K Number
K102189
Date Cleared
2012-03-14

(590 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim is a specialty medical glove which is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between examiner and patient. In addition these gloves are worn to protect the wearer against exposure to chemotherapy drugs. Tested chemotherapy drugs are as follows:
Cisplatin, 1.0 mg/ml ≥ 240
Cyclophosphamide (Cytoxan), 20.0 mg/ml ≥ 240
Cytarabine, 100 mg/ml ≥ 240
Cytovene, 10 mg/ml ≥ 240
Dacarbazine (DTIC), 10.0 mg/ml ≥ 240
Docetaxel, 10.0 mg/ml ≥ 240
Doxorubicin Hydrochloride, 2.0 mg/ml ≥ 240
Etoposide (Toposar), 20.0 mg/ml ≥ 240
Fluorouracil, 50.0 mg/ml ≥ 240
Ifosfamide, 50.0 mg/ml ≥ 240
Irinotecan Hydrochloride, 20 mg/ml ≥ 240
Mechlorethamine HCI, 1.0 mg/ml ≥ 240
Methotrexate, 25 mg/ml ≥ 240
Mitomycin C, 0.5 mg/ml ≥ 240
Mitoxantrone, 2 mg/ml ≥ 240
Oxaliplatin, 5mg/ml ≥ 240
Paclitaxel (Taxol), 6.0 mg/ml ≥ 240
ThioTEPA 75.8
Vincristine Sulfate, 1.0 mg/ml ≥ 240
Vinorelbine, 10 mg/ml ≥ 240
Low Permeation Time (minutes) 1.7
Carmustine (BCNU), 3.3 mg/ml
Please note that the following drug has an extremely low permeation time: Carmustine (BCNU) 1.7 (mins)

Device Description

Powder-Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim, meeting all the requirements of ASTM standards D6319-10 and ASTM D6124-06. The gloves are non-sterile, single-use, disposable devices.

AI/ML Overview

The provided text is a 510(k) summary for "Powder-Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim." This document describes the device's characteristics and performance to demonstrate substantial equivalence to a predicate device. It is not an AI/ML device, and therefore several of the requested sections (e.g., sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, training set size) are not applicable.

Here's the information that can be extracted:


Acceptance Criteria and Device Performance Study for Powder-Free Nitrile Examination Gloves

This device is not an AI/ML device, and thus many of the requested sections related to AI/ML study design are not applicable. The device is a physical medical glove.

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicStandard/Acceptance CriteriaDevice Performance
DimensionsASTM D6319-10Meets standard requirements
Physical PropertiesASTM D6319-10Meets standard requirements
ThicknessASTM D6319-10Meets standard requirements (Palm Thickness: ≥0.10 mm; Finger Thickness: ≥0.10 mm; Length: ≥270 mm)
BiocompatibilityPrimary Skin Irritation Study (ISO 10993-0:2002E); Dermal Sensitization (ASTM-F 720-81)Passes (Not a primary skin irritant); Passes (Not a contact sensitizer)
Watertight - 1000 ml (Freedom from Holes)21 CFR 800.20 (AQL 2.5 required per standard)Passes (Gloves Free of Holes at quality level of AQL 1.5, which is better than required)
Resistance to Permeation by Chemotherapy DrugsASTM D6978-05 (Specific breakthrough times for various drugs)Meets standard requirements for most drugs: - Cisplatin, 1.0 mg/ml: ≥ 240 minutes - Cyclophosphamide (Cytoxan), 20.0 mg/ml: ≥ 240 minutes - Cytarabine, 100 mg/ml: ≥ 240 minutes - Cytovene, 10 mg/ml: ≥ 240 minutes - Dacarbazine (DTIC), 10.0 mg/ml: ≥ 240 minutes - Docetaxel, 10.0 mg/ml: ≥ 240 minutes - Doxorubicin Hydrochloride, 2.0 mg/ml: ≥ 240 minutes - Etoposide (Toposar), 20.0 mg/ml: ≥ 240 minutes - Fluorouracil, 50.0 mg/ml: ≥ 240 minutes - Ifosfamide, 50.0 mg/ml: ≥ 240 minutes - Irinotecan Hydrochloride, 20 mg/ml: ≥ 240 minutes - Mechlorethamine HCl, 1.0 mg/ml: ≥ 240 minutes - Methotrexate, 25 mg/ml: ≥ 240 minutes - Mitomycin C, 0.5 mg/ml: ≥ 240 minutes - Mitoxantrone, 2 mg/ml: ≥ 240 minutes - Oxaliplatin, 5mg/ml: ≥ 240 minutes - Paclitaxel (Taxol), 6.0 mg/ml: ≥ 240 minutes - Vincristine Sulfate, 1.0 mg/ml: ≥ 240 minutes - Vinorelbine, 10 mg/ml: ≥ 240 minutes Exceptions: - ThioTEPA: 75.8 minutes - Carmustine (BCNU), 3.3 mg/ml: 1.7 minutes (noted as extremely low permeation time)
Design Specifications & Glove PerformanceASTM D6319-10 (Tensile Strength ≥ 14MPa; Elongation ≥400%)Meets ASTM D6319-10 (Tensile Strength ≥ 14MPa; Elongation ≥400%)
MaterialsN/A (Compared to predicate)Flexible Nitrile
ColorN/A (Compared to predicate)Synthetic Glove with embedded BLUE Biocompatible Colorant
Technology of Glove Design (Coatings/Donning Aid)N/A (Compared to predicate)Ambidextrous; Inner Glove Surface Chlorine Treated to Facilitate Donning; Fingertips Textured to Improve Tactility and Grip
Powder-FreeApplicable Definition for Powder Free (< 2 mg per glove)Meets Applicable Definition for Powder Free; < 2 mg per glove
SterilityNot Applicable (Non-Sterile)Not Applicable: Non-Sterile

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample sizes for the individual tests (e.g., number of gloves tested for each characteristic or drug permeation). The tests were performance-based, carried out according to ASTM standards for physical device properties and chemical resistance.

  • Data Provenance: The tests were conducted by Latexx Manufacturing Sdn. Bhd., located in Malaysia. The document doesn't specify if the data itself originated from Malaysia or another country, but the manufacturing site is in Malaysia. The studies are retrospective from the perspective of being submitted for regulatory approval, as they were completed prior to the submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

N/A. This is a physical device, and the "ground truth" is established by performance against pre-defined standardized physical and chemical tests (e.g., ASTM standards for tensile strength, elongation, dimensions, and chemotherapy drug permeation), not by expert interpretation of data.

4. Adjudication Method for the Test Set

N/A. Not applicable for a physical device undergoing standardized mechanical and chemical testing.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

N/A. This is not an AI/ML diagnostic device; it is a physical medical glove.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

N/A. This is not an AI/ML device. The tests performed are inherent to the material properties and design of the glove.

7. The Type of Ground Truth Used

The "ground truth" for the device's performance is established by standardized laboratory testing procedures defined by international and national standards (e.g., ASTM D6319-10, ASTM D6124-06, ASTM D6978-05, ISO 10993-0:2002E, ASTM-F 720-81, 21 CFR 800.20). These standards define the methodology for measuring characteristics like dimensions, physical properties, biocompatibility, watertight integrity, and resistance to chemotherapy drug permeation.

8. The Sample Size for the Training Set

N/A. This is a physical device, not an AI/ML algorithm that requires a training set. The device itself is manufactured, and its performance is tested against standards.

9. How the Ground Truth for the Training Set Was Established

N/A. Not applicable for a physical device.

{0}------------------------------------------------

MAR 1 4 2012

K 102189

510(k) SUMMARY

1.0 Submitter:

Latexx Manufacturing Sdn. Bhd. Name: Address: Lot 18374. Kamunting Industrial Estate 34600 Taiping, Perak Malaysia Phone: 605 891 1111 605 829 5590 Fax: Contact Person: Ooi Loon Seng (Madam)

Date of Preparation: February 10, 2012

2.0 Name of Device:

Powder-Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim

Common Name: Examination Gloves

Classification Name: Patient Examination Gloves, Specialty Chemotherapy (21 CFR 880.6250 Product Code LZC)

510(k) Number: K102189

Legally Marketed Device to Which Equivalence is being Claimed: 3.0

Primary Predicate: Cardinal Health Esteem Tru-Blu Stretchy Nitrile Exam Gloves 510(k) Number: K032444

Additional Predicate: Powder-Free Blue Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim 510(k) Number: K022765

4.0 Description of the Device:

Powder-Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim, meeting all the requirements of ASTM standards D6319-10 and ASTM D6124-06. The gloves are non-sterile, single-use, disposable devices.

Additional Information requested 24 February 2012

Page 4 of 10 27 February 2012

{1}------------------------------------------------

5.0 Indications for Use:

The Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim is a specialty medical glove which is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between examiner and patient. In addition these gloves are worn to protect the wearer against exposure to chemotherapy drugs. Tested chemotherapy drugs are as follows:

Test Chemical and ConcentrationAverage Breakthrough Detection Time (minutes)
Cisplatin, 1.0 mg/ml≥ 240
Cyclophosphamide (Cytoxan), 20.0 mg/ml≥ 240
Cytarabine, 100 mg/ml≥ 240
Cytovene, 10 mg/ml≥ 240
Dacarbazine (DTIC), 10.0 mg/ml≥ 240
Docetaxel, 10.0 mg/ml≥ 240
Doxorubicin Hydrochloride, 2.0 mg/ml≥ 240
Etoposide (Toposar), 20.0 mg/ml≥ 240
Fluorouracil, 50.0 mg/ml≥ 240
Ifosfamide, 50.0 mg/ml≥ 240
Irinotecan Hydrochloride, 20 mg/ml≥ 240
Mechlorethamine HCI, 1.0 mg/ml≥ 240
Methotrexate, 25 mg/ml≥ 240
Mitomycin C, 0.5 mg/ml≥ 240
Mitoxantrone, 2 mg/ml≥ 240
Oxaliplatin, 5mg/ml≥ 240
Paclitaxel (Taxol), 6.0 mg/ml≥ 240
ThioTEPA75.8
Vincristine Sulfate, 1.0 mg/ml≥ 240
Vinorelbine, 10 mg/ml≥ 240
Low Permeation Time (minutes)1.7
Carmustine (BCNU), 3.3 mg/ml

Please note that the following drug has an extremely low permeation time: Carmustine (BCNU) 1.7 (mins)

Latexx Manufacturing Sdn. Bhd.

Additional Information requested 24 February 2012

Page 5 of 10 27 February 2012

{2}------------------------------------------------

6.0 Summary of Technological Characteristics of the Device:

Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim have the following technological characteristics compared to ASTM or equivalent standards:

CharacteristicsStandardDevice Performance
DimensionsASTM D6319-10Meets standard requirements
Physical PropertiesASTM D6319-10Meets standard requirements
ThicknessASTM D6319-10Meets standard requirements
BiocompatibilityPrimary Skin IrritationStudy (ISO 10993-0:2002E)Dermal Sensitization(ASTM-F 720-81)Passes (Not a primary skinirritant)Passes (Not a contactsensitizer)
Watertight - 1000 ml(Freedom from Holes)21 CFR 800.20Passes
Resistance to Permeationby Chemotherapy DrugsASTM D6978-05Meets standard requirements

Latexx Manufacturing "Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim" demonstrated the following breakthrough detection times when tested according to ASTMD6978-05

Test Chemical and ConcentrationAverage BreakthroughDetection Time (minutes)
Cisplatin, 1.0 mg/ml≥ 240
Cyclophosphamide (Cytoxan), 20.0 mg/ml≥ 240
Cytarabine, 100 mg/ml≥ 240
Cytovene, 10 mg/ml≥ 240
Dacarbazine (DTIC), 10.0 mg/ml≥ 240
Docetaxel, 10.0 mg/ml≥ 240
Doxorubicin Hydrochloride, 2.0 mg/ml≥ 240
Etoposide (Toposar), 20.0 mg/ml≥ 240
Fluorouracil, 50.0 mg/ml≥ 240
Ifosfamide, 50.0 mg/ml≥ 240
Irinotecan Hydrochloride, 20 mg/ml≥ 240
Mechlorethamine HCl, 1.0 mg/ml≥ 240
Methotrexate, 25 mg/ml≥ 240
Mitomycin C, 0.5 mg/ml≥ 240
Mitoxantrone, 2 mg/ml≥ 240
Oxaliplatin, 5mg/ml≥ 240
Paclitaxel (Taxol), 6.0 mg/ml.≥ 240
Thiotepa75.8
Vincristine Sulfate, 1.0 mg/ml≥ 240
Vinorelbine, 10 mg/ml≥ 240
Low Permeation Time (minutes)Carmustine (BCNU), 3.3 mg/ml1.7

Latexx Manufacturing Sdn. Bhd.

Additional Information requested 24 February 2012

Page 6 of 10 27 February 2012

{3}------------------------------------------------

  • Substantial Equivalence Based on Assessment of Non-Clinical Performance Data: 7.0
    Comparison to Predicate and Non-Clinical Testing - The subject device (Powder-Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim) compares favorably to the predicate device (Cardinal Health Esteem Tru-Blu Stretchy Nitrile Exam Gloves) as indicated in the tabulated summaries to follow.

Substantial Equivalence Comparison of Indications for Use

The Indications for Use statements for the subject nitrile gloves and the cited predicate, as listed below, are substantially equivalent (SE).

Indications forUse –Subject Device:The Powder Free Nitrile Examination Gloves Tested for Use with ChemotherapyDrugs Labeling Claim is a specialty medical glove which is a disposable deviceintended for medical purpose that is worn on the examiner's hand or finger toprevent contamination between examiner and patient. In addition these gloves areworn to protect the wearer against exposure to chemotherapy drugs. Testedchemotherapy drugs are as follows:
Powder-FreeNitrileExaminationGloves Tested forUse withChemotherapyDrugs LabelingClaimTest Chemical and ConcentrationAverage Breakthrough Detection Time(minutes)
Cisplatin, 1.0 mg/ml≥ 240
Cyclophosphamide (Cytoxan), 20.0 mg/ml≥ 240
Cytarabine, 100 mg/ml≥ 240
Cytovene, 10 mg/ml≥ 240
Dacarbazine (DTIC), 10.0 mg/ml≥ 240
Docetaxel, 10.0 mg/ml≥ 240
Doxorubicin Hydrochloride, 2.0 mg/ml≥ 240
Etoposide (Toposar), 20.0 mg/ml≥ 240
Fluorouracil, 50.0 mg/ml≥ 240
Ifosfamide, 50.0 mg/ml≥ 240
Irinotecan Hydrochloride, 20 mg/ml≥ 240
Mechlorethamine HCI, 1.0 mg/ml≥ 240
Methotrexate, 25 mg/ml≥ 240
Mitomycin C, 0.5 mg/ml≥ 240
Mitoxantrone, 2 mg/ml≥ 240
Oxaliplatin, 5mg/ml≥ 240
Paclitaxel (Taxol), 6.0 mg/ml≥ 240
ThioTEPA75.8
Vincristine Sulfate, 1.0 mg/ml≥ 240
Vinorelbine, 10 mg/ml≥ 240
Low Permeation Time(minutes)Carmustine BCNU), 3.3 mg/ml1.7
Please note that the following drug has an extremely low permeation time: Carmustine(BCNU) 1.7 (mins)

Page 7 of 10 27 February 2012

{4}------------------------------------------------

Indications for Use -Predicate Device:These examination gloves are disposable devices intended formedical purposes that are worn on the examiner's hands orfingers to prevent contamination between patient and examiner.
Cardinal Health Esteem Tru-BluStretchy Nitrile Gloves (NitrilePowder-Free Examination Gloveswith Lotion Coating, Blue withUse for Chemotherapy)Cardinal HealthcareCorporation K032444)In addition, these gloves are worn to protect the wearer againstexposure to chemotherapy drugs.

Comparisons of non-clinical biocompatibility and engineering bench tests are included in the summary table below. ·

Subject DevicePowder-Free NitrileExamination Gloves Tested forUse with Chemotherapy DrugsLabeling ClaimPredicate DeviceCardinal Health Esteem Tru-Blu Stretchy Nitrile Gloves(Nitrile Powder-FreeExamination Gloves withLotion Coating, Blue with Usefor Chemotherapy)SubstantialEquivalence(SE)?
Cardinal HealthcareCorporation K032444)
ChemotherapyPermeationStandardMeets ASTM D6978-05Meets ASTM D6978-05Yes
DesignSpecifications &GlovePerformanceMeets ASTM D6319-10:-Tensile Strength ≥ 14MPa(≥14MPa per Standard)-Elongation ≥400% (≥400% perStandard)Meets ASTM D6319-00:-Tensile Strength 19MPa (≥14MPa per Standard)-Elongation 533% (≥400% perStandard)Yes
Freedom fromHolesMeets Requirements per21CFR800.20:Gloves Free of Holes at qualitylevel of AQL 1.5 (AQL 2.5required per standard)Meets Requirements per21CFR800.20:Gloves Free of Holes at qualitylevel of AQL 1.5 (AQL 2.5required per standard)Yes
Glove Thicknessand LengthMeets ASTM D6319-10:-Palm Thickness of ≥0.10 mm-Finger Thickness ≥0.10 mm-Length ≥270 mmMeets ASTM D6319-10:-Palm Thickness average 0.10mm-Finger Thickness average 0.13mm-Length average 246 mmYes
MaterialsFlexible NitrileFlexible NitrileYes

Substantial Equivalence Comparison Table – Engineering and Bench Testing

Latexx Manufacturing Sdn. Bhd.

Additional Information requested 24 February 2012

Page 8 of 10 27 February 2012

{5}------------------------------------------------

Substantial Equivalence Comparison Table - Engineering and Bench Testing (continued)

Subject DevicePowder-Free NitrileExamination Gloves Tested forUse with Chemotherapy DrugsLabeling ClaimPredicate DeviceCardinal Health Esteem Tru-Blu Stretchy Nitrile Gloves(Nitrile Powder-FreeExamination Gloves withLotion Coating, Blue with Usefor Chemotherapy)Cardinal HealthcareCorporation K032444)SubstantialEquivalence(SE)?
ColorSynthetic Glove with embeddedBLUE Biocompatible ColorantSynthetic Glove with embeddedBLUE Biocompatible ColorantYes
Technology ofGlove Design,i.e. Coatingsand/or DonningAidAmbidextrous;Inner Glove Surface ChlorineTreated to Facilitate Donning;Fingertips Textured to ImproveTactility and GripAmbidextrous;Lotion Coating on Inner Surfaceas Donning Aid:Outer Surface Textured toImprove GripYes
BiocompatibilityPasses tests for :-Primary Skin Irritation-Dermal SensitizationPasses tests for:-Primary Skin Irritation-Dermal SensitizationYes
Powder-FreeMeets Applicable Definition forPowder Free; < 2 mg per gloveMeets Applicable Definition forPowder Free; ≤ 2 mg per gloveYes
SterilityNot Applicable: Non-SterileNot Applicable: Non-SterileYes

Summary of Differences and Comparison of Safety and Effectiveness - The subject device differs from the predicate:

Difference in Donning Aid: The predicate device has an inner surface coated with lotion, and the subject device has a chlorine treated inner surface. Both the lotion coating and chlorine treatment serve as donning aids. Therefore, both gloves are functionally equivalent with regard to ease of user donning, and may be used to effectively keep users safe from exposure to chemotherapeutic agents.

Difference in Colorant: Latexx Manufacturing incorporates medical grade colorant into the subject device. The subject device has been tested and shown to function as an effective barrier to chemotherapeutic agents. The subject device has also passed biocompatibility tests for skin irritation and dermal sensitization; the subject gloves are non-toxic and may safely contact either users or patients.

  • 8.0 Animal Testing: No animal testing were required to demonstrate substantial equivalence.
  • Substantial Equivalence Based on an Assessment of Clinical Performance Data: 9.0 A clinical study was not conducted on the subject or predicate devices.

Additional Information requested 24 February 2012

P. 6 of 7

{6}------------------------------------------------

10.0 Conclusion:

"Powder-Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim" produced by Latexx Manufacturing Sdn. Bhd. are as safe and effective as the predicate device. The subject device has been tested against the applicable ASTM standards listed above, and meets the requirements set forth in those standards. Additional specific comparisons show the subject device substantially equivalent to the predicate.

Additional Information requested 24 February 2012

Page 10 of 10 27 February 2012

., 7 of 7

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, with three curved lines representing snakes or branches.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002

Latexx Manufacturing Sdn. Bhd. C/O Mr. Neil Burris Regulatory Affairs Consultant Reglera, LLC 11925 West 1-70 Frontage Road North, Suite 900 Wheat Ridge, Colorado 80033

MAR 1 4 2012

Re: K102189

Trade Name: Powder-Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim Regulation Number: 21 CFR 880. 6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZC Dated: February 27, 2012 Received: March 14, 2012

Dear Mr. Burris:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include " requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898.. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{8}------------------------------------------------

Page 2- Mr. Burris

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony D. Watson, B.S., M.S., M.Β.Α.
Director
Division of Anesthesiology, General Hospital,
Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and

Radiological Health

Enclosure

{9}------------------------------------------------

INDICATIONS FOR USE

Applicant: LATEXX MANUFACTURING SDN. BHD., LOT 18374, Kamunting Industrial Estate, 34600 Taiping, Perak, Malaysia

510(k) Number: K102189

Device Name: POWDER FREE NITRILE EXAMINATION GLOVES TESTED FOR USE WITH CHEMOTHERAPY DRUGS LABELING CLAIM

Indications For Use:

The Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs Labeling Claim is a specialty medical glove which is a disposable device intended for medical purpose that is wom on the examiner's hand or finger to prevent contamination between examiner and patient. In addition these gloves are worn to protect the wearer against exposure to chemotherapy drugs. Tested for use with chemotherapy drugs. Tested chemotherapy drugs are as follows:

Test Chemical and ConcentrationAverage Breakthrough Detection Time (minutes)
Cisplatin, 1.0 mg/ml≥ 240
Cyclophosphamide (Cytoxan), 20.0 mg/ml≥ 240
Cytarabine, 100 mg/ml≥ 240
Cytovene, 10 mg/ml≥ 240
Dacarbazine (DTIC), 10.0 mg/ml≥ 240
Docetaxel, 10.0 mg/ml≥ 240
Doxorubicin Hydrochloride, 2.0 mg/ml≥ 240
Etoposide (Toposar), 20.0 mg/ml≥ 240
Fluorouracil, 50.0 mg/ml≥ 240
Ifosfamide, 50.0 mg/ml≥ 240
Irinotecan Hydrochloride, 20 mg/ml≥ 240
Mechlorethamine HCI, 1.0 mg/ml≥ 240
Methotrexate, 25 mg/ml≥ 240
Mitomycin C, 0.5 mg/ml≥ 240
Mitoxantrone, 2 mg/ml≥ 240
Oxaliplatin, 5mg/ml≥ 240
Paclitaxel (Taxol), 6.0 mg/ml≥ 240
ThioTEPA75.8
Vincristine Sulfate, 1.0 mg/ml≥ 240
Vinorelbine, 10 mg/ml≥ 240
Low Permeation Time (minutes)1.7
Carmustine (BCNU), 3.3 mg/ml

Please note that the following drug has an extremely low permeation time: Carmustine (BCNU) 1.7 (mins)

Concurrence of CDRH Office of Device Evaluation (ODE)

Prescription Use
Per 21 CFR 801.109

OR Over-The-Counter

×

Elisha F. Marine-Willen

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: _______________________________________________________________________________________________________________________________________________________________

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.