K Number
K102356

Validate with FDA (Live)

Date Cleared
2010-09-20

(32 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

MD-534 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

Device Description

MD-534 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm. it is designed and manufactured according to ANSI/AAMI SP10--manual, electronic or automated sphygmanometers. The operational principle is based on oscillometric and silicon integrate pressure sensor technology. It can calculate the systolic and diast litegrac pressure, and display the result on the LCD. If any irregular heartbeat is detected, it can also be shown on the LCD. More over, it also has the function of averaging the last three measurements.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The provided text states that the device, MD-534 Fully Automatic Electronic Blood Pressure Monitor, conforms to several standards, implying these standards contain the acceptance criteria for performance. The most relevant standard for blood pressure measurement accuracy is:

  • AAMI SP10:2002, Manual, electronic or automated sphygmomanometers.
  • AAMI / ANSI SP10:2002/A1:2003 --, Amendment 1 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers.
  • AAMI / ANSI SP10:2002/A2:2006 --, Amendment 2 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers.

While the text states conformance, it does not explicitly list the specific numerical acceptance criteria from AAMI SP10 (e.g., mean difference and standard deviation between device and reference measurement), nor does it present specific device performance results in a table against those criteria. It only provides a general statement: "the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness." referring to differences between the new device and the predicate.

Therefore, a table cannot be fully completed with specific values from the provided text. However, we can infer the acceptance criteria are those outlined in AAMI SP10.

Acceptance Criterion (from AAMI SP10)Reported Device Performance (from submission)
Accuracy (Mean Difference)Confirms to AAMI SP10; specific value not reported
Accuracy (Standard Deviation)Confirms to AAMI SP10; specific value not reported
Mechanical SafetyConfirms to IEC 60601-1; "Identical" to predicate
Electrical SafetyConfirms to EN 60601-1-2; "Identical" to predicate
EMCConfirms to EN 60601-1-2; "Identical" to predicate
Biocompatibility"Identical" to predicate

2. Sample Size Used for the Test Set and Data Provenance

The document does not provide information on the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective). It only vaguely refers to "the test in this submission."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the number of experts used to establish ground truth or their qualifications. For blood pressure monitors, ground truth is typically established by trained technicians or clinicians using a calibrated reference sphygmomanometer (mercury or aneroid).

4. Adjudication Method for the Test Set

The document does not mention any adjudication method for the test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

A MRMC comparative effectiveness study was not conducted or reported. This type of study is more common for diagnostic imaging AI systems where human interpretation is a key component. Blood pressure monitors are standalone measurement devices.

6. Standalone (Algorithm Only) Performance

Yes, a standalone performance assessment was done. The entire submission focuses on the performance of the MD-534 device itself, comparing it against the AAMI SP10 standard and demonstrating substantial equivalence to the predicate device. The performance claims (conformance to standards) refer to the device's inherent accuracy.

7. Type of Ground Truth Used

The type of ground truth used, though not explicitly detailed in its method, would be comparison against a reference standard blood pressure measurement device (e.g., a mercury sphygmomanometer) as mandated by the AAMI SP10 standard. This standard requires direct comparison of the device's measurements against those taken simultaneously by trained observers using a reference method.

8. Sample Size for the Training Set

The document does not mention a "training set" or its sample size. This device is a traditional electronic medical device (blood pressure monitor) and not an AI/ML-based system that typically requires a separate training set. Its design is based on oscillometric principles, not machine learning algorithms that learn from data.

9. How the Ground Truth for the Training Set Was Established

As this is not an AI/ML device, there is no concept of a "training set" or its ground truth in this context.

{0}------------------------------------------------

510(k) Summary

SEP 4 7 2010 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.

1.0 submitter's information

Name: Andon Health Co., Ltd. Address: No 3, Jinping Street Ya An Road, Nankai District, Tianjin, P.R. China Phone number: 86-22-6052 6161 Fax number: 86-22-6052 6162 Contact: Liu Yi Date of Application: 08/16/2010

2.0 Device information

Trade name: MD-534 Fully Automatic Electronic Blood Pressure Monitor Common name: Noninvasive blood pressure measurement system Classification name: Noninvasive blood pressure measurement system

3.0 Classification

Production code: DXN- Noninvasive blood pressure measurement system. Regulation number: 870.1130 Classification: II Panel: Cardiovascular

4.0 Predict device information

Manufacturer: Andon Health Co., Ltd. KD-591 Fully Automatic Electronic Blood Pressure Monitor Device: 510(k) number: K080319

5.0 Device description

MD-534 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

{1}------------------------------------------------

2/3

it is designed and manufactured according to ANSI/AAMI SP10--manual, electronic or automated sphygmanometers.

The operational principle is based on oscillometric and silicon integrate pressure sensor technology. It can calculate the systolic and diast litegrac pressure, and display the result on the LCD. If any irregular heartbeat is detected, it can also be shown on the LCD. More over, it also has the function of averaging the last three measurements.

6.0 Intended use

MD-534 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

The intended use and the indication for use of MD-534, as described in its labeling are the same as the predict device KD-591.

Technological CharacteristicsComparison result
Design principleIdentical
AppearanceSimilar
Patients contact MaterialsIdentical
PerformanceSimilar
BiocompatibilityIdentical
Mechanical safetyIdentical
Energy sourceIdentical
Standards metIdentical
Electrical safetyIdentical
EMCIdentical
FunctionSimilar

7.0 Summary_comparing_technological_characteristics_with_predicate device

ર- 2

{2}------------------------------------------------

8.0 Performance summary

MD-534 Fully Automatic Electronic Blood Pressure Monitor conforms to the following standards:

  • · IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995.
  • · EN 60601-1-2, Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic Compatibility - Requirements and Tests, 2007
  • · AAMI SP10:2002, Manual, electronic or automated sphygmomanometers.
  • AAMI / ANSI SP10:2002/A1:2003 --, Amendment 1 to ANSI/AAM! SP10:2002 Manual, electronic, or automated sphygmomanometers.
  • AAMI / ANSI SP10:2002/A2:2006 --, Amendment 2 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers.

9.0 Comparison to the predict device and the conclusion

Our device MD-534 Fully Automatic Electronic Blood Pressure Monitor is substantially equivalent to the Fully Automatic Electronic Blood Pressure Monitor KD-591 whose 510(k) number is K080319.

The two devices are very similar in the intended use, the design principle, the material, the performance and the applicable standards. Only their appearsne and some functions such as memory times are different.

Compared to KD-591, the MD-534 does not have the display of Date, Time and Blood pressure classification, and has an additional display of "AVG" The memory time, MCU, and the humidity of operational parameter are also changed. More over, the systolic Range of MD-534 is narrow than KD-591

However, the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Andon Health Co., Ltd. c/o Mr. Liu Yi President No. 3 Jin Ping Street, Ya An Road, Nankai District Tianjin China 300190

SEP 1 7 2010

Re: K102356

Trade/Device Name: MD-534 Fully Automated Electronic Blood Pressure Monitor Regulatory Number: 21 CFR 870.1130 Regulation Name: Non-invasive Blood Pressure Measurement System Regulatory Class: II (two) Product Code: DXN Dated: August 16, 2010 Received: August 19, 2010

Dear Mr. Yi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 .- Mr. Liu Yi

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

17

Statement of Indications for Use

510(k) Number :K102356SEP 17, 2010
Device name:MD-534 Fully Automatic Electronic Blood Pressure Monitor

Indications for use:

MD-534 Fully Automatic Electronic Blood Pressure Monitor is for use by
medical professionals or at home and is a non-invasive blood pressure
measurement system intended to measure the diastolic and systolic blood
pressures and pulse rate of an adult individual by using a non-invasive
technique in which an inflatable cuff is wrapped around the upper arm. The
cuff circumference is limited to 22cm-48cm.

Prescription usePart 21 CFR 801 Subpart D)AND/OR Over-The-Counter Use YES(21 CFR 807 Subpart C)
-----------------------------------------------------------------------------------------------------------

(PLEASE DO NOT WRITE BELOW THIS LINE-COUNTINUE ON
ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) NumberK102356

יי

.

4-1

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).