K Number
K101566

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-09-28

(116 days)

Product Code
Regulation Number
866.1640
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

VITEK® 2 Yeast Caspofungin is designed for antifungal susceptibility testing of Candida species. VITEK 2 Yeast Caspofungin is a quantitative test intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antifungal agents. Caspofungin has been shown to be active both in vitro and in clinical infections against most strains of the following microorganisms according to the FDA label for the antifungal.

Candida albicans Candida glabrata Candida guilliermondii

Candida krusei Candida parapsilosis Candida tropicalis

. The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 System for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, S. pneumoniae and clinically significant yeast.

Device Description

VITEK® 2 Yeast Caspofungin is designed for antifungal susceptibility testing of Candida species. VITEK® 2 Yeast Caspofungin is a quantitative test intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antifungal agents. Caspofungin has been shown to be active against most strains of the following microorganisms listed below according to the FDA label for the antifungal.

Candida albicans Candida krusei Candida parapsilosis Candida tropicalis Candida lusitaniae Candida quilliermondii

The antifungal presented in VITEK® 2 AST Cards is in concentrations equivalent by efficacy to standard method concentrations in mcg/ml. The VITEK 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology.

The isolate to be tested is diluted to a standardized concentration in 0.45% saline before being used to TThe ISolate to be tested is different is a carded be normatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling and sealing operation. The VITEK® 2 monitors the growth of each well in the card over a defined period of time (up to 36 hours for yeast). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antifungal contained on the card.

AI/ML Overview

Here's an analysis of the acceptance criteria and study details for the VITEK® 2 Yeast Caspofungin device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document refers to the "FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA. Issued Mar. 5, 2007" as defining the acceptance criteria for substantial equivalence. While the document doesn't explicitly list numerical acceptance thresholds within this text, it does state the device's performance results in relation to those implied criteria.

MetricAcceptance Criteria (Implied by FDA Guidance)Reported Device Performance
Overall Essential Agreement(Defined by FDA Guidance, not specified here)99.5%
Overall Category Agreement(Defined by FDA Guidance, not specified here)99.8%
ReproducibilityAcceptableAcceptable
Quality ControlAcceptableAcceptable

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: The document mentions "fresh and stock clinical isolates, as well as a set of challenge strains." However, a specific numerical sample size for the test set is not provided in the document.
  • Data Provenance: The document does not specify the country of origin for the data. The data is described as an "external evaluation" using "fresh and stock clinical isolates" and "challenge strains," implying real-world samples and engineered samples, but does not explicitly state if it was retrospective or prospective. Given the mention of "fresh clinical isolates," it suggests a prospective component.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

The document does not provide details on the number of experts or their qualifications used to establish the ground truth for the test set. It refers to the "CLSI reference method" as the comparator for ground truth, which is a standardized laboratory method, not an expert panel.

4. Adjudication Method for the Test Set:

No adjudication method for a test set is described, as the ground truth relies on a standardized reference method (CLSI), not expert consensus that typically requires adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No MRMC comparative effectiveness study is mentioned, nor is there any discussion of human readers improving with AI assistance. The device is an automated in vitro diagnostic intended to determine antimicrobial susceptibility, not to assist human interpretation of images or other subjective data.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

Yes, the study described is a standalone performance study. The VITEK® 2 Yeast Caspofungin system is an automated device designed to produce quantitative and interpretive results (MIC values and interpretive categories) without direct human intervention in the result generation process once the sample is loaded. The performance metrics (Essential Agreement and Category Agreement) are direct comparisons between the device's output and the CLSI reference method.

7. Type of Ground Truth Used:

The ground truth used for the test set was the CLSI (Clinical and Laboratory Standards Institute) reference method. Specifically, it refers to the CLSI reference method incubated at 24 hours. The CLSI method for antimicrobial susceptibility testing is a highly standardized, laboratory-based method considered the gold standard for determining minimum inhibitory concentrations (MICs).

8. Sample Size for the Training Set:

The document does not provide any information regarding the sample size used for the training set.

9. How the Ground Truth for the Training Set Was Established:

The document does not provide information on how the ground truth for the training set was established. Given the nature of the device (an AST system), it's likely that extensive data from CLSI reference methods would have been used during development and internal validation, but this is not explicitly stated as for a "training set" in the context of this 510(k) summary.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for bioMérieux. The logo consists of the word "BIOMÉRIEUX" in a stylized font, with a graphic above it. The graphic is a circle bisected vertically, with the left half filled with horizontal lines and the right half solid black, and a curved line extending above and below the circle.

K101566
SEP 28 2010

510(k) SUMMARY

VITEK® 2 Yeast Caspofungin

510(k) Submission Information:

Submitter's Name:bioMérieux, Inc.
Address:595 Anglum RoadHazelwood, MO 63042
Contact Person:Nancy WeaverAssociate Director, Regulatory Affairs
Phone Number:314 -731-8695
Fax Number:314-731-8689
Date of Preparation:August 06, 2009
B. Device Name:
Formal/Trade Name:VITEK® 2 Yeast Caspofungin
Classification Name:21 CFR 866.1640Antimicrobial Susceptibility TestProduct Code NGZ
Common Name:VITEK® 2 AST-YS Caspofungin
C. Predicate Device:VITEK® 2 Yeast Fluconazole (K061945)

D. 510(k) Summary:

VITEK® 2 Yeast Caspofungin is designed for antifungal susceptibility testing of Candida species. VITEK® 2 Yeast Caspofungin is a quantitative test intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antifungal agents. Caspofungin has been shown to be active against most strains of the following microorganisms listed below according to the FDA label for the antifungal.

Candida albicans Candida krusei Candida parapsilosis Candida tropicalis Candida lusitaniae Candida quilliermondii

The antifungal presented in VITEK® 2 AST Cards is in concentrations equivalent by efficacy to standard method concentrations in mcg/ml. The VITEK 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology.

The isolate to be tested is diluted to a standardized concentration in 0.45% saline before being used to TThe ISolate to be tested is different is a carded be normatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling and sealing operation. The VITEK® 2 monitors the growth of each well in the card over a defined period of time (up to 36 hours for

bioMérieux, Inc. Phone: 314/731-8500 800/638-4835 Fax: 314/731-8700 595 Anglum Road, Hazelwood, Missouri 63042-2320, USA http://www.biomerieux-usa.com

{1}------------------------------------------------

2

veast). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antifungal contained on the card.

VITEK 2 Yeast Caspofungin demonstrated substantially equivalent performance when compared with the CLSI reference method, as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA. Issued Mar. 5, 2007.

The Premarket Notification (510[k]) presents data in support of VITEK® 2 Yeast Caspolungin. An external evaluation was conducted with fresh and stock clinical isolates, as well as a set of challenge strains. The external evaluations were designed to confirm the acceptability of VITEK® 2 Yeast Caspofungin by comparing its performance with the CLSI reference method incubated at 24 hrs. The data is representative of performance on both the VITEK® 2 and VITEK® 2 Compact instrument platforms VITEK 2 Yeast Caspofungin demonstrated acceptable performance of 99.5% overall essential Agreement and 99.8% overall category agreement with the reference method incubated for 24 hrs. Reproducibility and Quality Control demonstrated acceptable results.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract human profiles connected by flowing lines, creating a sense of unity and movement. The symbol is black, providing a strong contrast against the white background.

DEPARTMENT OF HEALTH & HUMAN SERVICES

. Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002

Ms. Nancy Weaver Associate Director, Regulatory Affairs bioMérieux, Inc. 595 Anglum Road Hazelwood, MO 63042

SEP 2 8 2010

Re: K101566

Trade Name: VITEK®2 Yeast Caspofungin Regulation Number: 21 CFR §866.1640 Antimicrobial susceptibility test powder Regulation Name: Regulatory Class: Class II Product Code: NGZ, LRG Dated: August 11, 2010 Received: August 12, 2010

Dear Ms. Weaver

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section

{3}------------------------------------------------

Page 2 - Nancy Weaver

510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Liddie Poole

for
Sally A. Hoivat, M.Sc., Ph.D.

Sally A. Hojvat, M.Sc., Ph.I Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): k101566_

SEP 2 8 2010

Device Name: VITEK 2 Yeast Caspofungin (0.25 - 4 ug/ml)

Indications For Use:

VITEK® 2 Yeast Caspofungin is designed for antifungal susceptibility testing of Candida species. VITEK 2 Yeast Caspofungin is a quantitative test intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antifungal agents. Caspofungin has been shown to be active both in vitro and in clinical infections against most strains of the following microorganisms according to the FDA label for the antifungal.

Candida albicans Candida glabrata Candida guilliermondii

Candida krusei Candida parapsilosis Candida tropicalis

. The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 System for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, S. pneumoniae and clinically significant yeast.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Freddie L. Poole

vision Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K/0 1566

Page 1 of 1

§ 866.1640 Antimicrobial susceptibility test powder.

(a)
Identification. An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.(b)
Classification. Class II (performance standards).