(29 days)
The GMK® Total Knee System is designed for cemented use in total knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components.
This knee replacement system is indicated in the following cases:
- Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid arthritis or polyarthritis
- avascular necrosis of femoral condyle .
- post traumatic loss of joint configuration .
- primary implantation failure. .
The tibial augments are to be attached to the tibial baseplate with both the fixing cylinders and bone cement.
The modification to the original Medacta GMK® (Global Medacta Knee) Total Knee System is a line extension to include the GMK® Revision femoral components (STD and PS), extension stems with offset adaptors, distal and posterior femoral wedges and tibial wedges.
GMK® Revision femoral components are based on the design of the GMK® Primary femoral components in the original 510(k) submission K090988. The GMK® Revision differs from GMK® Primary by an internal box that allows attachment of the extension stem and the femoral pegs are replaced by two threaded holes to attach femoral wedges.
GMK® Revision femoral components are available in two versions, standard and posterior stabilized femurs, left and right from size 1 to 6. The femoral posterior wedges include sizes 1- 6 with a thickness of 5mm and 10mm while the femoral distal wedges are available in the same size range (1- 6), in thicknesses of 4mm, 8mm and 12mm. Distal and posterior wedge screws are offered in 4mm, 8mm, and 12mm. GMK® Revision femoral components work with the same Tibial baseplates, UC tibial inserts, posterior stabilized tibial inserts and patellas cleared under the original GMK® Total Knee System, K090988.
The offset adaptors are available in 3mm and the associated extension stems are offered in diameters of 11mm, 16mm, 19mm and 20mm with each being 65mm, 105mm or 150mm in length. The tibial wedges are available in size 1- 6 at a thickness of 5mm or 10mm.
The provided text describes a 510(k) submission for a medical device, the GMK® Total Knee System - Revision. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel effectiveness. Therefore, the "study that proves the device meets the acceptance criteria" in this context refers to the performance testing conducted to show that the revised device is as safe and effective as its predicate.
Here's an analysis of the provided information:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria (Identified from Performance Testing) | Reported Device Performance |
|---|---|
| Biocompatibility: Meet recognized standards and have a long history of acceptable biocompatibility. | "Potential risks due to new materials were resolved by using only materials that meet recognized standards and have a long history of acceptable biocompatibility in orthopedic applications." (Materials for the revised components (CoCrMo, Ti6-Al4-V, Stainless Steel, UHMWPE) are stated to be the same as or similar to those in the original submission or commonly used in orthopedic implants, conforming to relevant ISO standards.) |
| Mechanical Integrity (Interlock Strength/Endurance): Interlocking mechanisms could survive 10 million cycles at physiological loads. | "Interlocking mechanisms could survive 10 million cycles at physiological loads..." Documentation states testing was conducted "on the worst case component size and option/design." Specific tests cited: ASTM F1814-97a, ASTM F897-02:2007, ASTM F1800-04. |
| Mechanical Integrity (Torsional and Disassembly Forces): Acceptable torsional and disassembly forces. | "...and had acceptable torsional and disassembly forces." (Results are stated to have met this criterion.) Specific tests cited: ASTM F1814-97a, ASTM F897-02:2007, ASTM F1800-04. |
| Overall Performance (Substantial Equivalence): Performance of the GMK® Total Knee System- Revision is substantially equivalent to the predicate device. | "The testing met all acceptance criteria and verifies that performance of the GMK® Total Knee System- Revision is substantially equivalent to the predicate device." "The results from testing and analysis provided in this submission support the conclusion that the GMK® Total Knee System- Revision is substantially equivalent to its predicate device with respect to indications for use and technological characteristics." (This is the overarching conclusion of the 510(k) submission based on the various tests and comparisons.) |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: Not explicitly stated as a number of devices. The document mentions "worst case component size and option/design" for mechanical testing, implying a limited, targeted selection of components for physical evaluation. This is typical for mechanical testing of orthopedic implants, where physical samples are manufactured and subjected to rigorous testing.
- Data Provenance: The testing was conducted by or on behalf of MEDACTA International SA, based in Castel San Pietro, Switzerland. The tests were performed to FDA guidance and international standards (e.g., ASTM, ISO), which are globally recognized. The data is prospective in the sense that the tests were performed specifically for this 510(k) submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- This information is not applicable to this type of submission. For a 510(k) for an orthopedic implant, "ground truth" is established through engineering and material science principles, based on recognized standards (ASTM, ISO) and laboratory testing. There are no clinical images or diagnostic cases requiring expert review for ground truth in this submission.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- This information is not applicable as the evaluation of this device involves mechanical and material testing against predefined engineering standards, not human interpretation of clinical data. The "adjudication" is met by the test results falling within the accepted parameters specified by the standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- This information is not applicable as this is a physical orthopedic implant, not a diagnostic imaging device or an AI-powered system designed to assist human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- This information is not applicable as this is a physical orthopedic implant, not a software algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The "ground truth" for this device's performance testing is established through recognized engineering standards (ASTM, ISO) and the physical properties of the materials and design. The acceptance criteria for mechanical integrity (e.g., surviving 10 million cycles under physiological loads, acceptable torsional forces) are derived from these established standards and real-world physiological demands on implants.
8. The sample size for the training set
- This information is not applicable. There is no "training set" in the context of mechanical performance testing for a physical orthopedic implant. This term is relevant for AI/machine learning models or statistical models that learn from data.
9. How the ground truth for the training set was established
- This information is not applicable, as there is no training set for this type of device submission.
{0}------------------------------------------------
510(k) Summary
Manufacturer:
MEDACTA International SA Strada Regina CH6874 Castel San Pietro Switzerland Phone (+41) 91 696 60 60 FAX (+41) 91 696 60 66
SEP 2 4 2010
(
Natalie J. Kennel Contact Person: Principal Consultant, NJK & Associates, Inc. 13721 Via Tres Vista San Diego, CA 92129 Phone (858) 705-0350 Fax (858) 764-9739 Nkennel@njkconsulting.com
Date Prepared: September 24, 2010
DEVICE INFORMATION
| Trade/Proprietary Name: | GMK® Total Knee System- Revision |
|---|---|
| Common Name | Total Knee Prosthesis |
| Classification Name: | Knee joint patellofemorotibial metal/polymer/metal,semiconstrained cemented prosthesis, |
| Classification: | Class II, 21 CFR 888.3560 |
| Product Code: | JWH |
Predicate Device: K090988 GMK® Total Knee System (Medacta International), cleared July 10, 2009
Product Description:
The modification to the original Medacta GMK® (Global Medacta Knee) Total Knee System is a line extension to include the GMK® Revision femoral components (STD and PS), extension stems with offset adaptors, distal and posterior femoral wedges and tibial wedges.
GMK® Revision femoral components are based on the design of the GMK® Primary femoral components in the original 510(k) submission K090988. The GMK® Revision differs from GMK® Primary by an internal box that allows attachment of the extension stem and the femoral pegs are replaced by two threaded holes to attach femoral wedges.
GMK® Revision Total Knee System 510(k)
Page I of 4
{1}------------------------------------------------
GMK® Revision femoral components are available in two versions, standard and posterior stabilized femurs, left and right from size 1 to 6. The femoral posterior wedges include sizes 1- 6 with a thickness of 5mm and 10mm while the femoral distal wedges are available in the same size range (1- 6), in thicknesses of 4mm, 8mm and 12mm. Distal and posterior wedge screws are offered in 4mm, 8mm, and 12mm. GMK® Revision femoral components work with the same Tibial baseplates, UC tibial inserts, posterior stabilized tibial inserts and patellas cleared under the original GMK® Total Knee System, K090988.
The offset adaptors are available in 3mm and the associated extension stems are offered in diameters of 11mm, 16mm, 19mm and 20mm with each being 65mm, 105mm or 150mm in length. The tibial wedges are available in size 1- 6 at a thickness of 5mm or 10mm.
Indications for Use:
.
The GMK® Total Knee System is designed for cemented use in total knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components.
This knee replacement system is indicated in the following cases:
- Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid arthritis or polyarthritis
- avascular necrosis of femoral condyle .
- post traumatic loss of joint configuration .
- primary implantation failure. .
Tibial augments are to be attached to the tibial baseplate with both the fixing cvlinders and bone cement.
Comparison to Predicate Devices
The indications for use for the modified system remain the same as the original 510(k), K090988.
GMK® Revision femoral components are manufactured of Cobalt Chromium Molybdenum (CoCrMo) according to ISO 5832-4:1996, Implants for Surgery-Metallic materials - Part 4: Cobalt-Chromium-Molybdenum Casting Alloy, the same as the GMK® Primary femoral components in the original submission. They have an asymmetric patellar groove (left and right) with different radii of curvature in coronal plane as well as non-parallel anterior and posterior cuts (wedge shape). The GMK® Revision posterior stabilized femoral components are design for use without cruciate ligaments when additional stability is required to prevent subluxation of the femur to the tibia in flexion, same as the GMK® Primary femoral components.
GMK® Revision femoral components are available in two versions, standard and posterior stabilized femurs, left and right from size 1 to 6. The sizes are the
{2}------------------------------------------------
same as GMK® Primary femoral components. The external shape of the GMK® Revision femoral component is identical to that of GMK® Primary. The GMK® Revision differs from GMK® Primary by an internal box that allows attachment of the extension stem and the femoral pegs are replaced by two threaded holes to attach femoral wedges.
GMK® Revision femoral components work with the same Tibial baseplates, UC tibial inserts, posterior stabilized tibial inserts and patellas cleared under the original GMK® Total Knee System, K090988. However, the GMK® Revision femoral components are not used with the standard tibial inserts. These system compatibilities are the same as the original devices, GMK® Total Knee System.
GMK® Revision femoral components are designed to add posterior and/or distal wedges as well as the 3mm offset with extension stems. In addition, tibial wedges, 3mm and 5mm offset with extension stems are designed for use with the same tibial baseplate cleared under the original GMK® Total Knee System, K090988. These additional system compatibilities are new to the original device however there are no new issues of safety and effectiveness.
GMK® Revision offset adaptors, extension stems, wedge screws and tibial wedges are manufactured of titanium alloy (Ti6-Al4-V) according to ISO 5832-3:1996, Implants for Surgery - Metallic materials - Part 3: Wrought titanium 6aluminum 4-vanadium alloy, the same as the extension stem in the original submission. The offset adaptors are available in 3mm and 5mm and the associated extension stems are offered in diameters of 11mm, 13mm, 16mm, 19mm and 20mm with each being 65mm, 105mm or 150mm in length. This is similar to the extension stem (Ø11mm, 64mm) in the original submission. Distal and posterior wedge screws are offered in 4mm, 8mm, and 12mm while the tibial wedges available in size 1- 6 at a thickness of 5mm or 10mm.
GMK® Revision femoral posterior and distal wedges are manufactured from. Stainless Steel according to ISO 5832-9:2007 Implants for surgery -- Metallic materials -- Part 9: Wrought high nitrogen stainless steel. The femoral posterior wedges include size 1- 6 with a thickness of 5mm and 10mm while the femoral distal wedges are available in the same size range (1- 6), in thicknesses of 4mm, 8mm and 12mm.
GMK® Revision fixing cylinder for the tibial wedges are manufactured from ultrahigh molecular weight polyethylene (UHMWPE) according to ISO 5834-2: 2006, Implants for Surgery - Ultra-High molecular weight polyethylene - Part 2: Moulded Forms, same as the UHMWPE components in the original submission. The GMK® Total Knee System- Revision components' design and technological characteristics are similar to the predicate device. The GMK® Total Knee System- Revision components are packaged and sterilized in the manner as the predicate, the GMK® Total Knee System- Primary.
{3}------------------------------------------------
Performance Testing
No performance standards applicable to this device have been adopted under Section 514 of the Food. Drug and Cosmetic Act.
Risk analysis was conducted on the impact of these changes and appropriate design verification and validation was conducted under the company's design controls. The protocols and predefined acceptance criteria were based on the FDA quidance and international standards.
Potential risks identified from the changes were changes in biocompatibility due to new materials and risk of breakage or reduced endurance of interlock strength of the following:
Extension stems and offset adaptors to the femoral components, Posterior and distal wedges to femoral components. Extension stems and offset adaptors to tibial baseplate Tibial wedges to tibial baseplate
Potential risks due to new materials were resolved by using only materials that meet recognized standards and have a long history of acceptable biocompatibility in orthopedic applications. Potential risks due to risk of breakage or reduced endurance of interlock strength of the various new components, extension stems, offset adaptors, femoral wedges, and tibial wedges, were resolved by conducting verification studies that demonstrated that interlocking mechanisms could survive 10 million cycles at physiological loads and had acceptable torsional and disassembly forces. The testing was conducted on the worst case component size and option/design. Testing was conducted according to ASTM F1814-97a Standard Guide for Evaluating Modular Hip and Knee Joint Components. ASTM F897-02:2007 Standard Test Method for Measuring Fretting Corrosion of Osteosynthesis Plates and Screws and ASTM F1800-04 Standard Test Method for Cyclic Fatique Testing of Metal Tibial Tray Components of Total Knee Joint Replacements. The testing met all acceptance criteria and verifies that performance of the GMK® Total Knee System- Revision is substantially equivalent to the predicate device.
Conclusion:
The results from testing and analysis provided in this submission support the conclusion that the GMK® Total Knee System- Revision is substantially equivalent to its predicate device with respect to indications for use and technological characteristics.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the agency's mission of promoting health, well-being, and human services. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Medacta International SA % Ms. Natalie J. Kennel Consultant, NJK & Associates, Inc. 13721 Via Tres Vista San Diego, California 92129
SEP 2 4 2010
Re: K102437
Trade/Device Name: GMK Total Knee System - Revision Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial metal/polymer/metal semi-constrained cemented prosthesis Regulatory Class: II Product Codes: JWH Dated: August 16, 2010 Received: August 26, 2010
Dear Ms. Kennel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 -- Ms. Natalie J. Kennel
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Вавaев вненит
Mark N. N Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use Statement
Image /page/6/Picture/1 description: The image shows the text '510(k) Number (if known):' followed by the handwritten text 'K10247'. This indicates that the image is likely part of a document or form where the 510(k) number, a premarket submission to the FDA, is being recorded. The handwritten entry suggests that the number was manually added to the document.
Device Name: GMK® Total Knee System- Revision
Indications for Use:
The GMK® Total Knee System is designed for cemented use in total knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components.
This knee replacement system is indicated in the following cases:
- Severely painful and/or disabled joint as a result of arthritis, traumatic . arthritis, rheumatoid arthritis or polyarthritis
- avascular necrosis of femoral condyle .
- post traumatic loss of joint configuration .
- primary implantation failure. .
The tibial augments are to be attached to the tibial baseplate with both the fixing cylinders and bone cement.
Prescription Use _ X _________________________________________________________________________________________________________________________________________________________
AND/OR Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ortu for mxm
(Division Sign d)
(Dission Sign-0 Division of Surgical. Orthopedic. and Restorative Devices
510(k) Number K102437
Page _ of _
§ 888.3560 Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis.
(a)
Identification. A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II.