(42 days)
Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
powder free vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.
The provided document describes the safety and effectiveness information for Powder Free Vinyl Patient Examination Gloves. Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristics | Acceptance Criteria (Standard) | Reported Device Performance |
|---|---|---|
| Dimension | ASTM standard D 5250-06 | Meets |
| Physical Properties | ASTM standard D 5250-06 | Meets |
| Freedom from pinholes | 21 CFR 800.20 | Meets |
| Powder Residual | ASTM standard D 5250-06 and D6124-06 | <2mg/glove |
| Biocompatibility | Primary Skin Irritation in rabbits (ISO 10993-10) | Passes (Not a Primary Skin Irritation) |
| Dermal sensitization in the guinea pig (ISO 10993-10) | Passes (Not a Dermal sensitization) | |
| Waterleak test performance | ASTM standard or equivalent standards, and FDA requirements for waterleak test | Meets (as stated in the conclusion statement) |
Note: The document explicitly states the device "meets" or "passes" all defined acceptance criteria.
2. Sample Size Used for the Test Set and the Data Provenance
The document does not explicitly state the sample size used for the test set for each characteristic. It refers to compliance with standards like ASTM D 5250-06, ASTM D6124-06, 21 CFR 800.20, and ISO10993-10. These standards typically define sampling plans for testing.
Data Provenance: The testing was presumably conducted by the manufacturer, SHIJIAZHUANG HYLAND PLASTIC CO., LTD, which is located in China. The nature of the tests (e.g., physical property tests, biocompatibility tests) suggests these are prospective tests performed specifically for this submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This information is not applicable and not provided in the document. The device is a patient examination glove, and its performance is evaluated against engineering and biological standards, not through the subjective interpretation of medical images or case decisions by human experts. The "ground truth" for this type of device is defined by the objective measurements and pass/fail criteria of the specified standards.
4. Adjudication Method for the Test Set
Adjudication methods are not applicable for this type of device and testing. Performance is determined by objective measurements compared against pre-defined thresholds in established standards, not by expert consensus or dispute resolution.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done
No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for diagnostic or screening devices where human interpretation is involved. This device is a medical accessory (glove) where performance is measured by physical and biological properties.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not Applicable. This device is a physical product (glove), not a software algorithm. Therefore, the concept of "standalone algorithm performance" does not apply.
7. The Type of Ground Truth Used
The "ground truth" used is based on objective engineering standards and biological testing standards. Specifically:
- ASTM standard D 5250-06 for dimensions, physical properties, and powder residual.
- 21 CFR 800.20 for freedom from pinholes.
- ASTM D6124-06 for powder residual.
- ISO 10993-10 for biocompatibility (Primary Skin Irritation and Dermal sensitization).
- FDA requirements for waterleak test.
8. The Sample Size for the Training Set
Not Applicable. This device is a physical product, not an AI/ML algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not Applicable. As there is no training set for this device, there is no ground truth to be established for it.
{0}------------------------------------------------
Summary
"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."
"The assigned 510(k) number is: ." (applicant leave blank)
OCT 2 1 2010
Premarket Notification [510(k)] Summary
{{a)(1)}. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :
| Submitter's name : | SHIJIAZHUANG HYLAND PLASTIC CO., LTD |
|---|---|
| Submitter's address : | NANZUO TOWN, YUANSHI COUNTY, SHIJIAZHUANG,050000, CHINA |
| Phone number : | (86) 311-83986806 |
| Fax number : | (86) 311-83986358 |
| Name of contact person: | Ms. Li Xiaojing |
| Date the summary was prepared: | September 8th, 2010 |
((a)(2)}. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known
| Device Name: | Powder Free Vinyl Patient Examination Gloves, Clear(non-colored) |
|---|---|
| Proprietary/Trade name: | Powder Free Vinyl Patient Examination GlovesOther clients private labeling |
| Common Name: | Patient examination glove |
| Classification Name: | Patient examination glove |
| Device Classification: | I |
| Regulation Number: | 21 CFR 880.6250 |
| Panel: | General Hospital (80) |
| Product Code: | LYZ |
((a)(3)). An identification of the legally marketed device to which your firm is claiming substantial equivalence .
Class I* powder free vinyl patient examination gloves , Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.
Predicate device : FUGUAN (Brand) Powder-Free Vinyl Patient Examination Gloves, Shijiazhuang Fuguan Plastic Products Co., Ltd.. K032908 .
[(a)(4)] A description of the device
{1}------------------------------------------------
Device Description : powder free vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.
{(a)(5)] The summary describes the intended use of the device
Device Intended Use: powder free vinyl patient examination glove, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
[(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.
The powder free vinyl patient examination gloves, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.
| Characteristics | Standard | Device performance |
|---|---|---|
| Dimension | ASTM standard D 5250-06 | Meets |
| Physical Properties | ASTM standard D 5250-06 | Meets |
| Freedom from pinholes | 21 CFR 800.20 | Meets |
| Powder Residual | ASTM standard D 5250-06 | Meets |
| Biocompatability | and D6124-06Primary Skin Irritation in rabbits | <2mg/glovePassesNot a Primary Skin Irritation |
| Dermal sensitization in the guinea pig | PassesNot a Dermal sensitization |
[(b)(1)] A brief discussion of the non-clinical testing data that was submitted, referenced or relied on to demonstrate that the Subject Device is safe and effective, and whose performance meets the requirements of its pre-defined acceptance criteria and intended uses.
Powder free vinyl patient examination gloves , Clear(non-colored) meet requirements per ASTM D5250-06, per ASTM D6124-06, per 21 CFR 800.20 and 1SO10993-10. , it is safe and effective, and it's performance meets the requirements of its pre-defined acceptance criteria and intended uses.
[(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .
Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.
[(b)(3)] The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performed as well or better than the legally marketed device identified in (a)(3).
It can be concluded that the Powder Free Vinyl Patient Examination Gloves , Clear(non-colored) It can be concluded mat the Fowber 1100 And FDA requirements for waterleak test on meet the AST M standard or oquiration stars as effective, and performed as well the legally marketed device identified in (a)(3).
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all caps. The text is in a bold, sans-serif font. The text is black and the background is white.
Image /page/2/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Shijiazhuang Hyland Plastic Company, Limited C/O Mr. Chu Xiaoan Room 1606 Building 1. Jianxiang, Yuan No. 209 Bei Si Huan Zhong Rd. Hai Di Beijing China 100083
OCT 2 1 2010
Re: K102611
Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, Clear (Non-Colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: September 8, 2010 Received: September 9, 2010
Dear Mr. Xiaoan;
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2- Mr. Xiaoan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.goy/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{4}------------------------------------------------
INDICATIONS FOR USE
Applicant: SHIJIAZHUANG HYLAND PLASTIC CO., LTD
OCT 2 1 2010
510(k) Number (if known): *
Device Name: Powder Free Vinyl Patient Examination Gloves, Clear(Non-colored)
Indications For Use:
Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use × (21 CFR 801 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth S. Clavins-Will
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesiology, General Devices Division of Antosthosis - 3.
Infection Control, Dental Devices
510(k) Number: K102611
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.