K Number
K102613
Date Cleared
2011-01-13

(125 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder free vinyl patient examination glove is non-sterile ,disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

powder free vinyl patient examination gloves, Clear(non-colored), that meets all of the requirements of ASTM standard D 5250-06.

AI/ML Overview

This is a Powder Free Vinyl Patient Examination Glove, Clear (non-colored).

1. Acceptance Criteria and Reported Device Performance:

CharacteristicsStandardAcceptance CriteriaDevice Performance:
DimensionASTM standard D 5250-06Meets D 5250-06 requirementsMeets
Physical PropertiesASTM standard D 5250-06Meets D 5250-06 requirementsMeets
Freedom from pinholes21 CFR 800.20Meets 21 CFR 800.20 requirementsMeets
Powder ResidualASTM standard D 5250-06, D6124-06< 2mg/glove< 2mg/glove
BiocompatibilityPrimary Skin Irritation in rabbitsNot a Primary Skin IrritationPasses (Not a Primary Skin Irritation)
Dermal sensitization in guinea pigNot a Dermal sensitizationPasses (Not a Dermal sensitization)

2. Sample size used for the test set and the data provenance:

The document does not specify exact sample sizes for each test in the test set. It refers to compliance with standards like ASTM D5250-06, ASTM D6124-06, 21 CFR 800.20, and ISO10993-10, which would inherently include testing procedures and sample size requirements within those standards.

The data provenance is not explicitly stated as retrospective or prospective, nor the country of origin. However, the manufacturer is HEBEI FAVOR PLASTIC CO., LTD from China.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

Not applicable. This device is a physical product (gloves) and its performance is evaluated against established physical and chemical standards, not through interpretation by medical experts.

4. Adjudication method for the test set:

Not applicable. The evaluation is based on objective measurements against engineering and medical standards, not on subjective interpretations requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device is a patient examination glove and does not involve AI or human "readers."

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This device is a patient examination glove and does not involve an algorithm.

7. The type of ground truth used:

The ground truth is established by adherence to recognized national and international standards for medical devices, specifically:

  • ASTM D5250-06 (Standard Specification for Vinyl Patient Examination Gloves)
  • ASTM D6124-06 (Standard Test Method for Residual Powder on Medical Gloves)
  • 21 CFR 800.20 (Pinhole, Tensile Strength, and Elongation Requirements for Surgeon's Gloves and Patient Examination Gloves)
  • ISO10993-10 (Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization)

8. The sample size for the training set:

Not applicable. This device is a physical product, not an AI/ML algorithm that requires a training set.

9. How the ground truth for the training set was established:

Not applicable. This device is a physical product, not an AI/ML algorithm.

{0}------------------------------------------------

Summary

pase lofz

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

"The assigned 510(k) number is: K 102613 ." (applicant leave blank)

JAN 1 3 2011

Premarket Notification [510(k)] Summary

[[a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :

Submitter's name :HEBEI FAVOR PLASTIC CO., LTD
Submitter's address :#79 SHENGLI ST. SHIJIAZHUANG, 050000, CHINA
Phone number :(86) 311-83985366
Fax number :(86) 311-83986358
Name of contact person:Mr. Wang Kai
Date the summary was prepared:September 8th, 2010

[{a)(2}]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known

Device Name:Powder Free Vinyl Patient Examination Gloves, Clear(non-colored)
Proprietary/Trade name:Powder Free Vinyl Patient Examination GlovesOther clients private labeling
Common Name:Patient examination glove
Classification Name:Patient examination glove
Device Classification:I
Regulation Number:21 CFR 880.6250
Panel:General Hospital (80)
Product Code:LYZ

[(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .

Class I* powder free vinyl patient examination gloves , Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.

Predicate device : FUGUAN (Brand) Powder-Free Vinyl Patient Examination Gloves, Shijiazhuang Fuguan Plastic Products Co., Ltd.. K032908 .

[(a)(4)] A description of the device

{1}------------------------------------------------

Device Description . powder free vinyl patient examination gloves, Clear(non-colored), that meets all of the requirements of ASTM standard D 5250-06.

[(a)(5)] The summary describes the intended use of the device

Device Intended Use: powder free vinyl-patient examination glove, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

[(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.

The powder tree vinyl patient examination gloves, non, sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

CharacteristicsStandardDevice performance:
DimensionASTM standard D 5250-06:Meets
Physical PropertiesASTM standard D 5250-06Meets
Freedom from pinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 5250-06Meets
and D6124-06<2mg/glove
BiocompatabilityPrimary Skin Irritation inrabbitsPasses
Not a Primary Skin Irritation
Dermal sensitization in theguinea pigPasses
Not a Dermal sensitization

[(b)(1) A brief discussion of the non-clinical testing data that was submitted, referenced or relied on to demonstrate that the Subject Device is safe and effective, and whose performance meets the requirements of its pre-defined acceptance criteria and intended uses.

Powder free vinyl patient examination gloves , Clear(non-colored) meet requirements per ASTM D5250-06, per ASTM D6124-06, per 21 CFR 800.20 and ISO10993-10. , it is safe and effective, and it's performance meets the requirements of its pre-defined acceptance criteria and intended uses.

[(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

[(b)(3)] The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performed as well or better than the legally marketed device identified in (a)(3).

It can be concluded that the Powder Free Vinyl Patient Examination Gloves , Clear(non-colored) meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL, meet labeling claims . It is as safe, as effective, and performed as well the legally marketed device identified in (a)(3).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the caduceus. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Hebei Favor Plastic Company, Limited C/O Mr. Chu Xiaoan Room 1606, Building 1 Jian Xiang Yuan, No. 209 Bei Si. Beijing China 100083

JAN 1 3 2011

Re: K102613

Trade/Device Name: Powder Free Vinyl Examination Gloves, Clear (Non-Colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: September 8, 2010 Received: September 10, 2010

Dear Mr. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Mr. Xiaoan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,
fous I. Dood

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

Applicant: HEBEI FAVOR PLASTIC CO., LTD

JAN 1 3 2011

510(k) Number (if known): *___________________________________________________________________________________________________________________________________________________

Device Name: Powder Free Vinyl Patient Examination Gloves, Clear(Non-colored)

Indications For Use:

Powder free vinyl patient examination glove is non-sterile ,disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

AND/OR Prescription Use . (Part 21 CFR 801 Subpart D) ·

Over-The-Counter Use _ × (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabeth D. Clamons-Will

Division Sian-Off Division of Anesthesiology, General Hospital Infection Control. Dental Devices

510/k) Number

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.