The 4CIS® SOLAR Spine System and 4CIS® APOLLON Spine System are a pedicle screw system indicated for the treatment of severe Spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bon graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
In addition, the 4CIS® SOLAR Spine System and 4CIS® APOLLON Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative Spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor and failed previous fusion (pseudarthrosis).
The 4C15® SOLAR Spine System and 4CIS® APOLLON Spine System are a toploading multiple component, posterior spinal fixation system which consists of pedicle screws, rods, nuts, and a transverse (cross) linking mechanism.
The 4CIS® SOLAR Spine System and 4CIS® APOLLON Spine System will allow surgeons to build a spinal implant construct to stabilize and promote spinal fusion. 4CIS® SOLAR Spine System and 4CIS® APOLLON Spine System implant components are supplied non-sterile are single use and are fabricated from titanium alloy (Ti-6AI-4V ELI) that conforms to ASTM F136. Various sizes of these implants are available. Specialized instruments are available for the application and removal of the 4CIS® SOLAR Spine System and 4CIS® APOLLON Spine System.
The provided text describes a medical device, the "4CIS® SOLAR Spine System and 4CIS® APOLLON Spine System," which is a pedicle screw spinal fixation system. However, the document does not contain any information regarding acceptance criteria, device performance metrics, or a study that specifically addresses these points as typically expected for software or AI-based devices.
The core of the document is a 510(k) summary for a spinal implant system, focusing on its substantial equivalence to predicate devices based on mechanical testing. It does not involve AI or software performance metrics that would require the detailed study information you've requested.
Therefore, I cannot populate the requested sections because the information is not present in the provided text. The document discusses:
- Device Identification: Trade name, common name, classification.
- Substantially Equivalent Predicate Devices: Identification of similar devices already on the market.
- Device Description: What the device is made of and its components.
- Indications for Use: What the device is intended to treat and where it can be used in the spine.
- Performance Data and Test Method: This section explicitly states: "Mechanical testing has been carried out as listed in APPENDIX 10 which includes the static compression test and static torsion test as well as the dynamic compressive fatigue testing in accordance with ASTM F1717 demonstrates equivalence to the above predicate devices." This refers to physical, mechanical tests of the implant's strength and durability, not performance criteria for a diagnostic/AI tool.
The question's structure (acceptance criteria, sample size, ground truth, expert opinions, MRMC studies, standalone performance, training set) is typically relevant for evaluating AI/machine learning algorithms or diagnostic devices. Since this document describes a physical surgical implant, these types of details are not applicable as presented.
In conclusion, the input text does not contain the information requested regarding the acceptance criteria and the study that proves the device meets those criteria, as understood in the context of device performance metrics for AI/software/diagnostic tools.
{0}------------------------------------------------
510(k) SUMMARY
The following 510(k) summary is being submitted as required by 21 CFR 807.92(a):
Submission Information
| Contact: | Seayoung Ahn7612 Barnum Road, Bethesda, MD 20817 USA |
|---|---|
| Sponsor: | 34-6 Keumam-ri, Seotan-myeon, Pyeongtaek,Gyeonggi-do, 451-852, Republic of Korea |
| Date Prepared: | February 28, 2011 |
| Tel No: | +82-31-664-4101 x110 |
Device Identification
| Trade Name: | 4CIS® SOLAR Spine System and4CIS® APOLLON Spine System |
|---|---|
| Common Name: | Pedicle Screw Spinal Fixation System |
| Classification Name: | Spondylolisthesis Spinal Fixation Device System(MNH)per 21 CFR § 888.3070,Spinal Pedicle Screw(MNI) per 21 CFR § 888.3070 |
Substantially Equivalent Predicate Devices
The subject devices, 4CIS® SOLAR Spine System and 4CIS® APOLLON Spine System, are substantially equivalent in function, design, composition, material and intended used to: Global Spinal Fixation System(K001668) and OPTIMAT™, Spinal System(K031585).
Device Description
The 4C15® SOLAR Spine System and 4CIS® APOLLON Spine System are a toploading multiple component, posterior spinal fixation system which consists of pedicle screws, rods, nuts, and a transverse (cross) linking mechanism.
The 4CIS® SOLAR Spine System and 4CIS® APOLLON Spine System will allow surgeons to build a spinal implant construct to stabilize and promote spinal fusion. 4CIS® SOLAR Spine System and 4CIS® APOLLON Spine System implant components are supplied non-sterile are single use and are fabricated from titanium alloy (Ti-6AI-4V ELI) that conforms to ASTM F136. Various sizes of these
Solco Biomedical Co.,Ltd
{1}------------------------------------------------
implants are available. Specialized instruments are available for the application and removal of the 4CIS® SOLAR Spine System and 4CIS® APOLLON Spine System.
Indications for Use
The 4CIS® SOLAR Spine System and 4CIS® APOLLON Spine System is a pedicle screw system indicated for the treatment of severe Spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bon graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
In addition, the 4CIS® SOLAR Spine System and 4CIS® APOLLON Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative Spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor and failed previous fusion (pseudarthrosis).
Performance Data and Test Method
Mechanical testing has been carried out as listed in APPENDIX 10 which includes the static compression test and static torsion test as well as the dynamic compressive fatigue testing in accordance with ASTM F1717 demonstrates equivalence to the above predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUN 1 3 2011
Solco Biomedical Co., Ltd. % Solco USA, Inc. Mr. Saeyoung Ahn 7612 Barnum Road Bethesda, Maryland 20817
Re: K102458
Trade/Device Name: 4CIS® SOLAR Spine System and 4CIS® APOLLON Spine System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNH, MNI Dated: June 3, 2011 Received: June 7, 2011
Dear Mr. Ahn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
Page 2 - Mr. Saeyoung Ahn
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm
Sincerely yours,
Mark N. Melkerson
Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known):
Device Name: 4CIS® SOLAR Spine System and 4CIS® APOLLON Spine System
Indications for Use:
The 4CIS® SOLAR Spine System and 4CIS® APOLLON Spine System are a pedicle screw system indicated for the treatment of severe Spondylolisthesis (Grade 3 and 4) of the LS-St vertebra in skeletally mature patients receiving fusion by autogenous bone grafic having implants attached to the lumbar and sacral spine (L3 to sacrum) with removel of the implants after the attainment of a solid fusion.
In addition, the 4CIS® SOLAR Spine System and 4CIS® APOLLON Spine System are intended to provide immobilization and stabilization of spinal segments in sketally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative Spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous fusion (pseudarthrosis).
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Dixision Sizn-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K102458