(266 days)
The golden gate™ anterior plate system is intended to provide stabilization during the development of a spinal fusion. The golden gate™ System is indicated for use in the anterior/anterolateral thoracolumbar spine (T1-L5) for the treatment of: DDD (back pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies), Spondylolisthesis, Spondylolysis, Fracture (including dislocation or subluxation), Spinal stenosis, Deformities (i.e., scoliosis, kyphosis, lordosis), Tumor, Pseudarthrosis, Revision of previous surgery.
The golden gate™ anterior plate system is used for the surgical stabilization and fixation of the thoracolumbar spine to provide a suitable environment for fusion to occur. The fixation components of the system include plates, screws and connectors (hex nuts and inlays). These are available in a variety of sizes and lengths to accommodate differing patient anatomy.
The provided text describes a 510(k) premarket notification for a medical device called the "golden gate™ System," a spinal intervertebral body fixation orthosis. The notification focuses on establishing substantial equivalence to previously cleared predicate devices through performance data, specifically mechanical testing.
Here's an analysis of the acceptance criteria and study information based on your request:
1. Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Mechanical performance for spinal intervertebral body fixation orthoses, as defined by ASTM F1717 guidelines (static and dynamic loads). | Static compression bending and torsion, and dynamic compression bending of the worst-case golden gate™ construct met or exceeded the performance of the predicate devices according to ASTM F1717. The system performed "as well as or better than the predicate devices." |
Note: The document does not explicitly state numerical acceptance criteria (e.g., minimum load bearing capacity in Newtons, or maximum displacement in millimeters). Instead, it relies on a comparative standard: performing as well as or better than the predicate devices when tested against the ASTM F1717 standard. This is a common approach for 510(k) submissions demonstrating substantial equivalence.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify a numerical sample size for the "worst case golden gate™ construct" tested. It refers to a "worst case" construct, implying that representative configurations of the device were tested. For mechanical testing, this typically involves a limited number of samples for each configuration to demonstrate compliance with a standard.
- Data Provenance: The study was a prospective mechanical laboratory study. The document does not provide details on the country of origin of the data beyond the sponsor being a German company (ulrich GmbH & Co. KG) with a US subsidiary (ulrich medical USA, Inc.). The testing was performed in accordance with ASTM F1717, an international standard.
3. Number of Experts Used to Establish Ground Truth and Qualifications
This section is not applicable to this submission. The device is a mechanical implant, and its performance is evaluated through objective mechanical testing against a recognized standard (ASTM F1717), not through subjective interpretation by human experts or clinical ground truth.
4. Adjudication Method for the Test Set
This section is not applicable. The "test set" in this context refers to the physical devices undergoing mechanical stress tests. There is no human adjudication involved in interpreting these objective mechanical measurements.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
This section is not applicable. An MRMC study is relevant for diagnostic devices or AI systems where human reader performance is being evaluated or compared. This submission pertains to a surgical implant, and its effectiveness is determined by mechanical stability and successful fusion, not diagnostic accuracy.
6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study
This section is not applicable. The "golden gate™ System" is a physical medical device (spinal implant), not a software algorithm or AI system. Therefore, standalone algorithm performance is not relevant.
7. Type of Ground Truth Used
The "ground truth" for this device's performance is established by objective mechanical test results directly measured during the tests (e.g., load-to-failure, displacement, fatigue cycles) compared against the benchmark performance of predicate devices as per the ASTM F1717 standard.
8. Sample Size for the Training Set
This section is not applicable. This is a physical medical device, not a machine learning model. Therefore, there is no "training set."
9. How the Ground Truth for the Training Set Was Established
This section is not applicable, as there is no training set for this device.
{0}------------------------------------------------
上101353
FEB - 4 2011
Image /page/0/Picture/2 description: The image shows the logo for Ulrich Medical USA. The word "Ulrich" is in a bold, sans-serif font, with the "U" partially overlapping a black square. Below "Ulrich" is the text "medical USA" in a smaller, sans-serif font.
510(k) Summary
Date: Sponsor: 21 December 2010 ulrich GmbH & Co. KG
Buchbrunnenweg 12 89081 Ulm Germany Phone: +49 (0) 731-9654-1304 Fax: +49 (0) 731-9654-2802
Contact Person:
Hans Stover ulrich medical USA, Inc. 612 Trade Center Blvd. Chesterfield, MO 63005 (636) 519-0268 Office (636) 519-0271 Fax
Proposed Trade
golden gate™ System
Device Classification Class II
Spinal intervertebral body fixation orthosis Classification Name:
Anterior thoracolumbar plate system Common Name:
Regulation:
Name:
Code:
Device Product
KWQ
888.3060
The golden gate™ anterior plate system is used for the surgical Device Description: stabilization and fixation of the thoracolumbar spine to provide a suitable environment for fusion to occur. The fixation components of the system include plates, screws and connectors (hex nuts and inlays). These are available in a variety of sizes and lengths to accommodate differing patient anatomy.
The golden gate™ anterior plate system is intended to provide Intended Use: stabilization during the development of a spinal fusion. The golden gate™ system is indicated for use in the anterior/anterolateral thoracolumbar spine (T1-L5) for the treatment of: DDD (back pain of discogenic origin with degeneration of the disc as confirmed by Spondylolisthesis, patient history and radiographic studies), Spondylolysis, Fracture (including dislocation or subluxation), Spinal stenosis, Deformities (i.e., scoliosis, kyphosis, lordosis), Tumor, Pseudarthrosis, Revision of previous surgery.
Materials:
The golden gate™ system components are manufactured from titanium alloy (Ti-6Al-4V) as described by ISO 5832-3 / ASTM F136.
Page 1 of 2
{1}------------------------------------------------
| Predicate Devices: | Kaneda SR (K971248)MACS TL (K002824 and K032059)Z-Plate (K922543) |
|---|---|
| TechnologicalCharacteristics: | The golden gate™ System possesses the same technologicalcharacteristics as the predicate devices. These include basic design(rod/plate-based screw system), material (titanium or titanium alloy),sizes (rod/plate lengths and screw diameters and lengths are withinthe range(s) offered by the predicate systems) and intended use (asdescribed above). The fundamental scientific technology of thegolden gate™ System is the same as previously cleared devices. |
| Performance Data: | Static compression bending and torsion, and dynamic compressionbending of the worst case golden gate™ construct was performedfollowing ASTM F1717. The mechanical test results demonstrate thatthe golden gate™ System performs as well as or better than thepredicate devices. |
| Conclusion: | The golden gate™ System is substantially equivalent to the devicesreferenced above and is therefore as safe and as effective for itsintended use. |
11 : 1 : 1 : 1
.
u
. .
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
FEB = 4 2011
ulrich GmbH & Co.. KG % ulrich medical USA. Inc. Mr. Hans Stover President and CEO 612 Trade Center Boulevard Chesterfield. Missouri 63005
Re: K101353
Trade/Device Name: golden gate™ System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: December 22, 2010 Received: December 23, 2010
Dear Mr. Stover:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in intestate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendmonts, or to devices that have been reclassified in accordance with the provisions of the Fedral Food, Dr to and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract links and warranties. We remind you; however, that device labeling must be truth tol and not muchading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your devices on be, found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
and the country of the county of the county of
{3}------------------------------------------------
Page 2 - Mr. Hans Stover
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours .-
At B. Rt
for.
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use Statement
510(k) Number:_K|01353
Device Name: golden gate™ system
Indications for Use:
The golden gate™ anterior plate system is intended to provide stabilization during the development of a spinal fusion. The golden gate™ System is indicated for use in the anterior/anterolateral thoracolumbar spine (T1-L5) for the treatment of:
- . DDD (back pain of discogenic origin with degeneration of the disc as confirmed by patient history and radiographic studies),
- Spondylolisthesis, .
- Spondylolysis, .
- Fracture (including dislocation or subluxation), .
- Spinal stenosis, .
- Deformity (i.e., scoliosis, kyphosis, lordosis), ●
- Tumor, .
- Pseudarthrosis, .
- Revision of previous surgery .
| Prescription Use | X |
|---|---|
| (21 CFR 801 Subpart D) |
AND/OR
| Over-the-Counter Use | |
|---|---|
| (21 CFR 807 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
Krol 353 510(k) Number_
N/A