(271 days)
Powdered vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
powdered vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.
This document describes the acceptance criteria and the study results for the Hebei Tengda Plastic Co., Ltd. Powdered Vinyl Patient Examination Gloves, Clear (Non-colored).
1. Table of Acceptance Criteria & Reported Device Performance
| Characteristics | Standard (Acceptance Criteria) | Device Performance |
|---|---|---|
| Dimension | ASTM standard D 5250-06 | Meets |
| Physical Properties | ASTM standard D 5250-06 | Meets |
| Freedom from pinholes | 21 CFR 800.20 | Meets |
| Powder Amount | ASTM standard D 5250-06 | Meets |
| Biocompatibility | Primary Skin Irritation in rabbits | Passes |
| Not a Primary Skin Irritation (<10mg/dm2) | ||
| Dermal sensitization in the guinea pig | Passes | |
| Not a Dermal sensitization |
2. Sample Size and Data Provenance for Test Set
The document does not explicitly state the specific sample sizes used for each of the tests (Dimension, Physical Properties, Freedom from pinholes, Powder Amount, Biocompatibility). However, it indicates compliance with recognized standards (ASTM D 5250-06, 21 CFR 800.20, ISO10993-10), which inherently define the testing methodologies and sample sizes required for compliance.
The data provenance is from nonclinical studies conducted to demonstrate substantial equivalence. The document does not specify the country of origin where these tests were performed, but it is submitted by a Chinese manufacturer (Hebei Tengda Plastic Co., Ltd.) to the US FDA. The studies would be considered retrospective in the context of this 510(k) submission, as they were conducted prior to the submission to demonstrate compliance.
3. Number of Experts and Qualifications for Ground Truth for Test Set
This type of device (patient examination gloves) does not typically involve human expert interpretation for establishing ground truth in the way medical imaging or diagnostic AI devices do. The "ground truth" for the performance criteria is established by the well-defined and objective metrics within the ASTM and FDA standards. Therefore, information about the number of experts or their qualifications for establishing ground truth is not applicable in this context.
4. Adjudication Method for Test Set
Adjudication methods like 2+1 or 3+1 are typically used in studies involving human interpretation (e.g., medical image reading) to resolve disagreements among experts. Since the tests for glove performance are objective and based on measurable physical and chemical properties, an adjudication method is not applicable. The results are directly compared against the quantitative thresholds defined by the standards.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC study was conducted or is applicable for this device. This device is a physical medical device (gloves), not an AI-assisted diagnostic tool that would involve human readers interpreting cases.
6. Standalone Performance Study (Algorithm Only)
Not applicable. This device is a physical product, not an algorithm.
7. Type of Ground Truth Used
The ground truth used is based on established industry standards and regulatory requirements, specifically:
- ASTM standard D 5250-06 (for Dimension, Physical Properties, Powder Amount)
- 21 CFR 800.20 (for Freedom from pinholes)
- ISO10993-10 (for Biocompatibility testing, although not explicitly listed in the performance table, it's mentioned as a standard met)
These standards define objective and measurable criteria that the device must meet.
8. Sample Size for the Training Set
Not applicable. This is a physical device, not an AI model that requires a training set.
9. How the Ground Truth for the Training Set was Established
Not applicable, as there is no training set for a physical device.
{0}------------------------------------------------
Attachment I.
og I of 2
Summary
APR 2 7 2011
"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."
"The assigned 510(k) number is: K102 151 - . " (applicant leave blank)
Premarket Notification [510(k)] Summary
I(a)(1)). The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :
| Submitter's name : | HEBEI TENGDA PLASTIC CO., LTD | |
|---|---|---|
| Submitter's address : | RM102, UNIT 2, #6 BUILDING, #118 DONGFENG RD.SHIJIAZHUANG, 050000, CHINA | |
| Phone number : | (86) - 311-83985266 | |
| Fax number : | (86) - 311-83986358 | |
| Name of contact person: | Ms. Sun Meng | |
| Date the summary was prepared: | July 28th, 2010 |
[(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known
| Device Name: | Powdered Vinyl Patient Examination Gloves,Clear (Non-colored) | |
|---|---|---|
| Proprietary/Trade name: | Powdered Vinyl Patient Examination GlovesOther clients private labeling | |
| Common Name: | Patient examination glove | |
| Classification Name: | Patient examination glove | |
| Device Classification: | I | |
| Regulation Number: | 21 CFR 880.6250 | |
| Panel: | General Hospital (80) | |
| Product Code: | LYZ |
[(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .
Class I* powdered vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.
Predicate device : FUGUAN (Brand) Powdered Vinyl Patient Examination Gloves, Shijiazhuang Fuguan Plastic Products Co., Ltd.. K032907.
[(a)(4)] A description of the device
Device Description : powdered vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.
{1}------------------------------------------------
{(a)(5)] The summary describes the intended use of the device
Device Intended Use: powdered vinyl patient examination glove, Clear(non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
[(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.
The powdered vinyl patient examination gloves, Clear (non-colored) non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.
| Characteristics | Standard | Device performance |
|---|---|---|
| Dimension | ASTM standard D 5250-06 | Meets |
| Physical Properties | ASTM standard D 5250-06 | Meets |
| Freedom from pinholes | 21 CFR 800.20 | Meets |
| Powder Amount | ASTM standard D 5250-06 | Meets |
| Biocompatability | <10mg/dm2 | |
| Primary Skin Irritation in rabbits | Passes | |
| Not a Primary Skin Irritation | ||
| Dermal sensitization in the guinea pig | PassesNot a Dermal sensitization |
[(b)(1)] A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .
Powdered vinyl patient examination gloves , Clear(non-colored) meet requirements per ASTM DS250-06, per ASTM D6124-01, per 21 CFR 800.20 and ISO10993-10.
[(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .
Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.
[(b)(3)] The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performed as well or better than the legally marketed device identified in (a)(3).
It can be concluded that the Powdered Vinyl Patient Examination Gloves, Clear(non-colored) meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL, meet labeling claims .
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The caduceus is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the circumference of the circle. The text is in all capital letters and is evenly spaced around the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Hebei Tengda Plastic Company, Limited C/O Mr. Chu Xiaoan Room 1606 Building Jianxiang Yuan No 209 Bei Si Huan Zhong Road Haidian District Beijing Shijiazhuang, China 100083 .
APR 2 7 2011
Re: K102151
Trade/Device Name: Powdered Vinyl Patient Examination Gloves, Clear (Non-colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: March 29, 2011 Received: March 29, 2011
Dear Mr. Xiaoan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2- Mr. Xiaoan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
fos 2
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital,
Infection Control and Dental Devices
Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{4}------------------------------------------------
INDICATIONS FOR USE
Applicant: HEBEI TENGDA PLASTIC CO., LTD
510(k) Number (if known): *K102 15 l
Device Name: Powdered Vinyl Patient Examination Gloves, Clear (Non-colored)
Indications For Use:
Powdered vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Rahush for EFe
sion of Anesthesiology, General Hospital ction Control, Dental Devices
:0(k) Number: K102151
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.