(90 days)
Flexitime Xtreme 2 is an addition-cross-linking polyvinyl siloxane impression material for all inlay, crown and bridge, edentulous and partial impressions.
The products are developed under the project name D 948. The Flexitime Xtreme 2 Light Flow , Medium Flow and Heavy Tray each are addition-crosslinking polyvinyl siloxane impression materials , all delivered in 50 ml cartridges, and are part of the Flexitime Xtreme -system. The Flexitime Xtreme 2 assortment is characterised by the addition of Blackbeery-aroma and is technically characterized by a short working time of appr. 1.5 min and a short time in mouth period of min. 1.5 min.. These products, which show a shorter working time and setting time, are highly desired for single step impressions especially in combination with the double arch impression technique. The materials were developed to ensure hydrophilic characteristics for optimal impression taking in the wet surroundings of the mouth combined with good mechanical properties. The materials are based on the existing materials D 919 (Flexitime Light Flow and Flexitime Medium Flow) and Flexitime Xtreme Heavy Tray to meet the requirements of the US market.
The provided text describes the safety and effectiveness of the dental impression material Flexitime Xtreme 2. However, it does not detail a study with the specific acceptance criteria, device performance metrics, and methodologies like sample size, ground truth establishment, expert qualifications, or adjudication methods in the way typically expected for a medical device AI/CAD software submission.
Instead, the document highlights compliance with existing standards and regulations for a conventional medical device (dental impression material). Therefore, I will extract information related to the device's characteristics and the evaluations conducted to demonstrate its safety and effectiveness, interpreting them as meeting acceptance criteria for compliance rather than specific performance metrics against a defined ground truth.
Here's a breakdown of the available information:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria (Interpreted from stated compliance) | Reported Device Performance (as per document) |
|---|---|
| Material Composition & Equivalence | Similar to Flexitime Xtreme (K042878) with an additional raw material (Heptamethyltrisloxane, Polyalkyleneoxid modified - CAS 27306-78-1) already used in other Flexitime products. |
| Functional Specifications | "The target-performance comparisons prove that the requirements of the functional specification are fulfilled by these laboratory products." |
| Physical Properties (EN ISO 4823:2007) | "The products fulfill all requirements of the EN ISO 4823: 2007." "The physical properties of Flexitime Xtreme 2 were determined in accordance with EN ISO 4823. The results have shown good properties of Flexitime Xtreme 2 in accordance to this standard." |
| Biocompatibility (EN ISO 10993-1) | "The biological compatibility of Flexitime Xtreme 2 was verified in accordance with the international standards." "A toxicological evaluation report has confirmed that the product Flexitime Xtreme 2 meets the requirements of the DIN EN ISO 10993 standard." |
| Clinical Performance & Safety (MEDDEV 2.7.1) | "A positive benefit versus risk ratio can be stated by the experts for Flexitime Xtreme 2, provided that the products applied in accordance with its intended use as outlined in the manufacturer's instruction for use." "Potential undesirable clinical effects and risks seem well controlled and acceptable, when weighed against their benefits in dentistry." |
| Risk Analysis (EN ISO 14971) | "The risk analysis (according to DIN EN ISO 14971) was carried out for Flexitime Xtreme 2 and showed that the application of Flexitime Xtreme 2 could be considered to be safe." |
| Intended Use | "Flexitime Xtreme 2 is an addition-cross-linking polyvinyl siloxane impression material for all inlay, crown and bridge, edentulous and partial impressions." |
| Working Time | "A short working time of appr. 1.5 min." |
| Time in Mouth Period | "Min. 1.5 min." |
| Hydrophilic Characteristics | Developed "to ensure hydrophilic characteristics for optimal impression taking in the wet surroundings of the mouth." |
| Mechanical Properties | Developed "combined with good mechanical properties." |
2. Sample size used for the test set and the data provenance:
The document does not specify a "test set" in the context of an AI/CAD device. The evaluations described are primarily focused on compliance with material standards and general clinical safety/effectiveness. There is no mention of specific sample sizes for clinical trials or material testing beyond the general statement that "laboratory prototypes have been compiled" and that clinical evaluation was performed according to MEDDEV 2.7.1.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Similar to point 2, the concept of establishing "ground truth" for a test set with expert consensus is not directly applicable here. The "experts" referred to are those who conducted the biocompatibility evaluation and the clinical evaluation.
- Clinical Evaluation: "A positive benefit versus risk ratio can be stated by the experts for Flexitime Xtreme 2..." The qualifications of these experts are not explicitly detailed beyond being involved in a "clinical evaluation report ... in accordance with MEDDEV 2.7.1."
- Biocompatibility Evaluation: Conducted according to EN ISO 10993-1. The individuals performing this are not specifically named or qualified in the summary, but it's implied they are qualified to interpret such standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
No adjudication method is described, as this is not a study involving human readers or AI in a diagnostic or prescriptive capacity.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This is not applicable as the device is a dental impression material, not an AI/CAD system. There are no "human readers" or "AI assistance" in the context of this submission.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This is not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
For the safety and effectiveness of the dental impression material, the "ground truth" is established through:
- Compliance with International Standards: EN ISO 4823 (physical properties), EN ISO 10993-1 (biocompatibility), EN ISO 14971 (risk analysis), and MEDDEV 2.7.1 (clinical evaluation).
- Laboratory Testing: "Target-performance comparisons" to functional specifications.
- Clinical Evaluation: Assessment of clinical performance and safety, leading to a conclusion of a positive benefit/risk ratio by experts. This is likely based on clinical observations and professional judgment rather than a single, objective "ground truth" like pathology.
8. The sample size for the training set:
Not applicable, as this is not an AI/CAD system.
9. How the ground truth for the training set was established:
Not applicable, as this is not an AI/CAD system.
Summary of the "Study" (Compliance and Evaluation instead of a traditional clinical study):
The device (Flexitime Xtreme 2) meets its "acceptance criteria" by demonstrating compliance with relevant international and European medical device standards and legislation. The "study" comprises various evaluations:
- Nonclinical Tests: Biocompatibility testing (EN ISO 10993-1) and physical property testing (EN ISO 4823). The biological compatibility was verified, and a toxicological evaluation confirmed it meets requirements, with a positive benefit/risk ratio. Physical properties were shown to be "good" according to the standard.
- Clinical Evaluation: Performed in accordance with MEDDEV 2.7.1. This involved critically evaluating the clinical benefits against potential risks. It concluded that the products exhibit the claimed technical performance, with acceptable risks when weighed against benefits, leading to a positive benefit-risk ratio.
- Risk Analysis: Conducted according to DIN EN ISO 14971, concluding that the application of Flexitime Xtreme 2 is safe.
These evaluations, rather than a single "study," collectively prove the device meets its acceptance criteria for safety and effectiveness within the regulatory framework for dental impression materials. No specific quantitative performance metrics against a defined ground truth (as would be seen for an AI diagnostic device) are provided, as the nature of the device does not necessitate such a study design.
{0}------------------------------------------------
Technical File
Summary of safety and effectiveness of the Impression material Flexitime Xtreme 2
-
- Submitter name
Heraeus Kulzer, LLC 300 Heraeus Way South Bend, Indiana 46614
- Submitter name
SEP 0 7 2010
Heraeus
Contact person: Cheryl V. Zimmerman Tel. +1 (574) 299 - 5444
Date summary prepared: April, 21, 2010
-
- Name of the device:
Flexitime Xtreme 2 Light Flow/Medium Flow/Heavy Tray The product is Regulatory Class II and the product code ELW It is used as dental impression material.
- Name of the device:
The GMDN number is 35866 and the description is Impression material, dental. silicon rubber , the UMDS number is 16-679 and the description is Dental-Abdruckmaterial, Silikongummi.
3. Substantially equivalence
Flexitime Xtreme 2 (Project Name D 948) is a revised version of the product Flexitime Xtreme (K042878). The main components and their ratio in Flexitime Xtreme and Flexitime Xtreme 2 are similar. Flexitime Xtreme 2 contains as an additional raw material Heptamethyltrisloxane, Polyalkyleneoxid modified(CAS 27306-78-1) (K091494), which has been already used in Flexitime Light Flow and Medium Flow.
4. Description of the device
The products are developed under the project name D 948. The Flexitime Xtreme 2 Light Flow , Medium Flow and Heavy Tray each are addition-crosslinking polyvinyl siloxane impression materials , all delivered in 50 ml cartridges, and are part of the Flexitime Xtreme -system. The Flexitime Xtreme
{1}------------------------------------------------
Seite 2
2 assortment is characterised by the addition of Blackbeery-aroma and is technically characterized by a short working time of appr. 1.5 min and a short time in mouth period of min. 1.5 min.. These products, which show a shorter working time and setting time, are highly desired for single step impressions especially in combination with the double arch impression technique. The materials were developed to ensure hydrophilic characteristics for optimal impression taking in the wet surroundings of the mouth combined with good mechanical properties. The materials are based on the existing materials D 919 (Flexitime Light Flow and Flexitime Medium Flow) and Flexitime Xtreme Heavy Tray to meet the requirements of the US market. Flexitime Light Flow and Flexitime Medium Flow are marketed since August 2009 and Flexitime Xtreme Heavy Tray is marketed since March 2006.
For the types D 948 Light Flow, D 948 Medium Flow and D 948 Heavy Tray the following accepted laboratory prototypes have been compiled: PDE 704-2/PDE 705-2, PDE 694-1/PDE 695-2 und PDE 702-1/PDE 703-1.
The target-performance comparisons prove that the requirements of the functional specification are fulfilled by these laboratory products. The prototypes were provided for further development and the upscaling by the process development department shows the functional specifications are fulfilled also for the scale up material.
The products fulfill all requirements of the EN ISO 4823: 2007.
5. Intended use
Flexitime Xtreme 2 is an addition-cross-linking polyvinyl siloxane impression material for all inlay, crown and bridge, edentulous and partial impressions.
6. Summaries
a: See point 3
b: (1) Nonclinical- tests: In accordance with the Medical Device Directive 93/42/EWG and national European medical device legislation, any medical device is requested to be evaluated by the legal medical device manufacturer regarding its clinical performance and safety. This includes an evaluation of biocompatibility in accordance with EN ISO 10993-1.
The biological compatibility of Flexitime Xtreme 2 was verified in accordance with the international standards.
{2}------------------------------------------------
Heraeus
Seite 3
The biocompatibility of Flexitime Xtreme 2 in the aforementioned indication was documented in a biocompatibility evaluation report and the benefit/riskrelation has been judged as positive.
(2) Clinical Evaluation
In accordance with the medical Device directive 93/42/EWG and national European medical device legislation, any medical device is requested to be evaluated by the legal medical device manufacturer regarding its clinical performance and safety. This includes a clinical evaluation in accordance with MEDDEV 2.7.1., which is intended to critically evaluate the clinical benefits of the medical device in comparison to its potential risks. Therefore, any clinical evaluation is part of the compulsory risk management process according to EN ISO 14971, and critical findings must further be considered in the current risk management process of the medical device manufacturer responsible for the evaluated device.
On this background, the clinical evaluation was performed in order to comply with the current European medical device legislation, in particular with MEDDEV 2.7.1. This critical evaluation followed the procedures outlined in the corresponding clinical evaluation plan.
Flexitime Xtreme 2 is an impression material system, which is generally classified as a class I medical device under the Medical Device Directive 93/42/EEC.
Considering the evaluated scientific data and technical results for Flexitime Xtreme 2 it is concluded that the products can be expected to exhibit the claimed technical performance and that potential undesirable clinical effects and risks seem well controlled and acceptable, when weighed against their benefits in dentistry. Therefore, a positive benefit versus risk ratio can be stated by the experts for Flexitime Xtreme 2, provided that the products applied in accordance with its intended use as outlined in the manufacturer's instruction for use.
The clinical evaluation report was prepared in accordance with MEDDEV 2.7.1 and followed the provisions of the corresponding clinical evaluation plan.
(3) Conclusion
A positive benefit versus risk ratio can be stated by the experts for Flexitime Xtreme 2. provided that the product applied in accordance with its intended use as outlined in the manufacturer's instruction for use for the clinical and the nonclinical test results.
(c) 510 (k) summary
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Heraeus" in a sans-serif font. The letters are outlined in black, giving them a slightly textured appearance. The word is presented in a straightforward, horizontal arrangement.
Seite 4
The risk potential of the Flexitime Xtreme 2 was proved considering the current composition. All properties of the product were verified successfully.
The biological compatibility of the impression material was investigated to evaluate the toxicological risk. A toxicological evaluation report has confirmed that the product Flexitime Xtreme 2 meets the requirements of the DIN EN ISO 10993 standard. The results were discussed in a Biocompatibility Evaluation Report and the benefit/risk-relation has been judged as positive.
The physical properties of Flexitime Xtreme 2 were determined in accordance with EN ISO 4823. The results have shown good properties of Flexitime Xtreme 2 in accordance to this standard.
Based on the results of the clinical evaluation report it is concluded that the product can be expected to exhibit the claimed technical performance and that potential undesirable clinical effects and risks seem well controlled and acceptable when weighted against their benefits in dentistry.
The risk analysis (according to DIN EN ISO 14971) was carried out for Flexitime Xtreme 2 and showed that the application of Flexitime Xtreme 2 could be considered to be safe.
Flexitime Xtreme meets all requirements relevant for dental impression material in accordance with the Medical Device directive 93/42/EWG and national European medical device legislation. Based on the actual facts Flexitime Xtreme 2 could be evaluated to be effective and safe with its intended use as outlined in the manufacturer's instruction for use.
Dormagen, 01.09.2010
Martin Sinwald
i.A.
Dr. Martin Grunwald
i.A. H. Hunter
Heike Jansen
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002
Ms. Cheryl Zimmerman Director, Quality Assurance & Regulatory Affairs Heraeus Kulzer, LLC 300 Heraeus Way South Bend, Indiana 46614
SEP 0 7 2010
Re: K101617
Trade/Device Name: Flexitime Xtreme 2 Regulation Number: 21 CFR 872.3660 Regulation Name: Impression Material Regulatory Class: II Product Code: ELW Dated: June 2, 2010 Received: June 9, 2010
Dear Ms. Zimmerman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranics. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2- Ms. Zimmerman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing.(21 CFR Part.807.); labeling (21 CFR.Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Jams D. Roberts for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page_1__of__1_
510(k) Number (if Known):
K10/617
Device Name: Flexitime Xtreme 2
SEP 0 7 2010
Indications For Use:
Flexitime Xtreme 2 is an addition-cross-linking polyvinyl siloxane impression material for all inlay, crown and bridge, edentulous and partial impressions.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device evaluation (ODE)
Prescription Use ﺮ OR
(Per 21 CFR 801.109)
Over-The-Counter Use
Sueom Runner
(Division Sign-Off Ontional Division of Anesthesiology, Infection Control, Dental Devices
510(k) Number: K101617
§ 872.3660 Impression material.
(a)
Identification. Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.(b)
Classification. Class II (Special Controls).