K Number
K102004
Device Name
ENDEAVOUR MONITORING SYSTEM
Date Cleared
2010-09-10

(57 days)

Product Code
Regulation Number
870.1025
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Endeavour Monitoring System is intended for intra hospital use under the direct supervision of a licensed healthcare practitioner. The Indications for Use for the Endeavour include the monitoring of the following human physiological parameters: - ECG waveform derived from 3, 5, 6 and 12 lead measurements - Heart Rate - Pulse Oximetry (SpO2) - ST Segment Analysis - Arrhythmia Detection - Non Invasive Blood Pressure (NIBP) - Invasive Blood Pressure (IBP) - Cardiac Output (CO) - Respiratory Gasses - Respiration Rate - Temperature The Endeavour Monitor has the capability of performing IV Drug and Hemodynamic Calculations and interfacing with network devices. The target populations are adult, pediatric and neonate with the exception of: - Arrhythmia detection and ST Segment Analysis, for which the target populations are adult and pediatric only, - IV Drug Calculations for which the target population is adult only, and - Cardiac Output for which the target population is adult and pediatric only.
Device Description
The Endeavour Monitoring System is a multi physiological parameter patient monitor. It is a modular system that allows users to customize monitored parameters based on a patients monitoring need or acuity level.
More Information

Not Found

No
The document explicitly states "Mentions AI, DNN, or ML: Not Found" and the description focuses on standard physiological monitoring parameters and calculations.

No.
The device is a monitoring system that measures physiological parameters. It does not provide any treatment or therapy.

No

The device is described as a "multi physiological parameter patient monitor" intended for "monitoring of the following human physiological parameters." It provides measurements and calculations (like ST Segment Analysis and Arrhythmia Detection) but the primary aim is monitoring, not making a definitive diagnosis. It aids in patient assessment, which can inform diagnosis, but it does not independently provide a diagnostic output.

No

The device description explicitly states it is a "multi physiological parameter patient monitor" and a "modular system," implying the presence of hardware components for acquiring and processing physiological data. The performance study also focuses on the "non-invasive blood pressure (NIBP) module," further indicating hardware is involved.

Based on the provided text, the Endeavour Monitoring System is not an IVD (In Vitro Diagnostic) device.

Here's why:

  • Intended Use: The intended use clearly states that the system is for monitoring human physiological parameters directly from the patient (ECG, SpO2, NIBP, etc.). IVD devices are used to examine specimens from the human body (like blood, urine, tissue) to provide information about a person's health.
  • Device Description: The description reinforces that it's a "multi physiological parameter patient monitor," which aligns with direct patient monitoring, not laboratory testing of specimens.
  • No mention of specimen analysis: The text does not mention any analysis of biological specimens.

While the device has the capability of performing "IV Drug and Hemodynamic Calculations," these are calculations based on the physiological parameters being monitored, not on the analysis of in vitro samples.

N/A

Intended Use / Indications for Use

The Endeavour Monitoring System is intended for intra hospital use under the direct supervision of a licensed healthcare practitioner. The Indications for Use for the Endeavour include the monitoring of the following human physiological parameters:

  • ECG waveform derived from 3, 5, 6 and 12 lead measurements .
  • Heart Rate .
  • Pulse Oximetry (SpO2) .
  • ST Segment Analysis
  • Arrhythmia Detection .
  • Non Invasive Blood Pressure (NIBP) ●
  • Invasive Blood Pressure (IBP) .
  • Cardiac Output (CO) .
  • Respiratory Gasses .
  • Respiration Rate .
  • Temperature

The Endeavour Monitor has the capability of performing IV Drug and Hemodynamic Calculations and interfacing with network devices.

Product codes (comma separated list FDA assigned to the subject device)

MHX, DSI, MLD, DSK, DXN, CCK, DQK, DRT, DQA, FLL, DXG

Device Description

The Endeavour Monitoring System is a multi physiological parameter patient monitor. It is a modular system that allows users to customize monitored parameters based on a patients monitoring need or acuity level.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

The target populations are adult, pediatric and neonate with the exception of:

  • Arrhythmia detection and ST Segment Analysis, for which the target populations are adult and pediatric only,
  • IV Drug Calculations for which the target population is adult only, and
  • Cardiac Output for which the target population is adult and pediatric only.

Intended User / Care Setting

intra hospital use under the direct supervision of a licensed healthcare practitioner.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

A non-significant risk study was conducted on the Endeavour Monitor to verify the clinical performance of the non-invasive blood pressure (NIBP) module to the predicate device, the Accutorr Plus Monitor (K983575). This clinical study was designed to compare the performance of the NIBP module in the Endeavour to that on the Accutorr Plus. The analysis demonstrated that there was no statistically significant difference in the mean readings obtained with the Endeavour and the Accutorr Plus for both systolic and diastolic blood pressure. No adverse events were reported.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K062098, K983575

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

0

K102004

510(k) Summary Endeavour Monitoring System

This 510(k) Summary is provided in accordance with the requirements of 21 CFR 807.92.

. .

Date:July 13, 2010
Submitter:Mindray DS USA, Inc.
800 MacArthur Blvd.
Mahwah, NJ 07430
Contact: Russell Olsen
Director, Quality Assurance and Regulatory Affairs
Telephone: 201-995-8391
Facsimile: 201-995-8605
SEP 10 2010
Device Trade Name:Endeavour Monitoring System
Common Name:Multi-parameter patient monitor (with arrhythmia detection or alarms)
Device Classification:§870.1025- MHX - Physiological, Patient Monitor, with
arrhythmia detector or alarms
§870.1025- DSI - Arrhythmia detector and alarm
§870.1025- MLD - ST Segment with alarm monitor
§870.1110- DSK- Blood Pressure computer
§870.1130- DXN - Non-invasive blood pressure measurement system
§868.1400- CCK - Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase
§870.1425- DQK - Programmable diagnostic computer
§870.2300- DRT - Cardiac Monitor (Incl. Cardiotachometer and rate
alarm)
§870.2700- DQA- Oximeter
§880.2910- FLL - Clinical electronic thermometer
§870.1435- DXG - Single-function, pre-programmed diagnostic
computer
Predicate Devices:Spectrum Monitor - K062098
Accutorr Plus Monitor - K983575
Device description:The Endeavour Monitoring System is a multi physiological
parameter patient monitor. It is a modular system that allows users
to customize monitored parameters based on a patients monitoring
need or acuity level.
Indications for Use:The Endeavour Monitoring System is intended for intra hospital use
under the direct supervision of a licensed healthcare practitioner. The
Mindray DS USA, Inc.
Endeavour Monitoring System
510(k) Summary

Page 1 OF 4

1

Indications for Use for the Endeavour include the monitoring of the following human physiological parameters:

-ECG waveform derived from 3, 5, 6 and 12 lead measurements -Heart Rate

-Pulse Oximetry (SpO2)

-ST Segment Analysis -Arrhythmia Detection

-Non Invasive Blood Pressure (NIBP)

-Invasive Blood Pressure (IBP)

  • -Cardiac Output (CO)
  • -Respiratory Gasses
  • -Respiration Rate
  • -Temperature

The Endeavour Monitor has the capability of performing IV Drug and Hemodynamic Calculations and interfacing with network devices.

The target populations are adult, pediatric and neonate with the exception of.

-Arrhythmia detection and ST Segment Analysis, for which the target populations are adult and pediatric only,

-IV Drug Calculations for which the target population is adult only, and

-Cardiac Output for which the target population is adult and pediatric onlv.

Technological Comparison to Predicate Device:

The Endeavour is equivalent to predicated devices respecting indications for use, basic operation, performance specifications and energy supply. The Endeavour incorporates the use of new materials for the handle, outer housing and front bezel. Additionally, the Endeavour supports a modular parameter design where as the predicate devices are internally configured.

Mindray DS USA, Inc. Endeavour Monitoring System 510(k) Summary

Page 2 OF 4

2

Summary of Performance Testing:

The Endeavour Monitoring System has been tested and found to be in compliance with the following recognized safety, performance and electromagnetic compatibility standards.

IEC 60601-1:1988 +A1:1991 +A2:1995Medical electrical Equipment - General Requirements for Safety
IEC 60601-1-1:2000Medical electrical equipment - Part 1-1: General requirements for safety - Collateral standard: Safety requirements for medical electrical systems
IEC 60601-1-2:2007Medical electrical equipment - Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests
IEC 60601-1-4:1996 + A1:1999Medical electrical equipment - Part 1-4: General requirements for safety - Collateral standard: Programmable electrical medical systems
IEC 60601-2-34:2000Medical electrical equipment - Part 2-34: Particular requirements for the safety, including essential performance, of invasive blood pressure monitoring equipment
IEC 60601-1-8:2003 +A1:2006Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
IEC 60601-2-27:2005Medical electrical equipment, Part 2: Particular requirements for the safety of electrocardiographic monitoring equipment
IEC 60601-2-25:1993 + A1:1999Medical electrical equipment, Part 2: Particular requirements for the safety of electrocardigraphs
IEC 60601-2-30:1999Medical electrical equipment, Part 2: Particular requirements for the safety, including essential performance, of automatic cycling non-invasive blood pressure monitoring equipment.
ISO 10993-1:2003Biological evaluation of medical devices - Part 1: Evaluation and testing.
AAMI / ANSI HE74:2001Human Factors Design Process for Medical Devices
ISO 15223: 2000Medical devices - Symbols to be used with medical device labels, labeling and information to be supplied
ISO 14971:2007Medical devices -- Application of risk management to medical devices
ISO 9919:2005Medical electrical equipment - Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use
ANSI/AAMI SP10:2002 + A1:2003 + A2:2006Manual, electronic or automated sphygmomanometers
ASTM E1112-00 (2006)Standard Specification for Electronic Thermometers for Intermittent Determination of Patient Temperature
EC13: 2002/(R)2007Cardiac monitors, heart rate meters, and alarms
EC11:1991/(R)2007Diagnostic electrocardiographic devices
EC57:1998/( R )2003Testing and Reporting Performance Results of Cardiac
Rhythm and ST-Segment Measurement Algorithms
EN ISO 21647:2004 + Cor 2006Medical electrical equipment - Particular requirements
for the basic safety and essential performance of
respiratory gas monitors

Mindray DS USA, Inc. Endeavour Monitoring System 510(k) Summary

Page 3 of 4

3

| | Additionally, a risk analysis has been developed to identify potential
hazards and document the mitigation of the hazards. The device's
software has been verified and validated in accordance with the
appropriate test requirements. |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Clinical Testing Summary: | A non-significant risk study was conducted on the Endeavour Monitor to
verify the clinical performance of the non-invasive blood pressure
(NIBP) module to the predicate device, the Accutorr Plus Monitor
(K983575). This clinical study was designed to compare the
performance of the NIBP module in the Endeavour to that on the
Accutorr Plus. |
| | The analysis demonstrated that there was no statistically significant
difference in the mean readings obtained with the Endeavour and the
Accutorr Plus for both systolic and diastolic blood pressure. No adverse
events were reported. |
| Conclusion: | Based on the technological comparison of the subject and predicate
devices, the results of clinical performance testing and compliance with
applicable performance, safety and electromagnetic compatibility
standards, the Endeavour Monitoring System shall be considered as safe,
as effective and substantially equivalent to the legally marketed predicate
devices. |

Mindray DS USA, Inc.
Endeavour Monitoring System 510(k) Summary

PAGE 4 OF 4

4

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the emblem.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002

Mindray DS USA, Inc. c/o Mr. Russell Olsen Director, Quality Assurance and Regulatory 800 MacArthur Blvd. Mahwah, NJ 0743

SEP 1 0 2010

Re: K102004

Trade/Device Name: Endeavour Monitoring System Regulatory Number: 21 CFR 870.1025 Regulation Name: Patient Physiological Monitor (with arrhythmia detection or alarms) Regulatory Class: II (two) Product Code: 74 MHX Dated: July 14, 2010 Received: July 15, 2010

Dear Mr. Olsen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

5

Page 2 – Mr. Russell Olsen

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D.

Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

6

Indications for Use

510(k) Number (if known):

KID2004

Device Name: ____ Endeavour Monitoring System SEP 1 0 2010

Indications for Use:

The Endeavour Monitoring System is intended for intra hospital use under the direct supervision of a licensed healthcare practitioner. The Indications for Use for the Endeavour include the monitoring of the following human physiological parameters:

  • ECG waveform derived from 3, 5, 6 and 12 lead measurements .
  • Heart Rate .
  • Pulse Oximetry (SpO2) .
  • ST Segment Analysis �
  • Arrhythmia Detection .
  • Non Invasive Blood Pressure (NIBP) ●
  • Invasive Blood Pressure (IBP) .
  • Cardiac Output (CO) .
  • Respiratory Gasses .
  • Respiration Rate .
  • Temperature �

The Endeavour Monitor has the capability of performing IV Drug and Hemodynamic Calculations and interfacing with network devices.

The target populations are adult, pediatric and neonate with the exception of:

  • Arrhythmia detection and ST Segment Analysis, for which the target populations are adult 트 and pediatric only,
  • IV Drug Calculations for which the target population is adult only, and 트
  • Cardiac Output for which the target population is adult and pediatric only. 합

Over-The-Counter Use AND/OR Prescription Use X (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division/Sign-Off)
Division of Cardiovascular Devices
510(k) NumberC162004

Page 1 of 1