(226 days)
The Mahe Fixation Plate and Screw System is indicated for fracture fixation and joint fusion in the pelvis, small bones and long bones. Examples include fractures of the tibia, fibula, femoral condyle, pelvis, acetabulum, metacarpals, metatarsals, clavicle, humerus, ulna, middle hand and middle foot bones; and treatment of the calcaneus. The system is indicated for use in adult patients. All implants are intended for single use only.
Mahe Fixation Plate and Screw System consists of various shape and sizes plates featuring compression and locking or non-locking holes, fullthreaded-cortical, short threaded-cancellous, locking or non-locking, cannulated, self-tapping or non-self-tapping screws, compression and dynamic screws, implantable K-wires, pins, washers, and appropriate instrumentation. Surgical instrumentation is provided to facilitate modification, insertion, and removal of implants. The plates and screws are fabricated from stainless steel and titanium. The system contains several modules based on the size of the device and application site such as fixation/reconstruction of small fragment bones, forefoot, mid-foot, rear-foot, ankle, or other bones appropriate for the size of the device. The plate implants are in following models available: Tubular Plates, Reconstruction Plates, Cloverleaf Plates, Calcaneus Plates, Anatomical Plates, Clavicula Hook Plates, Small Fragment, Locking Plates, and DHSP/DCSP Screw Plates. The thickness of the plates varies from 1.2mm to 10mm; the length varies from 26mm to 317mm; and the number of the holes varies from 2 to 22. The screw implants are in three different diameter sizes (2.7mm, 3.5mm, and 4.0mm) and the K-wire implants are in various lengths (from 70mm to 310mm) and diameter sizes (from 0.8 to 3mm) available. The system is sold non-sterile, the products have to be sterilized prior to use.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Standards Met) | Reported Device Performance (Summary) |
|---|---|
| ISO 5838 "Implants for surgery - Skeletal pins and wires" | The device demonstrates compliant performance with this standard. |
| ASTM F366-04 "Standard Specification for Fixation Pins and Wires" | The device demonstrates compliant performance with this standard. |
| ASTM F0543-07 "Standard Specification and Test Methods for Metallic Medical Bone Screws" | The device demonstrates compliant performance with this standard. |
| ASTM F 382-99 "Standard Specification and Test Method for Metallic Bone Plates" | The device demonstrates compliant performance with this standard. |
| Biocompatibility | Biocompatibility tests were performed, indicating no new safety concerns. |
| Mechanical Properties (compared to predicate devices) | Functional and mechanical testing demonstrates comparable mechanical properties to the predicate devices. |
| Safety & Effectiveness | The device is deemed safe and effective for its intended use, performing as well as or better than predicate devices, and uses well-researched materials. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample size used for the test set or the data provenance for the non-clinical performance data. It mentions that "All models that are covered by this 510(k) premarket notification have been on the market in Europe for many years with no device failures," suggesting real-world data from Europe, but this is not a formal "test set" in the context of a controlled study.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This information is not provided in the document. The data presented is primarily non-clinical "performance data" such as testing against industry standards, biocompatibility, and mechanical properties, rather than data requiring expert-established ground truth for a clinical test set.
4. Adjudication Method for the Test Set
This information is not applicable as the studies described are non-clinical performance data (e.g., mechanical testing against standards).
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. The document focuses on non-clinical performance data and a comparison of technological characteristics to predicate devices.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This is not applicable as the device is a medical implant (plates, screws, and pins), not an algorithm or AI system.
7. The Type of Ground Truth Used
For the non-clinical performance data, the "ground truth" is established by:
- Industry Standards: Compliance with established international and national standards (ISO 5838, ASTM F366-04, ASTM F0543-07, ASTM F 382-99).
- Predicate Device Performance: Direct comparison of functional and mechanical properties against legally marketed predicate devices.
- Biocompatibility Testing: Results of established biocompatibility tests.
There is no mention of clinical ground truth (e.g., pathology, outcomes data, expert consensus) in the context of a clinical study for this 510(k) summary.
8. The Sample Size for the Training Set
This is not applicable as the device is a medical implant, not an AI or algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established
This is not applicable as the device is a medical implant, not an AI or algorithm that requires a training set.
{0}------------------------------------------------
. (1/3)
hemedical
510(k) Summary K102845
510(k) Summary as required by section 807.92(c)
Submission Applicant: Mahe Medical GmbH Establishment Registration Number: 8010447 Phone: +49-7465-92750 Fax: +49-7465-927529 E-mail: ralf.rotter@mahe-med.de
Date: 5/11/2011
Application Correspondent/Contact: think ! Andrea Pecsi Schwarzwaldstraße 5 78532 Tuttlingen Germany Phone: +49-7462-924 051 Fax: +49-7462-924 128 E-mail: andrea@thinkworks.biz
Common name: Implant Plates/ Screws, Bone Plates/Screws, Bone Fixation System
Classification name:
21 CFR 888.3030 - Single/multiple component metallic bone fixation appliances and accessories.
Product Codes: HRS - Plate, Fixation, Bone HWC - Screw, Fixation, Bone
21 CFR 888.3040 - Smooth or threaded metallic bone fixation fastener
Product Code: HTY - Pin, Fixation, Smooth
Trade name: Mahe Fixation Plate and Screw System
Predicate Devices:
K083654 & K082527 - Zimmer® Universal Locking System: 2.7 mm and 3.5 mm Locking
Plates and Screws (Tivanium® Ti-6A1-4V Alloy, CP Grade Titanium) - Zimmer, Inc.
K091614 - OsteoMed Foot Plate and Screw Rigid Fixation System - OsteoMed L. P.
K083912 - Treu Bone Fixation Screws and Pins – Treu Instrumente GmbH
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for Mahe Medical. The text "mahemedical" is written in a lowercase, sans-serif font. To the right of the text is a stylized "M" logo, with a registered trademark symbol in the upper right corner.
510(k) Summary K102845
Description of the Device:
Mahe Fixation Plate and Screw System consists of various shape and sizes plates featuring compression and locking or non-locking holes, fullthreaded-cortical, short threaded-cancellous, locking or non-locking, cannulated, self-tapping or non-self-tapping screws, compression and dynamic screws, implantable K-wires, pins, washers, and appropriate instrumentation. Surgical instrumentation is provided to facilitate modification, insertion, and removal of implants.
The plates and screws are fabricated from stainless steel and titanium.
The system contains several modules based on the size of the device and application site such as fixation/reconstruction of small fragment bones, forefoot, mid-foot, rear-foot, ankle, or other bones appropriate for the size of the device. The plate implants are in following models available: Tubular Plates, Reconstruction Plates, Cloverleaf Plates, Calcaneus Plates, Anatomical Plates, Clavicula Hook Plates, Small Fragment, Locking Plates, and DHSP/DCSP Screw Plates. The thickness of the plates varies from 1.2mm to 10mm; the length varies from 26mm to 317mm; and the number of the holes varies from 2 to 22. The screw implants are in three different diameter sizes (2.7mm, 3.5mm, and 4.0mm) and the K-wire implants are in various lengths (from 70mm to 310mm) and diameter sizes (from 0.8 to 3mm) available.
The system is sold non-sterile, the products have to be sterilized prior to use.
Indications for Use:
The Mahe Fixation Plate and Screw System is indicated for fracture fixation and joint fusion in the pelvis, small bones and long bones. Examples include fractures of the tibia, fibula, femoral condyle, pelvis, acetabulum, metacarpals, metatarsals, clavicle, humerus, ulna, middle hand and middle foot bones; and treatment of the calcaneus. The system is indicated for use in adult patients. All implants are intended for single use only.
Technological characteristics compared to the Predicate Devices:
The features of the subject components are substantially equivalent to the predicate devices based on similarities in intended use, design, sizes and configurations. In addition, the classifications, manufacturing and sterilization methods of the predicate and subject components are identical. The Mahe Fixation Plate and Screw System and predicate devices are made from Titanium (alloy) and Stainless Steel. Functional and mechanical testing demonstrates the comparable mechanical properties of the Mahe Fixation Plate and Screw System to the predicate devices K083654 & K082527 Zimmer® Universal Locking System, K091614 OsteoMed Foot Plate and Screw Rigid Fixation System and K083912 Treu Bone Fixation Screws and Pins.
{2}------------------------------------------------
K102845 (3/3)
Image /page/2/Picture/1 description: The image contains the word "mahemedical" in a lowercase, italicized font. To the right of the word is a stylized logo that appears to be the letter "M". The logo is also in a bold, italicized font and has a small circle with an "R" inside of it in the upper right corner.
510(k) Summary K102845
The Mahe Fixation Plate and Screw System can be deemed substantially equivalent and safe and effective for its indicated use.
Non-clinical performance data:
Mahe certifies compliance with the requirements among others of following device relevant standards: ISO 5838 "Implants for surgery - Skeletal pins and wires", ASTM F366-04 "Standard Specification for Fixation Pins and Wires", ASTM F0543-07 "Standard Specification and Test Methods for Metallic Medical Bone Screws" and ASTM F 382-99 "Standard Specification and Test Method for Metallic Bone Plates". Moreover biocompatibility and mechanical tests have been performed on the Mahe Fixation Plate and Screw System.
Summary:
The presented data that was conducted on the Mahe Fixation Plate and Screw System shows in its results and in comparison that the products perform as well as or better than the predicate devices, safe and effective for their intended use and do not raise any questions regarding safety and effectiveness. All models that are covered by this 510(k) premarket notification have been on the market in Europe for many years with no device failures. The used materials are well researched and do not raise new questions regarding safety and effectiveness of the finished product.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
ﺳﺮ
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mahe Medical GmbH % Ms. Andrea Pecsi Schwarzwaldstraße 5 78532 Tuttlingen Germany
MAY 1 3 2011
Re: K102845
Trade/Device Name: Mahe Fixation Plate and Screw System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliance and accessories Regulatory Class: Class II Product Code: HRS, HWC, HTY Dated: April 8, 2011 Received: April 15, 2011
Dear Ms. Pecsi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Ms. Andrea Pecsi
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Aty B. R. h
for
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K102845(ili)
す。
Indications for Use Statement
510(k) Number (if known): K102845
Device Name: Mahe Fixation Plate and Screw System
Indications for Use:
The Mahe Fixation Plate and Screw System is indicated for fracture fixation and joint fusion in the pelvis, small bones and long bones. Examples include fractures of the tibia, fibula, femoral condyle, pelvis, acetabulum, metacarpals, metatarsals, clavicle, humerus, ulna, middle hand and middle foot bones; and treatment of the calcaneus. The system is indicated for use in adult patients. All implants are intended for single use only.
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number: K102845
§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.
(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.