(262 days)
Not Found
No
The device is a simple examination glove and the summary contains no mention of AI, ML, or any related technologies.
No.
The device, a patient examination glove, is intended to prevent contamination and is not used for treating or diagnosing a medical condition.
No
Explanation: The device is a glove, which is used to prevent contamination, not to diagnose a condition. Its intended use explicitly states "to prevent contamination between patient and examiner."
No
The device is a physical glove, not software. The description clearly states it is a "powder free vinyl patient examination glove" and describes its physical properties and testing against ASTM standards for gloves.
No, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states the device is a "disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner." This describes a barrier device used for protection during physical examination.
- IVD Definition: In vitro diagnostics (IVDs) are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections. They are used to provide information for diagnosis, monitoring, or screening.
- Lack of Diagnostic Function: The description of the vinyl glove does not involve any testing of biological samples or providing diagnostic information. Its function is purely a physical barrier.
Therefore, based on the provided information, the powder-free vinyl patient examination glove is a medical device, but it does not fit the definition of an In Vitro Diagnostic.
N/A
Intended Use / Indications for Use
Powder free vinyl patient examination glove is a Non-Sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Product codes (comma separated list FDA assigned to the subject device)
LYZ
Device Description
powder free vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
hand or finger
Indicated Patient Age Range
Not Found
Intended User / Care Setting
examiner
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Non-clinical testing demonstrated that the device "meets requirements per ASTM D5250-06, per ASTM D6124-06, per 21 CFR 800.20 and ISO10993-10." This includes performance criteria for Dimension, Physical Properties, Freedom from pinholes, Powder Residual (
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.
0
"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."
"The assigned 510(k) number is: K/02-085 ." (applicant leave blank)
Premarket Notification [510(k)] Summary
[(a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :
Submitter's name : | SHIJIAZHUANG GLORY PLASTIC CO., LTD |
---|---|
Submitter's address : | RM1502, UNIT 3, #76 HUAIAN WEST RD. |
SHIJIAZHUANG, 050000, CHINA | |
Phone number : | (86) 311-83985366 |
Fax number : | (86) 311-83986358 |
Name of contact person: | Ms. Zhang Lei |
Date the summary was prepared: | Mar 31th 2010 |
[(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known
Device Name: | Non-Sterile Powder Free Vinyl Examination Gloves, Clear (Non-colored). |
---|---|
Proprietary/Trade name: | Powder Free Vinyl Patient Examination Gloves |
Other clients private labeling | |
Common Name: | Patient examination glove |
Classification Name: | Patient examination glove |
Device Classification: | I |
Regulation Number: | 21 CFR 880.6250 |
Panel: | General Hospital (80) |
Product Code: | LYZ |
[(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .
Class I* powder free vinyl patient examination gloves , Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.
Predicate device : FUGUAN (Brand) Powder-Free Vinyl Patient Examination Gloves, Shijiazhuang Fuguan Plastic Products Co., Ltd.. K032908 .
[{a)(4)] A description of the device
Device Description : powder free vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.
1
pg 2 or 4
[(a)(5)] The summary describes the intended use of the device
Device Intended Use: powder free vinyl patient examination glove, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
[(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.
The powder free vinyl patient examination gloves, non sterile are summarized with the following technological characteristics compared 'to ASTM or equivalent standard.
Characteristics | Standard | Device performance |
---|---|---|
Dimension | ASTM standard D 5250-06 | Meets |
Physical Properties | ASTM standard D 5250-06 | Meets |
Freedom from pinholes | 21 CFR 800.20 | Meets |
Powder Residual | ASTM standard D 5250-06 | |
and D6124-06 |