(240 days)
The NucliSENS EasyQ® MRSA assay is a qualitative in vitro diagnostic test for the direct detection of methicillin-resistant Staphylococcus aureus (MRSA) from nasal swabs in patients at risk for nasal colonization. The NucliSENS EasyQ® MRSA assay is performed on the NucliSENS EasyQ® platform.
The test utilizes NASBA™ (nucleic acid sequence-based amplification) coupled with molecular beacons (sequence-specific fluorescent probes) to detect the presence of MRSA DNA.
The NucliSENS EasyQ® MRSA assay is used as a screening tool to aid in the prevention and control of MRSA infections in health care institutions.
NucliSENS EasyQ® MRSA is not intended to diagnose, quide or monitor treatment for MRSA infections, or provide results of susceptibility to methicillin. A negative result does not preclude MRSA nasal colonization. Concomitant cultures are necessary to recover organisms for epidemiological typing or for further susceptibility testing.
The NucliSENS EasyQ® MRSA assay is a qualitative in vitro diagnostic test for the direct detection of methicillin-resistant Staphylococcus aureus (MRSA) from nasal swabs. The test utilizes NASBA™ (nucleic acid sequence-based amplification) coupled with molecular beacons (sequence-specific fluorescent probes) to detect the presence of MRSA DNA. The assay is performed on the NucliSENS EasyQ® platform, which includes the NucliSENS EasyQ® Analyzer, NucliSENS EasyQ® Incubator II, NucliSENS EasyQ® Computer, Monitor, Printer, and NucliSENS EasyQ® Director Software 2.6. The assay involves manual specimen processing from a dry flocked nasal swab, followed by automated amplification and detection in a closed tube format. Result calculation is performed automatically via dedicated operator software based on the analysis of fluorescence signal curves. Qualitative output (MRSA positive, MRSA negative or invalid) is reported to users. The assay includes an inhibition control and run controls (blank and positive control). Specimen preparation controls are recommended but not provided in the kit.
Acceptance Criteria and Device Performance for NucliSENS EasyQ® MRSA Assay
1. Table of Acceptance Criteria and Reported Device Performance
The submission document primarily reports the performance metrics rather than explicitly stating pre-defined acceptance criteria with specific threshold values for each. However, based on the statistical results and the general context of medical device submissions, we can infer the implied acceptance criteria.
| Performance Metric | Implied/Inferred Acceptance Criterion | Reported Device Performance |
|---|---|---|
| Clinical Sensitivity (Overall) | High (e.g., >90%) | 95.8% (91.1-98.4% CI) |
| Clinical Specificity (Overall) | High (e.g., >90%) | 96.8% (95.5-97.7% CI) |
| Positive Predictive Value (PPV) (Overall) | Adequate, considering prevalence (e.g., >70% or >75% at typical prevalence) | 79.7% (72.9-85.4% CI) |
| Negative Predictive Value (NPV) (Overall) | Very High (e.g., >98%) | 99.4% (98.8-99.8% CI) |
| Clinical Sensitivity (Adults) | High (e.g., >90%) | 94.7% (88.8-98.0% CI) |
| Clinical Specificity (Adults) | High (e.g., >90%) | 96.5% (95.0-97.7% CI) |
| Clinical Sensitivity (Pediatrics) | High (e.g., >90%) | 100% (88.4-100% CI) |
| Clinical Specificity (Pediatrics) | High (e.g., >90%) | 97.3% (94.9-98.8% CI) |
| Initial Invalid Result Rate | Low (e.g., <5%) | 3.4% (42/1238) |
| Invalid Result Rate (after re-test) | Very Low | 1.1% (14/1238) |
| Analytical Sensitivity (LOD) | Must reliably detect MRSA at low bacterial concentrations. | 182-493 CFU/swab (varies by MREJ type and is effectively the acceptance criteria itself for analytical sensitivity) |
| Analytical Specificity | No false positives with related non-MRSA species or other nasal flora. | All strains (excluding one initial contamination) detected MRSA negative except two when tested with mecA positive clinical samples. |
| Analytical Reactivity (Inclusivity) | Cover a broad range of clinically relevant MRSA strains. | 95.3% of 193 tested MRSA strains accurately detected; 98.4% theoretical coverage of 319 database strains. |
| Precision (Within-Laboratory) | High agreement with expected outcome | 90.3% (High Negative), 98.6% (Low Positive), 100% (Moderate Positive) |
| Precision (Between-Laboratory) | High agreement with expected outcome across sites and lots | 93.5% (High Negative), 87.0% (Low Positive), 99.1% (Low Positive 2), 100% (Moderate Positive) |
| Carry-Over Contamination | No detectable cross-contamination | No cross-over contamination observed amongst 103 valid negative specimens. |
2. Sample Size Used for the Test Set and Data Provenance
-
Clinical Study Test Set:
- Total specimens collected: 1355 nasal specimens.
- Final evaluable specimens for analysis: 1238 specimens (117 excluded due to non-compliance with clinical protocol).
- Data Provenance:
- Country of Origin: US (7 geographically diverse institutions in the US).
- Retrospective or Prospective: Prospective investigational study.
-
Analytical Test Sets:
- Analytical Sensitivity (LOD): 24 replicates for each bacterial input for each of the seven MREJ types.
- Analytical Specificity: Fresh cultures from 100 phylogenetically related S. aureus and members of the nasal commensal flora (26 MRCoNS, 29 MSSCoNS, 25 MSSA, 20 other non-staphylococci).
- Analytical Reactivity: Fresh cultures from 193 MRSA strains.
- Precision (Within-laboratory): 3 replicates per run for three different bacterial inputs (High Negative, Low Positive, Moderate Positive) over 12 days.
- Precision (Between-laboratory): 3 replicates per run for four different bacterial inputs (High Negative, Low Positive, Low Positive 2, Moderate Positive) at 3 sites over 6 days.
- Carry-Over Contamination: 105 high MRSA positive and 105 MRSA negative samples (divided over 5 EasyQ runs).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
The ground truth for the clinical test set was established by a reference enriched culture method. The document does not specify the number or qualifications of experts involved in performing or interpreting these reference culture results. It only states that "suspect S. aureus colonies were confirmed with a Gram stain and latex agglutination test" and "Confirmed S. aureus colonies were tested for methicillin resistance using the cefoxitin disk test as described in CLSI M2A9 and CLSI M100 S17." This implies standard microbiological laboratory practices, but not explicitly "experts" in the sense of physicians or highly experienced scientists directly establishing ground truth beyond the protocol.
4. Adjudication Method for the Test Set
There is no mention of an adjudication method for the clinical test set. The comparison is directly between the NucliSENS EasyQ® MRSA assay results and the reference enriched culture method. Discrepancies are not stated to have undergone expert review or further adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not done. This device is an automated in vitro diagnostic test that provides a qualitative output (MRSA positive, MRSA negative, or invalid). Its performance is directly compared to a reference culture method, not to human readers' performance, either with or without AI assistance. Therefore, there is no effect size reported for human readers improving with AI vs. without AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, a standalone (algorithm only) performance study was done. The NucliSENS EasyQ® MRSA assay operates as an automated system, comprising reagents, instrumentation (EasyQ® Analyzer, Incubator), and dedicated software for result calculation. The "Performance Characteristics" section, including clinical performance, analytical sensitivity, specificity, reactivity, and precision, describes the performance of the device algorithm and system in isolation, without human interpretation or intervention in the final result determination from the EasyQ system beyond initial sample preparation.
7. The Type of Ground Truth Used
- Clinical Study Ground Truth: Reference enriched culture method. This method involved:
- Direct inoculation of a nasal swab into Trypticase Soy Broth (TSB) with 6.5% NaCl, incubated for 18-24 hours.
- Streaking the enrichment broth onto a Blood Agar Plate (BAP) and incubating for 24-48 hours.
- Confirming suspect S. aureus colonies with Gram stain and latex agglutination test.
- Testing confirmed S. aureus for methicillin resistance using the cefoxitin disk test (CLSI M2A9 and CLSI M100 S17).
- This is a labor-intensive, gold-standard microbiological method for MRSA detection.
8. The Sample Size for the Training Set
The document does not explicitly specify a sample size for a "training set" for the algorithm. This is common for older diagnostic devices, particularly PCR/NASBA-based assays, where the algorithm for calling positive/negative is typically rule-based on fluorescence curves and thresholding, rather than a machine learning model developed with a distinct training set. The "software design documentation" for the MRSA assay protocol is mentioned, indicating the development of the algorithm based on established principles rather than data-driven machine learning training.
9. How the Ground Truth for the Training Set Was Established
As there is no explicit mention of a training set or a machine learning algorithm that required one, the concept of establishing ground truth for a training set is not directly applicable in the terms usually associated with modern AI/ML device submissions. The assay relies on the well-understood biochemical principles of NASBA™ amplification and molecular beacon detection for targets (SCCmec cassette junction and mecA gene) that define MRSA. The "ground truth" for the development of the primers and probes would have been established through extensive genomic sequencing data ("Sequence alignments" section mentions 319 MRSA strains from public databases) and experimental validation in the laboratory to ensure accurate detection of target sequences.
{0}------------------------------------------------
NucliSENS Easy Q MRSA
MAY 2 0 2011
510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K102740
Name of device
NucliSENS EasyQ® MRSA Assay, including the following components:
Reagents
NucliSENS EasyQ® MRSA
Instrumentation and Software
NucliSENS EasyQ® MRSA assay protocol software included in the NucliSENS EasyQ® MRSA assay kit NucliSENS EasyQ® Director Software 2.6 NucliSENS EasyQ® Analyzer NucliSENS EasyQ® Incubator II
Classification
Device Class: Class 2
Assay Reagent System:
SYSTEM, NUCLEIC ACID AMPLIFICATION TEST, Common/Classification Name: DNA, METHICILLIN RESISTANT STAPHYLOCOCCUS AUREUS, DIRECT SPECIMEN
. Product code: NQX
Assay Instrumentation and Software System:
Common/Classification Name: INSTRUMENTATION FOR CLINICAL MULTIPLEX TEST SYSTEMS
Product code: OOI
Premarket Notification submitter:
| Company Name:Company Address: | bioMérieux, Inc.100 Rodolphe StreetDurham, NC 27712 |
|---|---|
| Contact:Telephone #:FAX #: | Jocelyn Jennings, Senior Manager, Regulatory Affairs919-620-2894919-620-2548 |
Preparation Date: September 20, 2010
{1}------------------------------------------------
NucliSENS Easy O MRSA
Image /page/1/Picture/1 description: The image shows the logo for bioMerieux. The logo consists of the word "BIOMERIEUX" in a blocky, sans-serif font. Above the text is a solid black circle with a thin line extending vertically upwards from the top of the circle and downwards through the middle of the text.
Device to which equivalence is claimed
The assay reagent system which is the subject of this submission is claimed equivalent to the BD GeneOhm test, which received 510(k) clearance from FDA as the IDI-MRSA™, #K033415, cleared on 18-Mar-2004.
The instrumentation and software in this submission are claimed equivalent to the instrumentation and software reviewed by FDA as the NucliSENS EasyQ Enterovirus v1.1, K093383, cleared on 6-July-2010.
Intended Use of the Device
The NucliSENS EasyQ® MRSA assay is a qualitative in vitro diagnostic test for the direct detection of methicillin-resistant Staphylococcus aureus (MRSA) from nasal swabs in patients at risk for nasal colonization. The NucliSENS EasyQ® MRSA assay is performed on the NucliSENS EasyQ® platform.
The test utilizes NASBA™ (nucleic acid sequence-based amplification) coupled with molecular beacons (sequence-specific fluorescent probes) to detect the presence of MRSA DNA.
The NucliSENS EasyQ® MRSA assay is used as a screening tool to aid in the prevention and control of MRSA infections in health care institutions.
The NucliSENS EasyQ® MRSA is not intended to diagnose, guide or monitor treatment for MRSA infections, or provide results of susceptibility to methicillin. A negative result does not preclude MRSA nasal colonization. Concomitant cultures are necessary to recover organisms for epidemiological typing or for further susceptibility testing.
Description of the Device
- A dry flocked swab, that is not pre-moistened, is used to collect a specimen from the 1. anterior nares. After collection the swab is transported to the laboratory in its container for specimen processing.
-
- The swab is introduced into the lysis tube and bacteria are recovered from the swab by rotating the swab manually.
- ತೆ. Bacteria in suspension in the lysis tube are concentrated by centrifugation.
- After carefully discarding the supernatant, the remaining bacterial pellet is subjected to র্বা mechanical lysis on a Genie 2 vortex machine. The lysate is subsequently heated at 95 °C for 2 minutes.
- Five microliters of the raw lysate are added to the NASBA™ amplification mixture that 5. contains the restriction enzyme, DNA sequence-specific primers and beacons.
- Next, the MRSA DNA is digested by the restriction enzyme and denatured in the EasyQ® રે. Incubator.
- The NASBA™ enzymes are then added for amplification of the DNA targets by NASBA™ 7. in the EasyQ® Analyzer.
-
- Amplification and detection are performed simultaneously in a dedicated fluorescence reader, the NucliSENS EasyQ® Analyzer.
- റ്റ് Amplification is detected by means of sequences-specific probes, which fluoresce upon binding of the probe to its target.
-
- The assay is run in a closed tube format minimizing any risk for amplicons contamination.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for bioMérieux. The logo consists of the word "BIOMÉRIEUX" in a sans-serif font, with a vertical line extending upwards from the middle of the word. At the top of the line is a black circle.
NucliSENS Easy O& MRSA
-
- The kit contains enough reagents for 48 reactions. Up to 44 specimens (plus 2 reguired NASBA™ controls and 2 recommended specimen processing controls) can be tested in a single run within 3 hours.
-
- Result calculation is performed automatically via dedicated operator software and is based on the analysis of fluorescence signal curves measured by the NucliSENS EasyQ® Analyzer. Qualitative output (MRSA positive, MRSA negative or invalid) is reported to users.
- An inhibition control (IC) is present in the reaction mix to detect inhibitory specimen. 13.
- The NucliSENS EasyQ® MRSA assay includes reagents for the run controls: two 14. NASBA™ controls (blank and positive control) that must be included in each run.
-
- Positive and neqative controls for specimen preparation are recommended for each EasyQ amplification run; these specimen controls are not provided in the kit. Strains suitable for use as specimen negative control (SNC) and the six specimen positive controls (SPC) are described. These strains allow the operator to perform quality control for all components of the test, including all primers and probes in the kit. The specimen positive controls may be rotated in series leading to one SPC and one SNC per run.
Reagents and Test Components
Assay Kit Components
| Component | Composition | Number of tubes and tubedescription | Caps Color code |
|---|---|---|---|
| LYS Tub(lysis tube) | Proclin solution with glassbeads* | 48 tubes1150 ul per tube | Uncolored captubes |
| REAGENT spheressachet(amplification/detectionreagents) | Lyophilized white sphere:primers, molecularprobesinhibition control:●non-infectious DNANASBA™ reagents●(nucleotides, DTT.KCI, MgCl2)carbohydrateLyophilized orange sphere:Restriction enzyme●Bovine SerumAlbumin**carbohydrate | 6 tubes in foil pack with silica geldesiccant2 spheres per tube (white andorange) | Blue cap tubes |
| ENZ I sphere sachet(enzyme reagents) | Lyophilized blue sphere :AMV-RT, RNase H.●T7 RNA polymeraseBovine Serum●Albumin** | 6 tubes in foil pack with silica geldesiccant1 sphere per tube (blue) | Red cap tubes |
| REAGENT dil | Reagent spheres diluent(Tris/HCI, KCI, DMSO,PVP ) | 2 tubes.600 ul per tube | Blue cap tubes |
Table 1: Components of NucliSENS EasyQ® MRSA
{3}------------------------------------------------
NucliSENS Easy Q* MRSA
Image /page/3/Picture/1 description: The image shows the logo for BIOMÉRIEUX. The logo consists of the company name in a stylized font, with a vertical line running through the middle of the name. Above the name is a solid black circle with a small line extending upwards from the top.
| Component | Composition | Number of tubes and tubedescription | Caps Color code |
|---|---|---|---|
| ENZdil | Enzyme sphere diluent (Sorbitol,Bovine Serum Albumin**, Proclin) | 2 tubes.350 μl per tube | Red cap tubes |
| NW | NASBATM water with proclin fordissolution of the CONTROL+sphere and for the NASBATM blank | 2 tubes.1500 μl per tube | White cap tubes |
| CONTROL+ spheresachet | NASBATM positive control DNA:lyophilized sphere containingnon infectious DNA plasmids | 4 tubes in foil pack with silica geldesiccant1 sphere per tube (white) | White cap tubes |
| TBSTRCPSTR | 8 tubes strip: 6 pieces8 cap strip: 12 pieces | Not applicable | |
| CD-ROM containingInstructions for use forNucliSENS EasyQ®MRSA assay, Assayprotocols, sample loginfile and worksheets | MRSA SPC US 1.0 for thespecimen positive control,MRSA SNC US 1.0 for thespecimen negative control,MRSA PC US 1.0 for the positivecontrol,MRSA BLK US 1.0 for the blankcontrol,MRSA US 1.0 for clinicalspecimen(s).Controls_MRSA_US_1.0.sl forthe sample login file.MRSA Worksheet SpecimenPreparation.MRSA Worksheet Amplificationand Detection.MRSA Worksheet Preparation ofSpecimen Controls |
(*) The volume of beads can vary which does not affect the test performance.
(**) Warnings and precautions: Bovine Serum Albumin is included in Enzyme diluents, Reagent sphere and Enzyme sphere. Certified knowledge of the origin and/or sanitary state of the animals does not totally guarantee the absence of transmissible pathogenic agents. It is therefore recommended that these products be treated as potentially infectious and handled observing the usual safety precautions (do not ingest or inhale).
Caution: Handle with care; treat as a potentially infectious material.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for bioMerieux. The logo consists of the word "BIOMERIEUX" in a sans-serif font, with a vertical line running through the middle of the word. Above the word is a black circle with a line extending upwards. The logo is simple and modern.
NucliSENS Easy Q& MRSA
| Product | Catalogue number |
|---|---|
| NucliSENS EasyQ® SystemIt includes the following:NucliSENS EasyQ® AnalyzerNucliSENS EasyQ® Incubator V2NucliSENS EasyQ® ComputerMonitorPrinterNucliSENS EasyQ® Director Software 2.6Mini MicrocentrifugeOperator Manuals | 4700016 |
Assay Components Sold Separately by bioMérieux
Additional Materials
| Product | Catalogue number |
|---|---|
| Micro tubes; 1.5 ml (500 x) | 200294 |
| 0.2 ml 8-tube strips (125 pcs) | 285048 |
| 0.2 ml 8-tube caps (125 pcs) | 285051 |
| EasyQ® Tube tray holder | 45685049 |
| EasyQ® Cap Installing Tool (Capping aid) | 45685050 |
| Pipetting aid | 45685207 |
Other
- . Sterile dry flocked swab with transport tube (bioMerieux reference 280101)
- Sterile water in ampoule compatible with densitometer ●
- 2 ml and 10 ml sterile tubes (for preparation of the specimen controls) ●
- Densitometer
- Standards for calibration of densitometer ●
- Vortex Genie 2 with MOBIO horizontal 24-positions adaptor only (bioMerieux reference ● for the MOBIO adapter 270677)
- . Vortex Genie 2 for mixing individual tubes
- Vortex with flathead dimpled adaptor
- Ice or cooling block for 1.5 ml tubes .
- . High speed centrifuge (capable of reaching speeds of up to 16 000 g) for 1.5 ml tubes
- . Calibrated monochannel micropipettes with variable volume: 5 to 1,000 ul delivery volume
- Sterile-packaged, disposable, aerosol resistant tips (with filter) ●
- Disposable gloves, powderless .
- . Tabletop centrifuge for 1.5 ml test tube
- Waste container with cap (e.g. 50 ml tubes or zip-bags)
- 95 °C dry bath (calibrated) for incubation of 1.5 ml tubes (range of calibration e.g.±1.7°C) ●
- . Timer
- Sterile 1.5 ml tubes ●
- Mucin from submaxillary gland (Mucin type I-S from Sigma reference M3895-1G)
{5}------------------------------------------------
- MRSA strains: MREJ type ii (ATCC 43300*), MREJ type iii (ATCC BAA-39*), MREJ type . iv (ATCC BAA-40*), MREJ type v (ATCC BAA-2096*), MREJ type vii (ATCC BAA-2095*) and MREJ type xii (ATCC BAA-2094*)
- MSSA strain (ATCC 29213*) ●
- TSA agar plate .
- . Blood agar plate
- Applicator .
- Strains can be purchased from ATCC or other vendors.
Assay Instruments and Software
The instrumentation and software for the NucliSENS EasyQ System, including the EasyQ Analyzer, EasyQ Incubator, and Director 2.5 software, received FDA clearance in June 2008 as part of the 510(k) submission for the NucliSENS EasyQ Enterovirus v1.0 test (submission K063261).
The instrumentation and software for the NucliSENS EasyQ System was re-submitted as K093383 in November 2009, to comply with an FDA request. In K093383, the EasyQ Incubator was replaced by the EasyQ Incubator II, and Director software 2.5 was replaced by Director software 2.6.
Instrumentation and software for the NucliSENS EasyQ MRSA test is largely unchanged from the system as reviewed in K093383.
Instrumentation: the use of the vortex adaptor plate is new for the EasyQ MRSA test and is documented in this submission (see BTL065234). No changes have been made to the EasyQ Incubator, to the PC (personal computer) or the PC monitor sold for use with the EasyQ Analyzer.
One change has been made to the EasyQ Analyzer. The Analyzer design includes the capacity for 8 filter pairs; however only 2 of the 8 positions were utilized in conjunction for the NucliSENS EasyQ Enterovirus v1.0 test (for detection of FAM and ROX beacons)'. For the NucliSENS EasyQ MRSA test, a third filter pair is used to enable detection of beacons labeled with Cy5.
Software: The Director 2.6 software is unmodified versus the version submitted in K093383. Note that Director software was already configured to enable use of up to 5 filter pairs, dependent on parameters incorporated in the assay protocol developed for each assay application
The assay protocol distributed with the EasyQ MRSA kit was developed using the same basic paradigm as for the Enterovirus v1.1 assay protocol in K093383. However whereas the Enterovirus v1.1 assay protocol was customized from the QL1 algorithm, the MRSA assay protocol was customized from a different base algorithm, T4. Full software design documentation to support the MRSA assay protocol is therefore included in this submission.
- 1 Per BTL036512 Product Design Requirements Labsystems Fluorescence Reader and related Software (Revision of Doc. 8730.99.5126 00), as submitted in K093383, the EasyQ Analyzer had 2 filtersets: PR23 Filters (Product Requirement 23, page 7) Current available filtersets: Fluorescein Ex 485(14±2) (half bandwidth) Em 538±(25±3) ROX Ex 584±(16±2) Em 620±(15±5) Blocking factor outside HBW 10-3 Filter specifications should enable to meet crosstalk & performance specification
- d Per BTL013130 PRD NucliSens EasyQ Director Software, NucliSens EasyQ Director IV (= Product Requirements Document, Director 2.5, as submitted in K093383), per requirement PRD1.1.4 on p. 10: "The software will offer support for assays that use more than two fluorescent labels up to 8 labels."
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for bioMerieux. The logo consists of the word "BIOMERIEUX" in a stylized font, with a vertical line running through the middle of the word. Above the word is a black circle with a small line extending upwards from the top.
NucliSENS Easy Q* MRSA
SUBSTANTIAL EQUIVALENCE INFORMATION
Predicate device name(s): BD GeneOhm MRSA Assay
Predicate device 510(k) number(s) K033415
Comparison with predicate
The EasyQ MRSA test is claimed substantially equivalent to the BD GeneOhm™ MRSA Assay (K033415).
| Similarities and Differences | |||
|---|---|---|---|
| ltem | EasyQ MRSA | Predicate | |
| Intended Use | Qualitative diagnostic screening forMRSA in nasal swab specimens | Same | |
| Specimen Type | Nasal swab | Same | |
| Technological Principles | Manual specimen processing withautomated amplification, detection,analysis and reporting of results | Same | |
| Amplification technology | NASBA (Nucleic Acid Sequence-Based Amplification) | PCR (Polymerase Chain Reaction) | |
| Detection technology | Fluorogenic target-specifichybridization | Same | |
| Instrument System | EasyQ® System | Cepheid SmartCycler® | |
| Specimen Processing | Manual processing in single 1.5-mLtube with glass beads using vortex andcentrifuge | Manual processing in two microtubesusing glass beads, vortex and centrifuge | |
| Analysis and reporting ofresults | Automated using diagnostic software(EasyQ Director and assay protocol)on the EasyQ® System | Automated using diagnostic software onthe Smart Cycler® | |
| DNA Target Sequence(s) | 1 - Sequence incorporating theinsertion site of StaphylococcalCassette Chromosome mec (SCCmec)at orfX junction, with 6 variant setsprimers and probes to target 7 MREJtypes i, ii, iii, iv, v, vii, xii (targetsequences for MREJ i and ii areidentical) | 1- Same | |
| 2 - mecA gene | 2- Not applicable | ||
| Probes | Molecular Beacons | Same | |
| Controls | Internal control in each test, plus 2external controls (positive and blank)supplied in kit, plus 2 specimencontrols (MRSA and MSSA)recommended for each run | One internal reagent control andexternal positive and negative controlsrequired per run |
Similarities and differences between the tests are outlined below:
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows a logo with the word "BIOMERIEUX" in a stylized font. Above the word is a symbol consisting of a black circle with a thin vertical line extending upwards from the top of the circle. The text is in all caps and appears to be a company logo.
Standard/Guidance Document Referenced (if applicable):
Not applicable
Test Principle:
Specimen Preparation:
Collection
` Trained personnel should perform the collection of swabs and care should be taken to obtain adequate specimen as outlined below:
Open the tube and remove the flocked swab from its empty tube.
Insert the flocked swab into the patient's nostril; rotate the swab 5 times to obtain a sample. Use swab dry without pre-moistening.
Insert the same swab into the second nostril and repeat the sampling.
Insert the swab back into its dry transport tube.
Label the plastic transport tube with patient identification.
Storage
Specimens once collected can be stored for a period:
- up to 4 days, both at room temperature or at 2-8 ℃ before testing .
- up to 1 day at room temperature followed by 3 days at 2-8 ℃ before testing.
Nucleic acid amplification and detection
DNA NASBA ™
In DNA NASBA™, the bacterial DNA is first digested with a restriction enzyme and then copies of RNA derived from a target DNA sequence are manifold multiplied by isothermal amplification by the combined activities of 3 enzymes: Avian Myeloblastosis Virus Reverse Transcriptase (AMV-RT). Bacteriophage T7 RNA Polymerase (T7 RNA Pol). and RNase H, along with both ribo- and deoxyribo-nucleotides (rNTPs and dNTPs) and the cofactors dithiothreitol (DTT), MgCl2, KCI and DMSO. This process is described below. The figure also shows the differences between NASBA™ targeted to DNA templates versus RNA templates.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for bioMérieux. The logo consists of a blue circle with a yellow line extending from the bottom. The text "bioMérieux" is written in a sans-serif font below the circle.
Image /page/8/Figure/2 description: The image shows the title of a figure. The title is "Figure 1. DNA vs RNA NASBA™". The title is written in a clear, sans-serif font and is centered on the image. The text is black and the background is white.
To start the DNA NASBA™ process, the restriction enzyme Hae III cuts the target DNA near the binding site of primer 1 (P1). Afterwards, the reaction mixture is heated to 95°C to denature the linearized DNA and to inactivate the restriction enzyme. Subsequently, the reaction mixture is cooled down to 41°C and the primer P1 anneals to the positive strand of the target DNA. This starts the "linear phase" of the NASBA™ reaction where double-stranded DNA amplicons with a T7 promoter tail are generated.
In the cyclic phase of the NASBA™ reaction, the T7 promoter tail on the cDNA target amplicon serves as a binding site for T7 RNA Pol. which then transcribes from the cDNA amplicon using rNTP's to generate a negative strand RNA copy. A single original positive strand target sequence (with T7 tail applied) will serve as a template for multiple RNA copies as the reaction proceeds.
Detection with Molecular Beacons
Molecular beacons are hairpin structure-forming single-stranded oligonucleotides labelled at one end with a fluorescent dye named the fluorophore and at the other end with a quencher. In the absence of complementary target amplicons, the hairpin structure closes, bringing the fluorophore and quencher into close proximity, and as a result no fluorescent signal is emitted. In the presence of target amplicons, the beacon hybridizes to its complementary target instead of forming an internal hairpin structure. Once the beacon opens, the quencher and fluorophore separate, quenching ceases and fluorescence is emitted.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for bioMerieux. The logo consists of the word "BIOMERIEUX" in a blocky, sans-serif font. Above the text is a stylized graphic that looks like a circle with a line extending upwards and downwards, creating a simple, abstract design.
Image /page/9/Figure/2 description: The image shows a diagram of a molecular beacon assay. Part A shows a molecular beacon with a fluorophore (F) and a quencher (Q). Part B shows the target RNA molecular beacon hybridization. Part C shows the target RNA molecular beacon hybrid with the fluorophore not quenched.
Figure 2: Principle of Molecular beacon detection
MRSA assay
The NucliSENS EasyQ® MRSA test utilizes NASBA™ (nucleic acid sequence-based amplification) coupled with molecular beacons (sequence-specific fluorescent probes) to detect the presence of MRSA DNA in nasal swabs from patients at risk for nasal colonization with MRSA. Amplification and detection occur simultaneously in the EasyQ Analyzer.
The NucliSENS EasyQ® MRSA is developed for amplification of two targets . simultaneously: the SCCmec cassette junction and the mecA gene.
False positive results have been reported in the literature with tests that detect the cassette junction only due to a phenomenon known as "mecA dropout" leading to methicillin-sensitive Staphylococcus aureus strains that lost the mecA gene but still harbour the cassette junction. Such strains are detected as MRSA positive by such tests. (See reference 1 in the Bibliography section).
The NucliSENS EasyQ® MRSA will report a MRSA positive result only if both . targets, the SCCmec cassette junction and the mecA gene, are detected, thus reducing the risk of false positive detection.
Primers and beacons in the kit are complementary to the SCCmec cassette junction sequences of MREJ types i-v, vii, and xii.
Explanation of the assay
A dry flocked swab, that is not pre-moistened, is used to collect a specimen from the anterior nares. After collection the swab is transported to the laboratory in its container for specimen processing.
- The swab is introduced into the lysis tube and bacteria are recovered from the swab by rotating the swab manually.
- . Bacteria in suspension in the lysis tube are concentrated by centrifugation.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for bioMerieux. The logo consists of the word "BIOMERIEUX" in a sans-serif font, with a vertical line extending from the top of the "E" in "MERIEUX" upwards. Above the word is a black circle with a thin vertical line extending upwards from the top of the circle.
After carefully discarding the supernatant, the remaining bacterial pellet is subjected to mechanical lysis on a Genie 2 vortex machine. The Ivsate is subsequently heated at 95 °C for 2 minutes.
Five microliters of the raw Ivsate are added to the NASBA™ amplification . mixture that contains the restriction enzyme. DNA sequence-specific primers and beacons.
Next. the MRSA DNA is digested by the restriction enzyme and denatured in . the EasyQ® Incubator.
The NASBA™ enzymes are then added for amplification of the DNA tarqets . bv NASBA™ in the EasyQ® analyzer.
Amplification and detection are performed simultaneously in a dedicated fluorescence reader, the NucliSENS EasyQ® Analyzer.
Amplification is detected by means of sequence-specific probes which . fluoresce upon binding of the probe to its target.
. The assay is run in a closed tube format minimizing any risk for amplicons contamination.
. The kit contains enough reagents for 48 reactions. Up to 44 specimens (plus 2 required NASBA™ controls and 2 recommended specimen processing controls) can be tested in a single run within 3 hours.
. Result calculation is performed automatically via dedicated operator software and is based on the analysis of fluorescence signal curves measured by the NucliSENS EasyQ® Analyzer. Qualitative output (MRSA positive, MRSA negative or invalid) is reported to users.
. An inhibition control (IC) is present in the reaction mix to detect inhibitory specimen.
The NucliSENS EasyQ® MRSA assav includes reagents for the run controls: . two NASBA™ controls (blank and positive control) that must be included in each run
. Positive and neqative controls for specimen preparation are recommended for each EasyQ® amplification run; these specimen controls are not provided in the kit. Strains suitable for use as specimen negative control (SNC) and the six specimen positive controls (SPC) are described in the package insert. These strains allow the operator to perform quality control for all components of the test, including all primers and probes in the kit. The specimen positive controls may be rotated in series with one SPC per run.
Performance Characteristics
Interfering substances
A study was conducted with potentially interfering substances encountered in nasal swabs. Substances tested were chemical substances that can either be naturally present in the nasal cavity or that can be artificially introduced into the nasal cavity.
Accordingly, the following substances were tested and evaluated: blood, mucin, azelastine HCI (Astelin). Eucalvptus globulus/ kalium bichromicus (Sino Fresh). phenylephrine HCI (Neo-synephrine), cromolyn sodium (NasalChrom), Zicum Gluconium (Zicam gel), oxymetazolinė HCl (Vicks Sinex), Triamcinolone acetonide (Nasacort), mometasone furoate (Nasonex).
The following drugs / biological substances have been shown to interfere with the performance of the assay:
- . Astelin (50 ul per swab)
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows a logo with the text "BIOMERIEUX" in a sans-serif font. Above the text is a circular shape with a line extending upwards from its center. The logo appears to be a stylized representation of a scientific or medical symbol.
- . Sino Fresh (50 ul per swab)
- Nasal chrom (50 ul per swab) .
- . Zicam (50 ul per swab)
- Blood (15 ul and 50 ul per swab).
In the context of the clinical trials, blood was reported on nasal swabs for 20/1238 (1,6%) specimens, one of which yielded an invalid result.
Biological interference
An excess of methicillin-sensitive Staphylococcus aureus or methicillin-resistant Staphylococcus epidermidis or methicillin-sensitive Staphylococcus epidermidis did not interfere with the detection of MRSA or with the reporting of MRSA negative swabs.
Analytical sensitivity: Limits of Detection (LOD) for strains of MRSA with different types of MREJ
The analytical sensitivity of the NucliSENS EasyQ® MRSA assay was determined using clinical nasal matrices and various strains of MRSA representing the seven types of MREJ detected by the test. Bacterial suspensions with a bacterial input ranging from 10 to 1000 CFU per swab were tested in 24 replicates of each input.
The limit of detection was defined as the input corresponding to the 95 % Hit Rate predicted by statistical analysis using a probit regression model.
The estimated type-specific LoDs were further confirmed with two batches of reagents. The claimed type-specific LoD summarized below correspond to the higher concentrations claimed from these two studies, i.e. either real input determined by plating and counting bacteria colony forming units in the confirmation study, or predicted concentrations by the probit regression model.
| MREJtype | SCCmectype | LOD(CFU/swab) |
|---|---|---|
| i | I | 410 |
| ii | II | 227 |
| iii | III | 493 |
| iv | III | 234 |
| V | IV | 200 |
| vii | III | 482 |
| xii | V | 182 |
LoD for seven MRSA types
Analytical specificity
Sequence alignments
Primers and probes were designed for optimal specificity by comparative nucleic acid sequence analysis of genetically related strains of Staphylococci and of unrelated species known to be present in nasal specimens.
Using this sequence analysis, the combination of MRSA primers and beacons was shown to be highly specific to MRSA strains.
Experimental data
Fresh cultures from 100 phylogenetically related S.aureus and members of the nasal commensal flora were tested at an input of 7.5X10E6 CFU/swab (26 strains of methicillin-resistant coagulase negative staphylococci, 29 strains of methicillin-
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image shows a logo with the word "BIOMERIEUX" in a stylized, sans-serif font. Above the text is a graphic element consisting of a solid black circle with a thin line extending upwards from it. A curved line extends downwards from the bottom of the letter "I" in "BIOMERIEUX", connecting to the circle above.
sensitive coaqulase-negative staphylococci. 25 strains of methicillin-sensitive coagulase-positive staphylococci including 20 strains of MSSA, 20 strains other than staphylococci belonging to the nasal commensal flora including 1 strain of Grampositive rods, 10 strains of Gram-negative rods, 7 strains of non staphylococci Grampositive cocci, 1 strain of Gram-negative cocci, 1 strain of yeast).
One strain of MSSA was initially detected positive for the cassette junction due to contamination. This MSSA strain was re-tested and found negative for the cassette junction.
All the strains were detected MRSA negative except two strains when tested with mecA positive clinical samples.
Analytical Reactivity
Sequence alignments
Genomic nucleic acid sequences of 319 strains of MRSA were found in public databases and were analyzed for inclusivity with the NucliSENS EasyQ® MRSA assay. Sequence analysis indicated that 314 strains were theoretically covered by the design of primers and beacons (98.4%).
Experimental data
Fresh cultures from 193 MRSA strains were tested at an input of 3 times the LOD. The 193 strains comprised:
- Strains of MRSA belonging to the following MREJ types: 21 MREJ type i, 119 . MREJ type ii, 16 MREJ type iii, 2 MREJ type iv, 1 MREJ type v, and 2 MREJ types vii. In addition, strains belonging to the new MREJ types ix, xiii, and xiv were also included.
- . Strains MRSA belonging to each of the SCCmec types I, II, III, IV, V and VI.
- Strains belonging to the worldwide most prevalent MRSA clones. .
- . Strains of MRSA belonging to the PFGE types USA 100, 200, 300, 400, 500, 600, 700, 800, 1000 and 1100.
- . Strains of MRSA characterized for MIC to oxacillin including both homogeneous highly resistant strains with high oxacillin MIC and heterogeneous resistant strains with lower oxacillin MIC (2-8 µg/ml) values.
- Strains of MRSA recently isolated in USA hospitals.
The NucliSENS EasyQ® MRSA assay accurately detected 184 of the 193 isolates of MRSA tested, 95.3%.
Precision within-laboratory and between-laboratory
The precision of the NucliSENS EasyQ MRSA assay was evaluated using artificial samples consisting of human nasal matrix spiked with MRSA strains at 3 different inputs. The precision study was determined with respect to the complete assay procedure, comprising all steps from the extraction of the specimen from the swab to the final result.
The panel of tested samples was comprised of three different bacterial inputs, tested in 3 replicates per run: "High Negative" samples corresponding to approximately 0.03 x LOD, "Low Positive" samples corresponding to approximately 1.00 x LOD and "Moderate Positive" samples corresponding to approximately 4.82 x LOD.
{13}------------------------------------------------
Image /page/13/Picture/1 description: The image shows the word "BIOMERIEUX" in all caps. Above the word is a black circle with a line extending from the top of the circle. The font of the word is sans-serif.
The precision was analyzed by determining for each bacterial input the percentage of correct NucliSENS EasyQ MRSA test results based on the expected outcome.
A "within-laboratory" precision study was performed at one internal site over 12 days. with 2 operators, using 2 instruments and 2 lots of reagents.
The overall percent agreement with the expected outcome was 90.3% for the high negative input. 98.6% for the low positive input and 100% for the moderate positive input in the table below. The table below also shows for the maxLambda mecA and maxLambda cassette junction values (internal measurements to determine the final assay result). the overall means, standard deviations, and coefficients of variation obtained for testing with the different bacterial inputs.
Within-Laboratory Variation
| Sample Input1 | % Agreement | 95% CI | MaxLambda mecA | MaxLambda cassette junction | ||||
|---|---|---|---|---|---|---|---|---|
| High Negative (HN) | 90.3% (65/72) | 81.0-96.0% | 2.669 | 0.0592 | 2.2 | 1.034 | 0.0723 | 7.0 |
| Low Positive (LP) | 98.6% (71/72) | 92.5-99.9% | 2.641 | 0.0782 | 3.0 | 1.578 | 0.2178 | 13.8 |
| Moderate Positive (MP) | 100% (72/72) | 95.0-100% | 2.680 | 0.0564 | 2.1 | 1.996 | 0.1498 | 7.5 |
| Total Agreement | 96.3% (208/216) | 92.8-98.4% |
approx. 0.03 x LOD 'HN =
LP= approx. 1.00 x LOD
MP= approx. 4.82 x LOD
A "between-laboratory" precision study was performed at 3 sites (1 internal site and 2 external sites in the US), over 6 days, with 2 operators at each site. 1 instrument at each site and 2 lots of reagents.
The overall percent agreement with the expected outcome was 93.5% for the high neqative input, 87.0% for the low positive input, and 100% for the moderate positive input in the table below. The table below also shows for the maxLambda mecA and maxLambda cassette junction values the overall means. standard deviations, and coefficients of variation obtained for testing with the different bacterial inputs.
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the logo for BIOMÉRIEUX. The logo consists of the company name in a stylized font, with a symbol above it. The symbol is a circle with a line extending upwards from the top.
NucliSENS Easy Q* MRSA
Between-Laboratory Variation
| SampleInput1 | Site 11 | Site 2 | Site 3 | TotalAgreement | MaxLambda mecA | MaxLambda cassette junction | ||||
|---|---|---|---|---|---|---|---|---|---|---|
| %Agreement | %Agreement | %Agreement | %Agreementand 95% CI | Mean | SD | %CV | Mean | SD | %CV | |
| HighNegative(HN) | 94.4%(34/36) | 100%(36/36) | 86.1%(31/36) | 93.5%(101/108)87.1 -97.4% | 2.650 | 0.0662 | 2.5 | 1.024 | 0.0558 | 5.4 |
| Low Positive(LP) | 85.7%(31/36) | 75.0%(27/36) | 100%(36/36) | 87.0%(94/108)79.2 -92.7% | 2.631 | 0.0837 | 3.2 | 1.347 | 0.2707 | 20.1 |
| Low Positive2 (LP2) | 100%(36/36) | 97.2%(35/36) | 100%(36/36) | 99.1%(107/108)94.9 -99.9% | 2.557 | 0.0995 | 3.9 | 1.735 | 0.2405 | 13.9 |
| ModeratePositive(MP) | 100%(36/36) | 100%(36/36) | 100%(36/36) | 100%(108/108)96.6 - 100% | 2.670 | 0.1119 | 4.2 | 1.816 | 0.2550 | 14.0 |
| TotalAgreement | 95.1%(137/144) | 93.1%(134/144) | 96.5%(139/144) | 94.9%(410/432)92.4 -96.8% |
'HN = approx. 0.03 x LoD
approx. 1.00 x LoD
approx. 2.00 x LoD LP= LP2=
MP= approx. 4.82 x LoD
Between-Lot Variation ( Between-Laboratory)
| Sample Input1 | Lot 1%Agreement | Lot 2%Agreement | TotalAgreement% Agreementand 95% CI | MaxLambda mecA | MaxLambdacassette junction | ||||
|---|---|---|---|---|---|---|---|---|---|
| High Negative(HN) | 94.4%(51/54) | 92.6%(50/54) | 93.5%(101/108)87.1 - 97.4% | 2.650 | 0.0662 | 2.5 | 1.024 | 0.0558 | 5.4 |
| Low Positive(LP) | 90.7%(49/54) | 83.3%(45/54) | 87.0%(94/108)79.2 - 92.7% | 2.631 | 0.0837 | 3.2 | 1.347 | 0.2707 | 20.1 |
| Low Positive2 (LP2) | 98.1%(53/54) | 100%(54/54) | 99.1%(107/108)94.9 - 99.9% | 2.557 | 0.0995 | 3.9 | 1.735 | 0.2405 | 13.9 |
| ModeratePositive (MP) | 100%(54/54) | 100%(54/54) | 100%(108/108)96.6 -100% | 2.670 | 0.1119 | 4.2 | 1.816 | 0.2550 | 14.0 |
| TotalAgreement | 95.8%(207/216) | 94.0%(203/216) | 94.9%(410/432)92.4 - 96.8% | ||||||
| 1HN = approx. 0.03 x LoD LP2= approx. 2.00 x LoD |
"HN = approx. 0.03 x LoD LP= approx. 1.00 x LoD approx. 2.00 x LoD approx. 4.82 x LoD
MP=
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for bioMerieux. The logo consists of the word "BIOMERIEUX" in a sans-serif font, with a vertical line running through the middle of the word. Above the word is a black circle with a line extending upwards from it. The logo is simple and modern.
| MaxLambda mecA | MaxLambdacassette junction | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Sample Input1 | Lot 1 | Lot 2 | TotalAgreement | Mean | SD | %CV | Mean | SD | %CV |
| %Agreement | %Agreement | % Agreementand 95% CI | |||||||
| High Negative(HN) | 91.7%(33/36) | 88.9%(32/36) | 90.3%(65/72)81.0 - 96.0% | 2.669 | 0.0592 | 2.2 | 1.034 | 0.0723 | 7.0 |
| Low Positive(LP) | 100%(36/36) | 97.2%(35/36) | 98.6%(71/72)92.5 - 99.9% | 2.641 | 0.0782 | 3.0 | 1.578 | 0.2178 | 13.8 |
| ModeratePositive (MP) | 100%(36/36) | 100%(36/36) | 100%(72/72)95.0 -100% | 2.680 | 0.0564 | 2.1 | 1.996 | 0.1498 | 7.5 |
| TotalAgreement | 97.2%(105/108) | 95.4%(103/108) | 96.3%(208/216)92.8 - 98.4% |
Between lot variation (Within -Laboratory)
HN = approx. 0.03 x LoD LP= approx. 1.00 x LoD
MP= approx. 4.82 x LoD
Carry-Over Contamination
To evaluate the risk of carry-over contamination for the NucliSENS EasyQ MRSA assay during the processes of sample preparation, amplification and detection, five MRSA NASBA runs were performed with alternating high MRSA positive (5x106 CFU/lysis tube) and MRSA negative samples (negative for both the mecA and the SCCmec cassette). Artificial samples were used consisting of lysis tubes with frozen matrix from clinical swabs, the positive artificial samples being spiked with a strain of MRSA at the concentration indicated above.
In total, 105 high positive and 105 MRSA negative samples were tested divided over 5 EasyQ runs.
The results of this carry-over contamination study showed that no cross-over contamination was observed amongst 103 valid negative specimens results surrounded by high positive specimens (105 MRSA negative samples were tested, but 2 were classified as invalid and thus discarded).
Thus, the workflow and design of the assay were shown to be robust against carry-over contamination.
Clinical Performance ·
Clinical Study Design
Performance characteristics of the NucliSENS EasyQ® MRSA assay were determined in a multicenter prospective investigational study employing 7 geographically diverse institutions in the US, eches propositive investigational study chiploying 7 geographically their species and 1980, 1998, 1998, 1998, 1998, 1998, 1998, 1998, 1998, 1998, 198 collection-only sites sent their specimens to one of the testing sites. The NucliSENS EasyQ MRSA assay was compared to the reference enriched culture, the most sensitive culture method. Subjects included patients from healthcare institutions at risk for colonization with MRSA and routine screening cultures. Each subject was enrolled in the study only once. Subjects taking antibiotics for eradication of MRSA colonization or for treatment of MRSA infection in the 7 days
{16}------------------------------------------------
Image /page/16/Picture/1 description: The image shows the logo for bioMerieux. The logo consists of the word "BIOMERIEUX" in a sans-serif font, with a stylized graphic above it. The graphic features a black circle with a line extending upwards and downwards, creating a simple, abstract design.
prior to specimen collection, patients that had contraindications to nasal swab collection or children under 2 years old were excluded from the trial.
Two (2) nasal swabs were prospectively collected from each patient; One (1) swab was tested
with the NucliSENS Easyand 1 swab with the reference culture method. For the enriched culture method, a swab was directly inoculated into an enrichment broth consisting of trypticase soy broth (TSB) with 6.5 % NaCl and incubated for 18-24 hours at 35 -37 ℃. The enrichment broth was then streaked onto a blood agar plate (BAP) and incubated for 24-48 hours at 35 -37 °C. The plate was then examined and suspect S. aureus colonies were confirmed with a Gram stain and latex agglutination test. Confirmed S. aureus colonies were tested for methicillin resistance using the cefoxitin disk test as described in CLSI M2A9 and CLSI M100 S17.
Performance (sensitivity and specificity) of the NucliSENS EasyQ® MRSA test was calculated relative to the reference culture results.
Overall Clinical Study Results: Agreement of NucliSENS EasyQ® Test with Reference Culture
Results from the NucliSENS EasyQ® MRSA assay compared with results from the reference culture method are presented in the tables below.
There were 144 final evaluable specimens that were positive for MRSA by the reference culture
method. There were 14' specimens that were invalid for NucliSENS EasyQ ® MRSA; 1 positive by reference culture and 13 were negative by reference culture.
NucliSENS EasyQ ® MRSA had a clinical sensitivity of 95.8%, a clinical specificity of 96.8%, a Positive Predictive Value (PPV) of 79.7% and a Negative Predictive Value (NPV) of 99.4%.
A total of 1355 nasal specimens were collected of which 1238 had reportable results and are included in the analysis. There were 117 specimens excluded due to non-compliance with the clinical protocol.
| Reference Culture* | |||||||
|---|---|---|---|---|---|---|---|
| Total | |||||||
| NucliSENSEasyQ® MRSA | 137 | રૂર્ણ | 172 | ||||
| 1046 | 1052 | ||||||
| Total | 143 | 1081 | 1224 |
Results Comparing NucliSENS EasyQ® MRSA with Reference Culture.
*mecA mediated oxacillin resistance using 30 µg cefoxitin disk
َ invalids not included in the calculation
Performance Parameters - NucliSENS EasyQ® MRSA versus Reference Culture
| ParameterEstimate1 | 95% ConfidenceInterval | |
|---|---|---|
| Sensitivity: EasyQ vs. Reference Culture (137/143) | 95.8% | 91.1-98.4% |
| Specificity: EasyQ vs. Reference Culture (1046/1081) | 96.8% | 95.5-97.7% |
| EasyQ Positive Predictive Value with Reference Culture (137/172) | 79.7% | 72.9-85.4% |
| EasyQ Negative Predictive Value with Reference Culture (1046/1052) | 99.4% | 98.8-99.8 % |
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the logo for bioMerieux. The logo consists of the word "BIOMERIEUX" in a blocky, sans-serif font. Above the text is a solid black circle with a thin line extending upwards and downwards from it.
NucliSENS Easy Q8 MRSA
Positive Predictive Value and Negative Predictive Value of the NucliSENS EasyQ MRSA Assay Compared to the Reference Culture Method and Prevalence of MRSA at Individual Sites.
| Site | MRSAPrevalence | PPVNucliSENSMRSA | 95% CI | NPVNucliSENS MRSA | 95% CI |
|---|---|---|---|---|---|
| 1 | 10.0% (20/200) | 67.9% (19/28) | 47.7-84.1% | 99.4% (168/169) | 96.8-99.9% |
| 2 | 9.1% (39/427) | 83.3% (35/42) | 68.6-93.0% | 99.0% (378/382) | 97.3-99.7% |
| 3 | 33.6% (50/149) | 85.7% (48/56) | 73.8-93.6% | 98.9% (90/91) | 94.0-99.9% |
| 4 | 10.3% (6/58) | 85.7% (6/7) | 42.1-99.6% | 100% (45/45) | 92.1-100% |
| 5 | 8.1% (24/296) | 77.4% (24/31) | 58.9-90.4% | 100% (265/265) | 98.6-100% |
| 6 | 6.5% (3/46) | 100% (3/3) | 29.2-100% | 100% (43/43) | 91.8-100% |
| 7 | 3.2% (2/62) | 40.0% (2/5) | 5.3-85.3% | 100% (57/57) | 93.7-100% |
| Total | 11.6% (144/1238) | 79.7% (137/172) | 72.9-85.4% | 99.4% (1046/1052) | 98.8-99.8% |
Clinical Study Results: Agreement of NucliSENS EasyQ® MRSA Test with Reference Culture for Adults
Of the 1238 nasal specimens that had reportable results, 877 were collected from adult patients.
Results from the NucliSENS EasyQ® MRSA assay with adult specimens compared wi from the reference culture method are presented in tables 13 and 14 below.
There were 114 final evaluable specimens that were positive for MRSA by the reference culture
method. There were 14' specimens that were invalid for NucliSENS EasyQ® MRSA; 1 positive by reference culture and 13 were negative by reference culture.
NucliSENS EasyQ® MRSA had a clinical sensitivity of 94.7% and a clinical specificity of 96.5%.
Results Comparing NucliSENS EasyQ® MRSA with Reference Culture for Adult Specimens
| Reference Culture* | Total | ||
|---|---|---|---|
| + | - | ||
| NucliSENS EasyQ® MRSA | 107 | 26 | 133 |
| - | 6 | 724 | 730 |
| Total | 113 | 750 | 863 |
- mecA mediated oxacillin resistance using 30µg cefoxitin disk
1 invalids not included in the calculation
{18}------------------------------------------------
Image /page/18/Picture/0 description: The image shows the logo for BioMerieux. The logo consists of the word "BIOMERIEUX" in all caps, with a vertical line running through the middle of the letters. Above the word is a black circle with a vertical line extending upwards.
| ParameterEstimate | 95% ConfidenceInterval | |
|---|---|---|
| Sensitivity: EasyQ vs.Reference Culture (107/113) | 94.7% | 88.8-98.0% |
| Specificity: EasyQ vs. Reference Culture (724/750) | 96.5% | 95.0-97.7% |
| EasyQ Positive Predictive Value with Reference Culture (107/133) | 80.5% | 72.7-86.8% |
| EasyQ Negative Predictive Value with Reference Culture (724/730) | 99.2% | 98.2-99.7% |
Performance Parameters - NucliSENS EasyQ® MRSA versus Reference Culture for Adult Specimens
Clinical Study Results: Agreement of NucliSENS EasyQ® MRSA Test with Reference Culture for Pediatrics
Pediatric samples included specimens from Child, Adolescent, and Transitional Adolescent patients. Of the 1238 nasal specimens that had reportable results, 361 were collected from pediatric patients. Results from the NucliSENS EasyQ® MRSA assay with pediatric specimens compared against results from the reference culture method are presented in the tables below. Age distribution is presented in the table below.
There were 30 final evaluable specimens that were positive for MRSA by the reference culture method. NucliSENS EasyQ® MRSA had a clinical sensitivity of 100% and a clinical specificity of 97.3%.
Results Comparing NucliSENS EasyQ® MRSA Assay with Reference Culture for Pediatric Specimens
| Reference Culture* | |||
|---|---|---|---|
| + | - | Total | |
| NucliSENSEasyQ® MRSA | 30 | 9 | 39 |
| 0 | 322 | 322 | |
| Total | 30 | 331 | 361 |
mecA mediated oxacillin resistance using 30μg cefoxitin disk
Performance Parameters - NucliSENS EasyQ® MRSA versus Reference Culture for Pediatric Specimens
| ParameterEstimate | 95% ConfidenceInterval | |
|---|---|---|
| Sensitivity: EasyQ vs. Reference Culture (30/30) | 100% | 88.4-100% |
| Specificity: EasyQ vs. Reference Culture (322/331) | 97.3% | 94.9-98.8% |
| EasyQ Positive Predictive Value with Reference Culture (30/39) | 76.9% | 60.7-88.9% |
| EasyQ Negative Predictive Value with Reference Culture (322/322) | 100% | 98.9-100% |
{19}------------------------------------------------
Image /page/19/Picture/0 description: The image shows the text "NucliSENS Easy Q¹ MRSA". The text is in a sans-serif font and is black. The background is white. The text is likely the name of a product or service.
Image /page/19/Picture/1 description: The image shows a logo with the word "BIOMERIEUX" in a stylized font. Above the word is a circular shape with a line extending upwards from it. The logo appears to be a company logo, possibly for a biotechnology or medical company.
Age Distribution of Patients
| Child (2years - <12) Adolescent (12vears) | Transitionalvears - < 18 years) vears - < 21 years) | Adolescents (18 Adults (≥ 21 years) | Total All Ages | ||
|---|---|---|---|---|---|
| 13% (158/1238) | 14% (178/1238) | 2% (25/1238) | 71% (877/1238) | 100% (1238/1238) | |
| MRSA Prevalence | 5.1% (8/158) | 10.7% (19/178) | 12.0% (3/25) | 13.0% (114/877) | 11.6% (144/1238) |
Clinical Study Results: Invalid Results
Testing of clinical specimens showed an initial invalid result rate of 3.4% (42/1238). There were 42 initial invalid results for NucliSENS EasyQ® MRSA from 114 valid runs of which 1.1% (14/1238) remained invalid after repeat testing from frozen lysate. The overall invalid run rate was 10.2% (13/127).
Proposed labeling
The proposed labeling is complete.
Conclusion
The information in this premarket notification is complete and supports a substantial equivalence decision.
Bibliography
1 J. Rupp et al. Be aware of the possibility of false positive result in single locus PCR assays for methicillin-resistant Staphylococcus aureus - J. Clin. Microbiol. – 2006, vol. 44, p. 2317.
{20}------------------------------------------------
Image /page/20/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, composed of three human profiles facing to the right. The symbol is stylized and simplified, giving it a modern and symbolic appearance.
Food and Druq Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
bioMérieux, Inc. c/o Jocelyn Jennings, R.A.C. Sr. Manager, Regulatory Affairs 100 Rodolphe Street Durham, NC 27712
MAY 2 0 2011
Re: K102740
Trade/Device Name: NucliSENS EasyQ® MRSA Assay Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial susceptibility test powder Regulatory Class: Class II Product Code: NQX, OOI Dated: May 10, 2011 Received: May 12, 2011
Dear Ms. Jennings:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce.prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket
{21}------------------------------------------------
Page 2 - Jocelyn Jennings
notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If vou desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{22}------------------------------------------------
INTENDED USE STATEMENT
510(k) Number (if known): K102740
Device Name: NucliSENS EasyQ® MRSA
Intended Use: The NucliSENS EasyQ® MRSA assay is a qualitative in vitro diagnostic test for the direct detection of methicillin-resistant Staphylococcus aureus (MRSA) from nasal swabs in patients at risk for nasal colonization. The NucliSENS EasyQ® MRSA assay is performed on the NucliSENS EasyQ® platform.
The test utilizes NASBA™ (nucleic acid sequence-based amplification) coupled with molecular beacons (sequence-specific fluorescent probes) to detect the presence of MRSA DNA.
The NucliSENS EasyQ® MRSA assay is used as a screening tool to aid in the prevention and control of MRSA infections in health care institutions.
NucliSENS EasyQ® MRSA is not intended to diagnose, quide or monitor treatment for MRSA infections, or provide results of susceptibility to methicillin. A negative result does not preclude MRSA nasal colonization. Concomitant cultures are necessary to recover organisms for epidemiological typing or for further susceptibility testing.
x Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety
(OIVD)
Laceddu M. Poole
Division Sign-Off
High 81°
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K102740
§ 866.1640 Antimicrobial susceptibility test powder.
(a)
Identification. An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.(b)
Classification. Class II (performance standards).