K Number
K102010
Device Name
VACUETTE PREMIUM SAFETY BLOOD COLLECTION SET
Date Cleared
2010-11-19

(126 days)

Product Code
Regulation Number
880.5570
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
VACUETTE® PREMIUM Safety Blood Collection Set is used in routine venipuncture procedures. It is used for blood collection. The safety shield is activated automatically to cover the needle immediately following blood collection thus preventing accidental needlestick injury.
Device Description
The Vacuette Premium Safety Blood Collection Set is a single use, individually wrapped, sterile winged blood collection needle with an integrated needle and safety shield bonded to a flexible tubing with a female luer adapter allowing the set to be used with a luer system (e.g. Holdex®). It is available with optional Luer Adapter and / or Luer Adapter + Holder. The device is free of PVC, DEHP, Latex.
More Information

No
The description focuses on mechanical safety features and standard blood collection components, with no mention of AI/ML terms or functionalities.

No.

The device is used for blood collection, a diagnostic procedure, and its primary function is to prevent needlestick injuries. It does not provide any treatment or therapy to the patient.

No
This device is used for blood collection, which is a sample collection procedure, not a diagnostic procedure itself.

No

The device description clearly outlines physical components like a needle, tubing, and safety shield, indicating it is a hardware device, not software-only.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is for "routine venipuncture procedures" and "blood collection." This describes a procedure for obtaining a sample from the body, not for performing a diagnostic test on a sample in vitro (outside the body).
  • Device Description: The description details a blood collection set, which is a tool for drawing blood. It doesn't mention any components or functions related to analyzing or testing the collected blood.
  • Lack of Diagnostic Function: There is no mention of the device being used to detect, measure, or analyze any substance or characteristic within the blood sample for diagnostic purposes.
  • Focus on Sample Collection and Safety: The emphasis is on the process of blood collection and the safety feature to prevent needlestick injuries.

IVD devices are specifically designed to perform tests on samples (like blood, urine, tissue) to provide information for diagnosis, monitoring, or screening. This device's function is solely to collect the sample.

N/A

Intended Use / Indications for Use

"VACUETTE® PREMIUM Safety Blood Collection Set is used in routine venipuncture procedures. It is used for blood collection. The safety shield is activated automatically to cover the needle immediately following blood collection thus preventing accidental needlestick injury."

Product codes

FMI, JKA

Device Description

Trade or propriety name or model of this device: VACUETTE® PREMIUM Safety Blood Collection Set; The Vacuette Premium Safety Blood Collection Set is a single use, individually wrapped, sterile winged blood collection needle with an integrated needle and safety shield bonded to a flexible tubing with a female luer adapter allowing the set to be used with a luer system (e.g. Holdex®). It is available with optional Luer Adapter and / or Luer Adapter + Holder. The device is free of PVC, DEHP, Latex.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

appropriately trained healtheare professionals

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Simulated use testing was performed by a panel of health care professionals to evaluate function and performance, ease of use and other functions of the safety feature.

Summary of Performance Studies

Simulated use testing was performed by a panel of health care professionals to evaluate function and performance, ease of use and other functions of the safety feature. Material testing for biocompatibility was also performed according to ISO 10993-1 requirements.

Key Metrics

Not Found

Predicate Device(s)

K011786, K011904

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).

0

NOV 1 9 2010

Kingalo

K162016
Section 5 - 510(k) SummaryPage 5-1

Date Prepared: July 13, 2009 Submitter / Contact: H. Carl Jenkins The Wood Burditt Group 10 E. Scranton Avenue, Suite 201 Lake Bluff, IL 60044 (ph) (847) 234-7500 x 205 (fax) (847) 574-0728 (email) hcjenkins@woodburditt.com Applicant / 510(k) Owner: MedPro Safety Products, Inc 817 Winchester Road, Suite 200 Lexington, KY 40505

Phone – (859) 225-5375 Fax - (859) 225-5347 Contact Person - Mrs. Tammy Nichols

Name of Device

Trade Name:VACUETTE® PREMIUM Safety Blood Collection Set
Common Name:Safety device for blood collection
Classification Name:Hypodermic Single Lumen Needle with antistick syringe
Regulations No:21 CFR 880.5570
Product Code:FMI

Predicate Information

Predicate Device:VACUETTE SAFETY BLOOD COLLECTION SET
510(k) Number:K011786
Date of Concurrence:07/12/2001
Predicate Device:BD VACUTAINER PUSH BUTTON BLOOD COLLECTION SET
510(k) Number:K011904

510(k) Number: K011984 Date of Concurrence: 08/29/2001

1

Section 5 - 510(k) SummaryPage 5-2
--------------------------------------

Substantial Equivalence Declaration:

The term "Substantial Equivalence" as used in this 510(k) Premarket Notification is limited to the definition of Substantial Equivalence found in the Federal Food, Drug, and Cosmetic Act, as amended and as applied under 21 CFR §807, Subpart E, under which a device can be marketed without premarket approval or reclassification. A determination of substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in support of, substantial equivalence herein shall be construed as an admission against interest under the US Patent Laws or their application by the courts.

Device Description:

Trade or propriety name or model of this device: VACUETTE® PREMIUM Safety Blood Collection Set;

GaugeTubingAccessoryAccessoryAccessory
2110Female LuerMale Luer
2310Female LuerMale Luer
2510Female LuerMale Luer
2119Female LuerMale Luer
2319Female LuerMale Luer
2519Female LuerMale Luer
2130Female LuerMale Luer
2330Female LuerMale Luer
2530Female LuerMale Luer
2110Female LuerMale LuerGreiner Bio-One Standard Holder
2310Female LuerMale LuerGreiner Bio-One Standard Holder

2

Section 5 - 510(k) Summary

GaugeTubingAccessoryAccessoryAccessory
2510Female LuerMale LuerGreiner Bio-One Standard Holder
2119Female LuerMale LuerGreiner Bio-One Standard Holder
2319Female LuerMale LuerGreiner Bio-One Standard Holder
2519Female LuerMale LuerGreiner Bio-One Standard Holder
2130Female LuerMale LuerGreiner Bio-One Standard Holder
2330Female LuerMale LuerGreiner Bio-One Standard Holder
2530Female LuerMale LuerGreiner Bio-One Standard Holder
2110Female LuerNA
2310Female LuerNA
2510Female LuerNA
2119Female LuerNA
2319Female LuerNA
2519Female LuerNA
2130Female LuerNA
2330Female LuerNA
2530Female LuerNA

Product Description:

The Vacuette Premium Safety Blood Collection Set is a single use, individually wrapped, sterile winged blood collection needle with an integrated needle and safety shield bonded to a flexible tubing with a female luer adapter allowing the set to be used with a luer system (e.g. Holdex®). It is available with optional Luer Adapter and / or Luer Adapter + Holder. The device is free of PVC, DEHP, Latex.

3

Indications for Use:

"VACUETTE® PREMIUM Safety Blood Collection Set is used in routine venipuncture procedures. It is used for blood collection. The safety shield is activated automatically to cover the needle immediately following blood collection thus preventing accidental needlestick injury."

Intended Use:

The Vacuette Premium Safety Blood Collection Set is used in routine venipuncture procedures. It is used for blood sollection. The winged needle is designed with a safety shield, which is activated automatically to cover the needle immediately following blood

eollection to aid in the protection against-accidental-needlestick injury-The product-is-to be-used by appropriately trained healtheare professionals only in accordance with these instructions.

Substantial Equivalence:

The VACUETTE® PREMIUM Safety Blood Collection Set is substantially equivalent to the Vacuette Safety Blood Collection Set (K011786) and the BD Pushbutton Butterfly (K011984).

Simulated Use Testing:

Simulated use testing was performed by a panel of health care professionals to evaluate function and performance, ease of use and other functions of the safety feature.

Material Testing:

The VACUETTE® PREMIUM Safety Blood Collection Set will meet ISO 10993-1 requirements for material safety and biocompatibility, per product category:

  • . External Communicating
  • . Blood Path Indirect
  • . Contact Duration "A" (Limited)

Conclusions:

Summary of Safety and Effectiveness -

The device has a release mechanism molded into the needle shield. During the blood collection process, as the user performs the initial puncture, release of the wings causes them to fall open and brings the needle shield into a staged position. When the user is ready to remove the device, they grasp the wings as they would during routine use, and upon closing the needle shield is deployed and covers the needle upon exit.

The device is substantially equivalent to the predicate devices (VACUETTE® Safety Blood Collection Set and BD Vacutainer Push Button Set) because it has a plastic safety shield that covers the needle

4

| Section 5 - 510(k) Summary | Page 5-5 | after
it
has
been
n
cont |

-------------------------------------------------------------------------------

aminated, protecting the healthcare worker from a potential needlestick. The predicate VACUETTE® Safety Blood Collection Set's safety shield

is activated when the user manually slides the shield over the needle after use. The predicate BD Vacutainer Push Button Blood Collection Set's safety shield is activated when the user manually presses the button located on the top of the device, activating the spring mechanism to deploy the safety shield. The VACUETTE® PREMIUM Safety Blood Collection Set's safety shield is automatically activated when the user closes the wings to withdraw the device from the patient. The spring loaded safety shield covers the needle after it has been contaminated, protecting the healthcare worker from a potential needlestick.

5

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features the department's emblem, which is a stylized representation of a human figure embracing a globe. The emblem is positioned to the right of the text, which is arranged in a circular pattern around the emblem.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

MedPro Safety Products, Incorporated C/O Mr. H. Carl Jenkins Wood Burditt Group 10 E. Scranton Avenue, Suite 201 · Lake Bluff, Illinois 60044

NOV 1 9 2010

Re: K102010

Trade/Device Name: VACUETTE® PREMIUM Safety Blood Collection Set Regulation Number: 21 CFR 880.5570 Regulation Name: Regulatory Class: II Product Code: JKA, FMI Dated: November 5, 2010 Received: November 8, 2010

Dear Mr. Jenkins:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

6

Page 2- Mr. Jenkins

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

7

Section 4 - Indications for Use StatementPage 4-1
-----------------------------------------------------

Indications for Use

510(k) number (if Known): K102010 NOV 1 9 2010

Device Name: VACUETTE® PREMIUM Safety Blood Collection Set

Indications for Use:

"VACUETTE® PREMIUM Safety Blood Collection Set is used in routine venipuncture procedures. It is used for blood collection. The safety shield is activated automatically to cover the needle immediately following blood collection thus preventing accidental needlestick injury."

Prescription Use __ X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

An Str for RZC

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

Page_1_of_1

K102010 510(k) Number: