analysis with an alternate assay.

& The FilmArray RP system detected PIV3 in the single false negative specimen when retested.

1 PIV3 viruses were identified in 10/10 false positive specimens using bi-directional sequence analysis.

8 RSV viruses were identified in 83/87 false positive specimens using bi-directional sequence analysis.

{7}------------------------------------------------

The FilmArray RP system detected a total of 81 mixed infections in the prospective clinical evaluation. This represents 16.1% of the total positive specimens (81/502). Seventy-six (76/81; 93.8%) were double infections, and 5 (5/81; 6.2%) were triple infections. The total number of test results comprising these co-infections was 167. The single most common co-infection (21/81; 25.9%) was Human Rhinovirus/Enterovirus with Respiratory Syncytial Virus. These viruses were the most prevalent in the tested population. Out of the 81 co-infections, 47 contained one or more analytes that had not been detected with the reference/comparator methods, i.e. discrepant co-infection.

Organism CombinationsNumberofPositiveSamplesPercentageof TotalSamplesTestedOrganismNumber ofMixedInfectionsPrevalencein MixedInfections
HRV/Entero + RSV212.46%Adenovirus1619.75%
Adenovirus + HRV/Entero91.06%CoVHKU11012.35%
HRV/Entero + PIV380.94%CoVNL631316.05%
hMPV + HRV/Entero70.82%hMPV2227.16%
hMPV + RSV40.47%HRV/Entero5669.14%
CoVNL63 + HRV/entero40.47%FluA/H100.00%
CoVHKU1 + hMPV30.35%FluA/H1-200900.00%
CoVHKU1 + HRV/Entero30.35%FluA/H300.00%
CoVHKU1 + RSV30.35%FluB00.00%
CoVNL63 + hMPV30.35%PIV31417.28%
CoVNL63 + RSV30.35%RSV3644.44%
hMPV + PIV330.35%
Adenovirus + HRV/Entero +PIV320.23%
Adenovirus + RSV20.23%
Adenovirus + CoVNL6310.12%
Adenovirus + hMPV10.12%
Adenovirus + PIV310.12%
CoVNL63 + HRV/Entero + RSV10.12%
CoVHKU1 + HRV/Entero +RSV10.12%
CoVNL63 + hMPV + RSV10.12%
Total Mixed Infections819.50%

Table 7. Mixed Infections Detected by FilmArray RP and Prevalence of Individual Analytes in Mixed Infections

{8}------------------------------------------------

Table 8 provides the prevalence and age distribution of FilmArray RP detected analytes in the clinical study. The most prevalent analytes were Human Rhinovirus/Enterovirus, Respiratory Syncytial Virus, and Human Metapneumovirus; these three analytes comprised 78% of all positive results. Positive results were obtained across all age groups tested; however, the majority was detected in children 5 years or younger.

AnalyteTotal(Prevalence)≤ 5 years6-21years22-49years≥ 50years
Adenovirus38 (4.5%)32231
Coronavirus HKU125 (2.9%)12184
Coronavirus NL6323 (2.7%)17222
Human Metapneumovirus94 (11%)764104
Human Rhinovirus/Enterovirus225 (26.4%)161242911
Influenza A (all subtypes)11 (1.3%)1172
Influenza A/H10 (0%)0000
Influenza A/2009 H111 (1.3%)1172
Influenza A/H30 (0%)0000
Influenza B0 (0%)0000
Parainfluenza Virus 333 (3.9%)31101
Respiratory Syncytial Virus139 (16.3%)127345

Table 8. Prevalence and Age Distribution of Analytes in the Clinical Study

Several analytes, such as influenza viruses were either not encountered in the clinical study or had a low prevalence. To supplement the results of the prospective clinical study, an evaluation of preselected archived samples was performed.

Testing of Preselected Archived Specimens

In addition to the prospective clinical study, archived clinical NPS specimens were also tested using the FilmArray RP. The specimens were selected because they had previously tested positive for one of the following organisms: Adenovirus, Enterovirus, Influenzas A/H1, 2009 H1N1, and H3, Influenza B, and Parainfluenza Virus 3, or had been negative by previous testing methods. Prior to testing with the FilmArray RP, the presence or absence of the analyte of interest was confirmed in each specimen using analyte specific PCR and bi-directional sequencing. Of 400 specimens, 349 were confirmed to contain the analyte of interest (or lack thereof for negative specimens). The specimens were organized into "test panels" and randomized such that the users testing the samples with the FilmArray RP were blinded as to the expected test result. Each panel contained specimens known to be positive and negative for the specific analyte being evaluated allowing the calculation of a positive percent agreement (PPA) and a negative percent agreement (NPA). A summary of the available demographic information of the tested samples is provided in Table 9 and the results of the FilmArray testing are presented in Table 10.

{9}------------------------------------------------

Total Specimens349
SexFemale (%)82 (23.5%)
Male (%)79 (22.6%)
Unknown188(53.9%)
AgeAvg14.1
Median4.0
Min0.5
Max83.0
Age Range≤589 (25.5%)
6-2135 (10.0%)
22-4923 (6.6%)
≥5014 (4.0%)
Unknowna188(53.9%)

Table 9. Demographic Summary of FilmArray RP Archived Specimen Study

4 Demographic information was not provided for specimens from one source. Because the specimens were provided by a pediatric hospital, it is understood that the age range of specimens was from <1 yrs to 21 yrs.

Table 10. FilmArray Archived Specimen Performance Data Summary
----------------------------------------------------------------------
Positive Percent Agreement (PPA)Negative Percent Agreement (NPA)
TP/TP+FNPercent95% CITN/TN+FPPercent95% CI
Adenovirus27/27100.0%87.2 - 100%28/28100.0%87.7 - 100%
Enterovirus22/2395.7%78.0 - 99.9%90/90100.0%96.0 - 100%
FluA/H132/32100.0%89.1 - 100%127/127100.0%97.1 - 100%
FluA/H1-200934/34100.0%89.7 - 100%125/125100.0%97.1 - 100%
FluA/H354/54100.0%93.4 - 100%105/105100.0%96.5 - 100%
Influenza B30/30100.0%88.4 - 100%129/129100.0%97.2 - 100%
PIV336/36100.0%90.3 - 100%93/93100.0%96.1 - 100%

{10}------------------------------------------------

Selected Analytic Studies

Limit of Detection

The analytical sensitivity or Limit of Detection (LoD) for each FilmArray RP analyte (except for Coronavirus HKU1) was determined by testing limiting dilutions of live, quantified viruses. The LoD for Coronavirus HKU1 was determined by testing limiting dilutions of a clinical specimen containing a high viral load of Coronavirus HKU1. LoD is defined as the lowest concentration at which the analyte is consistently detected (detection in ≥95% of samples tested). Simulated NPS sample matrix (cultured human cells in VTM) was spiked with one or more analytes and at least 20 replicates were tested at the LoD concentration. The LoD for each FilmArray RP analyte is listed in Table 11.

OrganismStrain IdentificationLimit of Detection
AdenovirusSerotype 1 (Species C)300 TCID50/mL
Coronavirus HKU1aClinical Specimen (Type B )1.9 x 106 RNA copies/mL
Coronavirus NL63NR-4705 TCID50/mL
Human MetapneumovirusType A1 (hMPV-16, IA10-2003)2 TCID50/mL
EnterovirusEchovirus 630,000 TCID50/mL
Human RhinovirusA11 TCID50/mL
Influenza A/H1A/Brisbane/59/07200 TCID50/mL
Influenza A/H1A/New Caledonia/20/992,000 TCID50/mL
Influenza A/2009 H1A/SwineNY/03/2009100 TCID50/mL
Influenza A/H3A/Wisconsin/67/20055 TCID50/mL
Influenza A/H3A/Port Chalmers/1/7350 TCID50/mL
Influenza BB/FL/04/0660 TCID50/mL
Influenza BB/Taiwan/2/6260 TCID50/mL
Parainfluenza Virus 3Type 310 TCID50/mL
Respiratory Syncytial VirusType A2 TCID50/mL

Table 11. LoD for Analytes Detected by FilmArray RP

" Coronavirus HKU1 was quantified by a non-FilmArray real-time PCR assay against a standard curve of synthetic Coronavirus HKUI RNA transcript to obtain quantification of the viral nucleic acid in the clinical specimen (RNA copies/mL).

NOTE: Most analytes were re-grown and quantified in TCID-s(50% Tissue Culture Infectious Dose). The unit TCID-so is a measure of infectivity or cytotoxicity rather than number of organisms or copies of nucleic acid. Variability in TCIDs/mL may not accurately reflect differences in the relative of detection between different organisms or different strains of the same organism.

{11}------------------------------------------------

Analytical Reactivity (Inclusivity)

The analytical reactivity of the FilmArray RP system assays was evaluated with an inclusivity panel consisting of 99 strains or isolates that represent the genetic, temporal. and geographic diversity of the FilmArray RP analytes. The tested organisms include: 17 Adenovirus, 4 Coronavirus (3 HKU1 and 1 NL63), 10 human Metapneumovirus, 12 Enterovirus, 14 Rhinovirus, 22 Influenza A (including 10 Influenza A/H1, 3 Influenza A/2009 H1 and 9 Influenza A/H3), 11 Influenza B, 3 Parainfluenza virus 3 and 6 Respiratory Syncytial Virus. Each organism was initially tested in a simulated NPS sample matrix at or near the system LoD. Higher concentrations were tested if the analyte was not detected at LoD.

Each of the 99 strains tested in this study were reactive with the FilmArray RP system. Reactivity of the FilmArray RP system with Adenovirus C serotype 6 was 10.000-fold less than the system LoD due to sequence variation in the region targeted by the FilmArray RP Adenovirus assay.

Additional clinical data analyses, and in silico analyses were also carried out to supplement the testing of the inclusivity panel.

Results from inclusivity testing are presented below. The concentration and multiple of LoD at which each strain was detected by the FilmArray RP system is indicated.

SpeciesSerotype (Isolate)ConcentrationMultiple of LoD
A31300 TCID50/mL1x
3300 TCID50/mL1x
7a300 TCID50/mL1x
7d2 (Iowa/2001)300 TCID50/mL1x
B7h (Iowa/1999)300 TCID50/mL1x
11 (Wisconsin/2005)3,000 TCID50/mL10x
14 (Missouri/2005)300 TCID50/mL1x
21 (Missouri/2005)300 TCID50/mL1x
34 (Texas/2005)300 TCID50/mL1x
C1300 TCID50/mL1x
2 (New York/2004)30,000 TCID50/mL100x
53,000 TCID50/mL10x
6 (Colorado/2005)*3,000,000 TCID50/mL10,000x
D83,000 TCID50/mL10x
E4a (S Carolina/2004)300 TCID50/mL1x
4p3 (New Jersey/2005)300 TCID50/mL1x

Table 12. Results of Inclusivity Testing for Adenovirus

{12}------------------------------------------------

SpeciesSerotype (Isolate)ConcentrationMultiple of LoD
F41 (Indiana/2004)300 TCID50/mL1x

*Due to sequence variation, the FilmArray RP Adenovirus assay reacts with adenovirus serotype 6 (species C) less efficiently than with other adenovirus serotypes.

Supplemental Adenovirus Reactivity Information (Clinical Data and in silico analyses):

A subset of prospective and retrospective archived clinical specimens that were FilmArray positive for Adenovirus was subjected to PCR and bi-directional sequence analysis. BLAST analysis of the sequence data obtained for a region of the Adenovirus polymerase gene identified adenovirus species B (serotypes 3, 3+11p, 7, 16 and 21), species C (serotypes 1, 2, and 5), and species E (serotype 4) in these clinical specimens. Species B, C and E are the most common Adenovirus species associated with respiratory illness. Species A, D, and F (often associated with conjunctivitis and gastroenteritis) were not identified in clinical specimens.

In addition to laboratory testing, bioinformatics resources were also used to predict reactivity of additional Adenovirus species and serotypes with the FilmArray RP System. Simulated reactivity was based on the number and location of mismatches between the target sequence and the assay primer(s). Table 13 lists the adenovirus types that were not tested by the FilmArray system either in analytical testing. Based on the bioinformatics analysis, the FilmArray RP system is predicted to react with all of the indicated Adenovirus species and serotypes.

VirusSpeciesSerotypeGenBank IDSimulated FilmArrayAdenovirus Result
HumanAdenovirusA12AB330093Positive
A18DQ149610Positive
16AB330097Positive
B35AB052912Positive
B50DQ149643Positive
D9AB330090Positive
D10DQ149615Positive
D13DQ149616Positive
D15DQ149617Positive
D17AB330098Positive
D19DQ149618Positive
D20DQ149619Positive
D22DQ149620Positive
D23DQ149621Positive
D24DQ149622Positive

Table 13. Simulated FilmArray RP Reactivity with Untested Adenovirus Serotypes

{13}------------------------------------------------

VirusSpeciesSerotypeGenBank IDSimulated FilmArrayAdenovirus Result
25DQ149623Positive
26DQ149624Positive
27DQ149625Positive
28DQ149626Positive
29DQ149627Positive
30DQ149628Positive
32DQ149629Positive
33DQ149630Positive
36DQ149631Positive
37DQ149632Positive
38DQ149633Positive
39DQ149634Positive
42DQ149635Positive
43DQ149636Positive
44DQ149637Positive
45DQ149638Positive
46DQ149639Positive
47DQ149640Positive
48AB330129Positive
49DQ149641Positive
51DQ149642Positive
53FJ169625Positive
F40AB330121Positive
G52DQ923122Positive

Table 13. Simulated FilmArray RP Reactivity with Untested Adenovirus Serotypes

Table 14. Results of Inclusivity Testing for Coronaviruses

TypeStrain / IsolateConcentration abMultiple of LoD
Clinical Sample #1120$2.08 x 10^6$RNA copies/mL1.1x
HKU1Clinical Sample # 6123$1.41 x 10^4$ TCID50/mL~ $1.9 x 10^6$RNA copies/mL1x
Clinical Sample #6213 (Type B)c$1.9 x 10^6$RNA copies/mL1x
NL63BEI Resources NR-47050 TCID50/mL1x

4 Virus contained in Clinical Sample #6123 was grown in culture and quantified (TCIDs) by infectivity assay. b Quantification of the viral RNA contained in clinical specimens containing Coronavirus HKU1 was performed using real-time RT-PCR against a standard curve generated from a synthetic RNA template.

6 Phylogenetic information for Coronavirus HKU1 Clinical Sample #6213 was obtained from bi-directional sequence analysis of the nucleocapsid (N) gene.

{14}------------------------------------------------

SubtypeStrain IDConcentrationMultiple of LoD
A19IA3-20022 TCID50/mL1x
16IA10-20032 TCID50/mL1x
A220IA14-20032 TCID50/mL1x
27IA27-20042 TCID50/mL1x
B13Peru2-20022 TCID50/mL1x
5Peru3-20032 TCID50/mL1x
13IA7-20032 TCID50/mL1x
18IA18-20032 TCID50/mL1x
B28Peru6-20032 TCID50/mL1x
22IA16-20032 TCID50/mL1x

Table 15. Results of Inclusivity Testing for Human Metapneumovirus

Table 16. Results of Inclusivity Testing for Enterovirus and Rhinovirus

SpeciesStrainConcentrationaMultiple of LoD
Enterovirus ACoxsackievirus A10 ATCC VR-16830,000 TCID50/mL1x
Enterovirus 71 ATCC VR-14321:30,000dilution of stockn/a
Enterovirus 719400 TCID50/mLb<1x
Enterovirus BCoxsackievirus A99400 TCID50/mLb<1x
Coxsackievirus B330,000 TCID50/mL1x
Coxsackievirus B430,000 TCID50/mL1x
Echovirus 630,000 TCID50/mL1x
Echovirus 99400 TCID50/mLb<1x
Echovirus 11300,000 TCID50/mL10x
Enterovirus CCoxsackievirus A21 /KuykendallATCC VR-85030,000 TCID50/mL1x
Coxsackievirus A24 DN-19ATCC VR-58330,000 TCID50/mL1x
Enterovirus DEnterovirus 68 (F02-3607 corn)ATCC VR-119730,000 TCID50/mL1x
Rhinovirus AA11 TCID50/mL1x
A2 (HGP) ATCC VR-48210 TCID50/mL10x
A7 (68-CV11) ATCC VR-16011 TCID50/mL1x
A16 (11757) ATCC VR-28310 TCID50/mL10x
A34 (137-3) ATCC VR-5071 TCID50/mL1x
A57 (Ch47) ATCC VR-1600100 TCID50/mL100x

{15}------------------------------------------------

SpeciesStrainConcentrationaMultiple of LoD
A77 (130-63) ATCC VR-11871 TCID50/mL1x
A85 (50-525-CV54) ATCC VR-119510 TCID50/mL10x
B3 (FEB) ATCC VR-4831 TCID50/mL1x
B14 (1059) ATCC VR-2841 TCID50/mL1x
Rhinovirus BB17 (33342) ATCC-282100 TCID50/mL100x
B27 (5870) ATCC VR-11371 TCID50/mL1x
B42 (56822) ATCC VR-33810 TCID50/mL10x
B83 (Baylor 7) ATCC VR-11931 TCID50/mL1x

ª The LoD for Enterovirus is 30,000 TCID50mL. The LoD for Rhinovirus is 1 TCID50mL.

b Strains were tested below the LoD concentration due to a lesser concentration of virus in the culture fluid.

Supplemental Human Rhinovirus/Enterovirus Reactivity Information

(clinical data and in silico analyses):

In addition to the analytical inclusivity testing, BLAST analysis was performed on sequence data (5' UTR) obtained from prospective and retrospective archived clinical specimens that were FilmArray positive for Human Rhinovirus/Enterovirus. The following species and subtypes were identified in clinical specimens:

Enterovirus Species A:Enterovirus serotype 71Coxsackievirus A2 and A6
Enterovirus Species B:Echovirus serotypes 3, 6, 7, 11, 15, 21 and 30Coxsackievirus B1, B3 and A9Enterovirus serotypes 81 and 88
Rhinovirus Species A:Human Rhinovirus serotypes 1B, 8, 9,10, 13, 19, 21, 22, 23,28, 30, 32, 34, 38, 39, 40, 46, 47, 49, 51, 54, 56, 58, 59, 61, 62,66, 68, 75, 77, 78, 80, 82, 98 and 100
Rhinovirus Species B:Human Rhinovirus serotypes 27, 69, 83 and 91
Rhinovirus Species C:at least 3 individual strains and 12 distinct isolates

ªHuman Rhinovirus species C (also known as Enterovirus species D) has not been classified into serotypes.

{16}------------------------------------------------

Enterovirus species C includes Poliovirus types 1-3 (PV1, PV2, and PV3). Simulated reactivity of the FilmArray RP Human Rhinovirus/Enterovirus assays with Enterovirus Species C Poliovirus sequences was generated using a bioinformatics approach. Alignment of assay primer sequences with the GenBank sequences indicates the FilmArray RP assay will react with Poliovirus types 1, 2 and 3, giving a Human Rhinovirus/Enterovirus result.

StrainGenBank IDSimulated FilmArray RP Result
Human poliovirus 1 strainCHN8264c/GZ/CHN/2004FJ769385Human Rhinovirus/Enterovirus
Human poliovirus 2, complete genomeAY177685Human Rhinovirus/Enterovirus
Human poliovirus 3 strain IRA10853, completegenomeEU684056Human Rhinovirus/Enterovirus

Table 17. Simulated Reactivity of the FilmArray RP HRV and Entero Assays with Poliovirus Sequences

TypeStrainConcentrationMultiple of LoD
Influenza A(H1N1)A/Brisbane/59/07200 TCID50/mL1x
A/Solomon Islands/3/2006200 TCID50/mL1x
A/Hawaii/15/01CDC#20017011171:300 ªn/a
A/New Caledonia/20/99200 TCID50/mL1x
A1/Denver/1/57200 TCID50/mL1x
A/Mal/302/54200 TCID50/mL1x
A1/FM/1/47200 TCID50/mL1x
A/Weiss/43200 TCID50/mL1x
A/PR/8/342000 TCID50/mL10x
A/NWS/33200 TCID50/mL1x
Swine NY/01/2009100 TCID50/mL1x
Influenza A(H1N1-2009)Swine NY/02/2009100 TCID50/mL1x
Swine NY/03/2009100 TCID50/mL1x
Swine NY/03/2009100 TCID50/mL1x
Influenza A(H3N2)A/Brisbane/10/075 TCID50/mL1x
A/Wisconsin/67/20055 TCID50/mL1x
A/NewYork/55/2005CDC#20057055611:300,000 ªn/a
A/Victoria/3/755 TCID50/mL1x

Table 18. Results of Inclusivity Testing for Influenza A

{17}------------------------------------------------

TypeStrainConcentrationMultiple of LoD
A/Port Chalmers/1/735 TCID50/mL1x
A/Aichi/2/6850 TCID50/mL10x
A/Hong Kong/8/685 TCID50/mL1x
Alice (vaccine) A/England/42/72ATCC VR-7765 TCID50/mL1x
MRC-2 Recombinant strainATCC VR-7775 TCID50/mL1x

4 Strain was not re-cultured and titered. Dilutions of the original culture/titer from the CDC were used. The HA titer for A/Hawaii/15/01 is unknown and the HA titer for A/NewYork/55/2005 was 256.

Supplemental Reactivity Information for Influenza Strains of Human, Swine and Avian Origin (analytical testing and in silico analyses):

Results of testing viral isolates or nucleic acid from viral culture indicate that the FilmArray RP pan-influenza and subtyping assays react with virus of swine and avian origin as expected (Table 19).

HostSubtypeIsolate / StrainTestConcentrationFilmArrayResult
SwineH1N1Influenza A/Swine/1976/31~1 x 106 EID50/mLInfluenza A H1 a
H1N1Influenza A/Swine/Iowa/15/30~1 x 107 EID50/mL
AvianH1N2Kilbourne F63A/NWS/34 (HA) x A/RockefellerInstitute/5/57 (NA)14.8 ng RNA bInfluenza A H1

Table 19. Results of Inclusivity Testing with Swine and Avian Isolates of Influenza A

ª No reactivity was observed with the 2009 H1 subtyping assay or other FilmArray RP assays.

b Purified and quantified RNA from avian influenza culture was obtained from BEI Resources.

Laboratory testing of influenza A strains was supplemented with in silico predictions of reactivity using bioinformatics and sequence alignments between FilmArray RP assay primers and sequences for influenza A strains of human, swine, and avian origin. For each strain. multiple (3) GenBank IDs were evaluated, corresponding to the gene segments targeted by the FilmArray RP assays (matrix (MA), non-structural (NS) and hemagglutinin(HA)). Simulated reactivity was determined based on the number and location of mismatches in the targeted region. The strains listed in Table 20 are predicted to react with the FilmArray RP pan-influenza A and H1, H1-2009 or H3 subtyping assays as indicated.

{18}------------------------------------------------

HostSubtypeStrainGenBankIDSimulatedFilmArray RPReactivity
HumanH1N1A/California/UR06-0393/2007(H1N1)CY026540CY026543CY026539Influenza A H1
H1N1-2009A/Aalborg/INS133/2009(H1N1)CY063606CY063610CY063607Influenza A H1-2009
H1N2A/New York/297/2003(H1N2)CY002668CY002665CY002664Influenza A H1
H2N2A/Albany/20/1957(H2N2)CY022013CY022014CY022017Influenza A(no subtypedetected)
H3N2A/Boston/38/2008(H3N2)CY044581CY044584CY044580Influenza A H3
H5N1A/Cambodia/R0405050/2007(H5N1)HQ200572HQ200573FJ225472Influenza A(no subtypedetected)
H5N1A/Hong Kong/486/97(H5N1)AF084281AF255368AF115289Influenza A(no subtypedetected)
H7N2A/New York/107/2003(H7N2)EU587368EU587373EU587374Influenza A(no subtypedetected)
H7N3A/Canada/rv504/2004(H7N3)CY015006CY015007CY015010Influenza A(no subtypedetected)
SwineH7N7A/Netherlands/219/03(H7N7)AY340089AY342422AY338459Influenza A(no subtypedetected)
H9N2A/Hong Kong/1073/99(H9N2)AJ404626AJ278647AJ278649Influenza A(no subtypedetected)
H1N1A/swine/Wisconsin/1/1971(H1N1)CY022414CY022417CY022413Influenza A H1
H1N2A/swine/Hong Kong/NS857/2001(H1N2)A/swine/Sweden/1021/2009(H1N2)GQ229348GQ229350GQ229347GQ495135GQ495136GQ495132Influenza A H1Influenza A(no subtypedetected)
AvianH5N1A/swine/East Java/UT6010/2007(H5N1)HM440124HM440111HM440123Influenza A(no subtypedetected)
H2N2A/chicken/New York/13828-3/1995(H2N2)A/Japan/305/1957(H2N2)CY014822CY014825CY014821CY045804CY014977CY014980Influenza A(no subtypedetected)
HostSubtypeStrainGenBankIDSimulatedFilmArray RPReactivity
A/Korea/426/1968(H2N2)CY031595Influenza A
CY031596(no subtypedetected)
CY031599
H3N1A/blue-winged teal/ALB/452/1983(H3N1)CY004635
CY004638Influenza A H3
CY005940
H3N2A/American black duck/North Carolina/675-075/2004(H3N2)GU051135Influenza A
GU051136(no subtypedetected)
GU051137
H3N5A/mallard/Netherlands/2/1999(H3N5)CY060261
CY060264Influenza A H3
CY060265
H3N6A/American black duck/NewBrunswick/25182/2007(H3N6)CY047696Influenza A
CY047697(no subtypedetected)
CY047700
H3N7A/northernshoveler/California/HKWF1367/2007(H3N7)CY033372
CY033375Influenza A H3
CY033376
H3N8A/American blackduck/Washington/699/1978(H3N8)GU052299
GU052302Influenza A H3
GU052300
H4N6A/blue-winged teal/Minnesota/Sg-00043/2007(H4N6)CY063977Influenza A
CY063978(no subtypedetected)
CY063981
A/rook/Rostov-on-Don/26/2007(H5N1)EU814503Influenza A(no subtypedetected)
EU814504
EU814507
H5N1A/turkey/VA/505477-18/2007(H5N1)GU186509Influenza A
GU186510(no subtypedetected)
GU186513
A/chicken/Bangladesh/1151-10/2010(H5N1)HQ156765Influenza A
HQ156766(no subtypedetected)
HQ156764
H5N2A/duck/Pennsylvania/10218/1984(H5N2)AB295603Influenza A
AB286120(no subtypedetected)
AB286652
H5N3A/duck/Singapore/F119/3/1997(H5N3)GU052802Influenza A
GU052803(no subtypedetected)
GU052805
H6N1A/duck/PA/486/1969(H6N1)EU743286Influenza A
EU743287(no subtypedetected)
EU743289
H6N2A/mallard/Czech Republic/15902-17K/2009(H6N2)HQ244430Influenza A
HQ244433(no subtypedetected)
HQ244434
H7N7A/mallard/Korea/GH171/2007(H7N7)FJ750872Influenza A(no subtypedetected)
FJ959087
FJ959090
HostSubtypeStrainGenBankIDSimulatedFilmArray RPReactivity
CY014664(no subtypedetected)
CY014667
H10N7A/chicken/Germany/N/1949(H10N7)GQ176136Influenza A
GQ176135(no subtypedetected)
GQ176132
H11N9A/duck/Memphis/546/1974(H11N9)CY014691Influenza A
GQ257441(no subtypedetected)
CY014687

Table 20. Simulated Reactivity of FilmArray Influenza A Assays with Human, Swine, and Avian Influenza Strains ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ -

{19}------------------------------------------------

Table 20. Simulated Reactivity of FilmArray Influenza A Assays with Human, Swine, and Avian Influenza Strains

{20}------------------------------------------------

Table 20. Simulated Reactivity of FilmArray Influenza A Assays with Human, Swine, and Avian Influenza Strains

Table 21. Results of Inclusivity Testing for Influenza B

StrainConcentrationMultiple of LoD
B/FL/04/0660 TCID50/mL1x
B/Ohio/01/2005CDC#20057433481:3,000,000 an/a
B/Florida/07/0460 TCID50/mL1x
B/Malaysia/2506/04600 TCID50/mL10x
B/Hong Kong/5/72 ATCC VR-82360 TCID50/mL1x
B/Taiwan/2/62 ATCC VR-29560 TCID50/mL1x
B/Maryland/1/59 ATCC VR-296600 TCID50/mL10x
B/GL/1739/54 ATCC VR-10360 TCID50/mL1x
B/Allen/45 ATCC VR-1026,000 EID50/mLn/a
B/Lee/40ATCC VR-10160 TCID50/mL1x
B/Brigit RecombinantATCC VR-78660 TCID50/mL1x

a Strain was not re-cultured and titered. Dilutions of the original culture/titer from the CDC were used. The HA titer for B/Ohio/01/2005 was 128.

Table 22. Results of Inclusivity Testing for Parainfluenza Virus 3
------------------------------------------------------------------------------
TypeStrain or SourceConcentrationMultiple of LoD
Zeptometrix#0810016CF10 TCID50/mLX
3C-243500 TCID50/mLATCC VR-9350x
NIH 47885BEI NR-3233100 TCID50/mL1 0x

{21}------------------------------------------------

TypeStrain or SourceConcentrationMultiple of LoD
Zeptometrix#0810040ACF2 TCID50/mL1x
AA/A2ATCC VR-15402 TCID50/mL1x
A/LongATCC VR-262 TCID50/mL1x
B/9320ATCC VR-9552 TCID50/mL1x
BB/Wash18537/62ATCC VR-15802 TCID50/mL1x
B/WV/14617/85ATCC VR-14002 TCID50/mL1x

Table 23. Results of Inclusivity Testing for Respiratory Syncytial Virus

Analytical Specificity (Cross-reactivity and Exclusivity)

The potential for cross-reactivity between assays contained in the FilmArray RP system was evaluated by testing simulated NPS samples containing high concentrations of respiratory panel viruses (tens to thousands-fold higher than LoD). No cross-reactivity was observed at the concentrations listed in Error! Reference source not found.24.

Virus or BacteriumType / StrainTest ConcentrationMultiple of LoD
AdenovirusSerotype 1(Species C)1.00x105 TCIDsc/mL333 x
CoronavirusHKU1 - Type BClinical specimen2.78x109 copies/mL1,463 x
NL63NR-4705.67x103 TCID50/mL1,134 x
HumanMetapneumovirusType A1 - hMPV-16IA10-2003 A18.17x103 TCID50/mL4.085 x
Human Rhinovirus /EnterovirusEchovirus 63.40x106 TCID50/mL113 x
Rhinovirus Al5.67x103 TCID50/mL5,670 x
Influenza AA/Brisbane/59/071.00x105 TCID50/mL500 x
HINIA/New Caledonia/20/991.00x105 TCID50/mL500 x
A/PR/8/341.00x106 TCID50/mL5,000 x
A1/FM/1/474.70x103 TCID50/mL24 x
A/NWS/334.70x103 TCID50/mL24 x
A1/Denver/1/574.70x103 TCID50/mL24 x
A/Solomon Islands/3/20061.39x104 TCID50/mL70 x

Table 24. Results of Testing for Cross-Reactivity with FilmArray RP Analytes

{22}------------------------------------------------

Virus or BacteriumType / StrainTest ConcentrationMultiple of LoD
Influenza AH1N1-2009A/Weiss/434.70x103 TCID50/mL24 x
A/Mal/302/541.39x104 TCID50/mL70 x
A/SwineNY/03/20094.00x105 TCID50/mL4,000 x
Influenza AH3N2A/Wisconsin/67/20058.17x103 TCID50/mL1634 x
A/Victoria/3/754.70x103 TCID50/mL940 x
A/Port Chalmers/1/735.67x103 TCID50/mL1,134 x
A/Aichi/2/681.00x105 TCID50/mL20,000 x
A/Hong Kong/8/681.00x105 TCID50/mL20,000 x
A/Alice4.70x103 TCID50/mL940 x
A/MRC 28.17x103 TCID50/mL1,634 x
A/Brisbane/10/078.17x103 TCID50/mL1,634 x
Influenza BB/FL/04/061.67x104 TCID50/mL278 x
B/Lee/408.17x103 TCID50/mL136 x
B/Taiwan/2/625.03x104 TCID50/mL838 x
B/GL/1739/548.17x103 TCID50/mL136 x
B/Maryland/1/598.17x103 TCID50/mL136 x
B/Florida/07/041.00x105 TCID50/mL1,667 x
B/Malaysia/2506/045.67x103 TCID50/mL95 x
B/Allen/451.00x105 TCID50/mL1,667 x
B/HongKong/5/728.17x103 TCID50/mL136 x
Parainfluenza VirusType 31.00x105 TCID50/mL10,000 x
Respiratory SyncytialVirusA1.39x104 TCID50/mL6,950 x
B2.14x104 TCID50/mL10,700 x

The potential for the FilmArray RP system to cross-react with non-FilmArray RP organisms was evaluated by testing an exclusivity panel consisting of 45 bacteria, 13 viruses, and 1 fungus. These organisms were selected based on their relatedness to FilmArray RP analytes, clinical relevance (cause respiratory symptoms or represent nasopharyngeal flora), or high prevalence within the population (e.g. Herpes Simplex Virus). Negative sample matrix was spiked with bacteria or fungi at a concentration of 106 CFU/mL and viruses at a concentration between 104 - 105 TCID50/mL, or the highest concentration possible. The FilmArray RP system did not cross-react with the exclusivity panel organisms listed in Table 25. One measles virus strain was found to contain Adenovirus, which was detected by the FilmArray RP.

Idaho Technology Inc. 510(k) FilmArray Respiratory Panel System

{23}------------------------------------------------

BacteriaStrain / IsolateVirusesStrain / Isolate
Bordetella bronchisepticaclinical isolateBocavirusType 1 - Clinical specimen
Bordetella holmesiiF061Coronavirus 229EATCC VR-740
Bordetella parapertussisA747Coronavirus SARSZeptometrix - Nucleic Acid
Bordetella pertussisE431Coronavirus OC43ATCC VR-759
Bordetella pertussisA639Cytomegalovirus (CMV)AD-169 (VR-538)
Bordetella pertussisATCC 8467Epstein-Barr Virus(EBV)B95-8
Bordetella pertussisATCC 9797Herpes Simplex VirusType 1
Bordetella pertussisATCC 51445Parainfluenza Virus 1Zeptometrix # 0810014CF
Bordetella pertussisATCC BAA-589Parainfluenza Virus 2Zeptometrix # 0810015CF
Bordetella pertussisATCC 9340Parainfluenza Virus 4Zeptometrix #0810060CF
Bordetella pertussisATCC 10380Measles VirusEdmonston
Bordetella pertussisATCC BAA-1335Measles VirusZeptometrix # 0810025CFª
Chlamydia trachomatisD-UW3MumpsZeptometrix # 0810079CF
ChlamydophilapneumoniaeTW183FungiStrain / Isolate
CorynebacteriumdiptheriaeATCC14779Candida albicansZeptometrix #0801504
Escherichia coliO157:H7
Haemophilus influenzaeMinnA
Lactobacillus acidophilusType strain
Lactobacillus plantarum17-5
Legionella longbeacheaeLong Beach 4
Legionella micdadeiTatlock
Legionella pneumophiliaPhiladelphia
Moraxella catarrhalisNe 11 (type strain)
MycobacteriumtuberculosisH37Ra-1
Mycoplasma hominisATCC 23114
Mycoplasma genitaliumATCC 33530
Mycoplasma pneumoniaeM129
Mycoplasma pneumoniaeATCC 15531
Mycoplasma pneumoniaeATCC 15293
Mycoplasma pneumoniaeATCC 15377
Mycoplasma pneumoniaeATCC 15492
Mycoplasma pneumoniaeATCC 29085
Mycoplasma pneumoniaeATCC 29342
Mycoplasma pneumoniaeATCC 39505
Mycoplasma pneumoniaeATCC 49894

.

Table 25. Non-FilmArray RP Exclusivity Panel

{24}------------------------------------------------

Neisseria elongatatype strain
Neisseria gonorrhoeaeATCC 700825
Neisseria meningitidisM1027 (typestrain)
Pseudomonas aeruginosaZeptometrix#0801519
Staphylococcus aureusCOL
Staphylococcus epidermidisRP62A
Streptococcus pneumoniaetype 59
Streptococcus pyogenesZeptometrix#0801512
Streptococcus salivariusATCC 13419
Ureaplasma urealyticumATCC 27618

4This viral stock produced one false positive Adenovirus result. The false positive was confirmed to be caused by Adenovirus contamination of the viral stock and was not due to cross-reactivity between the Adenovirus assay and Measles virus.

Supplemental Analytical Exclusivity Testing for Influenza Strains of Avian Origin:

Additional analytical exclusivity testing was carried out with either live isolates or purified genomic RNA of avian host influenza A strains with the following results:

HostSubtypeIsolate / StrainTest ConcentrationaFilmArray Result
AvianH2N2A/Japan/305/573.3 ng RNAInfluenza A(no subtype detected)
H2N2Kilbourne F386.3 ng RNA
H2N2A/Korea/426/68 (HA, NA) x A/Puerto Rico/8/346.3 ng RNA
H5N1A/Vietnam/1203/2004 R-H5N/Ab
H5N2A/duck/Pennsylvania/10218/842.5 ng RNA
H5N3Kilbourne F181A/duck/Singapore/645/97247 ng RNA
H7N2A/NewYork/107/2003N/Ab
H7N3A/Mallard/Netherlands/12/2000N/Ab
H10N7A/chicken/Germany/N/4968 ng RNA

Table 26. Results of Exclusivity Testing of Virus or Nucleic Acid from Culture of Avian Influenza A

4 Purified and quantified RNA from avian influenza cultures was obtained from BEI Resources

6 Stock virus HA titre from CDC = 128. Twenty microliters of virus stock tested.

{25}------------------------------------------------

Precision (Reproducibility)

A multicenter reproducibility study was performed to determine between-site and overall reproducibility of the FilmArray RP system.

Reproducibility testing occurred at three test sites utilizing a panel of twelve simulated NPS specimens spiked with various combinations of live respiratory pathogens (analytes) at three different test levels (high negative (LoD/10), low positive (1X LoD), and medium positive (3X LoD)). On each testing day, two operators at each site tested two aliquots of specimens on two different FilmArray instruments (six specimens per operator per instrument per day). Every specimen was tested four times a day on five days at the three testing sites. for a total of 60 tests per analyte per concentration. A total of 26 lots of reagents and 20 FilmArray instruments were utilized in the reproducibility study. Summary results for each analyte are summarized below.

Results for each analyte are summarized below:

AdenovirusSpecies C Serotype 1#Positive#Negative% AgreementwithExpectedResulta95%CIMeanTm%CVTmObservedTm Range
Medium PositiveSite A20/200/20100%83.2% - 100%83.740.2483.33 - 84.26
(3X LoD)Site B20/200/20100%83.2% - 100%84.230.2183.95 - 84.60
Site C20/200/20100%83.2% - 100%83.760.2883.03 - 84.13
900 TCID50/mLAll Sites60/600/60100%94.0% - 100%83.930.3683.03 - 84.60
Low PositiveSite A20/200/20100%83.2% - 100%83.420.2883.01 - 83.98
(1X LoD)Site B20/200/20100%83.2% - 100%83.900.2683.54 - 84.30
Site C20/200/20100%83.2% - 100%83.620.3883.05 - 84.64
300 TCID50/mLAll Sites60/600/60100%94.0% - 100%83.640.3983.01 - 84.64
High NegativebSite A18/202/2090.0%68.3% - 98.8%83.330.2683.02 - 83.76
(LoD/10)Site B16/204/2080.0%56.3%-94.3%83.710.3082.93 - 84.29
Site C10/2010/2050.0%27.2%-72.8%83.260.3182.61 - 83.86
30 TCID50/mLAll Sites44/6016/6073.3%60.3% - 83.9%83.430.3882.61 - 84.29
NegativeSite A0/180180/180100.0%98.0% - 100%
Site B0/180180/180100.0%98.0% - 100%
Site C0/180180/180100.0%98.0% - 100%
All Sites0/540540/540100.0%99.3% - 100%

Summary of Positive Agreement, Negative Agreement, and Tm Reproducibility Testing of Single Assay Analytes

4 Expected results for the "Medium Positive", and the "High Negative" panel members are positive. Expected result for the "Negative" panel member is negative.

{26}------------------------------------------------

Coronavirus HKU1Type BClinical Specimen 6123#Positive#Negative%AgreementwithExpectedResult *95%CIMeanTm%CVTmObservedTm Range
Medium PositiveSite A20/200/20100%83.2% - 100%75.610.2375.37 - 76.02
(3X LoD)Site B20/200/20100%83.2% - 100%76.020.2675.58 - 76.33
Site C20/200/20100%83.2% - 100%75.490.3274.96 - 75.90
5.7 x 10 6 RNAcopies/mLAll Sites60/600/60100%94.0% - 100%75.690.4174.96 - 76.33
Low PositiveSite A20/200/20100%83.2% - 100%75.470.2274.96 - 75.79
(1X LoD)Site B20/200/20100%83.2% - 100%75.890.2075.59 - 76.12
Site C20/200/20100%83.2% - 100%75.350.3074.83 - 75.81
1.9 x 10 6 RNAcopies/mLAll Sites60/600/60100%94.0% - 100%75.550.4074.83 - 76.12
High NegativebSite A15/205/2075.0%50.9% - 91.3%75.510.2875.06 - 75.90
(LoD/10)Site B17/203/2085.0%62.1% - 96.8%75.850.2275.26 - 76.23
Site C19/201/2095.0%75.1% - 99.9%75.330.2274.98 - 75.59
1.9 x 105 RNAcopies/mLAll Sites51/609/6085.0%73.4% - 92.9%75.550.3874.98 - 76.23
Site A0/180180/180100%98.0% - 100%
Site B0/180180/180100%98.0% - 100%
NegativeSite C0/180180/180100%98.0% - 100%
All Sites0/540540/540100%99.3% - 100%

4 Expected results for the "Medium Positive", the "Low Positive", and the "High Negative" panel members are possive. Expected result for the "Negative" panel member is negative.

6 High negative samples are targeted to be positive 20-80% of the time.

Coronavirus NL63BEI Resources NR-470#Positive#Negative%AgreementwithExpectedResulta95%CIMeanTm%CVTmObservedTm Range
Medium PositiveSite A20/200/20100%83.2% - 100%80.210.3179.64 - 80.90
(3X LoD)Site B20/200/20100%83.2%-100%80.550.3379.99 - 81.03
Site C20/200/20100%83.2%-100%80.040.2579.67 - 80.62
15 TCID50/mLAll Sites60/600/60100%94.0% - 100%80.290.4079.64 - 81.03
Low PositiveSite A20/200/20100%83.2%-100%80.080.2579.57 - 80.42
(1X LoD)Site B20/200/20100%83.2%-100%80.400.2479.98 - 80.89
Site C20/200/20100%83.2%-100%79.880.2779.36 - 80.40
5 TCID50/mLAll Sites60/600/60100%94.0% - 100%80.120.3879.14 - 80.89
High NegativebSite A13/207/2065.0%40.8% - 84.6%80.080.3479.21 - 80.82
(LoD/10)Site B14/206/2070.0%45.7% - 88.1%80.360.3079.98 - 80.91
Site C10/2010/2050.0%27.2% - 72.8%79.910.2679.24 - 80.30
0.5 TCID50/mLAll Sites37/6023/6061.7%48.2% - 73.9%80.110.3979.21 - 80.91
NegativeSite A0/180180/180100%98.0% - 100%
Site B0/180180/180100%98.0% - 100%
Site C0/180180/180100%98.0% - 100%
All Sites0/540540/540100%99.3% - 100%

4 Expected results for the "Medium Positive", the "Low Positive", and the "High Negative" panel members are positive. Expected result for the "Negative" panel member is negative.

{27}------------------------------------------------

Human MetapneumovirushMPV-16 (A1)#Positive#Negative%AgreementwithExpectedResulta95%CIMeanTm%CVTmObservedTm Range
Medium Positive(3X LoD)Site A20/200/20100%83.2%-100%77.610.2377.06-77.90
Site B20/200/20100%83.2%-100%78.070.2877.67-78.61
Site C19/201/2095.0%75.1%-99.9%77.730.2177.45-78.11
6 TCID50/mLAll Sites59/601/6098.3%91.1%-100%77.820.3677.06-78.61
Low Positive(1X LoD)Site A20/200/20100%83.2%-100%77.390.2177.04-77.79
Site B20/200/20100%83.2%-100%77.880.2277.37-78.11
Site C20/200/20100%83.2%-100%77.600.2477.16-77.99
2 TCID50/mLAll Sites60/600/60100%94.0%-100%77.620.3577.04-78.11
High Negativeb(LoD/10)Site A17/202/2085.0%62.1%-96.8%77.340.2077.05-77.59
Site B19/206/2095.0%75.1%-99.9%77.760.2277.06-78.11
Site C12/204/2060.0%36.1%-80.9%77.370.2976.74-77.79
0.2 TCID50/mLAll Sites48/6012/6080.0%67.7%-89.2%77.500.3576.74-78.11
Site A0/180180/180100%98.0%-100%
Site B0/180180/180100%98.0%-100%
NegativeSite C0/180180/180100%98.0%-100%
All Sites0/540540/540100%99.3%-100%

ª Expected results for the "Medium Positive", the "Low Positive", and the "High Negative" panel members are positive. Expected result for the "Negative" panel member is negative.

b High negative samples are targeted to be positive 20-80% of the time.

Influenza BB/FL/04/06#Positive#Negative%AgreementwithExpectedResulta95%CIMeanTm%CVTmObservedTm Range
Medium Positive(3X LoD)Site A20/200/20100%83.2% - 100%80.470.2679.88 - 80.93
Site B20/200/20100%83.2%-100%80.880.3180.30 - 81.30
Site C20/200/20100%83.2%-100%80.360.3279.78 - 80.80
180 TCID50/mLAll Sites60/600/60100%94.0% - 100%80.560.4079.78 - 81.30
Low Positive(1X LoD)Site A20/200/20100%83.2%-100%80.440.2780.00 - 80.92
Site B20/200/20100%83.2%-100%80.790.2980.40 - 81.33
Site C20/200/20100%83.2% - 100%80.340.2279.77 - 80.81
60 TCID50/mLAll Sites60/600/60100%94.0% - 100%80.510.3779.77 - 81.33
High Negativeb(LoD/10)Site A12/208/2060.0%36.1% - 80.9%80.420.3279.84 - 80.90
Site B10/2010/2050.0%27.2% - 72.8%80.780.2580.40 - 81.17
Site C8/2012/2040.0%19.1% - 64.0%80.300.2179.79 - 80.69
6 TCID50/mLAll Sites30/6030/6050.0%36.8% - 63.2%80.500.3679.79 - 81.17
NegativeSite A0/180180/180100%98.0% - 100%
Site B0/180180/180100%98.0% - 100%
Site C0/180180/180100%98.0% - 100%
All Sites0/540540/540100%99.3% - 100%

ª Expected results for the "Medium Positive", the "Low Positive", and the "High Negative" pancl members are possive. Expected result for the "Negative" panel member is negative.

{28}------------------------------------------------

Parainfluenza Virus 3Zeptometrix #0810016CF#Positive#Negative%AgreementwithExpectedResult *95%CIMeanTm%CVTmObservedTm Range
Medium PositiveSite A20/200/20100%83.2% - 100%81.170.3680.71 - 81.86
(3X LoD)Site B20/200/20100%83.2% - 100%81.480.3581.03 - 81.89
Site C20/200/20100%83.2% - 100%80.940.2880.63 - 81.37
30 TCID50/mLAll Sites60/600/60100%94.0% - 100%81.220.4180.63 - 81.89
Low PositiveSite A20/200/20100%83.2%-100%80.970.3680.36 - 81.52
(1X LoD)Site B20/200/20100%83.2% - 100%81.350.2680.93 - 81.79
Site C17/203/2085.0%62.1% - 96.8%80.860.2880.10 - 81.21
10 TCID50/mLAll Sites57/603/6095.0%86.1% - 99.0%81.080.4080.10 - 81.79
High Negative®Site A10/2010/2050.0%27.2% - 72.8%80.990.2680.30 - 81.34
(LoD/10)Site B7/2013/2035.0%15.4% - 59.2%81.290.2880.61 - 81.77
Site C5/2015/2025.0%8.7% - 49.1%80.840.2480.41 - 81.24
1 TCID50/mLAll Sites22/6038/6036.7%24.6% - 50.1%81.050.3480.30 - 81.77
Site A0/180180/180100%98.0% - 100%
NegativeSite B0/180180/180100%98.0% - 100%
Site C0/180180/180100%98.0% - 100%
All Sites0/540540/540100%99.3% - 100%

4 Expected results for the "Medium Positive", the "Low Positive", and the "High Negative" panel members are positive. Expected result for the "Negative" panel member is negative.

b High negative samples are targeted to be positive 20-80% of the time.

Respiratory Syncytial VirusType A#Positive#Negative%AgreementwithExpectedResult a95%CIMeanTm%CVTmObservedTm Range
Medium Positive(3X LoD)Site A60/600/60100%94.0% - 100%80.440.3579.46 - 80.83
Site B60/600/60100%94.0% - 100%80.860.2580.41 - 81.56
Site C60/600/60100%94.0% - 100%80.390.3380.08 - 80.91
6 TCID50/mLAll Sites180/1800/180100%97.8% - 100%80.580.4079.46 - 81.56
Low Positive(1X LoD)Site A40/400/40100%91.2% - 100%79.820.5078.93 - 80.62
Site B40/400/40100%91.2% - 100%80.400.4679.47 - 81.03
Site C40/400/40100%91.2% - 100%80.130.4779.13 - 80.79
2 TCID50/mLAll Sites120/1200/120100%97.0% - 100%80.100.5778.93 - 81.03
High Negativeb(LoD/10)Site A18/202/2090.0%68.3% - 98.8%79.630.5078.72 - 80.72
Site B17/203/2085.0%62.1% - 96.8%80.120.5079.26 - 80.89
Site C11/209/2055.0%31.5% - 76.9%79.970.5778.83 - 80.84
0.2 TCID50/mLAll Sites46/6014/6076.7%64.0% - 86.6%79.900.5878.72 - 80.89
NegativeSite A0/120120/120100%97.0% - 100%
Site B0/120120/120100%97.0% - 100%
Site C0/120120/120100%97.0% - 100%
All Sites0/360360/360100%99.0% - 100%

ª Expected results for the "Medium Positive", the "Low Positive", and the "High Negative" panel members are positive. Expected result for the "Negative" panel member is negative.

{29}------------------------------------------------

Summary of Positive Agreement, Negative Agreement, and Tm Reproducibility Testing of Multi-Assay Analytes

EnterovirusEchovirus 6 (Species B)#Positive#Negative%Agreement withExpected Result *95%CI
Medium PositiveSite A20/200/20100%83.2% - 100%
(3X LoD)Site B20/200/20100%83.2% - 100%
Site C20/200/20100%83.2%-100%
90,000 TCID50/mLAll Sites60/600/60100%94.0% - 100%
Low PositiveSite A20/200/20100%83.2% - 100%
(1X LoD)Site B20/200/20100%83.2% - 100%
Site C20/200/20100%83.2%-100%
30,000 TCID50/mLAll Sites60/600/60100%94.0% - 100%
High NegativebSite A20/200/20100%83.2% - 100%
(LoD/10)Site B20/200/20100%83.2%-100%
Site C20/200/20100%83.2% - 100%
3,000 TCID50/mLAll Sites60/600/60100%94.0% - 100%
Site A0/6060/60100%94.0% - 100%
NegativeSite B0/6060/60100%94.0% - 100%
Site C0/6060/60100%94.0% - 100%
All Sites0/180180/180100%97.8% - 100%

Reproducibility Agreement Summary for Enterovirus (Human Rhinovirus/Enterovirus)

  • Expected results for the "Medium Positive", the "Low Positive", and the "High Negative" panel members are positive. Expected result for the "Negative" panel member is negative.

b High negative samples are targeted to be positive 20-80% of the time.

Reproducibility Tm Summary (by Assay) for Enterovirus

AssayEnterovirusEchovirus 6 (Species B)MeanTm% CVTmObservedTm Range
Medium PositiveSite A87.120.2486.79 - 87.73
3x LoDSite B87.510.3086.99 - 88.05
Site C86.980.3386.37 - 87.64
90,000 TCID50/mLAll Sites87.180.3986.37 - 88.05
Low PositiveSite A87.000.3386.17 - 87.64
1x LoDSite B87.360.2986.80 - 87.85
Entero 1Site C86.810.3586.15 - 87.41
30,000 TCID50/mLAll Sites87.050.4286.15 - 87.85
High NegativeSite A86.890.2986.06 - 87.40
0.1x LoDSite B87.340.3186.67 - 87.86
Site C86.670.2985.97 - 87.30
3,000 TCID50/mLAll Sites86.960.4485.97 - 87.86
Entero 2Medium PositiveSite A87.090.2886.68 - 87.73
3x LoDSite B87.470.3086.82 - 88.00
Site C86.930.3686.16 - 87.64
90,000 TCID50/mLAll Sites87.140.4186.16 - 88.00
Low PositiveSite A86.980.3086.28 - 87.53
1x LoDSite B87.340.2886.89 - 87.82
Site C86.770.3586.05 - 87.52
30,000 TCID50/mLAll Sites87.020.4186.05 - 87.82

Idaho Technology Inc. 510(k) FilmArray Respiratory Panel System

{30}------------------------------------------------

AssayEnterovirusEchovirus 6 (Species B)MeanTm% CVTmObservedTm Range
High Negative0.1x LoDSite A86.860.2986.17 - 87.54
Site B87.260.3586.59 - 87.94
Site C86.650.2786.27 - 87.20
3,000 TCID50/mLAll Sites86.920.4286.17 - 87.94
Medium Positive3x LoDSite A85.700.3185.21 - 86.18
Site B86.190.3085.35 - 86.77
Site C85.590.3484.91 - 86.19
90,000 TCID50/mLAll Sites85.810.4484.91 - 86.77
HRV4Low Positive1x LoDSite A85.450.2684.81 - 86.06
Site B85.870.2485.44 - 86.36
Site C85.320.4084.69 - 86.26
30,000 TCID50/mLAll Sites85.540.4084.69 - 86.36
High Negative0.1x LoDSite A85.370.2684.80 - 86.04
Site B85.820.2385.43 - 86.23
Site C85.220.2284.82 - 85.58
3,000 TCID50/mLAll Sites85.460.3984.80 - 86.23
Reproducibility Agreement Summary for Rhinovirus (Human Rhinovirus/Enterovirus)
Human RhinovirusA1#Positive#Negative%Agreement withExpected Resulta95%CI
Medium PositiveSite A60/600/60100%94.0% - 100%
(3X LoD)Site B60/600/60100%94.0% - 100%
Site C60/600/60100%94.0% - 100%
3 TCID50/mLAll Sites180/1800/180100%97.8% - 100%
Low PositiveSite A20/200/20100%83.2%-100%
(1X LoD)Site B20/200/20100%83.2%-100%
Site C20/200/20100%83.2%-100%
1 TCID50/mLAll Sites60/600/60100%94.0% - 100%
High NegativebSite A40/400/40100%91.2% - 100%
(LoD/10)Site B40/400/40100%91.2% - 100%
Site C32/408/4080.0%64.4% - 91.0%
0.1 TCID50/mLAll Sites112/1208/12093.3%87.3% - 97.1%
Site A0/6060/60100%94.0% - 100%
Site B0/6060/60100%94.0% - 100%
NegativeSite C0/6060/60100%94.0% - 100%
All Sites0/180180/180100%97.8% - 100%

" Expected results for the "Medium Positive", and the "High Negative" panel members are positive. Expected
result for the "Negative" panel member is negative.
" High negativ

^bHigh negative samples are targeted to be positive 20-80% of the time.

{31}------------------------------------------------

Reproducibility Tm Summary (by Assay) for Rhinovirus

AssayHuman Rhinovirus A1Mean Tm% CV TmObserved Tm Range
HRV1Medium Positive 3x LoDSite A83.790.4483.25 - 85.09
Site B84.060.3183.43 - 84.56
Site C83.680.2383.22 - 84.09
3 TCID50/mLAll Sites83.840.3883.22 - 85.09
HRV1Low Positive 1x LoDSite A83.710.2983.07 - 84.35
Site B84.070.3483.44 - 84.66
Site C83.930.3483.24 - 84.71
1 TCID50/mLAll Sites83.900.3783.07 - 84.71
HRV1High Negative 0.1x LoDSite A83.560.2883.02 - 84.26
Site B83.910.3483.22 - 84.52
Site C83.760.3183.04 - 84.48
0.1 TCID50/mLAll Sites83.740.3683.02 - 84.52
HRV2Medium Positive 3x LoDSite A83.330.4882.55 - 84.67
Site B83.650.2983.11 - 84.17
Site C83.300.2882.70 - 83.77
3 TCID50/mLAll Sites83.420.4182.55 - 84.67
HRV2Low Positive 1x LoDSite A83.280.3182.57 - 83.86
Site B83.660.3683.02 - 84.37
Site C83.520.3982.82 - 84.29
1 TCID50/mLAll Sites83.480.4082.57 - 84.37
HRV2High Negative 0.1x LoDSite A83.170.3482.42 - 83.88
Site B83.580.3682.80 - 84.31
Site C83.370.3382.51 - 83.92
0.1 TCID50/mLAll Sites83.370.4082.42 - 84.31
HRV3Medium Positive 3x LoDSite A82.730.5381.99 - 83.94
Site B83.250.3682.49 - 83.88
Site C82.890.4382.10 - 83.88
3 TCID50/mLAll Sites82.960.5181.99 - 83.94
HRV3Low Positive 1x LoDSite A82.670.4981.76 - 83.65
Site B83.240.4482.40 - 84.08
Site C83.070.4182.13 - 83.76
1 TCID50/mLAll Sites82.980.5481.76 - 84.08
HRV3High Negative 0.1x LoDSite A82.680.5081.78 - 83.67
Site B83.210.4682.18 - 84.37
Site C82.970.4281.85 - 83.64
0.1 TCID50/mLAll Sites82.960.5281.78 - 84.37
HRV4Medium Positive 3x LoDSite A83.810.3883.28 - 84.98
Site B84.140.3483.43 - 84.83
Site C83.800.2583.44 - 84.20
3 TCID50/mLAll Sites83.900.3783.28 - 84.98
HRV4Low Positive 1x LoDSite A83.790.3183.18 - 84.45
Site B84.080.3183.55 - 84.61
Site C84.040.3883.35 - 84.93
1 TCID50/mLAll Sites83.940.3783.18 - 84.93
HRV4High Negative 0.1x LoDSite A83.580.2683.11 - 84.05
Site B83.880.3183.34 - 84.49
Site C83.860.2783.24 - 84.58
0.1 TCID50/mLAll Sites83.760.3383.11 - 84.58

Idaho Technology Inc. 510(k)
FilmArray Respiratory Panel System

{32}------------------------------------------------

Influenza A H1N1A/Brisbane/59/07#Positive#Equivocal#Negative%Agreement withExpected Result a95%CI
Medium PositiveSite A20/200/200/20100%83.2%-100%
(3X LoD)Site B20/200/200/20100%83.2%-100%
Site C20/200/200/20100%83.2%-100%
600 TCID50/mLAll Sites60/600/600/60100%94.0% - 100%
Low PositiveSite A20/200/200/20100%83.2%-100%
(1X LoD)Site B19/201/200/2095.0%75.1% - 99.9%
Site C17/202/201/2085.0%62.1%-96.8%
200 TCID50/mLAll Sites56/603/601/6093%83.8% - 98.2%
High NegativebSite A20/200/200/20100%83.2%-100%
(LoD/10)Site B17/202/201/2085.0%62.1%-96.8%
Site C14/205/201/2070.0%45.7% - 88.1%
20 TCID50/mLAll Sites51/607/602/6085.0%73.4% - 92.9%
Site A0/1800/180180/180100%98.0% - 100%
NegativeSite B0/1800/180180/180100%98.0% - 100%
Site C0/1800/180180/180100%98.0% - 100%
All Sites0/5400/540540/540100%99.3% - 100%

Reproducibility Agreement Summary for Influenza A/H1

4 Expected results for the "Medium Positive", and the "High Negative" panel members are positive. Expected result for the "Negative" panel member is negative.

6 High negative samples are targeted to be positive 20-80% of the time.

Reproducibility Tm Summary (by Assay) for Influenza A/H1

AssayInfluenza A H1N1A/Brisbane/59/07MeanTm% CVTmObservedTm Range
FluApan1Moderate Positive3x LoDSite A84.670.2484.27-85.12
Site B85.100.2484.78-85.56
Site C84.640.3483.76-85.23
600 TCID50/mLAll Sites84.800.3783.76-85.56
FluApan1Low Positive1x LoDSite A84.570.2784.17-85.31
Site B85.010.2884.59-85.75
Site C84.680.2684.16-85.18
200 TCID50/mLAll Sites84.750.3484.16-85.75
FluApan1High Negative0.1x LoDSite A84.270.2683.85-84.81
Site B84.750.2384.29-85.32
Site C84.460.3583.89-85.48
20 TCID50/mLAll Sites84.480.3783.85-85.48
FluApan2Moderate Positive3x LoDSite A80.420.2579.78-80.63
Site B80.850.2780.39-81.26
Site C80.310.2879.89-80.72
600 TCID50/mLAll Sites80.520.3979.78-81.26
FluApan2Low Positive1x LoDSite A80.360.2379.99-80.73
Site B80.800.2180.42-81.15
Site C80.520.2280.19-80.89
200 TCID50/mLAll Sites80.570.3279.99-81.15
FluApan2High Negative0.1x LoDSite A79.910.3779.15-80.41
Site B80.490.3079.67-80.83
Site C80.100.3579.56-80.73
20 TCID50/mLAll Sites80.170.4579.15-80.83
FluAH1-panModerate Positive3x LoDSite A78.790.2578.31-79.25
Site B79.200.3178.30-79.57
Site C78.760.3977.79-79.25
600 TCID50/mLAll Sites78.910.4277.67-79.57

{33}------------------------------------------------

AssayInfluenza A H1N1A/Brisbane/59/07MeanTm% CVTmObservedTm Range
Low Positive1x LoDSite A78.770.2578.42 - 79.34
Site B79.200.2778.72 - 79.67
Site C78.800.2578.30 - 79.26
200 TCID50/mLAll Sites78.930.3678.30 - 79.67
High Negative0.1x LoDSite A77.650.3377.15 - 78.21
Site B78.180.4577.47 - 79.26
Site C77.930.4977.43 - 79.04
20 TCID50/mLAll Sites77.920.5177.15 - 79.26

Reproducibility Agreement Summary for Influenza A/2009 H1

Influenza A 2009 H1N1A/Swine NY/03/2009#Positive#Equivocal#Negative% Agreement withExpected Resulta95%CI
Medium Positive(3X LoD)Site A20/200/200/20100%83.2%-100%
Site B20/200/200/20100%83.2%-100%
Site C20/200/200/20100%83.2%-100%
300 TCID50/mLAll Sites60/600/600/60100%94.0%-100%
Low Positive(1X LoD)Site A20/200/200/20100%83.2%-100%
Site B20/200/200/20100%83.2%-100%
Site C20/200/200/20100%83.2%-100%
100 TCID50/mLAll Sites60/600/600/60100%94.0%-100%
High Negativeb(LoD/10)Site A20/200/200/20100%83.2%-100%
Site B19/201/200/2095.0%75.1%-99.9%
Site C20/200/200/20100%83.2%-100%
10 TCID50/mLAll Sites59/601/600/6098.3%91.1%-100%
NegativeSite A0/1800/180180/180100%98.0%-100%
Site B0/1800/180180/180100%98.0%-100%
Site C0/1800/180180/180100%98.0%-100%
All Sites0/5400/540540/540100%99.3%-100%

4 Expected results for the "Medium Positive", the "Low Positive", and the "High Negative" panel members are positive. Expected result for the "Negative" panel member is negative.

b High negative samples are targeted to be positive 20-80% of the time.

Reproducibility Tm Summary (by Assay) for Influenza A/2009 H1

AssayInfluenza A 2009 H1N1A/Swine NY/03/2009MeanTm% CVTmObservedTm Range
FluApan1Moderate Positive3x LoD Site A84.670.2484.27 - 85.12
Site B85.100.2484.78 - 85.56
Site C84.640.3483.76 - 85.23
300 TCID50/mL All Sites84.800.3783.76 - 85.56
Low Positive1x LoD Site A84.570.2784.17 - 85.31
Site B85.010.2884.59 - 85.75
Site C84.680.2684.16 - 85.18
100 TCID50/mL All Sites84.750.3484.16 - 85.75
High Negative0.1x LoD Site A84.270.2683.85 - 84.81
Site B84.750.2384.29 - 85.32
Site C84.460.3583.89 - 85.48
10 TCID50/mL All Sites84.480.3783.85 - 85.48
FluApan2Moderate Positive3x LoD Site A80.620.2080.31 - 80.83
Site B81.010.1780.70 - 81.36
Site C80.690.2380.19 - 81.02
300 TCID50/mL All Sites80.810.3280.19 - 81.36

{34}------------------------------------------------

Influenza A 2009 H1N1A/Swine NY/03/2009MeanTm% CVTmObservedTm Range
Assay
Low Positive1x LoDSite A80.390.2979.87 - 80.73
Site B80.910.2780.20 - 81.24
Site C80.300.1780.05 - 80.61
100 TCID50/mLAll Sites80.620.4479.87 - 81.24
High Negative0.1x LoDSite A80.430.2780.08 - 81.03
Site B80.720.2780.11 - 81.14
Site C80.410.3479.86 - 80.82
10 TCID50/mLAll Sites80.540.3479.86 - 81.14
Moderate Positive3x LoDSite A78.870.4078.20 - 79.56
Site B79.440.3578.94 - 79.99
Site C78.620.4377.92 - 79.44
300 TCID50/mLAll Sites78.970.5877.92 - 79.99
FluAH1-panLow Positive1x LoDSite A78.370.3077.89 - 79.24
Site B78.900.3778.21 - 79.76
Site C78.160.2577.83 - 78.69
100 TCID50/mLAll Sites78.470.5177.78 - 79.76
High Negative0.1x LoDSite A78.340.3777.90 - 79.37
Site B78.830.3777.99 - 79.68
Site C78.080.2777.68 - 78.51
10 TCID50/mLAll Sites78.400.5377.68 - 79.68
Moderate Positive3x LoDSite A78.730.2478.31 - 79.14
Site B79.200.2478.84 - 79.67
Site C78.540.3077.89 - 78.92
300 TCID50/mLAll Sites78.810.4477.89 - 79.67
FluAH1-2009Low Positive1x LoDSite A78.640.2678.10 - 79.14
Site B79.030.2578.53 - 79.48
Site C78.490.2478.12 - 78.82
100 TCID50/mLAll Sites78.710.3978.10 - 79.48
High Negative0.1x LoDSite A78.600.2877.90 - 79.04
Site B79.000.2378.52 - 79.36
Site C78.520.2878.10 - 78.93
10 TCID50/mLAll Sites78.700.3877.90 - 79.36

Reproducibility Agreement Summary for Influenza A/H3

Influenza A H3N2A/Wisconsin/67/2005#Positive#Equivocal#Negative%Agreement withExpected Result a95%CI
Medium Positive(3X LoD)Site A20/200/200/20100%83.2%-100%
Site B20/200/200/20100%83.2%-100%
Site C20/200/200/20100%83.2%-100%
15 TCID50/mLAll Sites60/600/600/60100%94.0%-100%
Low Positive(1X LoD)Site A20/200/200/20100%83.2%-100%
Site B20/200/200/20100%83.2%-100%
Site C20/200/200/20100%83.2%-100%
5 TCID50/mLAll Sites60/600/600/60100%94.0%-100%
High Negativeb(LoD/10)Site A3/2011/206/2015.0%3.2% - 37.9%
Site B4/2012/204/2020.0%5.7%-43.7%
Site C3/208/209/2015.0%3.2% - 37.9%
0.5 TCID50/mLAll Sites10/6031/6019/6016.7%8.3%-28.5%
Site A0/1800/180180/180100%98.0%-100%
Site B0/1800/180180/180100%98.0%-100%
NegativeSite C0/1800/180180/180100%98.0%-100%
All Sites0/5400/540540/540100%99.3%-100%

ª Expected results for the "Medium Positive", the "Low Positive", and the "High Negative" panel members are positive. Expected result for the "Negative" panel member is negative.

{35}------------------------------------------------

AssayInfluenza A H3N2A/Wisconsin/67/2005MeanTm% CVTmObservedTm Range
FluApan1Moderate Positive3x LoDSite A85.330.2684.79 - 85.73
Site B85.480.4284.91 - 86.27
Site C85.060.3884.50 - 85.40
15 TCID50/mLAll Sites85.360.3884.50 - 86.27
Low Positive1x LoDSite A84.950.4184.19 - 86.03
Site B85.310.3184.79 - 86.02
Site C84.830.2684.37 - 85.25
5 TCID50/mLAll Sites85.030.4184.19 - 86.03
High Negative0.1x LoDSite A84.910.3984.22 - 85.60
Site B85.240.3084.78 - 85.77
Site C84.860.2784.48 - 85.33
0.5 TCID50/mLAll Sites85.010.3784.22 - 85.77
FluApan2Moderate Positive3x LoDSite A79.750.2579.51 - 79.99
Site B79.940.3079.40 - 80.37
Site C79.600.2479.14 - 79.86
15 TCID50/mLAll Sites79.810.3379.14 - 80.37
Low Positive1x LoDSite A79.420.3179.00 - 80.30
Site B79.690.3579.14 - 80.29
Site C79.250.2078.82 - 79.59
5 TCID50/mLAll Sites79.450.3778.82 - 80.30
High Negative0.1x LoDSite A79.220.3578.64 - 79.75
Site B79.590.2979.05 - 79.99
Site C79.240.2078.84 - 79.54
0.5 TCID50/mLAll Sites79.350.3678.64 - 79.99
FluAH3Moderate Positive3x LoDSite A82.580.3581.87 - 83.01
Site B82.890.1982.60 - 83.11
Site C82.440.3381.98 - 82.81
15 TCID50/mLAll Sites82.620.3981.87 - 83.11
Low Positive1x LoDSite A82.320.2981.88 - 82.69
Site B82.690.3482.16 - 83.32
Site C82.210.2581.76 - 82.73
5 TCID50/mLAll Sites82.390.4181.67 - 83.32
High Negative0.1x LoDSite A82.280.3681.71 - 82.91
Site B82.610.2982.17 - 83.05
Site C82.200.2381.87 - 82.57
0.5 TCID50/mLAll Sites82.370.3681.71 - 83.05

Reproducibility Tm Summary (by Assay) for Influenza A/H3

{36}------------------------------------------------

Precision (Repeatability)

The repeatability of the FilmArray RP System results was evaluated by repeated testing of the same 12 specimens tested in the reproducibility study while minimizing as many sources of variability as possible. The in-house repeatabilty testing was performed at Site C over the course of 12 testing days for a total of 48 test results per specimen. On each day, all 12 specimens were tested 4 times by two operators on two FilmArray instruments.

Spiked OrganismModerate Positive(3x LoD)Low Positive(1x LoD)High Negative(0.1x LoD)
# Positive /Total% PositiveResults# Positive /Total% PositiveResults# Positive /Total% PositiveResults
Adenovirus48/48100.0%48/48100.0%34/4870.8%
Coronavirus HKU148/48100.0%48/48100.0%44/4891.7%
Coronavirus NL6348/48100.0%48/48100.0%27/4856.3%
Human Metapneumovirus47/4897.9%48/48100.0%32/4866.7%
Enterovirus48/48100.0%48/48100.0%48/48100%
Human Rhinovirus144/144100.0%47/4897.9%83/9686.5%
Influenza A/H148/48100.0%43/48a89.6%a39/48b81.3%b
Influenza A/2009 H1N148/48100.0%48/48100.0%47/48c97.9%c
Influenza A/H348/48100.0%48/48100.0%7/48d14.6%d
Influenza B48/48100.0%48/48100.0%25/4852.1%
Parainfluenza Virus 348/48100.0%41/4885.4%14/4829.2%
Respiratory Syncytial Virus144/144100.0%96/96100.0%32/4866.7%
% positive (all analytes)per test level767/76899.9%611/62497.9%432/62469.2%
Table 27. Summary of Positive Agreement Results for Repeatability Testing

4 The five (5) non-positive results for Influenza A/H1 at LoD include: (1) Influenza A/H1 equivocal and (3) Influenza A equivocal results.

6 The nine (9) non-positive results for Influenza A/H1 at 0.1 x LoD include: (1) Negative, (3) Influenza A (no subtype detected), (1) Influenza A/H1 equivocal, and (4) Influenza A equivocal results.

€ One (1) equivocal Influenza A/H1-2009 result at the 0.1 x LoD test level.

d The 42 non-positive results for Influenza A/H3 at 0.1 x LoD include: (18) Negative, (2) Influenza A (no subtype detected) (15) Influenza A H3 equivocal, and (7) Influenza A equivocal results.

{37}------------------------------------------------

Interference

Substances that could be present in NPS samples or introduced during sample handling were evaluated for their potential to interfere with assay performance. Four different organism mixes containing FilmArray RP analytes were spiked into a simulated NPS (sNPS) sample matrix (human epithelial cells in VTM) at 5 x their respective LoDs. The 5x LoD organism concentration was chosen to be near the analyte LoD but also to provide consistent results for sample-to-sample comparison. Each FilmArray RP analyte was tested in the presence of each potentially interfering substance listed in Table . None of the substances tested were found to compete or interfere with the control or analyte assays in the FilmArray RP.

Endogenous SubstancesCompetitive / Interfering Microorganisms
Human Blood (with Na Citrate) (1% v/v)Respiratory Syncytial Virus A $2.8 x 10^4$ TCID50/mL
Mucin (bovine submaxillary gland) (1% v/v)Human Rhinovirus $1.1 x 10^4$ TCID50/mL
Human Genomic DNA: 0.2 ng/µLInfluenza A 2009 H1N1 $1.0 x 10^5$ TCID50/mL
2 ng/µLStaphylococcus aureus $1.0 x 10^6$ CFU/mL
20 ng/µLNeisseria meningitides $1.0 x 10^6$ CFU/mL
Corynebacterium diptheriae $1.0 x 10^6$ CFU/mL
Exogenous Substances
Saline Nasal Spray with Preservatives (1% v/v)Analgesic ointment (1% w/v)
Nasal Decongestant Spray (Oxymetazoline HCl) (1%v/v)Petroleum Jelly (1% w/v)
Tobramycin (0.6 mg/mL)Smokeless Tobacco (1% w/v)
Mupirocin (2% w/v)
Technique Specific Substances
Laboratory Reagents:Viral Transport Media:Swabs:
Bleach (1%, 2%, 5% v/v)Remel M4Copan 168C (rayon / twisted aluminum shaft)
Disinfecting wipesRemel M4-RTCopan FloQ (flocked nylon / plastic shaft)
Ethanol (7% v/v)Remel M5Copan 175KS01 (polyester / aluminum shaft)
DNAzap (1% v/v)Remel M6Millipore 519CS01M (flocked nylon / plastic shaft)
RNaseOut (1% v/v)Copan UTM
Table 28. List of Potentially Interfering Substances Evaluated
------------------------------------------------------------------------

Evaluation of the FilmArray RP system was not performed using clinical NPS specimens obtained from individuals who had recently received the FluMist® nasal influenza vaccine (MedImmune). However, analytical testing was performed with simulated samples containing various concentrations of the 2009-2010 formulation of the vaccine material. The FilmArray RP assays react with the Influenza A H1, Influenza A H3 and Influenza B viral material contained in the vaccine (Table 29). No cross-reactivity was observed with other, non-influenza FilmArray RP assays.

{38}------------------------------------------------

FluMist®2009-2010(% v/v)AdenovirusCoronavirusHKUICoronavirusNL63HumanMetapneumovirusHuman Rhinovirus /EnterovirusInfluenza AInfluenza BParainfluenzavirus 3Respiratory SyncytialVirus
H1H1-2009H3
10%-----+-++--
1%-----+-++--
0.1%-----+-++--
0.01%-----+-++--
0.001%-----+-++--
0.0001%-------Equiva+--
0.00001%--------+--
0.000001%-----------

·

:

:

Table 29. Evaluation of FluMist® Nasal Vaccine as a Potentially Interfering Substance

ª Equivocal Influenza A/H3 result

.

{39}------------------------------------------------

Image /page/39/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. To the left of the bird, there is a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

Beth Lingenfelter, M.S.
Director, Regulated Products
Idaho Technology Inc.
390 Wakara Way
Salt Lake City, UT 84108

FEB 17 2011

Re:K103175
Trade/Device Name:FilmArray Respiratory Panel (RP) System
Regulation Number:21 CFR §866.3980
Regulation Name:Respiratory Viral Panel Multiplex Nucleic Acid Assay
Regulatory Class:Class II
Product Codes:OCC, OEM, OOU, OEP, OTG, NXD, OOI
Dated:February 2, 2011
Received:February 3, 2011

Dear Ms. Lingenfelter:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{40}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1.1.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Uwe Schif for

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{41}------------------------------------------------

Indications for Use

510(k) Number: K103175 Device Name: FilmArray Respiratory Panel (RP) System

Indications for Use:

The FilmArray Respiratory Panel (RP) is a multiplexed nucleic acid test intended for use with the FilmArray instrument for the simultaneous qualitative detection and identification of multiple respiratory viral nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals suspected of respiratory tract infections. The following virus types and subtypes are identified using the FilmArray RP: Adenovirus, Coronavirus HKU1, Coronavirus NL63, Human Metapneumovirus, Influenza A, Influenza A subtype H1, Influenza A subtype H3, Influenza A subtype 2009 H1, Influenza B, Parainfluenza virus 3, Rhinovirus/Enterovirus, and Respiratory Syncytial Virus. The detection and identification of specific viral nucleic acids from individuals exhibiting signs and symptoms of a respiratory infection aids in the diagnosis of respiratory infection if used in conjunction with other clinical and epidemiological information. Negative results do not preclude respiratory infection and should not be used as the sole basis for diagnosis, treatment or other management decisions. Positive results do not rule out bacterial infection or co-infection with other organisms. The agent detected may not be the definite cause of disease. The use of additional laboratory testing (e.g. bacterial and viral culture, immunofluorescence, and radiography) and clinical presentation must be taken into consideration in order to obtain the final diagnosis of respiratory infection.

Due to seasonal prevalence, performance characteristics for Influenza A/H1, Influenza A/H3, Influenza A/2009 H1, and Influenza B were established primarily with retrospective clinical specimens.

Due to the genetic similarity between human Rhinovirus and Enterovirus, the FilmArray RP cannot reliably differentiate them. A positive FilmArray RP Rhinovirus/Enterovirus result should be followed-up using an alternate method (e.g. cell culture or sequence analysis).

The FilmArray RP detects Adenovirus species C serotype 6 with reduced sensitivity. It is recommended that specimens found to be negative for Adenovirus after examination using FilmArray RP be confirmed by an alternate method (e.g. FDA cleared molecular test or cell culture).

Performance characteristics for influenza A were established when influenza A/2009 H1N1, A/H1, and A/H3 were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary. If infection with a novel influenza

{42}------------------------------------------------

A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Districting Officer

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) < 103175

§ 866.3980 Respiratory viral panel multiplex nucleic acid assay.

(a)
Identification. A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B;
(2) Influenza A subtype H1 and Influenza A subtype H3;
(3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B;
(4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus;
(5) Human Metapneumovirus;
(6) Rhinovirus; and
(7) Adenovirus.
(b)
Classification. Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.